Non-inferiority to HCP-fit strategy on APHAB subscales
Non-inferior (p<0.001 for each subscale)
—
—
Pivotal clinical study: 28 subjects
—
Indications for Use
The Vibe SF self-fitting hearing aid is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. It is adjusted by the user to meet the user's hearing needs through software tools. The device is intended for direct-to-consumer sale and use without the assistance of a hearing care professional. The EasyFit self-fitting web application is intended to support self-fitting and fine-tuning of the Vibe SF hearing aid.
Device Story
Vibe SF is an in-the-ear (ITE) air-conduction hearing aid for self-fitting by users with mild-to-moderate hearing loss. System includes wireless hearing aids and EasyFit web application (mobile/tablet browser). Users perform acoustic profiling via the app to derive initial settings; fine-tuning is achieved by selecting improvement areas in the app, which generates new settings. Devices communicate via magnetic inductive ear-to-ear (e2e) link. Powered by zinc-air 10A batteries. Used in home environments; no professional involvement required. Output is amplified sound adjusted to user preference. Benefits include improved speech understanding and reduced communication difficulties in daily life.
Clinical Evidence
Single-arm randomized crossover study (n=28) comparing Vibe SF self-fitting to audiologist-fit (HCP) using the same hardware. Primary endpoint: APHAB subscales (Ease of Communication, Background Noise, Reverberant Room). Results: Vibe SF non-inferior to HCP fit (p<0.001). Secondary endpoints: QuickSIN, SSQ-12, and EMA showed no significant differences between fitting strategies. No adverse device effects reported.
Technological Characteristics
ITE form factor; zinc-air 1.3V 1.2mAh batteries; magnetic inductive e2e wireless link. Biocompatibility per ISO 10993-1, -5, -10. Software-based fitting via web app. Electroacoustic performance per ANSI/ASA S3.22. Non-sterile.
Indications for Use
Indicated for individuals 18+ with perceived mild to moderate hearing impairment. Intended for direct-to-consumer use without professional assistance.
Regulatory Classification
Identification
A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. A self-fitting air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable.
Special Controls
In combination with the general controls of the FD&C Act, the self-fitting air-conduction hearing aid is subject to the following special controls:
- 1. Clinical data must evaluate the effectiveness of the self-fitting strategy.
- 2. Electroacoustic parameters, including maximum output limits, distortion levels, selfgenerated noise levels, latency, and frequency response, must be specified and tested.
- 3. Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.
- 4. Software verification, validation, and hazard analysis must be performed.
- 5. If the device incorporates wireless technology:
- (A) Performance testing must validate safety of exposure to non-ionizing radiation;
- (B)Performance data must validate wireless technology functions; and
- (C) Labeling must specify instructions, warnings, and information relating to wireless technology and human exposure to non-ionizing radiation.
- 6. Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.
- 7. Patient labeling must include the following:
- (A)Information on how a patient can self-identify as a candidate for the device;
- (B) Information about when to seek professional help;
- (C) A warning about using hearing protection in loud environments;
- (D) A warning about staying alert to sounds around the user of the device;
- (E) Technical information about the device, including information about electromagnetic compatibility; and
- (F) Information on how to correctly use and maintain the device.
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical data must evaluate the effectiveness of the self-fitting strategy.
(2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested.
(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) If the device incorporates wireless technology:
(i) Performance testing must validate safety of exposure to non-ionizing radiation; and
(ii) Performance data must validate wireless technology functions.
(6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 12, 2022
WSAUD A/S % John Smith Partner Hogan Lovells US LLP Columbia Square, 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K220403
Trade/Device Name: Vibe SF Self-Fitting Hearing Aid Regulation Number: 21 CFR 874.3325 Regulation Name: Self-Fitting Air-Conduction Hearing Aid Regulatory Class: Class II Product Code: QDD Dated: July 14, 2022 Received: July 14, 2022
Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2 1 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K220403
Device Name
Vibe SF Self-Fitting Hearing Aid Indications for Use (Describe)
The Vibe SF seff-fitting hearing aid is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. It is adjusted by the user's hearing needs through software tools. The device is intended for direct-to-consumer sale and use without the assistance of a hearing care professional.
The EasyFit self-fitting web application is intended to support self-fitting and fine-tuning of the Vibe SF hearing aid.
