Concha Sol Hearing Aids (CL-1001)

K233747 · Concha Labs · QUH · Mar 15, 2024 · Ear, Nose, Throat

Device Facts

Record IDK233747
Device NameConcha Sol Hearing Aids (CL-1001)
ApplicantConcha Labs
Product CodeQUH · Ear, Nose, Throat
Decision DateMar 15, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3325
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Concha Sol Hearing Aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. No pre-programming or hearing test is necessary. The device is intended for use without the assistance of a hearing care professional.

Device Story

Concha Sol (CL-1001) is a receiver-in-canal (RIC), self-fitting, wireless air-conduction hearing aid for adults with perceived mild-to-moderate hearing loss. The system includes the hearing aids, ear tips, carry case, and the Soundscope mobile application. Users operate the device independently; the app assesses hearing levels across frequencies and user preferences to customize amplification using the NAL-NL2 formula. The device features omni-directional/directional microphone modes, volume control via rocker switch, and Bluetooth for telephony and audio streaming. Powered by size 312 zinc-air batteries. Used in daily life to improve environmental and directive listening. Output affects clinical decision-making by providing personalized amplification without professional intervention, increasing accessibility and affordability for the target population.

Clinical Evidence

Within-subject, randomized, crossover study (n=21; 13 men, 8 women; mean age 66.2) compared self-fitting (via Soundscope app) vs. clinician-fitting. Primary endpoint: QuickSIN (speech-in-noise). Secondary endpoint: APHAB. Results showed comparable performance and satisfaction between self-fitting and clinician-fitting methods. No device-related adverse events reported. Human factors validation (n=15) and self-selection study (n=27) confirmed usability for intended users.

Technological Characteristics

RIC form factor; Zinc Air 312 batteries; Bluetooth connectivity. Electroacoustic specs: 6.7 ms latency, <1% harmonic distortion, 18.6 dB SPL EIN. Complies with ANSI S3.22/ASA S3.22-2014, CTA 2051, IEC 60601-1, IEC 60601-2-66, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Software developed per IEC 62304.

Indications for Use

Indicated for individuals 18+ with perceived mild to moderate hearing impairment. Intended for self-fitting and use without professional assistance.

Regulatory Classification

Identification

A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. A self-fitting air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable.

Special Controls

In combination with the general controls of the FD&C Act, the self-fitting air-conduction hearing aid is subject to the following special controls: - 1. Clinical data must evaluate the effectiveness of the self-fitting strategy. - 2. Electroacoustic parameters, including maximum output limits, distortion levels, selfgenerated noise levels, latency, and frequency response, must be specified and tested. - 3. Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device. - 4. Software verification, validation, and hazard analysis must be performed. - 5. If the device incorporates wireless technology: - (A) Performance testing must validate safety of exposure to non-ionizing radiation; - (B)Performance data must validate wireless technology functions; and - (C) Labeling must specify instructions, warnings, and information relating to wireless technology and human exposure to non-ionizing radiation. - 6. Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use. - 7. Patient labeling must include the following: - (A)Information on how a patient can self-identify as a candidate for the device; - (B) Information about when to seek professional help; - (C) A warning about using hearing protection in loud environments; - (D) A warning about staying alert to sounds around the user of the device; - (E) Technical information about the device, including information about electromagnetic compatibility; and - (F) Information on how to correctly use and maintain the device.

