The EarLens™ Contact Hearing Device transmits amplified sound by vibrating the eardrum through direct contact. It is indicated for individuals 18 years and older with a mild to severe sensorineural hearing impairment who can benefit from amplification. The device can provide the full spectrum of amplification that includes 125 Hz - 10,000 Hz.
Device Story
EarLens™ CHD is a tympanic membrane contact hearing aid for adults with mild-to-severe sensorineural hearing loss. System comprises an external Audio Processor (BTE unit + Ear Tip) and a patient-matched Tympanic Membrane Transducer (TMT). BTE captures sound via microphones, processes it via DSP, and transmits signals/power wirelessly to the TMT using a Class 1 laser emitter in the Ear Tip. TMT, placed on the umbo by a physician, converts light signals into mechanical vibrations, directly stimulating the eardrum. Used in home/clinic settings; programmed by hearing professionals using EarLens™ Fitting Software (ELF). Output provides amplification across 125 Hz–10,000 Hz. Benefits include improved speech recognition and functional gain, particularly in high frequencies. Device is MR unsafe; requires professional fitting and deep ear canal impression.
Clinical Evidence
Prospective, single-arm, open-label study (n=48 subjects, 96 ears) evaluated safety and effectiveness over 4 months. Primary effectiveness endpoint: mean word recognition (NU-6) improved by 33% (p<0.0001) at soft levels. Secondary endpoint: mean functional gain of 30.5 dB (p<0.0001) from 2000-10000 Hz. Safety: no decrease in hearing sensitivity >10 dB observed. Adverse events (n=31) were primarily minor (abrasions, blood blisters) and resolved. Pilot study (n=5) confirmed <1°C temperature rise in ear canal after 8 hours of use. Bench testing validated mechanical integrity, electrical safety, and EMC.
Technological Characteristics
Tympanic membrane contact hearing aid; transducer-based direct vibration of the tympanic membrane. Requires biocompatible patient-contacting components. Includes software for signal processing. Subject to non-clinical performance testing for mechanical integrity, electrical/thermal safety, and electromagnetic compatibility. Frequency range: 125 Hz – 10,000 Hz.
Indications for Use
Indicated for individuals 18+ years with mild to severe sensorineural hearing impairment. Contraindicated for patients with abnormal tympanic membrane, middle ear abnormalities, history of chronic ear infections (past 24 months), rapidly progressive/fluctuating hearing loss, or compromised immune system/recent head radiation or chemotherapy.
Regulatory Classification
Identification
A tympanic membrane contact hearing aid is a prescription wearable device that compensates for impaired hearing. Amplified sound is transmitted by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane. A tympanic membrane contact hearing aid is subject to the requirements in § 801.422 of this chapter.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, and must include:
(i) Mechanical integrity testing;
(ii) Electrical and thermal safety testing;
(iii) Software verification, validation, and hazard analysis;
(iv) Reliability testing consistent with expected device life;
(v) Electromagnetic compatibility testing; and
(vi) Validation testing of device output and mechanical force applied to the tympanic membrane in a clinically appropriate model.
(3) Clinical performance testing must characterize any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
(4) Professional training must include the ear impression procedure, correct placement, fitting, monitoring, care, and maintenance of the device.
(5) Labeling must include the following:
(i) A detailed summary of the adverse events and effectiveness outcomes from the clinical performance testing;
(ii) Detailed instructions on how to fit the device to the patient;
(iii) Instructions for periodic cleaning of any reusable components;
(iv) Information related to electromagnetic compatibility; and
(v) Patient labeling that includes:
(A) A patient card that identifies if a patient has been fitted with any non-self- removable components of the device and provides relevant information in cases of emergency;
(B) Information on how to correctly use and maintain the device;
(C) The potential risks and benefits associated with the use of the device; and
(D) Alternative treatments.
In combination with the general controls of the FD&C Act, the tympanic membrane contact hearing aid is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR EARLENS™ CONTACT HEARING DEVICE (CHD)
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Tympanic membrane contact hearing aid: A tympanic membrane contact hearing aid is a prescription device that compensates for impaired hearing. Amplified sound is transmitted by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.
NEW REGULATION NUMBER: 874.3315
CLASSIFICATION: CLASS II
PRODUCT CODE: PLK
#### BACKGROUND
DEVICE NAME: EARLENS™ CONTACT HEARING DEVICE (CHD)
SUBMISSION NUMBER: DEN150002
DATE OF DE NOVO: JANUARY 6, 2015
CONTACT: EARLENS™ CORPORATION 4045A Campbell Avenue MENLO PARK, CA 94025
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
The EarLens™ Contact Hearing Device transmits amplified sound by vibrating the eardrum through direct contact. It is indicated for individuals 18 years and older with a mild to severe sensorineural hearing impairment who can benefit from amplification. The device can provide the full spectrum of amplification that includes 125 Hz - 10,000 Hz.
#### LIMITATIONS
Prescription Use only: Federal (USA) law restricts this device to sale by or on the order of a physician.
Limitations on device use are included in the Instructions for Use as Contraindications. Warnings, and Precautions
{1}------------------------------------------------
#### Contraindications
The Patient must not have known or active medical issues that would preclude having a hearing device, including:
- an abnormal tympanic membrane (TM) (deemed perforated, inflamed or has a . dimeric or monomeric area, or in any other way abnormal);
- an abnormal middle ear or a history of prior middle ear surgery other than . tympanostomy tubes:
- an ear canal anatomy that prevents the physician from seeing an adequate amount of the tympanic membrane;
- an anatomical configuration of the external auditory canal that prevents satisfactory ● placement of the Tympanic Membrane Transducer (TMT);
- a history of chronic and recurrent ear infections in the past 24 months; ●
- a rapidly progressive or fluctuating hearing impairment;
- diagnosed with having a compromised immune system which may impact the tissue ● of the auricle or ear canal, such as keratosis obturans, ichthyosis, eczema of the auricle or ear canal, or received radiation of the head ever or chemotherapy for cancer within the last six years.
### Warnings
Before using the EarLens™ CHD, read and make sure you understand each of the following safety warnings:
- . The EarLens™ Contact Hearing Device is considered MR unsafe. The CHD should be removed prior to an MRI exam or MRI exposure.
- . Do not use therapeutic or medical diathermy using electromagnetic radiation (magnetic induction coils or microwave) from the shoulders up with the EarLens™ CHD in place.
- . The Behind-the-Ear (BTE) Unit and Ear Tip contain a Class 1 laser product. It is safe to use under normal operating conditions. The Class I laser light is NOT visible. DO NOT look directly into the Laser or aim it directly into your eye. Should the device become damaged, stop using the device and contact your hearing professional.
- If you experience discomfort or pain in your ear, contact your ENT physician ● immediately.
- . Do not insert foreign objects into the ear, such as O-tips, bobby pins or fingernails. Insertion of foreign objects could result in pain and damage to the ear, damage to the TMT or cause it to operate improperly.
- . Contact your hearing professional if you experience discharge from the ear or persistent discomfort or any other problems.
- Should the BTE become unusually warm or hot, promptly remove the BTE and . discontinue use. Contact your hearing healthcare professional.
{2}------------------------------------------------
## Precautions
- Individuals with known nickel sensitivity/allergy should be informed that the TMT component contains nickel that is coated with a parylene barrier. If an allergic reaction develops, the TMT should be promptly removed.
- The EarLens™ CHD is a patient-matched design and intended to be used for a single . patient.
