QUH · Self-Fitting Air-Conduction Hearing Aid, Over The Counter

Ear, Nose, Throat · 21 CFR 874.3325 · Class 2

Overview

Product CodeQUH
Device NameSelf-Fitting Air-Conduction Hearing Aid, Over The Counter
Regulation21 CFR 874.3325
Device ClassClass 2
Review PanelEar, Nose, Throat

Identification

A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. A self-fitting air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the self-fitting air-conduction hearing aid is subject to the following special controls: - 1. Clinical data must evaluate the effectiveness of the self-fitting strategy. - 2. Electroacoustic parameters, including maximum output limits, distortion levels, selfgenerated noise levels, latency, and frequency response, must be specified and tested. - 3. Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device. - 4. Software verification, validation, and hazard analysis must be performed. - 5. If the device incorporates wireless technology: - (A) Performance testing must validate safety of exposure to non-ionizing radiation; - (B)Performance data must validate wireless technology functions; and - (C) Labeling must specify instructions, warnings, and information relating to wireless technology and human exposure to non-ionizing radiation. - 6. Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use. - 7. Patient labeling must include the following: - (A)Information on how a patient can self-identify as a candidate for the device; - (B) Information about when to seek professional help; - (C) A warning about using hearing protection in loud environments; - (D) A warning about staying alert to sounds around the user of the device; - (E) Technical information about the device, including information about electromagnetic compatibility; and - (F) Information on how to correctly use and maintain the device.

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical data must evaluate the effectiveness of the self-fitting strategy. (2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested. (3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device. (4) Software verification, validation, and hazard analysis must be performed. (5) If the device incorporates wireless technology: (i) Performance testing must validate safety of exposure to non-ionizing radiation; and (ii) Performance data must validate wireless technology functions. (6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.

Recent Cleared Devices (10 of 10)

RecordDevice NameApplicantDecision DateDecision
K254108Jabra Enhance Select Self-fit (Jabra Enhance Select 550)GN Hearing A/SJun 25, 2026SESE
K233747Concha Sol Hearing Aids (CL-1001)Concha LabsMar 15, 2024SESE
K233464WSA Self-Fitting Hearing Aid Gen 2Wsaud A/SFeb 9, 2024SESE
K231550Sontro® OTC Hearing Aids (AI)Soundwave Hearing, LLCOct 26, 2023SESE
K221052Self-Fitting hearing aidsHuizhou Jinghao Medical Technology Co., Ltd.Sep 25, 2023SESE
K223848Intrisound™ Tuned Lumen® 155 Hearing AidsTuned , Ltd.Sep 8, 2023SESE
K230538All-Day Clear Slim ADCS1 Hearing Aid. All-Day Clear ADC1 Hearing AidSonova AGJun 16, 2023SESE
K223137Lexie Lumen Self-Fitting OTC Hearing Aids with Lexie ApplicationHearx SA (Pty) , Ltd.Mar 14, 2023SESE
K221698Eargo 5 , Eargo 6Eargo, Inc.Dec 21, 2022SESE
K221064Nuheara IQbuds 2 PRO Hearing AidNuheara LimitedOct 30, 2022SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...