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical data must evaluate the effectiveness of the self-fitting strategy. (2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested. (3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device. (4) Software verification, validation, and hazard analysis must be performed. (5) If the device incorporates wireless technology: (i) Performance testing must validate safety of exposure to non-ionizing radiation; and (ii) Performance data must validate wireless technology functions. (6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. March 15, 2024 Concha Labs % Joseph Burke Regulatory Consultant for Concha Labs Joseph Burke Consulting Inc. 16 Rebekah Lane Sutton, Massachusetts 01590 # Re: K233747 Trade/Device Name: Concha Sol Hearing Aids (CL-1001) Regulation Number: 21 CFR 874.3325 Regulation Name: Self-Fitting Air-Conduction Hearing Aid Regulatory Class: Class II Product Code: QUH Dated: February 12, 2024 Received: February 12, 2024 # Dear Joseph Burke: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shuchen Peng -S Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Submission Number (if known) K233747 Device Name Concha Sol Hearing Aids (CL-1001) #### Indications for Use (Describe) The Concha Sol Hearing Aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. No pre-programming or hearing test is necessary. The device is intended for use without the assistance of a hearing care professional. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k)#: K233747 | 510(k) Summary | Prepared 12 Feb 202- | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------------------| | 1. CONTACT DETAILS | | 21 CFR 807.92 (a)(1) | | Concha Labs<br>55 East 3rd Avenue<br>San Mateo, CA, 94401<br>Company Contact: Carlos Bullock<br>Official Correspondent: Joseph Burke | | | | 2. DEVICE NAME | | 21 CFR 807.92 (a)(2) | | Name of Device: Concha Sol OTC Hearing Aids (CL-1001)<br>Common Name: Self-fitting OTC Hearing Aid<br>Classification Name: Self-fitting Air Conduction Hearing Aid<br>Regulation: 21 CFR §874.3325<br>Regulatory Class: II<br>Product Code: QUH<br>Review Panel: Ear, Nose and Throat Devices | | | ### 3. LEGALLY MARKETED PREDICATE DEVICE Predicate #: K211008 Predicate Trade Name: Bose SoundControl Hearing Aids Product Code: QDD #### DEVICE DESCRIPTION SUMMARY 4. The Concha Sol OTC Hearing Aids (Model CL-1001) are receiver-in-the ear, self-fitting, wireless air conduction hearing aids (§ 874.3325) intended to amplify sounds in the environment for adult users who have a perceived mild-to-moderate hearing loss. The hearing aids include Ear Tips, Carry Case and the Concha Labs mobile application. The device is indicated for over-thecounter sale (product code QUH) and integrates user input with a self fitting strategy, enabling users to independently derive and customize their hearing aid fitting and settings. Self-fitting wireless air-conduction hearing aids are Class II medical devices. The Self-Fitting hearing aids incorporate microphones for audio input, and sound is delivered to the ear via a receiver that can be coupled with domes. The mobile application contains a self-fitting software known as Soundscope, an innovative software that uses the NAL-NL2 programming formula used globally by hearing aid manufacturers. The hearing profile/personalization system in the Concha Sol Hearing Aid assesses the individual's hearing levels across a frequency spectrum in order to generate and customize the device's amplification specific to their hearing needs. The self-fitting software also assesses the individual's hearing preferences to help fine-tune the device's amplification. In addition to hearing aid functionality for environmental and directive listening; the hearing aids can also be used for telephony and streaming audio from a Bluetooth® compliant mobile device. The hearing aids are powered using Zinc Air size 312 cell batteries. **Prepared 12 Feb 2024** 55 East 3rd Avenue San Mateo, CA 94401 21 CFR 807.92 (a)(3) 21 CFR 807.92 (a)(4) **510(k) Sur** {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Concha Labs. The logo consists of a yellow stylized seashell-like icon on the left, followed by the text "CONCHA LABS" in a bold, sans-serif font. The text is in a dark gray or black color, providing a strong contrast against the white background. ## 5. INTENDED USE # 21 CFR 807.92 (a)(5) 21 CFR 807.92 (a)(5) 21 CFR 807.92 (a)(6) The Concha Sol Hearing Aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. No pre-programming or hearing test is necessary. The device is intended for use without the assistance of a hearing care professional. ## 6. INDICATIONS FOR USE COMPARISON The indications for use of the subject device are "The Concha Sol Hearing Aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. No preprogramming or hearing test is necessary. The device is intended for use without the assistance of a hearing care professional." A comparison of the indications for use of the Concha Sol Hearing