- You may shower, bathe or swim with the TMT in place. Ear plugs may be used to prevent water from entering the ears so long as care is taken to not over-insert them. Removing water from your ears may be more difficult with the TMT in place.
- . Avoid getting the BTE unit wet, as this may damage the device. Remove the BTE unit prior to showering, swimming, or bathing.
- . You may experience a reduction in your hearing levels when the TMT is in place but the BTE is not activated.
- . Do not direct streams of liquid (i.e. isopropyl alcohol. hydrogen peroxide. DeBrox) into your ears, as this may cause the TMT to become dislodged or cause damage to the devices.
- Failure to oil the ear canal weekly may result in TMT displacement.
- If the EarLens™ BTE fails to operate or if it appears damaged, including the presence of battery leakage or swelling, promptly remove the BTE and discontinue use. Contact your hearing professional.
- . Electromagnetic fields produced by other electrical equipment such as cell phones. metal detectors microwaves, RFID systems and commercial theft detection systems (also known as electronic article surveillance [EAS]) may interfere with the CHD. In the event that the patient perceives unexpected noise or interference in the presence of these devices, move away from the source to mitigate the potential interference. Remove the BTE and if you have further concerns, contact your hearing healthcare professional.
### PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The EarLens™ Contact Hearing Device (CHD) transmits amplified sound to compensate for hearing impairment by direct vibration of the tympanic membrane (eardrum). The EarLens™ CHD is composed of an external Audio Processor, which includes a Behind-the-Ear (BTE) Unit and an Ear Tip, a Tympanic Membrane Transducer (TMT), the EarLens™ Fitting Software (ELF), and a Charger with a Power Adapter (Figure 1). In this device, light is used to wirelessly transmit both signal and power from the Audio Processor to the TMT. The BTE sits behind the outer ear, housing the rechargeable battery, digital signal processor (DSP), microphones and drive electronics. The Ear Tip contains the light emitter and directs the light signal down the ear canal.
The TMT resides at the end of the ear canal on the skin around the tympanic membrane. The TMT receives the light signal and converts it into direct vibration of the umbo of the tympanic
{3}------------------------------------------------
membrane. The EarLens™ Charger charges two BTEs at the same time when connected to either the wall power adapter or from the internal battery contained in the Charger. The CHD is patientmatched for single patient use. The ELF enables the hearing professional to program the device specific to the patient's hearing needs. The EarLens™ Impression System is provided to the physician to enable the collection of a deep ear canal impression to create patient-matched TMTs for each patient.
Image /page/3/Picture/1 description: This image shows a diagram of the inner ear. The diagram includes labels for the ear tip and TMT. The diagram also shows a BTE hearing aid.
Figure 1: The EarLens™ Contact Hearing Device
# Tympanic Membrane Transducer
The Tympanic Membrane Transducer (TMT) is designed to receive light signals from the Ear Tip and convert the light signals into mechanical vibrations of the tympanic membrane (Figure 2). The TMT is matched for each patient and is placed into position by a physician using the Grasping Tab. It is positioned in the ear at the end of the ear canal on the skin around the tympanic membrane. The Bias Springs, the patient-matched Sulcus Platform, and the Chassis stabilize the TMT in the ear canal, enabling the Umbo Platform to remain in contact with the umbo area of the tympanic membrane. The Photodetector captures the light signals and converts the light energy into current, which activates the Microactuator. The Microactuator vibrates the Umbo Platform, which vibrates the umbo of the tympanic membrane, causing sound to be perceived.
{4}------------------------------------------------
Image /page/4/Figure/0 description: The image shows a device with several labeled components. The components labeled are a photodetector, grasping tab, microactuator, bias spring, chassis, umbo platform, and sulcus platform. The device appears to be a small, possibly medical, device. The device has a gray and white color scheme.
#### Figure 2: Tympanic Membrane Transducer
### The Audio Processor
The Audio Processor consists of the Behind-the-Ear Unit (BTE) and the Ear Tip. The BTE is directly connected to the Ear Tip via the Ear Tube (Figure 3). The BTE is designed to pick up sounds via the microphone, apply signal processing, and transmit the electrical signal via the Ear Tube to the Ear Tip. The BTE can be removed, replaced, and recharged by the patient. The BTE Case contains two microphones, a digital signal processor, light drive electronics, a Program Selection Button, and a rechargeable battery. The battery is recharged by connecting the Charging Ports to the EarLens™ Charger. The Programming Port enables a hearing professional to program the BTE. The Ear Tube connects the BTE case to the
Image /page/4/Figure/4 description: This image is a diagram of an audio processor. The diagram labels the different parts of the audio processor, including the microphones, ear tip, BTE (behind-the-ear), ear tube, and program selection button. The audio processor is a small electronic device that is worn behind the ear and is used to amplify sound for people with hearing loss.
Ear Tip and can be adjusted by a hearing professional. The Ear Tip receives an electrical signal from the BTE and converts this to a light signal using a light emitter. The Ear Tip is a shell with a large opening or vent. The Ear Tip also stabilizes and aims the light emitter at the Photodetector of the TMT.
### The EarLens™ Charger
The EarLens™ Charger is designed to recharge the BTEs and act as a storage and travel case. The Charger incorporates Case Wings, BTE Charging Slots, Charging Status LEDs and an AC Adapter Port. When connected to the wall power adapter, the Charger houses and charges either a single BTE or two BTEs simultaneously. If desired, the Charger can be unplugged from the wall and used as a travel case. The Charger incorporates an internal battery, which will charge the BTEs when the Charger is unplugged. An AC wall power adapter is included.
{5}------------------------------------------------
#### EarLens™ Fitting Software (ELF)
The EarLens™ Fitting Software (ELF) is used to program the BTE. The ELF enables the hearing professional to calibrate and program the device specific to the patient's needs with up to four programs. In addition to the ELF software, the following devices are required (Figure 4):
- . Personal Computer
- . HiPro 2 Box with PC communication cable
- . Left/Right programming cables for connecting the hearing devices to the HiPro 2
#### EarLens™ Impression System
Figure 4: EarLens™ Fitting Accessories
Image /page/5/Picture/7 description: The image shows a laptop, a CD, and a device labeled "5x faster". The laptop is black and has a Dell logo on it. The CD is silver and is lying on the laptop's keyboard area. The "5x faster" device is connected to the laptop with several wires.
The EarLens™ Impression System is used by the physician to collect a deep ear canal impression. The impression is used to manufacture the patient-matched components of the EarLens™ CHD. Mineral oil (White Mineral Oil. Food Grade) is used to provide the surface tension forces to keep the TMT in position.
The EarLens™ Impression System includes:
- Reusable dispenser .
- . Lubricating mineral oil (single patient use)
- EarLens™ Impression Material 50 mL dual cartridge (contains enough material for . approximately three impressions)
- . Mixing tip with stainless steel extension (single use)
- Impression Container (single use)
The EarLens™ Impression System is regulated under 21 CFR Part 874.3300, product code LDG, which is Class I exempt. However, given the risks associated with the deep ear impression procedure, this procedure was assessed to ensure that it can generate reasonably safe and accurate ear impressions that resulted in appropriately fitted EarLens™ CHD device components (TMT and Ear Tip).
### SUMMARY OF NONCLINICAL/BENCH STUDIES
Non-clinical/bench studies conducted on the EarLens™ CHD to demonstrate a reasonable assurance of safety and effectiveness of the device are summarized in the sections below.