Restricted Device (per 21 CFR 801.420 and CFR 801.421).
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### WSAUD A/S's Vibe SF Self-Fitting Hearing Aid
#### Submitter
WSAUD A/S Nymøllevej 6 DK-3540 Lynge Denmark Phone: 0045 21370938 Contact Person: Kristine Klitgaard Pedersen
Date Prepared: August 12, 2022
Name of Device: Vibe SF Self-Fitting Hearing Aid
Common or Usual Name: Hearing Aid
Classification Name: Self-Fitting Air-Conduction Hearing Aid (21 CFR 874.3325)
Regulatory Class: Class II
Product Code: QDD
| Predicate Device: | Bose Corporation | Bose Hearing Aid | DEN180026 |
|-------------------|------------------|------------------|------------------------------------------------|
| Reference Device: | Sivantos, Inc. | Signia Silk 1X | Class II 510(k)-Exempt<br>Wireless Hearing Aid |
#### Device Description
The Vibe SF Self-Fitting Hearing Aid ("Vibe SF") is a self-fitting air conduction hearing aid that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.
The Vibe SF hearing aid is based on WSAUD's wireless hearing aid series Signia Silk 1X (regulated under 21 C.F.R. § 874.3305). However, unlike standard wireless air-conduction hearing aids, the Vibe SF is intended to be sold directly to the end-user without involvement of a hearing care professional. Briefly, the self-fitting hearing aid system consists of two components:
- a) A wireless hearing aid with a self-fitting feature (Vibe SF); and
- b) A mobile web application (EasyFit) to support the self-fitting and fine-tuning of the hearing aid.
The Vibe SF wireless hearing aid hardware includes: Ear pieces (click sleeves), housing, circuit board, battery, chip, and electroacoustic components. The subject device is fitted for
{4}------------------------------------------------
bilateral use with a left and right device. Left and right hearing aids are able to communicate with each other over a magnetic inductive wireless link. The instant fit in-the-ear (ITE) style of the Vibe SF hearing aid uses Click Sleeves in 4 different sizes to couple with the ear canal. The hearing aids are marked with 'R' for right and 'L' for left, and are powered by standard (non-rechargeable) Zinc-Air batteries (size 10A). The Vibe SF hearing aid includes software that communicates with the EasyFit app to provide self-fitting functionality.
The Vibe SF is fitted by the user through the EasyFit self-fitting web application, which is run on an Internet browser on the user's mobile device (smartphone or tablet with iOS or Android operating system). The EasyFit self-fitting web application guides the user through the selffitting procedure and also allows the user to set audiological gain parameters and preferred settings on the Vibe SF Hearing Aid. In addition, a software application (app for iOS and Android) called Vibe app is available as remote control for adjusting volume during daily use.
# Indications for Use
The Vibe SF self-fitting hearing aid is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. It is adjusted by the user to meet the user's hearing needs through software tools. The device is intended for direct-to-consumer sale and use without the assistance of a hearing care professional.
The EasyFit self-fitting web application is intended to support self-fitting and fine-tuning of the Vibe SF hearing aid.
# Summary of Technological Characteristics
Both the subject (Vibe SF) and the predicate device (Bose Hearing Aid) are self-fit direct-toconsumer hearing aids indicated for individuals 18 and older with perceived mild to moderate hearing impairment. At a high level, the subject and predicate devices are based on the following same technological elements:
- Self-fit hearing aid.
- Home healthcare environment use. ●
- Software to support self-fitting and fine-tuning of the hearing aid. ●
- Software platform compatibility (iOS, Android). .
The following technological differences exist between the subject (Vibe SF) and predicate device (Bose Hearing Aid):
- . Form factor. Vibe SF has an in-the-ear design while Bose Hearing Aid has a head band design.
- Replaceable battery. Vibe SF uses zinc air batteries and is non-rechargeable. Bose Hearing Aid is rechargeable.
- Proprietary fitting algorithm for self-fitting. .
- . Wireless technologies. Vibe SF has a proprietary e2e (ear-to-ear) wireless technology. Bose Hearing Aid uses Bluetooth.
A table comparing the key features of the subject and predicate devices is provided below.