Aids are similar to those of the predicate device (Bose SoundControl Hearing Aid (K211008)); with the exception of the point of sale. The Bose SoundControl is described as a Direct to Consumer (DTC) device (Product code QDD); whereas the Concha Sol Hearing Aids are indicated for Over the Counter (OTC) sale (Product code QUH). The Concha Sol is also in compliance with all elements of the OTC Hearing Aid Final Rule, as well as compliant with OTC requirements for labeling. ## 7. TECHNOLOGICAL COMPARISON The technological characteristics of the Concha Sol Hearing Aids are similar to those of the predicate device Bose SoundControl Hearing AID (K211008). The table below provides a comparison of the subject and predicate devices. A thorough review of these products revealed the devices are equivalent from a technological perspective. | | Subject Device | Predicate | Discussion | |-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | Concha Sol<br>Hearing Aids | Bose SoundControl<br>Hearing Aid | | | 510k Number | K233747 | K211008 | | | Product code | QUH | QDD | Similar to the predicate with<br>exception of the point of sale:<br>OTC (Concha) vs. DTC<br>(Bose). Subject device<br>complies with OTC hearing<br>aid requirements outlined in<br>21 CFR 800.30. | | Regulation number | 21 CFR 874.3325 | 21 CFR 874.3325 | Same as predicate | | Regulation name | Self-fitting air<br>conduction hearing<br>aid | Self-fitting air conduction<br>hearing aid | Same as predicate | | Intended Use | The Concha Sol<br>Hearing Aid is<br>intended to amplify<br>sound for individuals<br>with perceived mild to<br>moderate hearing<br>impairment. | The Bose SoundControl<br>Hearing Aid is intended to<br>amplify sound for<br>individuals with perceived<br>mild to moderate hearing<br>impairment. | Same as predicate | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a yellow symbol that appears to be a stylized representation of a flame or a drop of liquid with a jagged edge at the top. The symbol is enclosed by two curved lines on either side, resembling parentheses or sound waves. The overall design is simple and abstract, with a focus on the central flame-like shape. | Indications for Use | The Concha Sol<br>Hearing Aid is<br>intended to amplify<br>sound for individuals<br>18 years of age or<br>older with perceived<br>mild to moderate<br>hearing impairment.<br>They are adjusted by<br>the user to meet the<br>user's hearing needs.<br>No pre-programming<br>or hearing test is<br>necessary. The<br>device is intended for<br>use without the<br>assistance of a<br>hearing care<br>professional. | The Bose SoundControl<br>Hearing Aids are intended<br>to amplify sound for<br>individuals 18 years of<br>age or older with<br>perceived mild to<br>moderate hearing<br>impairment. They are<br>adjusted by the user to<br>meet the user's hearing<br>needs. No pre-<br>programming or hearing<br>test is necessary. The<br>device is intended for<br>direct-to-consumer sale<br>and use without the<br>assistance of a hearing<br>care professional. | The subject device has<br>similar Indications for Use<br>as the predicate device<br>,with the appropriate<br>adjustment in accordance<br>with the OTC Hearing Aid<br>Final Rule. | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Battery | Single use 312 Zinc<br>Air Batteries | Single use 312 Zinc Air<br>Batteries | Same as predicate | | | Subject Device | Primary Predicate | Discussion | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | Concha Sol OTC<br>Hearing Aids | Bose SoundControl<br>Hearing Aid | | | Housing | Behind The Ear (BTE)<br>Receiver In Canal<br>(RIC) hearing aid<br>housing. Separate left<br>and right ear units<br>with a separate Ear<br>Tip for each unit. On<br>the device, 'Rocker<br>Switch' used to<br>control volume.<br>Mobile device<br>streaming and phone<br>audio via Bluetooth. | Behind The Ear (BTE)<br>Receiver In Canal (RIC)<br>hearing aid housing.<br>Separate left and right<br>ear units with a separate<br>Ear Tip for each unit. On<br>device, 'Rocker Switch'<br>used to control volume. | Devices are similar with<br>only slight dimensional<br>differences. | | Input signal<br>compression | Wide dynamic range<br>compression | Wide dynamic range<br>compression | Same as predicate | | Microphones | Microphones may,<br>during use, be<br>configured by the<br>user in omni-<br>directional or<br>directional modes | Microphones may,<br>during use, be<br>configured by the user in<br>omni-directional or<br>directional modes | Same as predicate | | Battery Lifetime | Per CTA 2051 (4.1-<br>4.17)<br>Battery life of 5 days,<br>13 hours per day | Per CTA 2051 (4.1-4.7)<br>Battery life of 4 days, 14<br>hours per day | The battery life is similar to<br>that of the predicate and<br>differences raise no new<br>questions of safety and<br>effectiveness. | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Concha Labs. The logo consists of a yellow abstract symbol on the left, followed by the words "CONCHA LABS" in a bold, sans-serif font. The text is