# BIOCOMPATIBILITY / MATERIALS
The EarLens™ CHD contacts the patient's intact skin surfaces and is therefore characterized as a skin surface contacting device. The TMT is categorized as a permanent contact device as the device is intended to be used more than 30 days. The Audio Processor and Charger are categorized as temporary contact devices intended to be used less than 24 hours at a time. In accordance with ISO 10993-1: Biological Evaluation of
{6}------------------------------------------------
Medical Devices. Part 1: Evaluation and Testing within a Risk Management Process, the following biocompatibility testing was conducted on the EarLens™ CHD. The biocompatibility assessment was deemed adequate.
| Component Tested | Test Requirement | Test Results |
|--------------------------------------------------------|------------------------------------------------|--------------|
| EarLens™ CHD System<br>(TMT, BTE, Ear Tip) and Charger | MEM elution<br>Cytotoxicity per ISO<br>10993-5 | Pass |
| EarLens™ Ear Tip (soft) | | Pass |
| EarLens™ System<br>(TMT, BTE, Ear Tip) | Tests skin sensitization<br>per ISO 10993-10 | Pass |
| EarLens™ System<br>(TMT, BTE, Ear Tip) | Tests for irritation per<br>ISO 10993-10 | Pass |
Table 1: Summary of Biocompatibility Test Results for EarLens™ CHD
Furthermore, the EarLens™ TMT was tested for Nickel leaching per the requirements of EN1811, which is a European standard and the levels of Nickel were found to be within the safe levels per the standard. Although no nickel leaching was noted, results of the 14 dav soak test in simulated sweat solution suggest that traces of oxidation (discoloration) may be visible on the TMT surface following prolonged wear. Performance testing indicated that this is not likely to affect the structural integrity of the TMT within the 1 year expected device life. Because the oxidation (discoloration) was contained beneath the parylene coating, this should not pose a safety risk.
### SHELF LIFE/STERILITY
The EarLens™ CHD is provided non-sterile. The expected service life of the EarLens™ CHD and all accompanying components of the system is one year, based on the results of the mechanical and reliability testing. The BTE and Ear Tip can be cleaned with a soft cloth. Liquid cleaners should not be used on the BTE as these can damage the device.
### ELECTROMAGNETIC COMPATIBILITY TESTING
The requester conducted electromagnetic compatibility testing on the EarLens™ CHD and its respective components and subcomponents, as applicable, to provide a reasonable assurance of safety of the device.
During all the electromagnetic compatibility (EMC) tests, the BTE and light emitting Ear Tip were coupled to a TMT placed in a mold representative of the anatomical shape of the ear. The Ear Tip was placed at a fixed distance from the TMT and optically coupled to it. The BTE was programmed to output a 1000 Hz acoustic tone at a level 3dB below the full scale. The digital representation of the 1000 Hz tone was encoded into light pulses by the laser housed in the Ear Tip and emitted onto the TMT. The audio signal received at the TMT was monitored. During the EMC tests, continuous monitoring of this signal ensured that the laser output remained stable in the presence of external EM fields generated during the testing and that these external EM fields did not induce any energy in the TMT. The acceptance criteria
{7}------------------------------------------------
for the variation of this signal was selected to be less than 3dB because this represents a typical volume adjustment step size in audio instruments, including common hearing aids. No variation of more than 3dB was observed during the EMC testing, demonstrating adequate EMC performance of the device system.
## Emissions and Immunity Testing
EMC for the EarLens™ CHD was tested for both the normal use of the device and the charging operation. The testing was done in accordance with the FDA recognized standard. IEC 60601-1-2 Ed. 3.0. 2007-03-30. Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests. This testing also was conducted at higher compliance limits per the FDA guidance document. Design Considerations for Devices Intended for Home Use, (issued on November 24, 2014), to demonstrate immunity to the following levels for the home environment:
- . Electrostatic Discharge (ESD): +/- 8kV contact discharge and +/- 15kV air discharge
- Power frequency magnetic fields: 30 A/m at 50 Hz or 60 Hz
- . Conducted RF: 3 V r.m.s outside of ISM and amateur radio bands between 0.15 MHz and 80 MHz: 6 V r.m.s in ISM and amateur radio bands between 0.15 MHz and 80 MHz
- . Radiated immunity (RF EM fields): 10 V/m 80 MHz - 2.6 GHz
All tests passed at the higher test levels stated above. which supports immunity at these levels and use of the device in the home environment. Furthermore, in order to assess immunity of EarLens™ CHD on board an aircraft, testing was conducted per the requirements of RTCA DO-160 Section 20 (RF Immunity) Category T, Radio Technical Commission for Aeronautics: Radio Frequency Susceptibility - Well Protected Location, with passing results.
### Immunity Tests with Intentional Radiators
In addition, the performance of the EarLens™ CHD was tested in close proximity with intentional radiators, such as cell phones and metal detectors (Table 2). The EarLens™ CHD was placed within 5 cm of each intentional radiator individually to simulate user interaction with the equipment. The EarLens™ CHD was rotated to expose all six sides of the device to the intentional radiator. The BTE was programmed to transmit a 1 kHz tone and the level of tone was continuously monitored at the TMT in presence of the intentional radiators. Acceptance level was set such that the audio level at the TMT remained within +/-3dB of the original level when the intentional radiator was not present.
The following equipment was tested with passing results supporting immunity:
- . Cell phones (800MHz/1900MHz)
- . Wireless 2.4GHz router
- . Cordless phones (900MHz/2.4GHz)
- CB radio 27MHz
- . Family Radio 460MHz
{8}------------------------------------------------
- Microwave oven
- Portable radio 150MHz
- Land line phone receiver •
- . RFID readers
- . Metal detectors (pulse induction)
- . Electronic article surveillance (EAS) system
Table 2: Description of Intentional Radiators Used for Testing
| Item<br># | Equipment Description | Manufacturer | Model | Serial number | RF Output Power | Operating<br>Frequency |
|-----------|---------------------------------------------|--------------|-------|---------------|-------------------------------------------|------------------------|
| 1 | Wireless Broadband<br>Router - Access point | | | | 0.083 Watts | 2.4 GHz |
| 2 | Portable 2-Way Radio | | | | 2 Watts | 150 MHz -<br>170 MHz |
| 3 | Family Radio Service<br>Band Walkie Talkie | | | | 0.46 Watts | 460 MHz |
| 4 | Cordless Phone | | | | 88 dBuV/m @ 3m | 2.4 GHz |
| 5 | Citizen Band CB Radio | | | | 4 Watts | 27 MHz |
| 6 | Cordless Phone | | | | 88 dBuV/m @ 3m | 900 MHz |
| 7 | Cell Phone | | | | 1.8 Watts (800MHz)<br>1.4 Watts (1900MHz) | 800 MHz,<br>1900 MHz |
| 8 | Cell Phone | | | | 0.7 Watts (800MHz)<br>1.6 Watts (1900MHz) | 800 MHz,<br>1900 MHz |
| 9 | Microwave Oven | | | | 800 Watts | 2.4 GHz |
| 10 | RFID Tag/ Reader | | | | N/A | 125 kHz |
| 11 | Land Line Phone<br>Receiver | | | | N/A | N/A |
| 12 | Metal Detector | | | | N/A | 7 kHz |
| 13 | EAS System | b(4) | | | N/A | 58 kHz |
| 14 | EAS System | b(4) | | | N/A | 58 kHz |
| 15 | EAS System | b(4) | | | N/A | 58 kHz |
| 16 | EAS System | b(4) | | | N/A | 58 kHz |
During the immunity testing, no degradation of performance was observed, indicating that the EarLens™ CHD has high immunity against a range of potential interferers that may be encountered in daily use. Higher intensity or different configurations of potential interferers may negatively impact coexistence in ways not observed here. As an added warning, the device labeling (including patient manual) includes the recommendation from IEC60601-1-2:2007 on the minimal distance to avoid different interferers of different frequencies and output levels.