{5}------------------------------------------------
| | Subject Device:<br>Vibe SF Self-Fitting<br>Hearing Aid (K220403) | Predicate Device:<br>BOSE® Hearing Aid<br>(DEN180026) | Discussion |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | WSAUD A/S | Bose Corporation | Different legal entities |
| FDA Product<br>Code | QDD | QDD | Same |
| Classification<br>Regulation | 21 CFR 874.3325 | 21 CFR 874.3325 | Same |
| Indication for<br>use | The Vibe SF self-fitting<br>hearing aid is intended to<br>amplify sound for<br>individuals 18 years of age<br>or older with perceived<br>mild to moderate hearing<br>impairment. It is adjusted<br>by the user to meet the<br>user's hearing needs<br>through software. The<br>device is intended for<br>direct-to-consumer sale<br>and use without the<br>assistance of a hearing<br>care professional. | The Bose Hearing Aid is<br>intended to amplify sound<br>for individuals 18 years of<br>age or older with perceived<br>mild to moderate hearing<br>impairment. It is adjusted<br>by the user to meet the<br>user's hearing needs. No<br>pre-programming or<br>hearing test is necessary.<br>The device is intended for<br>direct-to-consumer sale<br>and use without the<br>assistance of a hearing<br>care professional. | Same intended use.<br>Nearly identical indication<br>for use. |
| Physical<br>Description | In-The-Ear (ITE) | Flexible neckband housing<br>Earbuds are connected to<br>the neckband by flexible<br>wires and on each earbud<br>is mounted a Bose<br>StayHear+ eartip; three<br>sizes of tips are available. | Different form factor.<br>Biological safety and<br>human factors testing<br>support substantial<br>equivalence. |
| Maximum<br>Output<br>Characteristics | Less than or equal to 114<br>dB SPL (2 cc coupler) | Less than or equal to 120<br>dB SPL | Adequate for fitting<br>moderate hearing loss<br>(55 dBHL) as prescribed<br>by NAL-NL2 |
| HFA-FOG | 45 dB | 43 dB | Adequate for fitting<br>moderate hearing loss<br>(55 dBHL) as prescribed<br>by NAL-NL2 |
| Power Source | Powered by a zinc-air 1.3<br>V, 1.2 mAh battery | Powered by a<br>rechargeable 3.7 V, 250<br>mAh Li-ion battery pack | Predicate was<br>rechargeable; subject<br>device uses off-the-shelf<br>disposable hearing aid<br>batteries. Acoustic<br>measurements support<br>substantial equivalence. |
| Labelling | Includes applicable<br>requirements of 21 CFR<br>801.420 and 801.421. | Includes applicable<br>requirements of 21 CFR<br>801.420 and 801.421. | Same |
| Reprocessing | Non-sterile. The hearing | Non-sterile. The Earbud | Same |
| | Subject Device:<br>Vibe SF Self-Fitting<br>Hearing Aid (K220403) | Predicate Device:<br>BOSE® Hearing Aid<br>(DEN180026) | Discussion |
| | sleeve) may be wiped daily<br>with a soft, dry tissue. | may be wiped with a soft,<br>dry cloth. StayHear+ tips<br>may be rinsed with warm<br>water and thoroughly dried<br>before attaching them to<br>the earbuds. | |
| Biological<br>evaluation | A biological evaluation<br>covering all components<br>with skin contact has been<br>conducted per ISO 10993-<br>1, ISO 10993-5 and ISO<br>10993-10. | Biocompatibility tested:<br>MEM elution Cytotoxicity<br>per ISO 10993-5 and<br>Irritation and Skin<br>sensitization per ISO<br>10993-10. | Same |
| Self-fitting<br>strategy | For the initial setup, users<br>run an acoustic profiling<br>test using the hearing aid<br>devices. The initial setting<br>is selected based on the<br>identified profile. The user<br>adjusts the setting for<br>comfortable loudness<br>impression.<br>For the fine tuning the user<br>can select from some pre-<br>defined improvement areas.<br>The application running on<br>a mobile phone web<br>browser uses the current<br>setting and the desired<br>improvement to generate a<br>new setting. The user is<br>able to acoustically<br>compare both settings and<br>select the one which is<br>preferred for further use. | Users can control and<br>customize signal<br>processing parameters to<br>their hearing needs via the<br>Bose user interface, which<br>consists of two Dimension-<br>Reduced Controllers<br>(DRCs) ("Loudness" and<br>"Fine Tuning")<br>implemented in an<br>application running on a<br>mobile device. | Both use proprietary<br>algorithms for self-fitting.<br>Clinical testing supports<br>substantial equivalence. |