in a dark gray or black color. The logo appears to be clean and modern. | | Subject Device | Primary Predicate | Discussion / Results | |-------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | Concha Sol OTC<br>Hearing Aids | Bose<br>SoundControl<br>Hearing Aid | | | Self-fitting method | Subject device uses a<br>proprietary<br>Soundscope<br>technology, with initial<br>gain parameters<br>based on NAL- NL2. | Loudness and Fine-<br>tuning app-based<br>Controllers. Utilizes<br>a proprietary fitting<br>method. | Performance testing<br>evaluated these differences<br>in technology and the results<br>support substantial<br>equivalence. | | Mobile App | Mobile application on<br>a smart device (phone<br>or tablet) with an iOS<br>operating platform. | Mobile application on a<br>smart device (phone or<br>tablet) with either iOS or<br>Android platforms. | Similar to predicate but<br>currently subject device only<br>uses iOS operating platform.<br>This does not raise new<br>issues regarding<br>safety and effectiveness. | | Streaming /<br>Telephony | Subject device<br>provides audio<br>streaming and<br>telephone via<br>Bluetooth. | N/A | Predicate does not include<br>the features. They do not<br>raise new issues of safety or<br>effectiveness Performance<br>testing evaluated the<br>differences in technology<br>and the support substantial<br>equivalence. | | Electro-Acoustic<br>Characteristics | Subject Device | Primary Predicate | Discussion | | | Concha Sol OTC<br>Hearing Aids | Bose SoundControl<br>Hearing Aid | Compliance to \$800.30 | | Latency Clause | 6.7 ms | Less than or equal to 15<br>ms | Both devices meet<br>requirements | | Frequency<br>response | <200Hz ->10000 Hz | <200Hz -> 8000 Hz | Both devices meet<br>requirements | | Input Distortion<br>Clause | 1.20% | 5% | Both devices meet<br>requirements | | Equivalent Input<br>Noise (EIN) | 18.6 dB SPL | < 27 dB SPL | Both devices meet<br>requirements | | Harmonic<br>Distortion | Less than or equal to<br>1% | Less than or equal to 1% | Same as predicate<br>and meet<br>requirements | | Max OSPL90 | 115.2 dB SPL | 113 dB SPL | Both devices meet<br>requirements | | HFA OSPL90 | 113 dB SPL | 106 dB SPL | Both devices are<br>adequate for fitting mild<br>to moderate hearing<br>loss as prescribed by<br>NAL- NL2. | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Concha Labs. The logo consists of a stylized yellow seashell on the left, followed by the text "CONCHA LABS" in a bold, sans-serif font. The text is in a dark color, contrasting with the yellow seashell. | HFA FOG | 48.4 dB | 30 dB | Subject device will<br>provide more high-<br>frequency average full-<br>on gain (HFA-FOG) and<br>therefore more<br>amplification for soft<br>sound inputs. No<br>specific gain limit was<br>identified or required in<br>the Final OTC Hearing<br>Aid rule. Although this<br>value is higher than the<br>predicate Bose<br>SoundControl device by<br>18.4 dB, the subject<br>device features output<br>limiting (WDRC). | |---------|---------|-------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | RTG | 36.3 dB | 29 dB | The subject device is<br>adequate for fitting mild<br>to moderate hearing<br>loss as prescribed by<br>NAL-NL2. The data from<br>a clinical validation<br>study supports the<br>substantial equivalence<br>between the subject<br>device and the<br>predicate. | These findings, along with the results of performance testing, indicate that the Concha Sol Hearings Aids are substantially equivalent to the Bose SoundControl (K211008) from a technological perspective. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for CONCHA LABS. The logo consists of a stylized yellow seashell on the left, followed by the text "CONCHA LABS" in a bold, sans-serif font. The text is black and evenly spaced. #### 8. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS 21 CFR 807.92 (b) ### Biocompatibility Compliance to the testing requirements for a limited duration (less than 24 hrs.), direct skin contact device was based on the FDA recognized consensus standards ISO 10993-1 and FDA quidance on application of said standard to medical devices for U.S. market. ISO 10993-1: Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing within a Risk Management Process and FDAs 2020 final guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Biocompatibility assessment outcomes were deemed adequate based on intended use. ### Shelf Life/Sterility The non-sterile hearing aid is maintained by the user by wiping with a dry, soft cloth on the outside surface of the housing; a provided brush assists in removal of visible debris from vents and ear tip. Device is intended to have a 2 year service life while none of the device components are subject to degradation that could impact an 'endpoint' of shelf-life. #### Electrical Safety: The Concha Sol Hearing Aid electrical circuitry was tested extensively to verify design criteria and performance with respect to electrical system specifications and properties in support of its safety and