K161671 · Oticon Medical AB · LXB · Sep 29, 2016 · Ear, Nose, Throat
Device Facts
Record ID
K161671
Device Name
Ponto 3, Ponto 3 Power and Ponto 3 SuperPower
Applicant
Oticon Medical AB
Product Code
LXB · Ear, Nose, Throat
Decision Date
Sep 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ponto 3 sound processors are intended for improvement of hearing for patients with conductive or mixed hearing losses, whether unilaterally fitted, or for those with single sided deafness.
Device Story
Bone-anchored hearing system; sound processor connects to skull-implanted abutment. Input: ambient sound via microphones. Processing: Inium Sense platform performs signal processing, microphone directionality, and binaural coordination. Output: vibrations transmitted through skull to cochlea via bone conduction. Used in clinic/home; operated by patient. SuperPower model uses UltraDrive amplifier/transducer for higher output. Wireless transmission enables binaural processing. Healthcare provider uses computer-based fitting software to adjust settings to patient requirements. Benefits: improved hearing for conductive/mixed loss or single-sided deafness.
Clinical Evidence
Bench testing only. Included functional testing, firmware validation, EMC emissions/immunity, ESD, and electroacoustical verification (battery, frequency range, Peak OFLs, THD, equivalent input noise). Binaural coordination and microphone directionality verified. MFO and gain of Ponto 3 SuperPower compared to Baha Cordelle II.
Technological Characteristics
Bone-anchored sound processor; Inium Sense platform; UltraDrive amplifier (SuperPower); wireless connectivity for binaural processing; computer-based fitting system; coupling for skin-penetrating abutment or headband accessory.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (PTA BC threshold ≤45 dB HL for Ponto 3, ≤55 dB HL for Ponto 3 Power, ≤65 dB HL for Ponto 3 SuperPower) or single-sided deafness (PTA AC threshold of hearing ear ≤20 dB HL). Bilateral fitting for symmetric loss. Contraindicated for patients under age 5.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings and body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2016
Oticon Medical AB Ms. Satu Hjartstam Quality Assurance and Regulatory Affairs Manager Datavagen 37b Askim, SE-436 32 SE
Re: K161671
Trade/Device Name: Ponto 3, Ponto 3 Power and Ponto 3 SuperPower Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: August 31, 2016 Received: September 2, 2016
Dear Ms. Hjartstam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161671
#### Device Name
Ponto 3, Ponto 3 Power, Ponto 3 SuperPower
Indications for Use (Describe)
Intended use:
The Ponto 3 sound processors are intended for improvements with conductive or mixed hearing losses, whether unilaterally or bilaterally fitted, or for those with single sided deafness.
Indications for use:
Ponto 3, Ponto 3 Power and Ponto 3 SuperPower are intended for the following patients and indications:
· Patient with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto 3 sound processor, 55 dB HL for use with the Ponto 3 Power sound processor and 65 dB HL for use with the Ponto 3 SuperPower sound processor.
· Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
· Indicated for patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
· Also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
The placement of a bone anchored implant is contraindicated for patient below the age of 5.
The Ponto 3, Ponto 3 Power and Ponto 3 SuperPower sound processors are intended to be used with either the Ponto implant system or with specific compatible BAHA abutments/implants from Cochlear Bone Anchored Solutions (BAS) (refer to the Ponto 3 labeling for specific compatible Cochlear models). In addition, selected Cochlear Baha sound processors can be used with the Ponto implant/abutment system (refer to the Ponto labeling for compatible Baha sound processor models).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K161671
## Ponto 3, Ponto 3 Power and Ponto 3 SuperPower
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
| Submitter name: | Oticon Medical AB |
|-----------------|---------------------------------------------|
| Address: | Datavägen 37 B<br>SE-436 32 Askim<br>Sweden |
| Phone: | +46 31 748 6100 |
| Facsimile: | +46 31 687 756 |
| Contact Person: | Satu Hjärtstam |
| Mobile phone: | +46 735 44 98 13 |
| Date Prepared: | Sep 15, 2016 |
### Name of Device and Name/Address of Manufacturer
Ponto 3, Ponto 3 Power and Ponto 3 SuperPower
Oticon Medical AB Datavägen 37 B SE-436 32 Askim Sweden
Common or Usual Name: Ponto bone anchored hearing system
Classification Name: Hearing aid, bone conduction
Classification Regulation: 21 C.F.R. §874.3300 (Product codes LXB, MAH)
#### Predicate Devices
| Device | 510(k) no. | Manufacturer |
|----------------------------------------------------------------------|-------------------|-------------------|
| Ponto Plus and Ponto Plus Power<br>(Primary predicate device) | K132775 | Oticon Medical AB |
| Baha Cordelle II<br>(Additional predicate device) | K080363 | Cochlear BAS AB |
| Alta 2 (Air conduction hearing aid)<br>(Additional predicate device) | Exempt from 510k. | Oticon A/S |
{4}------------------------------------------------
## Intended Use / Indications for Use
#### Intended use
The Ponto 3 sound processors are intended for improvement of hearing for patients with conductive or mixed hearing losses, whether unilaterally fitted, or for those with single sided deafness.
### Indications for use
Ponto 3, Ponto 3 Power and Ponto 3 SuperPower are intended for the following patients and indications:
- . Patient with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto 3 sound processor, 55 dB HL for use with the Ponto 3 Power sound processor and 65 dB HL for use with the Ponto 3 SuperPower sound processor.
- Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed ● hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
- Indicated for patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
- Also indicated for any patient who is indicated for an air-conduction contralateral routing of . signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
The placement of a bone anchored implant is contraindicated for patient below the age of 5.
The Ponto 3, Ponto 3 Power and Ponto 3 SuperPower sound processors are intended to be used with either the Ponto implant system or with specific compatible BAHA abutments/implants from Cochlear Bone Anchored Solutions (BAS) (refer to the Ponto 3 labeling for specific compatible Cochlear models). In addition, selected Cochlear Baha sound processors can be used with the Ponto implant/abutment system (refer to the Ponto labeling for compatible Baha sound processor models).
## Technological Characteristics
The Ponto 3, Ponto 3 Power and Ponto 3 SuperPower are modifications of the previously cleared Ponto Plus and Ponto Plus Power (K132775) and represent three latest sound processor models in the Ponto bone anchored hearing system. (Hereinafter referred to as the Ponto 3 sound processors or as the Ponto 3 family when referring to all three models and Ponto 3, Ponto 3 Power and Ponto 3 SuperPower when referring to the 3 different models.) A bone anchored hearing system consists of a sound processor connected to an implant with a skin penetrating abutment. The implant is surgically anchored in the skull bone behind the ear. Vibrations generated by the sound processor are transmitted via the implant directly through the skull bone to the cochlea as bone conduction sound. The sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. Alternatively, it can be connected to head band accessories, to function as a conventional bone conductor. Using a computer based fitting system the sound processor can be adjusted to the patient's individual hearing requirements. The Ponto 3
{5}------------------------------------------------
family is intended to be used with either the Ponto implant system or with specific compatible BAHA abutments/implants from Cochlear Bone Anchored Solutions (BAS).
The Ponto 3 family incorporates Inium Sense platform that is also used in the predicate Oticon A/S air conduction hearing aids legally marketed as class II. 510(k)-exempt devices. With the Injum Sense platform the microphone directionality system is updated. Furthermore, for Ponto 3 wireless transmission is enabled allowing for binaural signal processing.
In addition to the regular and power version sound processors, the Ponto 3 family introduces a SuperPower version with UltraDrive, consisting of a redesigned amplifier and optimized transducer. Ponto 3 SuperPower provides equivalent maximum force output (MFO) as the Baha Cordelle II, previously cleared through K080363, and thus expanded indication up to and including 65 dB HL is proposed.
This submission also includes minor modifications to the previously cleared accessories Genie Medical fitting software, Soft band and Streamer.
## Performance data
Testing of the Ponto 3 sound processors includes functional testing and firmware validation.
Ponto 3 SuperPower sound processors have been tested to (1) not emit excessive amounts of electromagnetic energy (EMC emissions); (2) operate as intended without performance degradation in the presence of an electromagnetic disturbance (EMC immunity) and ESD. Also electroacoustical verification has been conducted including tests for battery voltage and current consumption, frequency range, Peak OFLs at 90, 60 and 50 dB SPL, total harmonic distortion and equivalent input noise. The current consumption and battery voltage have been verified for the Ponto 3 and Ponto 3 Power sound processors. The Ponto 3 Power hardware is identical to Ponto Plus and Ponto Plus Power hardware. Update from the previously cleared Inium platform to Inium Sense and addition of the new sound processing features do not affect the electromagnetic compatibility or acoustical performance and thus no new verification tests regarding EMC or electroacoustical performance have been carried out on the Ponto 3 and Ponto 3 Power sound processors. Furthermore, bench tests were conducted to verify the binaural coordination and updated microphone directionality system.
The above mentioned tests verify that the Ponto 3 sound processors are equivalent to the Ponto Plus sound processors and that the binaural signal processing and directionality features are functionally equivalent to the same of the predicate Oticon A/S air conduction hearing aids legally marketed as class II, 510(k)-exempt devices. Moreover, the maximum output force measurements show that the MFO and gain of the Ponto 3 SuperPower is equivalent or higher when compared to the output of the predicate Baha Cordelle II sound processor. In all instances, the Ponto 3 sound processors functioned as intended and the performance observed was as expected. Hence we have come to the conclusion that further testing will not raise new issues of safety and efficacy.
## Substantial Equivalence
The Ponto 3 sound processors have the same intended use as the Ponto Plus sound processors and Ponto 3 and Ponto 3 Power have the same indications as the Ponto Plus and Ponto Plus Power, respectively. Ponto 3 SuperPower has the same indications as Ponto Plus Power, expect the bone-conduction hearing threshold that is 55 dB HL for Ponto Plus Power and 65 dB HL for Ponto 3 SuperPower. Indications of the Ponto 3 SuperPower are the same as those of the Baha Cordelle II. The principles of operation of the Ponto 3 sound processors are the same as those of the Ponto Plus sound processors.
{6}------------------------------------------------
The sound processing platform and the wireless technology in the Ponto 3 sound processors are the same as those used in the legally marketed, class II 510(k)-exempt Oticon A/S air conduction hearing aids. The maximum force output and gain of the Ponto 3 SuperPower are equivalent or higher in comparison to those provided by the Baha Cordelle II. The minor technological differences between the Ponto 3 sound processors and their predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Ponto 3 sound processors are as safe and effective as the Ponto Plus sound processors and that the Ponto 3 SuperPower is as safe and effective as the Baha Cordelle II sound processor. Thus the Ponto 3 sound processors are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.