| 21 CFR<br>874.3325<br>Special controls | Includes applicable<br>requirements for special<br>controls for 21 CFR<br>874.3325 | Includes applicable<br>requirements for special<br>controls for 21 CFR<br>874.3325 | Same |
| 21 CFR<br>874.3305<br>Special controls | Includes applicable<br>requirements for special<br>controls for 21 CFR<br>874.3305 | Includes applicable<br>requirements for special<br>controls for 21 CFR<br>874.3305 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# Performance Data
# Non-Clinical Testing Summary
Performance testing was conducted to demonstrate that the Vibe SF Self-Fitting Hearing Aid is as safe and effective as the Bose Hearing Aid. WSAUD conducted a series of non-clinical tests on the Vibe SF to assess electrical safety. EMC, wireless coexistence, electroacoustic performance, usability, biocompatibility, and software. The results are summarized in the table below:
| Test Standard/Method | Test Purpose/Description | Result |
|----------------------------------------|-------------------------------|--------|
| ISO 10993-1:2018 | Biocompatibility | Pass |
| ISO 10993-5:2009 | | Pass |
| ISO 10993-1:2010 | | Pass |
| IEC 62304:2006+A1:2015 | Software | Pass |
| IEC 60601-1:2005/AMD1:2012/AMD2:2020 | Electrical Safety | Pass |
| IEC 60601-2-66:2019 | | |
| IEC 60601-1-11:2015/AMD1:2020 | | |
| IEC 60601-1-2:2014/AMD1:2020 | Electromagnetic Compatibility | Pass |
| AAMI TIR 69:2017 | Wireless Coexistence | Pass |
| ANSI C63.27:2017 | | |
| ANSI/ASA S3.22:2014 | Electroacoustic | Pass |
| ANSI/CTA 2051:2017 | | |
| IEC 60601-1-6:2010/AMD1:2013/AMD2:2020 | Usability Engineering | Pass |
| IEC 62366-1:2015/AMD1:2020 | | |
In all instances, the Vibe SF Self-Fitting Hearing Aid functioned as intended and results observed were as expected.
As noted in the above table, a human factors validation test was performed to demonstrate that the Vibe SF can be used safely and effectively, per its labeling, by representative intended users, for the intended uses and in the intended use environments. The results indicated that any remaining residual use-related risks were acceptable and were outweighed by the benefits derived from use of the device.
Labeling mitigations were implemented on the Vibe SF box labeling and in the Vibe SF Instructions for Use ("IFU") to help potential users with the self-selection process. These included clarification of the intended users of the device, identifications in which the device may and may not be appropriate, and informing users of the device's limitations and the value of a professional hearing loss evaluation.
# Clinical Testing Summary
WSAUD performed a pivotal clinical study to assess the patient-reported benefit from using the self-fitted Vibe SF hearing aid versus the same form factor hearing aid fitted by a hearing care professional ("HCP"). The study was designed as a Single-arm randomized crossover study with patients serving as their own control. It was blinded for some of the secondary endpoints. There was 1 study site in the United States.
{8}------------------------------------------------
Primary Effectiveness Endpoint: The primary effective of the study was to demonstrate that the Vibe self-fitting (SF) strategy is non-inferior to the audiologist-fit (HCP fit) strategy in subject's perceived hearing aid benefit after using the Vibe SF (investigational device) and Silk 1X HCP fit (comparator device) hearing aids in real-life conditions. Hearing aid benefit was measured by 3 sub-scales of the Abbreviated Profile of Hearing Aid Benefit (APHAB): ease of communication ("EC"), [ability to hear in] background noise ("BN"), and [ability to hear in] a reverberant room ("RV"). The primary endpoint result was comprised of the combined results from these subscales; success required simultaneous non-inferiority of the Vibe SF as compared to HCP fit hearing aid on all three subscales.
Primary Safety Endpoint: The safety endpoint was the rate of adverse device effects.
