EASYLISTENER 2 FM, RADIUM FM AND VOCALIGHT INFRARED SOUND FIELD SYSTEMS

K043090 · Phonic Ear, Inc. · LZI · Dec 23, 2004 · Ear, Nose, Throat

Device Facts

Record IDK043090
Device NameEASYLISTENER 2 FM, RADIUM FM AND VOCALIGHT INFRARED SOUND FIELD SYSTEMS
ApplicantPhonic Ear, Inc.
Product CodeLZI · Ear, Nose, Throat
Decision DateDec 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing on he aring impairments. The device is used with an associated transmitter microphone and speaker systems and does not provide coupling to the ear through either earphones or earmolds. These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTE hearing aids and/or personal FM devices. The talker's voice is mildly amplified and dispersed throughout the room to assist listeners, regardless of seating location, to consistently beap what the talker is saying.

Device Story

Sound Field systems (EasyListener 2, Radium, Vocalight) function as group auditory trainers; input consists of voice signals captured via transmitter microphone; signals transmitted via FM or infrared wireless technology to receiver/amplifier; processed signals broadcast through strategically placed wall/ceiling loudspeakers. System installed by audiologist or sound contractor; operated by speaker/instructor. Provides mild voice amplification to assist listeners in classrooms or group settings; improves speech perception regardless of seating location. Does not couple to ear; compatible with BTE hearing aids or personal FM devices. Benefits include consistent speech audibility for all listeners.

Clinical Evidence

No clinical data provided. Evidence relies on field studies conducted in accordance with ANSI S12.60 Acoustical Performance Criteria, Design Requirements, and Guidelines for Schools, and compliance with FCC Rules Part 15 for wireless transmission.

Technological Characteristics

Wireless sound field amplification system using FM and/or infrared technology. Components include transmitters, receivers, and wall/ceiling-mounted loudspeakers. Features include on/off switch and volume controls. Operates as a group auditory trainer; non-ear-coupled. Complies with ANSI S12.60 and FCC Part 15, Subpart A.

Indications for Use

Indicated for use as group auditory trainers for listeners with normal hearing or hearing impairments to assist in speech perception by amplifying and dispersing a talker's voice in a room. No contraindications.

Regulatory Classification

Identification

A group hearing aid or group auditory trainer is a hearing aid that is intended for use in communicating simultaneously with one or more listeners having hearing impairment. The device is used with an associated transmitter microphone. It may be either monaural or binaural, and it provides coupling to the ear through either earphones or earmolds. The generic type of device includes three types of applications: hardwire systems, inductance loop systems, and wireless systems.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K043090 ### DFC 2 3 2004 Phonic Ear, Inc. Office of Regulatory Affairs & QMS Sound Field Group Amplification Systems - 510(k) Section C ## 510(k) Summary Phonic Ear, Inc. 3880 Cypress Drive Petaluma, California 94954-7600 Establishment Registration No. 2918633 Mr. Michael Gibbs Telephone Number: (707)769-1110 Ext. 247 Fax Number: (707) 769-9624 | Device Trade or Proprietary Names: | EasyListener 2 FM Sound Field System<br>Radium FM Sound Field System<br>Vocalight Infrared Sound Field System | |------------------------------------|---------------------------------------------------------------------------------------------------------------| | Device Common Name: | Sound Field Amplification Systems | | Classification Name: | Group Hearing Aid or Group Auditory Trainer | | Device Classification: | Class II | | Regulation Number: | 874.3320 | | Product Code: | EPF | #### Intended Use of the Device: Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing on he aring impairments. The device is used with an associated transmitter microphone and speaker systems and does not provide coupling to the ear through either earphones or earmolds. These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTE hearing aids and/or personal FM devices. The talker's voice is mildly amplified and dispersed throughout the room to assist listeners, regardless of seating location, to consistently beap what the talker is saying. #### Identification of predicate devices : Sound Field systems are substantially equivalent to various types of approved Group Hearing-Aids or Group Auditory Trainers that are directly connected or coupled to the listener as described above with similar results: K974287 – FM Receiver Group Auditory Trainer-Solaris Binaural Hearing System {2}------------------------------------------------ Phonic Ear, Inc. Office of Regulatory Affairs & OMS Sound Field Group Amplification Systems - 510(k) Section C #### Description of the candidate devices : Sound Field group amplification systems consist of FM or IR transmitters and microphone used by a speaker/instructor to send auditory signals to FM or IR receiver/amplifier which noreesses the voice signals and broadcast the signals through one or more loudspeakers. These systems are generally installed under the guidance of an audiologist or a sound contractor. Sound Field systems consist of an on/off switch and volume controls. Specifications and user guider in cre provided for Sound Field installation and use. Sound Field systems are assembled from standard electronic components widely used by Group Hearing Aid and Group Auditory Trainer manufacturers. Sound Field wireless ampification systems amplify and broadcast the speaker/instructor's voice through strategically-mnounted.will and/or ceiling loud speakers placed throughout the listening area to distribute the final moder want signal (sound) of the speaker/instructor and are not directly connected or coupled to the piroener. Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing on he aring impairments. The device is used with an associated transmitter microphone and speakers systems and does not provide coupling to the ear through either earphones or earmolds. These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTE hearing aids and/or personal FM devices. The talker's voice is mildly amplified and dispersed throughout the room to assist listeners, regardless of seating location, to consistently bear what the talker is saying. There are no known contraindications associated with the use of Sound Field group amplification systems. Many case studies, experiments, and trials exist that document the benefite and effectiveness of speech perception from Sound Field amplification. Field studies were conducted in accordance with ANSI S12.60 Acoustical Performance Criteria, Design Requirements, and Guidelines for Schools. Prior to the issuance of ANSI S12.60 in 2002 there were no known performance standards or special controls for Sound Field ampification systems. Test procedures to obtain FCC (Federal Communications Commission) approval are in accordance with good engineering practices and in accordance with FCC Rules, Part 15, Subpart A, 15, 19 (a) (3) and (C) Code of Federal Regulations (CFR 47). {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 3 2004 Phonic Ear, Inc. c/o Mr. Michael L. Gibbs Regulatory Affairs & QMS Manager Corporate Headquarters 3880 Cypress Drive Petaluma, CA 94928 Re: K043090 Trade/Device Name: Phonic Ear Sound Field Group Amplification Systems Regulation Number: 21 CFR 874.3320 Regulation Name: Group Hearing Aid, Group Auditory Trainer Regulatory Class: Class II Product Code: LZI Dated: November 30, 2004 Received: December 2, 2004 Dear Mr. Gibbs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, A halpi korenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Phonic Ear, Inc. Office of Regulatory Affairs & OMS Sound Field Group Amplification Systems - 510(k) Section D # Indications for Use #### 510(k) Number (if known): K043090 Device Name: Phonic Ear Sound Field Group Amplification Systems Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing or hearing impairments. The device is used with an associated transmitter microphone and speaker systems and does not provide coupling to the ear through either earphones or earmolds. These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTF hearing aids and/or personal FM devices. The talker's voice with or which bI E throughout the room to assist listeners, regardless of seating location, to consistently hear what the talker is saying. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) **Over-the-Counter Use** sion of Ophthalr se and Throat De **510(k) Number** K043090
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...