K121317 · Cochlear Americas · MAH · Oct 12, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K121317
Device Name
COCHLEAR BAHA IMPLANT SYSTEM
Applicant
Cochlear Americas
Product Code
MAH · Ear, Nose, Throat
Decision Date
Oct 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Baha implant system is intended for treatment of patients who have conductive or mixed hearing loss as a result of certain medical conditions such as bilateral atresia and chronic supportive otitis media, and for those who have Single-Sided Deafness (SSD) caused by a congenital condition, surgery, trauma, or disease. The intended use of the system is to provide an osseointegrated fixation point for connection of an external sound processor. The abutment piece, the subject of this 510(k), is intended to provide a connection between the osseointegrated implant and the external sound processor.
Device Story
System comprises titanium implant, percutaneous abutment, and external sound processor. Implant surgically placed in skull; osseointegrates with bone. Abutment connects implant to external sound processor. Processor captures sound; transmits vibrations through abutment to implant; implant vibrates skull; stimulates inner ear nerve fibers. Bypasses external auditory canal and middle ear. Used in clinical settings by physicians; patient wears processor daily. Benefits patients with conductive/mixed hearing loss or SSD by providing direct bone-conducted sound transmission. Modification involves BA400 abutment design change and addition of hydroxyapatite coating to improve integration/stability.
Clinical Evidence
Bench testing only. Performance testing compared proposed BA400 abutment system against predicate system. Results met or exceeded established acceptance criteria.
Technological Characteristics
Titanium (ASTM F 67-06 Grade 4, USN R50700, ISO 5832-2) and Hydroxyapatite (ISO 13779-2). Percutaneous abutment design with hydroxyapatite coating. Mechanical bone-conduction transmission principle. No software or electronics in the abutment component.
Indications for Use
Indicated for patients aged 5+ with conductive/mixed hearing loss (bone-conduction thresholds ≤45-65 dB HL depending on processor) or unilateral sensorineural deafness (SSD) with normal hearing in the contralateral ear (air-conduction threshold ≤20 dB HL). Also indicated for patients who are candidates for AC CROS hearing aids but cannot/will not use them. Bilateral fitting indicated for symmetric moderate-to-severe conductive/mixed loss.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
K121317
OCT 1 2 2012
Cochlear™
Hear now. And abouts
### 6. 510(k) Summary A. Submitter Information
Submitted by:
On behalf of:
Contact:
B. Date Prepared
். Device Class
D. Device Name Trade/Proprietary Name: Common/Usual Name: Classification Name:
> Classification Panel: Product Code:
E. Predicate Device: Trade/Proprietary Name: Common/Usual Name: Classification Name:
> Classification Panel: Product Code: 510(k):
Cochlear Americas 13059 East Peakview Ave. Centennial, CO 80111
Cochlear Bone Anchored Solutions AB Konstruktionsvägen 14 SE-435 33 Mölnlycke Sweden (Establishment Number 9616024)
Sean Bundy Director of Quality and Regulatory Cochlear Americas sbundy@cochlear.com (303) 524-7139 (o) (303) 524-6825 (f)
## 1-May-2012
ll
Cochlear™ Baha® Implant System Auditory Osseointegrated Implant Hearing Aid, Bone Conduction, Implanted 21 CFR 874.3300, Class II Ear Nose and Throat MAH
Cochlear™ Baha® Implant System Auditory Osseointegrated Implant Bone Conduction Hearing Aid 21 CFR 874.3300, Class II Ear Nose and Throat гхв K100360
{1}------------------------------------------------
Cochlear™
Hear now. And abuays
### F. Purpose of Submission
The Baha implant system works by combining an external sound processor with an abutment and a small titanium implant placed in the skull behind the ear through a simple surgical procedure. The purpose of this Traditional 510(k) is to introduce slight modifications to the BA300 abutment (part of The Cochlear Baha BIA300 Series Implant System) cleared under 510(k) K100360. The modified abutment will be known as model BA400, and the system containing the abutment will be known as the 400 series. The other two parts that make the system function, the implant and the sound processors, will remain unchanged from what is legally cleared. The BA400 will be available as a standalone part or pre-mounted onto a Baha B1300 series titanium implant.