Patient Population: The devices were tested on 28 subjects. Table 1 shows the baseline demographics of the study subjects. All subjects reported a self-perceived hearing loss, with 71.4% describing their hearing loss with having "A little trouble" and 28.6% with having "A lot of trouble." 4 of the 28 subjects were experienced hearing aid users with a mean of 33 months of hearing aid use.
| Characteristic | Summary |
|-----------------------------------------|----------------|
| Gender | |
| Female | 32.1% (9/28) |
| Male | 67.9% (19/28) |
| Age Group | |
| 25-34 | 3.6% (1/28) |
| 35-44 | 7.1% (2/28) |
| 45-54 | 14.3% (4/28) |
| 55-64 | 25.0% (7/28) |
| 65-74 | 28.6% (8/28) |
| 75-84 | 21.4% (6/28) |
| Ethnicity | |
| Not Hispanic/Latino | 100.0% (28/28) |
| Race | |
| American Indian or Alaska Native | 3.6% (1/28) |
| American Indian or Alaska Native; Other | 3.6% (1/28) |
| White | 92.9% (26/28) |
Table 1. Demographics of the study subjects
The average baseline air conduction audiograms are very similar to the results obtained in a published study with the predicate device (compare Figure 1, which shows the average air conduction audiograms of all participants of the Vibe SF study, and Figure 2 copied from Sabin et al. 20201).
<sup>1</sup> A. V. Sabin, D. J. Van Tassel, B. Rabinowitz and S. Dhar (2020). Validation of a self-fitting method for over-thecounter hearing aids. Trends in Hearing, 1-19.
{9}------------------------------------------------
Image /page/9/Figure/1 description: This image is a graph showing hearing threshold in decibels (dB HL) as a function of frequency in hertz (Hz). The x-axis represents frequency, with values ranging from 250 Hz to 8k Hz. The y-axis represents hearing threshold, with values ranging from -10 dB HL to 100 dB HL. The graph shows that hearing threshold increases as frequency increases, with the lowest threshold at 250 Hz and the highest threshold at 8k Hz.
Figure 1. Average air conduction audiograms for participants in the Vibe SF Study. Error bars reflect standard deviation across all ears.
Image /page/9/Figure/3 description: This image shows a graph comparing the air conduction threshold in dB HL for two groups, labeled 'ABP Group' and 'Self Group', across different frequencies in kHz. The x-axis represents the frequency in kHz, ranging from 0.25 to 8, while the y-axis represents the air conduction threshold in dB HL, ranging from -10 to 100. Both groups show a decrease in air conduction threshold as the frequency increases, indicating better hearing at higher frequencies. Error bars are included to show the standard deviation.
Figure 2. From Sabin et al. 2020.1 Average air conduction audiograms for participants in the ABP (squares) and Self (circles) groups. Error bars reflect standard deviation across all ears in the group. ABP = Audiologist Best Practices.
| | | | Table 2. Patient Accountability |
|--|--|--|---------------------------------|
|--|--|--|---------------------------------|
| Stage | Investigational<br>Device Arm Total | Control<br>Arm Total | Total |
|-----------------------------------------|-------------------------------------|----------------------|-------|
| Enrollment | 28 | 28* | 28 |
| Treatment started | 14 | 14 | 28 |
| Treatment finished | 13 | 13 | 26 |
| Primary Safety Endpoint Analysis | 28 | 28* | 28 |
| Primary Effectiveness Endpoint Analysis | 28 | 28* | 28** |
* Cross-over design, patients are the same ones as for investigational device
** 2 subjects had missing HCP fit and Vibe data due to cause but unaided data was collected. Imputed '0' for difference of benefits for those 2 subjects for the intent to treat (ITT) summary.
{10}------------------------------------------------
# Results:
The study met the primary endpoint. The primary effectiveness endpoint results demonstrated that the Vibe SF strategy is non-inferior to the HCP fit strategy in participants' perceived hearing aid benefit. The Vibe SF strategy was non-inferior to the HCP fit strategy in each of the 3 benefit scores for each subscale in the APHAB (p<0.001 for each subscale/coprimary endpoint). Therefore, the Vibe SF strategy demonstrated non-inferiority to the HCP fit strategy (combined p-value <0.001).
