K953872 · Wordcomp International Communication Group (Wicg) · LXB · Apr 12, 1996 · Ear, Nose, Throat
Device Facts
Record ID
K953872
Device Name
SECOND EAR BONE CONDUCTION HEARING AID
Applicant
Wordcomp International Communication Group (Wicg)
Product Code
LXB · Ear, Nose, Throat
Decision Date
Apr 12, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic
Intended Use
A hearing aid is wearable sound-amplifying device that is intended to compensate for impaired hearing.
Device Story
The EC-1000 Second Ear® is a bone conduction hearing aid designed to bypass outer and middle ear obstructions. It consists of a transducer unit, a pocket-sized audio processor/driver module, and a connecting cable. The device captures ambient sound via a built-in microphone in the processor module, processes the signal, and drives the transducer. The transducer, held against the skull (typically behind the ear) via a strap or headband, converts electrical signals into mechanical vibrations transmitted directly to the cochlea. The system features an adjustable tone control, a voice filter for reducing exterior noise, and an auxiliary input jack for external audio sources (TV, radio, telephone). It is powered by a rechargeable nickel-metal-hydride battery. The device is intended for use by patients with conductive hearing loss to improve auditory perception.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and performance specification comparisons.
Technological Characteristics
Materials: Medical grade plastic. Sensing/Actuation: Electromechanical transducer (low-impedance). Energy Source: 4.8 VDC Nickel-Metal-Hydride rechargeable battery. Form Factor: Two-piece system (transducer and pocket-worn processor). Connectivity: Auxiliary input jack for external audio sources. Frequency Response: 150 Hz–8 kHz (unfiltered), 300 Hz–3 kHz (filtered). Max Gain: 57 dB. S/N Ratio: 40 dB.
Indications for Use
Indicated for persons with moderate to severe conductive hearing losses, particularly those compounded by congenital or secondary obstruction of auditory air conduction mechanisms.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Predicate Devices
Starkey Model BC1 Bone Conduction CROS Hearing Aid (K923784)
K161123 — Baha 5 Power Sound Processor · Cochlear Americas · May 20, 2016
K153245 — Baha 5 Super Power Sound Processor · Cochlear Americas · Mar 2, 2016
K082108 — OBC BONE ANCHORED HEARING AID SYSTEM · Oticon Medical AB · Nov 14, 2008
K050653 — TRANSEAR BONE CONDUCTION HEARING AID · United Hearing Systems, Inc. · Jul 1, 2005
K062404 — TRANSEAR BONE CONDUCTION HEARING AID · Ear Technology Corp. · Sep 8, 2006
Submission Summary (Full Text)
{0}
510(k) Premarket Notification - Original Submission
Second Eat® Bone Conduction Hearing Aid
March 26, 1996
K953872
# APR 12 1996
## Section 510(k)
## Premarket Notification
## Summary of Safety and Effectiveness Information
**Regulatory Authority:** Safe Medical Devices Act of 1990, 21 CFR 807.92
1. **Device Trade Name:** Second Eat® Bone Conduction Hearing Aid
2. **Common / Classification Name:**
- Common Name: Bone Conduction Hearing Aid
- Classification Name: Hearing Aid
3. **Applicant / Sponsor Name / Address:**
WORDCOMP INTERNATIONAL COMMUNICATION GROUP
4700 Von Karman, Suite 115
Newport Beach, California 92660
(714) 752-6677 - (714) 752-6769 Fax
4. **Contact Person:**
Mr. Masanori Tada
WORDCOMP INTERNATIONAL COMMUNICATION GROUP
4700 Von Karman, Suite 115
Newport Beach, California 92660
(714) 752-6677 - (714) 752-6769 Fax
5. **Establishment Registration Number:**
- Number: 9021402
6. **Classification:**
§ 874.3300 Hearing Aid. (a) Identification. A hearing aid is wearable sound-amplifying device that is intended to compensate for impaired hearing. This generic type of device includes the air-conduction hearing aid and the bone-conduction hearing aid, but excludes the group hearing aid or group auditory trainer (874.3320), master hearing aid (874.3330), and tinnitus masker (874.3400). (b) Classification. (1) Class I for the air-conduction hearing aid. (2) Class II for the bone-conduction hearing aid.
| Device Class: | Class II |
| --- | --- |
| Classification Panel: | Ear, Nose & Throat Devices Panel |
| Classification Code: | 77LXB |
david/ear3.doc
11
{1}
510(k) Premarket Notification - Original Submission
Second Ear® Bone Conduction Hearing Aid
March 26, 1996
## 7. Substantially Equivalent Device(s):
1. Starkey Model BC1 Bone Conduction CROS Hearing Aid
2. Unitron® Bone Conduction Hearing Aid
3. Radioear B70 Series Audio Transducers
## 8. Device Description:
The EC-1000 Second Ear® is a bone conduction-type hearing aid. Unlike conventional products which depend upon acoustic coupling through air, the EC-1000 Second Ear® is based on bone conduction technology. Sound is transmitted directly through the bones of the skull to the cochlea, bypassing the outer and middle ear.