{9}------------------------------------------------
## SOFTWARE
### Level of Concern
The requester identified the Level of Concern for the EarLens™ CHD as Minor based on their answers to the questions listed in Table 1 (Major Level of Concern) and Table 2 (Moderate Level of Concern) of the FDA guidance document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Consequently, the requester states that any failure or design flaws in the software are unlikely to cause any injury to the patient or to the (clinician) operator. Given that the device maximum equivalent pressure output (MEPO, discussed below) is appropriately limited based on the device design and the selection and tolerance of the various device hardware components including the BTE (e.g., battery, DSP, and laser diode) and TMT (photodetector and microactuator), this assessment for software risk is reasonable.
#### Software Description
EarLens™ software is designed in accordance with the guidance of ISO 13485 and ISO 62304, Medical devices -- Quality management systems -- Requirements for regulatory purposes and ISO 62304, Medical device software -- Software life cycle processes, and is very similar to software used in fitting air conduction hearing aids. The fitting software controls the parametric settings of the hearing aid and DSP firmware applies these values in the fitting of the EarLens™ CHD. During the fitting process, clinicians specify the parametric values of the device programs by entering values from their keyboards of the PC. The DSP firmware applies these values by writing these values into the hardware.
The EarLens™ Fitting Software (ELF) is used to program the BTE. The ELF software enables the hearing professional to calibrate and program the device specific to the patient's needs. In addition to the ELF software, the following devices are required:
- Personal Computer, Windows 7 platform (details below)
- . HI-PRO 2 Box with PC communication cable
- . The HI-PRO 2 box [GN Otometrics A/S] is a standard programming interface for many hearing aid systems. It connects to a computer using a USB interface and to the hearing aid using programming cables. The HI-PRO 2 is powered from the USB interface. The HI-PRO 2 has the following additional characteristics:
- Safety: EN 60601-1, Type BF
- . Electromagnetic Compatibility: EN 60601-1-2
- Systems: EN 60601-1-1 or IEC 60601-1, 3rd ed.
- I Hearing Instrument Interface: EN 60118-14
- Left/Right programming cables for connecting the EarLens™ BTEs to the HiPro ● 2 via the programming ports. These cables are available from a number of hearing aid industry supplier b(4) ]. The 4 pin CS44 is compatible with the EarLens™ CHD.
CS44 is compatible with the EarLens™ CHD
{10}------------------------------------------------
The ELF software is stored on a Compact Disc (CD) and is installed using an industry standard installation package and can be run from an icon on the desktop.
### Device Software Hazard Analysis
The potential hazard associated with the use of the device is the potential occurrence of uncomfortably loud sounds. The hearing aid may cause momentary discomfort at the time of the occurrence. This hazard is mitigated by limiting the hearing aid output to below the patient's loudness discomfort level.
Software Requirement Specification
The fitting software is used to program the parameters and the DSP firmware applies these values in the hardware. b(4)
Standard PC peripherals include a monitor with 1024x768 resolution, mouse and a keyboard. b(4)
#### Verification and Validation
The software is verified using functional testing against requirements, unit tests and moreover, the clinical study validates that the compression algorithms functions as designed.
#### Revision Level History
The software is version controlled and the revision history is maintained. The latest version is 1.0.3.0 and version 1.1.20 for the DSP firmware.
### PERFORMANCE TESTING - BENCH
The requester conducted performance testing on the EarLens™ CHD to support a reasonable assurance of safety and effectiveness of the device. All of the testing described below was conducted on final versions of the TMT, BTE, Ear Tips and Chargers (or final subassemblies) which were manufactured in the final form using representative production processes. Appropriate sample sizes were justified. Table 3 provides a summary of each of the bench test purpose, methods, acceptance criteria, and results. In sum, the bench testing was deemed adequate and supports an expected device life of 1 year.
Table 3: Summary of Bench Testing
| Test Standard/Method | Test Purpose/<br>Description | Component(s)<br>Tested | Acceptance<br>Criteria | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 Standard<br>for medical electrical<br>equipment: General<br>requirements on basic<br>safety and essential<br>performance | To<br>demonstrate<br>that<br>EarLens™<br>CHD meets<br>general safety<br>requirements | • TMT<br>• BTE<br>• Ear Tip<br>• Charger | Per the<br>standard<br>requirements | All tests Passed<br>and requirements<br>of IEC 60601-1<br>Edition 3.0, 2005-<br>12 were met. |
| IEC 60601-1-11<br>Collateral standard for<br>medical electrical<br>equipment and medical<br>electrical systems used<br>in the home healthcare<br>environment | To<br>demonstrate<br>that<br>EarLens™<br>CHD is safe<br>to use in the<br>home<br>environment | • TMT<br>• BTE<br>• Ear Tip<br>• Charger | Per the<br>standard<br>requirements | All tests Passed<br>and requirements<br>of ANSI/AAMI<br>(IEC) 60601-1-<br>11:2010 were met. |
| IEC 60825-1<br>Standard for safety of<br>laser products:<br>equipment classification<br>and requirements | To<br>demonstrate<br>that the laser<br>used in<br>EarLens™<br>CHD is safe<br>and classified<br>as Class I<br>laser product | • BTE<br>• Ear Tip | Per the<br>standard<br>requirements | All tests Passed<br>and the output of<br>the laser used was<br>found to be well<br>within the safe<br>accessible<br>emission levels<br>(AEL) and<br>maximum<br>permissible<br>exposure (MPE)<br>levels for eye and<br>skin. EarLens™<br>CHD was<br>classified as a<br>Class I laser<br>product per the<br>IEC 60825-1. |
| ASTM D4169<br>Standard practice for<br>performance testing of<br>shipping containers and<br>systems | To<br>demonstrate<br>EarLens™<br>CHD can be<br>reliably<br>shipped | • TMT<br>• BTE<br>• Ear Tip<br>• Charger | Per the<br>standard<br>requirements | All tests Passed<br>and requirements<br>of ASTM D4169<br>were met. |
| Electrical Testing | | | | |
| Output Limiter Test<br><br>Measure worst case<br>battery current and the<br>laser output | To<br>demonstrate<br>the output<br>limiter limits<br>the laser<br>output for safe | • BTE<br>• Ear Tip | Battery<br>current is<br>limited to 12<br>mA ± 1 mA<br>at battery<br>voltage 4.2V. | Passed |
| | operation | | | |
| Harmonic Distortion<br>Test<br>Measure harmonic<br>distortion at the BTE<br>output with a 1kHz tone<br>at -3dB re FS<br>Measure harmonic<br>distortion at the TMT<br>output with a frequency<br>sweep up to 10,000 Hz<br>at -3dB re FS light input | To<br>demonstrate<br>the distortion<br>of the<br>EarLensTM<br>CHD is low | • BTE<br>• Ear Tip<br>• TMT | Distortion is<br>less than 5% | Passed |
| Maximum Output Test<br>Measure average<br>maximum equivalent<br>pressure output (MEPO)<br>in human temporal<br>bones | To<br>demonstrate<br>that the<br>maximum<br>output of<br>EarLensTM<br>CHD does not<br>exceed 132dB<br>SPL | • BTE<br>• Ear Tip<br>• TMT | Average<br>maximum<br>equivalent<br>pressure<br>output<br>(MEPO) is<br>$≤$ 132 dB SPL | Passed |