effectiveness (IEC 60601-1:2012). The Concha Sol Hearing Aid was tested in finished form and passed testing for battery life, and runtime (ANSI S3.22:2014). #### 60601 The Concha Sol Hearing Aid was tested per the requirements of IEC 60601-1:2012, standard for medical electrical equipment: General requirements on basic safety and essential performance to provide reasonable assurance of the basic safety and essential performance of the device. The Concha Sol Hearing Aid was also tested per the requirements of IEC 60601-2-66:2019, standard for medical electrical equipment: Practical requirements for basic safety and essential performance of Hearing Instruments and Hearing Instrument Systems to provide reasonable assurance of the basic safety and performance of the device. ### EMC and Wireless The Concha Sol Hearing Aid was tested per the requirements of IEC 60601-1-2, 2014, AMD1:2020 standard for medical electrical equipment: General requirements on basic safety and essential performance on Electromagnetic Compatibility to provide reasonable assurance that the device performs all/remain functions that are within its specified range of operation in concert with its intended use. ### Electroacoustic Testing The Concha Sol Hearing Aid electroacoustic design was tested extensively to verify design criteria and performance with respect to electroacoustic system specifications and properties in support of its safety and effectiveness. The Concha Sol Hearing Aid passed the following tests not limited to: Monaural/Binaural Directions, Battery Life, and ANSI S3.22/ASA S3.22-2014 and CTA 2051 related tests. Electro Acoustic Measurement Verification was completed and found in compliance with §800.30 requirements and similar to the predicate. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for CONCHA LABS. The logo consists of a yellow seashell-like icon on the left, followed by the text "CONCHA LABS" in a bold, sans-serif font. The text is in a dark gray or black color, contrasting with the yellow icon. ## Battery Life Verification Testing demonstrated that Concha Sol Hearing Aid device meets ANSIS3.22/CTA 2051 standard for manufacturer to report the estimated battery life values requirements and IEC 60601-2-66:2019. ## Mechanical Testing The Concha Sol Hearing Aid mechanical design was tested extensively to verify its design criteria and performance with respect to the mechanical system specifications and properties in support of reasonable assurance of safety and effectiveness. The Concha Sol Hearing Aid, including Ear Tips were tested in finished form and passed Packaging and Environmental testing. ## Transportation Simulations of transport of the Concha Sol Hearing Aids' packaging to remain intact during testing per ASTM 599, ASTM 5276 and ASTM 4169 DC-13. All requirements were met. ## Human Factors/Usability Concha Labs assessed its hearing aid for usability and human factors to provide reasonable assurance of the usability engineering process and to mitigate risks caused by usability problems per its intended use. The company performed a series of three iterative usability evaluations (Formative Studies 1, 2 and 3) during the development of the hearing aid and companion Concha Labs mobile app. Participants met the device criteria for an intended user. Key findings and device/labeling modifications were implemented to the user interface design in response to the studies. Summative Human factors were validated across two primary studies. - Human factors/usability validation test in which 15 representative users attempted all critical and essential tasks in the context of realistic use scenarios; - · Self-selection validation study in which 27 potential consumers determined if the hearing aid was appropriate for their needs based on pre-screening information (e.g., outside package labeling, etc.). The validation of human factors/usability aspects of the device included 15 representative users ranging in age from 25 – 76 (median age of 70) and included 10 females/5 males with 2/3 of all participants aged 60 and above. Five participants self-described a mild hearing loss and 10 indicated a moderate with 9/15 indicating no prior hearing aid experience. Assessment to determine if potential users can make a correct self-selection decision (suitability of Concha Sol for their personal use) based on their perceived level of hearing loss, presence or absence of contraindications after only having access to review Concha Labs website or product packaqing. After analysis of data per pre-specified 'success' outcomes it was determined that the Concha Sol Hearing Aid and Concha app are able to be used correctly by the intended users under the anticipated condition of use. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Concha Labs. The logo consists of a yellow conch shell-like symbol on the left, followed by the text "CONCHA LABS" in a sans-serif font. The text is in a dark gray or black color, contrasting with the white background. # Clinical Performance Testing The objective of the within-subject, randomized, crossover study design of the clinical investigation was to assess and evaluate the performance and clinical effectiveness of the Concha Sol self-fitting hearing aid, according to two distinct fitting methods. The first method involved users fitting the hearing aid devices themselves by relying exclusively on the provided iQS smartphone application (Concha Labs App), without any support from a hearing care professional (SELF-FIT). The second method used the same hearing devices, but it was clinically fitted by a hearing care professional following industry best practices (CLIN-FIT). ## STUDY PARTICIPANT DEMOGRAPHICS Demographically, the study participants included a representative mix of female/male, age, and prior experience with hearing aids. | SEX | NUMBER | |---------------------|--------| | MEN | 13 | | WOMEN | 8 | | AGE, | YEARS | | MEAN (SD) | 66.2 | | MEDIAN (IQR) | 73.0 | | HA EXPERIENCE | NUMBER | | INEXPERIENCED USERS | 17 | | EXPERIENCED USERS | 4 | Additionally, the 'indications for use' of the Concha Sol Hearing Aid stated the intended users have a perceived mild to moderate hearing loss. As part of the study protocol, audiometric measures were taken and indicated that their hearing was consistent with that target population (perceived mild-to-moderate) hearing status. ## STUDY ENDPOINTS AND SUCCESS CRITERIA ### Effectiveness: A standardized measure of speech communication in noise, the Quick Speech-in-Noise (QuickSIN) Test was used as the primary effectiveness endpoint. A secondary effectiveness endpoint, the Abbreviated Profile of Hearing Aid Benefit (APHAB) was used to evaluate the effect on hearing impairment or benefit in daily life. Note that for each intervention, endpoints were measured at the end of the two-week take-home field trial. Following previous studies investigating the effectiveness of self-fitting relative to clinician-fitting (including the predicate device study; Sabin et al., 2020), a margin of 1.8 dB was defined for the QuickSIN benefit, and a margin of 9.9 for the APHAB. For the clinical study to be deemed successful, the SELF-FIT method needs to yield outcomes (namely QuickSIN and APHAB scores) which are comparable to the CLIN-FIT method. For both assessed endpoints, the study results indicated comparable performance and satisfaction outcomes as revealed from clinical data as compared to the same hearing aid fit by a clinician. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for Concha Labs. The logo consists of a stylized yellow seashell on the left, followed by the words "CONCHA LABS" in a simple, sans-serif font. The text is in a dark gray or black color, providing a clear contrast against the white background. ## Safety: No device-related or other adverse events were observed over the course of the study. Also, no health-related issues due to device fit or discomfort were observed, and no device failures leading to potential safety issues were observed. These results support that the subject hearing aids are safe for its intended users. The clinical study has demonstrated that the Concha Sol hearing aids are comparably safe and effective for their intended use and are substantially equivalent to the Bose predicate device. The Concha Sol hearing aids have the same intended use and indications with similar technological characteristics and principles of operation as its predicate device and any differences raise no new questions of safety and effectiveness when compared to the predicate. ### Conclusion: The Concha Sol OTC Hearing Aid has the same intended use and indications for use compared to the predicate device with comparable technological characteristics and principles of operation that do not raise different questions of safety and effectiveness compared to the predicate device. Nonclinical performance testing demonstrated comparable safety and effectiveness with respect to electrical safety, EMC and electroacoustic testing as compared to the primary predicate, Bose SoundControl Hearing Aid (K211008). Similarly, software was developed, tested, and documented in accordance with software development and testing pursuant to IEC 62304:2006+A1:2015, FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices" (2005), and demonstrated that the Concha Sol Hearing Aid should perform as intended in the specified use conditions consistent with that reported by the predicate, the Bose SoundControl Hearing Aid. The usability testing and clinical study provided evidence of clinically meaningful benefit to a population experiencing obstacles to accessibility and affordability of hearing aids. The evidence of clinical effectiveness (verified both quantitatively) validates that the Concha Sol Hearing Aid performs comparably to the clinical effectiveness of the predicate device. The Concha Sol Hearing Aid is substantially equivalent in intended use to the Bose SoundControl Hearing Aid (K211008).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...