As can be seen in Figure 3 below, there is a benefit (positive mean and median values) for both fitting strategies in all domains, suggesting that participants had fewer communication problems with the devices on average. The improvements relative to the unaided condition were significant, but the differences between the HCP fit and Vibe SF devices was not significant. For all subscales, the differences in perceived mean benefit between the two fitting strateqies were very small and significantly lower than the non-inferiority marqins. This means that there should be no clinical difference between the two fitting strategies.
Image /page/10/Figure/4 description: The image is a boxplot comparing APHAB benefit for HCP fit and Vibe SF across three conditions: EC, BN, and RV. The y-axis represents APHAB benefit, ranging from -80 to 80. For each condition, there are two boxplots side by side, one for HCP fit and one for Vibe SF, allowing for a visual comparison of the distributions of APHAB benefit between the two fitting methods.
Figure 3. Results of the APHAB Benefit shown for EC, BN and RV. Boxplots show the mean, median, interquartile range and minimum/maximum. Outliers are indicated with a dot.
To be able to compare the data to the results obtained with the predicate device (Sabin et al. 20201), the global APHAB scores were calculated. The comparison can be seen in Table 3 below. The baseline number of problems is slightly higher in the study population of the Vibe SF Clinical study, but the benefit with the hearing aid is very similar.
| Table 3. Results of APHAB and SSQ Questionnaires, comparison to the data shown in Sabin |
|-----------------------------------------------------------------------------------------|
| et al. 2020¹ for the predicate device |
| | APHAB | | | | SSQ | |
|---------|-----------|---------|---------|---------|-----------|---------|
| | predicate | | actual | | predicate | actual |
| | Unaided | Benefit | Unaided | Benefit | Benefit | Benefit |
| HCP Fit | | | | | | |
| Mean | 31 | 11.94 | 37.53 | 12.63 | 0.91 | 1.03 |
| SD | 15.22 | 15.14 | 15.37 | 14.05 | 1.93 | 1.4 |
| Min | 12 | -9 | 14.67 | -13.11 | -7.5 | -2.33 |
| Max | 82 | 58 | 71.83 | 37.83 | 4.6 | 3.67 |
{11}------------------------------------------------
| | APHAB | | | | SSQ | |
|------|-----------|---------|---------|---------|-----------|---------|
| | predicate | | actual | | predicate | actual |
| | Unaided | Benefit | Unaided | Benefit | Benefit | Benefit |
| N | 34 | 34 | 28 | 26 | 37 | 26 |
| Self | | | | | | |
| Mean | 31.67 | 14.57 | 37.53 | 11.77 | 1.21 | 1.15 |
| SD | 16.02 | 18.31 | 15.37 | 19.26 | 1.7 | 1.63 |
| Min | 7 | -27 | 14.67 | -19.94 | -1.8 | -2.83 |
| Max | 67 | 54 | 71.83 | 61.44 | 5.3 | 3.67 |
| N | 30 | 30 | 28 | 26 | 28 | 26 |
# Secondary Effectiveness Endpoint Results:
EMA Scores: Ecological Momentary Assessment (EMA) scores were similar for the self-fit strategy and HCP fit strategy in each listening situation. Both fitting strategies provided good sound quality and high speech understanding. Sound quality was rated fair to excellent in over 90% of the questionnaires. Median speech understanding was 8 or above on a scale of 0 – 10 (10 signifying "understand every word"), and satisfaction was neutral or better in over 90% of the questionnaires, with 65.5% being satisfied or very satisfied during the HCP fit trial and 75.5% being satisfied or very satisfied during the Vibe SF trial.
QuickSIN Scores: The mean and median results of the QuickSIN test for the HCP fit and for self-fit strategies were very similar, with no significant differences observed (Figure 4). With the devices, programmed to either fitting strategy, on average the participants performed in what would be considered the "normal/near normal" category for SNR-Loss (0-3 dB SNR Loss; e.g., only slightly poorer than someone with normal hearing). The QuickSIN was administered at what was considered to be an ecologically appropriate level of 50 dB HL (~62 dB SPL). At this level, slightly above average speech, we would not expect to see differences between the fitting strategies given the small differences between aided ear canal SPL, and the fact that the participants had mild-to-moderate hearing losses.