| Mechanical and Reliability Testing | | | | |
| Ear Tip Pull Test<br>Check the Ear Tip cable<br>integrity during and<br>after repeated<br>application of 8oz. of<br>force at the connector<br>end of the cable | To assess<br>reliability of<br>the Ear Tip | • Ear Tip | Ear Tip cable<br>withstands 8<br>oz. of pull<br>force at the<br>connector for<br>1095 cycles | Passed |
| Ear Tip Bend Test<br>Check the Ear Tip cable<br>integrity during and<br>after repeated<br>application of a 150°<br>bend to the cable from<br>its nominal position at | To assess<br>reliability of<br>the Ear Tip | • Ear Tip | Ear Tip cable<br>withstands<br>150° bend<br>for 1095<br>cycles | Passed |
| the point where the<br>cable is expected to<br>enter the ear canal | | | | |
| Ear Tip Twist Test<br>Check the Ear Tip cable<br>integrity during and<br>after repeated<br>application of a 150°<br>twist to the cable from<br>its nominal position at<br>the connector to the<br>BTE | To assess<br>reliability of<br>the Ear Tip | • Ear Tip | Ear Tip cable<br>withstands<br>150° twist<br>for 1095<br>cycles | Passed |
| Accelerated Aging Test<br>- TMT<br>Test performance of<br>TMT after 1 year of<br>simulated aging at 65°C | To assess<br>reliability of<br>adhesive<br>bonds and<br>components | • TMT | No<br>performance<br>degradation<br>in TMT<br>output and<br>adhesive<br>bond<br>strengths | Passed |
| Accelerated Aging Test<br>- Ear Tip<br>Test performance of Ear<br>Tip after 1 year of<br>simulated aging at 75°C | To assess<br>reliability of<br>adhesive<br>bonds and<br>components | • Ear Tip | Less than<br>10%<br>degradation<br>in laser<br>output<br><br>No<br>degradation<br>in adhesive<br>bond strength<br>when Ear Tip<br>cable is<br>repeatedly<br>subjected to<br>10N force at<br>the connector<br>end for 26<br>cycles | Passed |
| Maximum force on the<br>Umbo of the eardrum<br>Measure bias force<br>exerted by the Umbo<br>Platform of the TMT<br>under worst case<br>orientation of the device<br>where effects of gravity<br>are additive | To ensure safe<br>operation of<br>TMT | • TMT | Maximum<br>force exerted<br>on the Umbo<br>of the<br>eardrum by<br>the TMT is ≤<br>6mN | Passed |
| Mechanical strength –<br>grasping tab<br>Apply and measure<br>force using a force<br>gauge coupled to the<br>grasping tab while the<br>chassis is held in fixed<br>position and increase<br>force until component<br>failure is observed | To ensure<br>grasping tab<br>remains intact<br>with forces<br>arising from<br>manipulation<br>by the<br>physician | • TMT | Grasping tab<br>will<br>withstand<br>pull forces of<br>at least 1N | Passed |
| Mechanical strength –<br>Umbo Platform<br>Apply and measure<br>force using a force<br>gauge coupled to the<br>drive-post , while the<br>Umbo Platform is held<br>in fixed position and<br>increase force until<br>component failure is<br>observed | To ensure the<br>umbo<br>platform does<br>not separate | • TMT | Umbo<br>Platform will<br>withstand<br>forces of at<br>least 0.25N | Passed |
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
### IEC 60601-1 Testing
The EarLens™ CHD was tested per the requirements of IEC 60601-1, Edition 3.0, 2005-12, Standard for medical electrical equipment: General requirements on basic safety and essential performance to provide reasonable assurance of the basic safety and essential performance of the device. These tests included power consumption test, humidity preconditioning, determination of applied parts and accessible parts, durability and legibility of markings test, leakage current test, patient leakage test, dielectric strength test, ball pressure test, creepage distances and air clearances, sharp edges test, instability in transport position test, instability excluding transport test, normal heating test, ingress of liquids,
{15}------------------------------------------------
hazardous situations and fault conditions test, rigidity of enclosure (push) test, strength of enclosure (impact) test, drop impact, and mold stress relief test. All tests passed.
## IEC60601-1-11 Testing
The EarLens™ CHD is expected to be used predominately by an older population (>60 years of age) based on the ages of the study subjects in the clinical studies. It was therefore determined that compliance with IEC 60601-1-11:2010 is necessary. This standard applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems which are intended by the manufacturer for use in the home healthcare environment, regardless of whether the medical electrical equipment or medical electrical system is intended for use by a lay operator or by trained healthcare personnel. The object of IEC 60601-1-11:2010 is to specify general requirements that are in addition to those of the general standard IEC 60601-1:2005 and to serve as the basis for particular standards. Compliance testing to this standard was successfully completed and EarLens™ CHD demonstrated basic safety and essential performance as it pertains to medical equipment in the home healthcare environment following the guidelines of the ANSI/AAMI HA606061-1-11:2010 version of this home healthcare environment standard.
These tests include environmental conditions of transport and storage between uses, environmental operating conditions, shock test for mechanical strength for transit-operable ME equipment, and broad-band random vibration test for mechanical strength for transitoperable ME equipment.
# IEC60825-1 Testing
The EarLens™ CHD was tested per the requirements of IEC 60825-1. Safety of Laser Products, and was classified as a Class I laser product. The output of the laser used in the EarLens™ CHD was found to be well within the safe accessible emission levels (AEL) and maximum permissible exposure (MPE) levels for eye and skin.
These tests include measurement of laser radiation and measurement of pulse width and demonstrate safe operation of the laser used in the EarLens™ CHD. Additionally, as part of the clinical study described below, a 'roll-in cohort' of 10 ears (5 subjects) underwent multiple temperature measurements within their ear canal after at least 8 hours of continuous use with the CHD and exposure to the laser. The results showed that there was less than 1°C temperature rise in the ear canal, further demonstrating the safe operation of the laser used in the EarLens™ CHD.
### Electrical Testing
Extensive testing was conducted on the EarLens™ CHD to verify the design criteria and device performance with respect to the electrical system specifications and properties in support of its safety and effectiveness. The BTE Audio Processors, Ear Tips and Chargers were tested in finished form and passed the following tests: total harmonic distortion, maximum output, output noise, microphone directionality, programmable settings, dielectric
{16}------------------------------------------------
strength and leakage current, battery charge time, light output, electromagnetic emissions and immunity. The TMTs were tested in finished form and passed the following tests: total harmonic distortion and vibrational response at -3dB re F.S. and -20dB re F.S. input levels, electromagnetic emissions and immunity.
### Mechanical and Reliability Testing
Extensive testing was conducted on the EarLens™ CHD to verify the design criteria and device performance with respect to the mechanical system specifications and properties in support of reasonable assurance of safety and effectiveness.