The EC-1000 Second Ear® offers an innovative tone control system, providing flexible audio control adjustments. Also, features such as a voice filter which when selected, reduces additional exterior sound. As an added feature an auxiliary input jack allows the user to connect Second Ear® directly to TV, CD players, portable radios, tape players or even telephones.
State of the art nickel-metal-hydride rechargeable battery technology is used for maximum battery life while minimizing size and weight.
## Transducer
The EC-1000 Second Ear® consists of a transducer unit, a shirt pocket sized audio processor/driver module and a wire cable for connecting the transducer with the audio processor/driver. The transducer is a convenient sized unit that is held against the head and is driven electrically to transmit audio vibrations to the underlying bones of the skull. It is a low-impedance device, similar to a loudspeaker.
## Audio Processor/Driver
The audio processor/driver module connects the transducer through the wire cable and usually would be carried in the users pocket. The wire cable connecting the audio processor/driver and the transducer can be placed comfortably at sensitive bone-conductive location such as behind-ear.
The audio processor/driver module is completely self-contained and is battery powered. It also has a built-in microphone for picking up sound. A rechargeable battery pack is the power source.
## Battery Pack
The battery pack can be charged in place in the unit or snapped out for easy exchanging. By purchasing a spare battery pack, the user can be using one battery while charging the other, guaranteeing uninterrupted use of the EC-1000. The user should be able to obtain 24 to 48 hours of operation in normal use from fully charged batteries. Volume setting and environmental conditions affect battery life.
david/ear3.doc
12
{2}
510(k) Premarket Notification - Original Submission
Second Eat® Bone Conduction Hearing Aid
March 26, 1996
## Packaging:
Standard packaging is used in the production of this device. Packaging quality is sufficient to protect the device from damage during transportation and storage. When not in use, the device should be stored in the same package as when it was supplied.
## Equivalence:
The Second Eat® Bone Conduction Hearing Aid is equivalent to the Starkey Labs, the Unitron brand bone conduction hearing aids and the Radioear Audio Transducer. The referenced units provide amplified auditory information via bone conduction as does the Second Eat®. All units are intended for use by persons with moderate to severe conductive hearing losses compounded by congenital or secondary obstruction of auditory air conduction mechanisms.
The frequency response of the Second Eat® Bone Conduction Hearing Aid without the voice filter is 150 Hz to 8 KHz. When the voice filter is engaged, response is 300 Hz to 3 KHz. The frequency response of the Unitron® brand bone conduction hearing aids is between 100 Hz and 6.6 KHz depending upon the unit selected. The Second Eat® is also equipped with treble and bass controls.
Power requirements for the Second Eat® Bone Conduction Hearing Aid are supplied by a Nickel-Metal-Hydride battery. The Second Eat® Bone Conduction Hearing Aid is provided with a 12 V, 750mA AC adapter and battery charger while the competitive brands typically utilize disposable batteries. Fully charged battery capacity is 600mA-hrs for the Second Eat® and is approximately the same for the competitive bone conduction hearing aids.
The sound conduction transducers of the Second Eat® and the comparison units are placed against the users skin behind the ear or other bony location of the skull with the aid of a strap or headband. Maximum voltage gain for the Second Eat® Bone Conduction Hearing Aid is 57dB. Overall S/N ratio for the Second Eat® is 40dB.
## Conclusion:
A review of the provided performance specifications for the Second Eat® and the listed comparison devices shows that all three devices are substantially the same in terms of indications for use, function, form, design, materials and performance. The feature comparison table below summarizes this information.
david/ear3.doc
{3}
510(k) Premarket Notification - Original Submission
Second Ear® Bone Conduction Hearing Aid
March 26, 1996
FEATURE COMPARISON TABLE
| FEATURE | Second Ear® | Unitron | Radioear | Starkey Labs. | S. E.? |
| --- | --- | --- | --- | --- | --- |
| Materials: | Medical grade plastic | Same | Same | Same | Yes |
| Indications for Use: | Moderate to severe conductive hearing losses.
Particularly useful for conductive losses compounded by congenital or secondary obstruction of auditory air conduction mechanisms | Same | Same | Same | Yes |
| Power Requirement: | 4.8 VDC Nickel-Metal-Hydride Battery | R675 & R6-AA Nickel Metal & Alkaline Battery | NiCad, R675 & R6-AA Nickel Metal & Alkaline Battery | NiCad. | Yes |
| Frequency Response | 150Hz to 8 KHz
300Hz to 3 Khz w/filter | 100 Hz to 5.5 KHz | 200Hz to 4KHz | Not Stated | Yes |
| Max. Gain | 57 dB | 60-86 dB depending on model | Not Stated | Not Stated | Yes |
| Device Class | II | II | II | II | Yes |
| Manufacturer: | WORDCOMP INTERNAT. | Unitron | Radioear | Starkey Labs, Inc. | Yes |
| Classification Code: | 77LXB | Same | Same | Same | Yes |
| K-Number | Pending | K884288 | Unknown | K923784 | NA |
david/ear3.doc
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.