Image /page/11/Figure/5 description: The image is a boxplot comparing QuickSiN [dB SNR loss] for unaided, HCP fit, and Vibe SF conditions. The y-axis represents QuickSiN loss, ranging from -4 to 10. The boxplots show the distribution of QuickSiN loss for each condition, with the median marked by an 'X'. The HCP fit condition appears to have the highest median QuickSiN loss, while the unaided condition has the lowest.
Figure 4. QuickSIN results for unaided, HCP fit and Vibe SF, speech presented at 50 dB HL. Boxplots showing the mean ("x"), median, interquartile range and minimum/maximum.
SSQ-12 Scores: In all of the aspects of hearing assessed by the SSQ-12 (domains included following speech, location and distance hearing, qualities of speech), results showed a benefit compared to unaided hearing for both fitting strategies. There was virtually no difference in
{12}------------------------------------------------
SSQ-12 scores between the HCP fit and the self-fit results. As can be seen in Table 1 (above with APHAB), SSQ-12 Benefit was also very similar to that observed with the predicate device.
Gain Selection: Real ear measurements of both the self-fit and the HCP fit devices showed similar amplification across all measured frequencies, consistent with the National Acoustic Laboratories' Nonlinear Version 2 (NAL-NL2) targets. The HCP-adjusted devices provided slightly greater amplification at higher frequencies. These results are similar to those reported for the predicate device in Sabin et al. 2020, 1 where the self-selected settings correlated with, but were lower than, the HCP-selected settings.
It is also noteworthy that the NAL-NL2 algorithm offers target calculation corrections for new users. 24 of the 28 participants in the study were new users. Several of them had hearing losses to the degree that their fitting targets would have been lowered by hearing care professionals by 2-3 dB, or maybe more. Hence, the real-life mismatch from the NAL-NL2 targets for the SF approach shown in our mean data would not be quite as large as suggested. Furthermore, the Vibe SF hearing aid does provide for user adjustments to increase or decrease gain. Most appropriate for matching to NAL-NL2 targets are adjustment options including overall master gain adjustment for the entire frequency response as well as adjustments for speech intelligibility.
Image /page/12/Figure/4 description: The image contains two line graphs comparing frequency in Hz to decibels. The left graph compares "Target NAL-NL2" and "REAR Vibe SF" in terms of "Rear Ear Target / REAR 65 [dB SPL]" from 250 Hz to 8k Hz. The "Target NAL-NL2" line ranges from approximately 58 dB to 50 dB, while the "REAR Vibe SF" line ranges from approximately 58 dB to 42 dB. The right graph shows the "Difference from Target [dB]" from 250 Hz to 8k Hz, ranging from approximately 4 dB to -8 dB.
Figure 5. (Left) NAL-NL2 Tarqet and measured REAR (real ear aided response) for medium input averaged across all subjects for Vibe SF. (Right) Average and standard deviation of the difference between NAL-NL2 target and measured responses computed across all subjects.
Fitting Preference: An equal number of participants preferred using the HCP fit hearing aids as the self-fit hearing aids (each around 40%). The remaining 20% of participants had no preference between the two fitting strategies.
Safety Outcomes: No adverse device effects were reported during the study. Two adverse events were reported which were not device or study procedure related. One event was reported as a food-related stomach issue and the other, a cold virus. Both were non-device related and resolved without medical intervention. Additionally, one use error was reported but was resolved without medical incident or untoward clinical signs; thus, it was not recorded as an adverse device effect.
{13}------------------------------------------------
Summary: This study evaluated the effectiveness of the Vibe self-fitting (SF) strategy for the Vibe SF hearing aids as measured by perceived hearing aid benefit. The data showed strong statistical evidence (p<0.001) that the Vibe SF strategy was non-inferior to an HCP fit strategy that used recognized prescriptive targets from the NAL-NL2. Therefore, this study demonstrates the clinical effectiveness of the Vibe SF strategy. Furthermore, the results of the primary and secondary endpoints are similar to those reported for the predicate device (ABHAB, SSQ, QuickSIN). No adverse device effects were reported during the study.
#### Conclusion
The Vibe SF Self-Fitting Hearing Aid is as safe and effective as the predicate Bose Hearing Aid. The Vibe SF Self-Fitting Hearing Aid has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. In addition, the minor technological differences between the Vibe SF Self-Fitting Hearing Aid and its predicate device raise no new issues of safety or effectiveness. Thus, the Vibe SF Self-Fitting Hearing Aid is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.