← Product Code [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW) · K141456

# THE FIAGON NAVIGATION - EXTENDED INSTRUMENT SET ENT, REGISTRATIONPOINTER, VENTERAPOINTER, GUIDEWIRE, POINTERSHELL (K141456)

_Fiagon GmbH · PGW · Oct 8, 2014 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K141456

## Device Facts

- **Applicant:** Fiagon GmbH
- **Product Code:** [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW.md)
- **Decision Date:** Oct 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Fiagon Navigation – Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access

## Device Story

Extended Instrument Set ENT consists of reusable, electromagnetically navigated surgical instruments (suction tubes, pointers, registration probes, and adapters) for use with the Fiagon Navigation system. Instruments feature integrated sensors at the tip, tracked within a low-energy magnetic field generated by the navigation system. The system overlays instrument position onto preoperative CT/MRI scans. Used in ENT procedures (e.g., sinus surgery, skull base access) by surgeons to precisely locate anatomical structures. Malleable suction instruments allow for simultaneous suction and navigation. Instrument adapters enable tracking of standard surgical tools. Output is visual guidance on the navigation display, assisting surgeons in navigating complex anatomy, potentially improving surgical precision and reducing risk of injury to critical structures.

## Clinical Evidence

Bench testing only. Evaluated device precision and accuracy. Mean bench accuracy measured at 0.7 mm - 1.2 mm (SD 0.29 mm - 0.42 mm). 95% confidence levels were < 2 mm, comparing favorably to predicate values (0.9 mm mean, < 3 mm 95% confidence).

## Technological Characteristics

Reusable, electromagnetically tracked surgical instruments. Includes suction tubes, pointers, and adapters. Sensors integrated at instrument tips. Operates via low-energy magnetic field generator. Compatible with CT/MRI-based navigation software. Materials are user-sterilizable. Dimensions vary by component type.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Fiagon Navigation System ([K133573](/device/K133573.md))
- Malleable Suction™ Instruments ([K133665](/device/K133665.md))
- Instatrak® ([K960330](/device/K960330.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2014

Fiagon GmbH % Dr. Dirk Mucha Manager, Regulatory Affairs Neuendorfstrasse 23b 16761 Hennigsdorf, Germany

Re: K141456

Trade/Device Name: Fiagon Navigation - Extended Instrument Set ENT Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: September 5, 2014 Received: September 10, 2014

Dear Dr. Mucha:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known): K141456

Device Name: Fiagon Navigation – Extended Instrument Set ENT

Indications for Use:

The Fiagon Navigation – Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to:

ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

And / Or

Over-The-Counter-Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

## 510(k) Summary

### May 28, 2014

#### 1. Submitter Information

| Submitter: | Fiagon GmbH                |
|------------|----------------------------|
| Address:   | Neuendorfstrasse 23b       |
|            | 16761 Hennigsdorf, Germany |

| <i>Telephone:</i> | +49 3302 201 21 10 |
|-------------------|--------------------|
| <i>Telefax:</i>   | +49 3302 201 21 15 |

Contact: Mr. Dirk Mucha, Manager Regulatory Affairs

#### Device Information 2.

| Trade Name:     | Fiagon Navigation - Extended Instrument Set ENT |
|-----------------|-------------------------------------------------|
| Common Name:    | Image guided surgery system                     |
| Classification: | Class II per 21 CFR 882.4560                    |
| Device:         | Ear, Nose, and Throat Stereotaxic Instrument    |
| Product Code:   | PGW                                             |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for a new set of instruments for the Image Guided Surgery System, Fiagon Navigation system (K133573).

#### 4. Predicate Device Information

The Fiagon Navigation System described in this submission is substantially equivalent to the following predicates:

|   | Predicate Device               | Manufacturer                      | 510(k) No. |
|---|--------------------------------|-----------------------------------|------------|
| 1 | Fiagon Navigation System       | Fiagon GmbH                       | K133573    |
| 2 | Malleable Suction™ Instruments | Medtronic<br>Navigation, Inc.     | K133665    |
| 3 | Instatrak®                     | Visualization<br>Technology, Inc. | K960330    |

{4}------------------------------------------------

### 5. Device Description

The Fiagon Navigation - Extended Instrument Set ENT are reusable instruments intended to be used with the Fiagon Navigation system. The instruments in the Set are electromagnetically navigated devices that are