The BTE Audio Processors, Ear Tips and Chargers were tested in finished form and passed the following tests: operating conditions, storage conditions, Ear Tip pull test, Ear Tip bend test, Ear Tip twist test, accelerated aging test, liquid/dust ingress test, BTE mechanical strength test. BTE dimensional measurements. Ear Tip vent hole dimensional check, BTE connector torque test, and shipping and environmental tests per ASTM D4169-09.
The TMTs were tested in finished form and passed the following tests: TMT dimensional check, TMT mass, TMT bias force test, TMT adhesive bond strength tests, accelerated aging and fatigue test, shipping and environmental tests per ASTM D4169-09. The maximum worst-case force applied to the umbo under normal conditions was measured and determined to be at a safe limit. The information provided is adequate validation of the force applied to the tympanic membrane (< 6mN) using an appropriate model. Accelerated fatigue testing included a finite element analysis of the microactuator reed and a cyclic fatigue analysis of the bias springs to show that the device is likely to maintain its structural integrity and function for at least 1 year of use under worst case conditions. Tensile testing was performed after accelerated aging in order to confirm long term adhesive bond strength.
#### Temporal Bone Testing
As part of the non-clinical validation, a temporal bone model was used to ensure that the output of the system was within the expected range. In this validation, a calibrated Laser-Doppler Vibrometer (LDV) was used to measure the amplitude of motion of the footplate of the stapes on four temporal bone samples. In each case, the tympanic membrane was exposed to acoustic energy at frequencies between 100Hz and 24,000Hz, and a probe microphone was used to measure the sound pressure level at the tympanic membrane. The sound pressure level and stapes motion were measured simultaneously, and a transfer function determined for each frequency measured (Figure 5a).
{17}------------------------------------------------
Image /page/17/Picture/0 description: This image shows a diagram of the inner ear. The diagram shows the sound pressure input, the sound pressure measurement via probe tube microphone, and the stapes velocity via laser-Doppler vibrometer (LDV). The sound pressure input is shown as blue waves, the sound pressure measurement is shown as a black line, and the stapes velocity is shown as a red line.
Figure 5a: Stapes Velocity Measurements with an acoustic input
Next, for each temporal bone sample, a CHD system was placed and activated in a normal fashion (see Figure 5b). The stapes motion and CHD output signal were recorded and a second transfer function recorded. The two measurements can be correlated to determine the output at full scale, or Maximum Equivalent Pressure Output (MEPO).
{18}------------------------------------------------
Image /page/18/Picture/0 description: This image shows a diagram of the inner ear. The diagram shows the light signal driven by the ear tip. The stapes velocity is measured via laser-Doppler vibrometer (LDV).
Figure 5b: Stapes velocity measurement with a vibratory input
In order to determine the equivalent output of the system in a worst case condition, the emitter was positioned so that all of the light energy was focused on the photodetector of the TMT and the MEPO again determined. In addition to the MEPO calculation based on moderate sound pressure levels, the CHD was driven to full scale output for both the nominal and worst case emitter positions. The testing demonstrated that the overall mean maximum equivalent pressure output (MEPO) of the system on four temporal bone samples was 127dB SPL. This is an acceptable maximum equivalent pressure output for the device. The information provided is adequate validation of the output applied to the tympanic membrane in a clinically appropriate model.
### Magnetic Resonance Imaging (MRI) Safety
The TMT contains an electromagnetic actuator and ferromagnetic materials that could lead to patient injury or device damage during exposure to a magnetic resonance environment. To mitigate this risk, the device is labeled "MR Unsafe." In addition, a patient card is required to be provided to the patient. which instructs physicians to remove the TMT from the eardrum prior to performing an MRI examination.
### SUMMARY OF CLINICAL INFORMATION
{19}------------------------------------------------
The following is a summary of two clinical studies performed by the sponsor to support a reasonable assurance of safety and effectiveness for the EarLens™ CHD.
## Ear Impression Study
EarLens™ conducted an impression study to collect ear impressions from many different ear canal anatomies, which provided validation of the deep canal impression procedure itself. This ear impression procedure involves making a mold deep into the canal and including the tympanic membrane and thus, is considered to involve slightly more risk than typical ear impression procedures for conventional types of hearing aids. The procedure will be performed by a healthcare professional trained in this particular ear impression procedure.
A total of 78 subjects had bilateral ear canal impressions taken (37 female, 41 male). The total number of ears that had impressions taken was 154 (2 subjects had unilateral impressions). Subject age ranged from 32 to 82 years with an average of 66.1 years. The subjects were seen across three clinical sites.
A total of 200 ear impressions were taken on 154 ears (78 subjects). Fourteen subjects (18%) required more than one clinic visit to obtain adequate ear impressions. Multiple impressions were required on some subject ears due to the presence of voids in the impression of anatomical areas that are critical to the patient-matched transducer. The average number of impressions per subject was 2.6, or 1.3 per ear. Thirty-eight percent (38%) of subjects required three or more impressions; 17% required four or more impressions.
Based on a subject survey, a total of 84% of impressions were rated as either no discomfort or mild discomfort. While 80% of the study ears were observed to be normal (unremarkable otoscopic inspection) after the impression procedure, the remaining ears were reported with minor skin contact findings. The most common observation was a micro-hematoma (16%), while petechia (1%), ecchymosis (2%) and abrasion (1%) were reported less frequently. All otologic findings resolved without treatment and without sequelae. Micro-hematomae and ecchymoses can take 2-4 weeks to resolve. Abrasion and petechia, depending on extent and location, can be a more minor finding and may or may not require delay of subsequent otologic procedures, including transducer placement.
Given the nature of this ear impression procedure, the numbers of adverse events reported above are anticipated and acceptable. No major or permanent trauma to the ear canal or tympanic membrane was noted. In order to minimize adverse events, training in this type of ear impression procedure will be necessary and designated as a special control to fit the EarLens™ CHD.
### Pivotal Clinical Study
A clinical study of the EarLens™ CHD confirmed a reasonable assurance of safety and effectiveness of the device for individuals with a mild to severe sensorineural hearing impairment between the frequencies of 125 Hz to 10.000 Hz. This prospective, single arm, open label study assessed 48 subjects (96 ears) who wore the fully activated system in both ears in their daily lives for four months per study protocol. The subjects served as their own controls, with outcomes
{20}------------------------------------------------
comparing the unaided (no treatment) condition to the aided (treatment with the EarLens™ CHD) condition. Safety and effectiveness were assessed during the four months.
The average age of the study population was 69 years with gender ratio at 60% for males and 40% for females. The subjects were seen across three clinical sites with the largest enrollment at Site 1 (Site 1= 21, Site 2=15, Site 3=12). All subjects were experienced hearing aid users who met the inclusion and exclusion criteria.
The inclusion criteria were:
- 1. Age 18 to 85 years;
- 2. Mild to severe hearing impairment within the fitting range shown below: One frequency may fall outside range for each ear;
| Freq (Hz) | 125 | 250 | 500 | 1000 | 2000 | 4000 | 6000 | 8000 | 10000 |
|-------------|-----|-----|-----|------|------|------|------|------|-------|
| HL min (dB) | 0 | 0 | 0 | 0 | 15 | 30 | 30 | 30 | 30 |
| HL max (dB) | 50 | 60 | 60 | 70 | 75 | 80 | 80 | 80 | 80 |
Image /page/20/Figure/6 description: This image is an audiogram, which is a graph that shows a person's hearing thresholds at different frequencies. The x-axis represents frequency in Hertz (Hz), ranging from 125 to 10k. The y-axis represents hearing level in decibels (dB), ranging from -10 to 100. The shaded area on the audiogram represents the range of sounds that the person can hear, showing a hearing loss at lower frequencies.