### G. Device Description
The technology base for the Baha implant was originally derived from the 1952 discovery by Dr. Per-Ingvar Brânemark that titanium was biocompatible with bone, leading to the term "osseointegration". The Brånemark System, as it would later be known, formed the basis for the rapid development and widespread implementation of root form endosseous dental implants, now safely and effectively used by many hundreds of thousands of people worldwide. Root form dental implants share many characteristics with the Baha implant component including material, conformation, and surgical implantation procedures.
The Baha implant system functions by combining 3 parts: a titanium implant, a percutaneous abutment, and a sound processor. The system works by utilizing natural bone transmission as a pathway for sound to travel to the inner ear, bypassing the external auditory canal and middle ear. After surgical placement, the titanium implant naturally integrates with the skull bone over time through a process known as osseointegration. The external sound processor transmits sound vibrations through the percutaneous abutment to the titanium implant. The vibrating implant creates vibrations within the skull that stimulate the nerve fibers of the inner ear, allowing hearing.
# 17 of 317
{2}------------------------------------------------
Hear now, And always
The minor changes to this device are only to the abutment, the piece that serves as the connection between the osseointegrated implant and the external sound processor. The changes include a modified design and the addition of a hydroxyapatite coating, neither of which modify the intended functionality or fundamental operating principles of the implant/abutment system.
#### H. Intended Use
The Baha implant system is intended for treatment of patients who have conductive or mixed hearing loss as a result of certain medical conditions such as bilateral atresia and chronic supportive otitis media, and for those who have Single-Sided Deafness (SSD) caused by a congenital condition, surgery, trauma, or disease. The intended use of the system is to provide an osseointegrated fixation point for connection of an external sound processor. The abutment piece, the subject of this 510(k), is intended to provide a connection between the osseointegrated implant and the external sound processor. The intended use remains unchanged from the predicate device.
#### l. Technological Characteristics
The Baha implant system with the hydroxyapatite coated abutment has the same intended use, the same basic mechanical design, the same functional characteristics, the same fundamental operating principles, and is made of the same titanium as the predicate device.
#### J. Materials
Titanium according to ASTM F 67-06 Grade 4 USN R50700 (ISO 5832-2)
Hydroxyapatite: Cas(PO4)3(OH) according to ISO 13779-2
#### K. Performance Data
Performance testing was conducted based on a comparison between proposed and predicate implant/abutment systems. The results were equal or better than established acceptance criteria, where predicate testing was used as a baseline.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cochlear company logo. The logo consists of a stylized graphic above the word "Cochlear" in a sans-serif font. The graphic appears to be an abstract representation of a cochlea, the spiral-shaped part of the inner ear. The letters "TM" are written in superscript to the right of the company name.
Hear now. And always
### Conclusion L.
Based on the indications for use, technological characteristics, and substantial equivalence comparison to the predicate device, the Baha Implant system with the hydroxyapatite coated abutments has been shown to be safe and effective for its intended use.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains an image of an eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 12 2012
Cochlear Americas c/o Mr. Sean Bundy Director of Regulatory Affairs and Quality 13059 East Peakview Avenue Centennial, CO 80111
Re: K121317
Trade/Device Name: Cochlear Baha Implant System – BA400 Abutment Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: MAH Dated: September 21, 2012 Received: September 24, 2012
Dear Mr. Bundy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eun. R. Atkinson, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a black and white drawing of a circular object. The object appears to be a ball of yarn or a similar tangled mass of string. The drawing is simple, with rough lines and shading to indicate the texture and form of the object.
Hear now And always
# 5. Indications for Use Statement
510(k) Number (if known): 1121317
Device Name: Cochlear ™ Baha® BA400 Abutment
Indications for Use:
The Cochlear Baha® auditory osseointegrated implant system using model BA400 abutment is intended for the following patients and indications for use:
- Patients aged 5 and older
- Patients who have a conductive or mixed hearing loss and can still benefit . from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3kHz) should be better than or equal to 45 dB HL for use with the BP100 sound processor, 55 dB HL for use with the Intenso sound processor, 55 db HL for use with the BP110 Power sound processor, and 65 db HL for use with the Cordelle II Sound Processor.
- Bilateral fitting is intended for patients who meet the above criterion in both . ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conductive thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with . normal hearing in the other ear (i.e. Single-sided deafness: SSD™). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 ав нг
- Baha for SSD is also indicated for any patient who is indicated for an air-. conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Daviel Cluffn
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
121317
15 of 317
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.