- Navigated suction instruments (malleable, designed to be bendable by hand, sensor a. within the tip, attachable to standard surgical vacuum suction systems)
- b. Navigated pointing devices (flexible, sensor within the tip)
- c. Registration probe (designed for non.sterile patient registration)
- d. Instrument adaptors (designed for mechanical connection to cylindric schaped surgical instruments of different diameters)

Each device incorporates a sensor device which is tracked by the navigation system with in the low-energy magnetic field of a field generator (part of the navigation system)

The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy)

| Components                     | Grouping                                                          | Sterility State                                          | Patient contact                                                   |
|--------------------------------|-------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------|
| FlexTube 3 mm                  | Suction and navigation<br>Instrument                              | user sterilized, reusable, delivered in non steril state | limited duration (<24h) External communicating Tissue/bone/dentin |
| FlexTube 4 mm                  | Suction and navigation<br>Instrument                              | user sterilized, reusable, delivered in non steril state | limited duration (<24h) External communicating Tissue/bone/dentin |
| PointerTube<br>Straight        | Suction and navigation<br>Instrument                              | user sterilized, reusable delivered in non steril state  | limited duration External communicating Tissue/bone/dentin        |
| PointerTube<br>Stammberger     | Suction and navigation<br>Instrument                              | user sterilized, reusable delivered in non steril state  | limited duration External communicating Tissue/bone/dentin        |
| PointerTube Sinus<br>Frontalis | Suction and navigation<br>Instrument                              | user sterilized, reusable delivered in non steril state  | limited duration External communicating Tissue/bone/dentin        |
| RegistrationPointer            | Pointing Instrument<br>for patient registration<br>only, ball tip | reusable not to be sterilized outside sterile field      | limited duration Surface communicating Skin                       |

### List of components

{5}------------------------------------------------

| Components        | Grouping            | Sterility State                                                         | Patient contact                                                                   |
|-------------------|---------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| VenteraPointer    | Pointing Instrument | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>(< 24h)<br>• External communicating<br>• Tissue/bone/dentin |
| GuideWire         | Pointing Instrument | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>(< 24h)<br>• External communicating<br>• Tissue/bone/dentin |
| PointerShell 4 mm | Instrument adapter  | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin                           |
| PointerShell 5 mm | Instrument adapter  | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin                           |
| PointerShell 3 mm | Instrument adapter  | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin                           |

### 6. Intended Use

The Fiagon Navigation - Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to:

ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access.

{6}------------------------------------------------

### 7. Comparison of Technological Characteristics

The substantial equivalence of the Fiagon Navigation – Extended Instrument Set ENT to the predicates is shown by similarity in intended use, indications for use, materials, and performance. The Fiagon Navigation System and its predicates utilize:

- Electromagnetic tracking technology for navigation attached or included in surgical ● instruments (all predicates)
- Malleable suction instruments enabled for image guidance (Predicate 2)
- Adapters for Instruments for enabling image guidance (Predicate 3)
- . Instruments with tracking sensor at the instrument tip, precalibrated (Predicate 1 and 2)

The primary difference between the Fiagon Navigation - Extended Instrument Set ENT and it predicates Fiagon Navigation System is that the Fiagon Navigation - Extended Instrument Set ENT provides electromagnetic tracking of the same kind in instruments of different geometry. This different does not raise new issues of safety and effectiveness, with respect to the following:

Cannulated instruments have been demonstrated by predicate 2. Therefore, the suction instruments of the Extended Instrument Set ENT do not raise new issues of safety and effectiveness.

Instrument adapters have been demonstrated by predicate 3. Therefore, the instrument adapters of the Extended Instrument Set ENT do not raise new issues of safety and effectiveness

### 8. Performance Data

Testing was performed in order to determine device precision and accuracy. The following nonclinical tests were performed to determine substantial equivalence:

Bench testing was conducted to determine the device accuracy and the performance of the electromagnetic field distortion mechanism.

A mean bench accuracy of 0.7 mm - 1.2 mm (Standard deviation 0.29 mm - 0.42 mm) was measured for the different instrunents. All 95% confidence levels were < 2mm which compares to the values 0.9 mm (mean) resp. < 3 mm (95% confidence) reported for the predicate devices.

The results supports the claim of substantial equivalence to the predicate devices.

Testing was completed to ensure functionality and compatibility with the Fiagon Navigation system.

{7}------------------------------------------------

### 9. Conclusion

Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the Fiagon Navigation – Extended Instrument Set ENT has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K141456](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K141456)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