- 3. Symmetric hearing impairment: no more than a 15 dB difference in thresholds between the ears at five of eight tested frequencies (125, 500, 1000, 2000, 4000, 8000, and 10.000 Hz).
- 4. The difference between the diagnostic word recognition scores (Northwestern Auditory Test No. 6 (NU-6)) between the two ears for a given subject must not exceed 25% to reduce the possibility that a subject may have retrocochlear involvement.
- 5. No significant conductive hearing impairment;
- a. No more than a 10 dB air-bone gap at three of four tested frequencies (500, 1000, 2000. or 4000 Hz):
- b. Normal Type A tympanometry (indicating normal mobility of the tympanic membrane (TM) and middle ear ossicles).
{21}------------------------------------------------
- 6. Greater than or equal to 60% on maximum word recognition score using a 50-word NU-6 list for each ear measured at a presentation level determined by the threshold at 2000 Hz and presented over inserts or earphones demonstrating an ability to benefit from amplification.
- 7. Current unilateral or bilateral user of hearing aids for at least 6 weeks prior to study enrollment.
- 8. Able and willing to commit to the travel and time demands of the study (available for six months or longer) and able to comprehend and comply with the study materials and instructions.
- 9. Native speaker of American English due to use of American English study materials.
The exclusion criteria were:
- 1. Must not have known or active medical issues that would preclude having the study device, including:
- a. abnormal TM (deemed perforated, inflamed, scelerotic, or has a dimeric or monomeric area, or in any other way abnormal);
- b. Abnormal middle ear or a history of prior middle ear surgery other than tympanostomy tubes:
- c. Ear canal anatomy that prevents the physician from seeing an adequate amount of the TM, or an anatomical configuration of the external auditory canal that prevents satisfactory placement of the TMT (examples include a large anterior canal bulge and exostoses of the ear canal).
- 2. Must not have other known or active medical issues including:
- a. History of chronic and recurrent ear infections in the past 24 months;
- b. History of dizziness and/or vertigo in the past 24 months:
- Currently taking medications/treatments with known ototoxic effects; C.
- d. Rapidly progressive or fluctuating hearing impairment:
- e. Diagnosis of a compromised immune system which may impact the tissue of the auricle or ear canal, keratosis obturans, ichthyosis, eczema of the auricle or ear canal, or received radiation of the head ever or chemotherapy for cancer within the last six months.
- 3. Must not fit the definition of a vulnerable subject, as per FDA regulations 21 CFR Parts 50 and 56.
- 4. Must not have an ear canal anatomy that precludes manufacture of the TMT as determined by EarLens™ manufacturing personnel. The manufacturing personnel will be independent from the personnel evaluating the audiological data.
# Safety Results
The primary safety endpoint was intended to demonstrate that use of the EarLens™ CHD did not result in a change in residual hearing function. The objective was to identify any change in baseline hearing after four months of device usage using a four frequency threshold criteria (500, 1000. 2000. and 4000 Hz. referred to as PTA4). A determination of clinically non-significant hearing threshold change was made if the calculated PTA4 hearing change of the subject population was less than 10 dB. After wearing the TMT for 4 months, no decrease in hearing
{22}------------------------------------------------
sensitivity of more than 10 dB was observed. A secondary safety endpoint assessed any decrease in hearing sensitivity of >10 dB by subject per ear at each test frequency. After four months of use, no subjects exhibited a decrease of >10 dB at either ear by frequency. In addition, no serious device or procedure-related adverse events (AEs) were reported during the trial. There were 31 AEs reported in 20 subjects at 22 ears. All but one of the AEs was temporary and resolved. One subject reported a 'fullness' sensation when wearing the CHD, which did not change during the trial but effectiveness outcomes for this subject were not impacted. The subject continued use throughout the study period of four months. The event is unresolved because the subject elected to participate in the protocol extension and had the devices replaced after the 4-month study period. The following table (Table 4) identifies the AEs by type, frequency of occurrence and resolution status at the active study end.
| Adverse Event Category | Number<br>Occurring | Serious AE | Status |
|---------------------------------------------------------|---------------------|------------|----------|
| Abrasion/blood blister in ear<br>canal | 17 | No | Resolved |
| Ear discomfort/pain | 5 | No | Resolved |
| Inflammation/granulation tissue<br>on tympanic membrane | 3 | No | Resolved |
| Abrasion/blood blister on<br>tympanic membrane | 2 | No | Resolved |
| Ear tip-related: | | | |
| Ear canal swelling, itching, etc | 2 | No | Resolved |
| Pain upon eructation & valsalva | 1 | No | Resolved |
| Sensation of fullness | 1 | No | Ongoing |
| Table 4: Adverse events across study period | | | | | | |
|---------------------------------------------|--|--|--|--|--|--|
|---------------------------------------------|--|--|--|--|--|--|
In addition, images of the subjects' tympanic membranes were taken pre- and post-4 months of EarLens™ CHD wear. These images (n=78 ears) were reviewed and were unremarkable. No significant safety concerns were noted.
In order to assess the risk of overheating of the ear canal from the light source transducer (laser diode), 5 roll-in subjects were enrolled in a pilot phase of the study for this safety analysis. Ear canal temperature measures in the 5 roll-in subjects (10 ears) showed no rise in ear canal temperatures after 8 continuous hours of device use. This clinical validation, in conjunction with the non-clinical testing of the laser diode, provides a reasonable assurance of the safety of this component of the EarLens™ CHD.
{23}------------------------------------------------
Hearing thresholds were measured with the TMTs in place, but without the audio processor, in order to assess the potential temporary tympanic membrane damping that may occur as a result of having a device in continuous contact with the surface of the tympanic membrane. The TMTs are designed to remain in place even when the Audio Processor is not worn. When the audio processor is removed (swimming, bathing, sleeping), users may experience tympanic membrane damping, which would be interpreted as slight reduction of sound that already caused by their hearing loss, due to the loading effect of the TMT. The amount of damping observed varied, but the effect on PTA (500Hz, 1000 Hz, and 2000 Hz) averaged 4 dB of damping, which is immediately reversed when the TMT is removed. The maximum amount of damping observed was 20 dB at a single frequency. The amount of damping stayed essentially the same for the duration of the 4 month study. When the audio processor is in place, the gain delivered by the EarLens™ CHD more than overcomes the TM damping effect. Figure 6 shows the average amount of damping observed across the frequency range 125-10000 Hz (n=90 ears).
Image /page/23/Figure/1 description: The image is a graph that shows hearing change in decibels (dB) as a function of frequency in Hertz (Hz). The x-axis is on a logarithmic scale, ranging from 100 to 100000 Hz. The y-axis represents the hearing change in dB, ranging from -10 to 10. The graph shows a curve that fluctuates between approximately -6 dB and 0.5 dB.
Figure 6: Damping of hearing from presence of TMT compared to unaided baseline
#### Effectiveness Results
The primary efficacy endpoint was intended to demonstrate device effectiveness by improving speech recognition using the Northwestern Auditory Test No.6 (NU-6) of word recognition with the EarLens™ CHD at a speech level of 45 dB HL. The objective was to show that the EarLens™ CHD provides a statistically significant improvement in mean aided word recognition at 30 days post placement when compared to the baseline unaided condition measured prior to placement. The average baseline unaided score was 52% and the average aided score was 85% (see Figure 7); this improvement was statistically significant (p<0.0001) and very clinically
Frequency (Hz)
{24}------------------------------------------------
meaningful with an average of 33% improvement in word recognition ability at a soft listening level.
A secondary measure of device effectiveness was defined as more than 10 dB improvement (functional gain) in thresholds over the range of frequencies from 2000 to 10000 Hz for the aided condition measured at 30 days post placement when compared to unaided condition measured prior to placement of the EarLens™ CHD. Mean functional gain was 30.5 dB (p<0.0001). indicating that the EarLens™ CHD was able to deliver significant functional gain (see Figure 8). Functional gain reached a maximum of 68 dB at 9-10k Hz, which is not typically achieved with conventional air-conduction hearing aids.
These primary and secondary measures of device effectiveness were repeated at later time intervals in order to demonstrate stability of the subjects' performance over time. The repeated measures provide evidence that the gains made by 30 days persist through the 120-day interval, suggesting that the EarLens™ CHD remains effective while being worn over time.
The signal processing of the EarLens™ CHD includes algorithms (Wide Dynamic Range Compression (WDRC), Fixed Directional Microphone (DDM), Environmental Noise Reduction (ENR), Dynamic Feedback Cancellation (DFC), Automatic Gain Control Output (AGCO), Sliding Bias, and dual microphone processing) that were worn by subjects throughout the study, and thus were included as part of the effectiveness outcomes and perceived benefit.
Additional effectiveness measures included speech in noise testing with the Hearing in Noise Test (HINT) and Quick SIN. The aided 30-day scores trend toward improvement but statistically significant benefit was not reached on either of these tests in the omni-directional microphone listening mode. When directionality was active, the directional microphone provided an average of 3.14 dB of benefit over the unaided baseline condition. This is a statistically significant (p<0.0001) and clinically meaningful improvement relative to the baseline unaided scores.
{25}------------------------------------------------
An additional measure of device effectiveness was perceived benefit as measured by the Abbreviated Profile of Hearing Aid Benefit (APHAB). The APHAB is a validated and widely used clinical tool for assessing hearing aid benefit. The APHAB was completed for the EarLens™ CHD and the subjects' own conventional hearing aids. The average baseline unaided percentage of communication difficulties was 58% (standard deviation = 16%), the percentage of difficulties decreased to 30% (standard deviation = 13%) with the subject's own air conduction hearing aid, and for EarLens™ it was 29% (standard deviation = 14%). Ninety-two percent (92%) of subjects completing the study (35 out of 38) perceived a statistically significant improvement for EarLens™ relative to unaided as measured by APHAB. based on normative data for this questionnaire.' This amount of benefit is predictive of noticeable benefit and success with amplification. ' The global APHAB scores for the subjects (n=36) using their own hearing aids and the EarLens™ CHD were essentially the same (1% difference) suggesting that there was not a clinically significant difference in perception of overall benefit for the EarLens™ CHD compared to the subjects' own hearing aids as measured by APHAB.
All subjects completed a Study Exit Questionnaire (developed by EarLens™) upon exiting the study after 4 months of EarLens™ CHD wear. This is a 15 item questionnaire that asks subjects to rate their satisfaction with the EarLens™ CHD in the following domains: quality, naturalness, and clarity of sound, understanding speech in noise, ability of the device to address their hearing problems, quality of music, device comfort, cosmetics, etc. For these items, subjects were asked to select one of 6 ratings that ranged from Very Satisfied to Very Dissatisfied. In addition, the subjects were asked to rate the performance of the EarLens™ CHD in terms of how well it improved their experience in several areas (e.g., hearing in noisy environments, ability to participate in group conversations) relative to the unaided condition. Subjects were asked to select one of 5 ratings that ranged from Verv Much Improved to Worse. The majority of subjects were satisfied with their ability to understand speech in noisy environments with the EarLens™ CHD (90% responded Satisfied or Very Satisfied) and found that the EarLens™ CHD improved their ability to participate in group conversations (80% responded Improved or Very Much Improved). Additionally, most subjects found that the EarLens™ CHD reduced the level of effort required to carry on conversations (87.5% responded Improved or Very Much Improved). Overall, most subjects were satisfied with the quality of sound delivered by the EarLens™ CHD (90% responded Satisfied or Very Satisfied) and found that EarLens™ CHD improved their overall quality of life (77.5% responded Improved or Very Much Improved).
'Cox, R., & Alexander, G. (1995). The abbreviated profile of hearing aid benefit. Ear and Hearing, 16(2), 176-186.
### Summary of Extended Study
The safety and effectiveness of the EarLens™ CHD is currently being monitored beyond the 4 months of the Definitive Study. In the Extended Study, 24 subjects (48 ears) have opted to continue wearing the device after completing the Definitive study. As of April 30, 2015, 92% of ears have 6 or more months cumulative TMT wear, 50% have 9 or more months, and 4% (2 ears) have 12 or more months of cumulative TMT wear. Of the 24 active subjects in the Extended Wear Study, 11 related AEs were experienced by 8 subjects in 10 ears. All events were temporary and resolved. Nine of 11 AEs were related to ear cleaning pre-impression (3 AEs), the
{26}------------------------------------------------
impression procedure (4 AEs), or the inspection process pre-impression (2 AE). Two of the related AEs were attributed to Ear Tip fit and both were resolved after Ear Tip modification. One subject continues to report a sensation of fullness.
Based on the results of these clinical studies, the EarLens™ CHD has been shown to be reasonably safe and effective in delivering the full spectrum of amplification from 125 Hz to 10.000 Hz for at least 4 months of wear.
## LABELING
The sponsor provided labeling information, which includes Physician Instructions for Use, Patient Instructions for Use, and Hearing Professional Instructions for Use. In addition, the sponsor provides an Ear Impression Kit that includes instructions for the physician regarding how to obtain an ear impression for the EarLens™ CHD. Because the TMT component cannot be self-removed, a patient card is required that can be carried with the patient to provide information about the device (e.g., MR Unsafe, who should remove it, who to contact) in cases of emergency.
The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 Prescription devices. The patient labeling also follows the principles identified in FDA's guidance entitled "Medical Device Patient Labeling" (April 2001). The labeling includes a summary of the clinical studies, detailed instructions on how to fit the device to the patient, instructions for periodic cleaning of the BTE and Ear Tip, and information related to electromagnetic compatibility. In addition, the patient instructions for use includes information on how to correctly use and maintain the device. the potential risks and benefits associated with the use of the device, and alternative treatments.
### RISKS TO HEALTH
Table 5 below identifies the risks to health that may be associated with use of tympanic membrane contact hearing aid and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Method |
|-------------------------------------------|------------------------------------------------------------------------|
| Adverse Tissue Reactions | Biocompatibility Labeling |
| Electromagnetic Incompatibility | Non-Clinical Performance Testing Labeling |
| MRI Incompatibility | Labeling |
| Overheating of Ear Canal or Skin | Non-Clinical Performance Testing Clinical Performance Testing Labeling |
| Damage to Eyes from Direct Laser Exposure | Labeling |
Table 5: Identified Risks to Health and Mitigation Methods
{27}------------------------------------------------
| Trauma/Damage to the Ear Canal, Tympanic<br>Membrane, or Middle Ear System | • Non-Clinical Performance Testing<br>• Clinical Performance Testing<br>• Training<br>• Labeling |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Residual Hearing Loss…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.