K090996 · Oticon Medical AB · LXB · Jul 29, 2009 · Ear, Nose, Throat
Device Facts
Record ID
K090996
Device Name
PONTO PRO
Applicant
Oticon Medical AB
Product Code
LXB · Ear, Nose, Throat
Decision Date
Jul 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ponto Pro sound processor is intended for improvement of hearing for patients with conductive and mixed hearing losses, bilateral fitting and single sided deafness.
Device Story
Ponto Pro is a bone conduction hearing aid. It captures sound via a processor, which the user connects to a surgically implanted titanium abutment behind the ear. The processor converts sound into mechanical vibrations, transmitted through the skull to the cochlea. Alternatively, it can be used with a headband. An audiologist uses a computer-based fitting system to adjust the device to the patient's specific audiogram and hearing needs. The device improves hearing for patients with conductive, mixed, or single-sided deafness. It is designed for use by patients in daily life, with professional fitting in a clinical setting. It is compatible with Oticon Medical and Cochlear BAS bone-anchored implant systems.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by functional testing and software validation comparing the Ponto Pro to the predicate OBC device.
Technological Characteristics
Bone conduction sound processor; PEEK coupling; titanium implant; zinc-air battery; 125 Hz-8 kHz frequency response; 33 dB max gain; digital sound processing; computer-based fitting system; compatible with snap-coupling abutments.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (PTA BC threshold better than 45 dB HL at 0.5, 1, 2, 3 kHz); bilateral fitting for patients with symmetric conductive/mixed loss (BC threshold difference <10 dB avg at 0.5, 1, 2, 4 kHz or <15 dB at individual frequencies); and patients with profound sensorineural hearing loss in one ear and normal hearing in the other (AC threshold of hearing ear better than 20 dB HL at 0.5, 1, 2, 3 kHz). Contraindicated for patients under age 5.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
K090196
# SECTION 2. SUMMARY AND CERTIFICATION
### 510(K) SUMMARY ব
JUL 2 9 2002
### Summary of Safety and Effectiveness
In accordance with 21 CFR 807.92, the following information constitutes the Oticon Medical summary for the Ponto Pro bone anchored sound processor
| SUBMITTER'S NAME: | Oticon Medical AB |
|---------------------|------------------------------------------|
| ADDRESS: | Ekonomiv. 2<br>SE-436 33 Askim<br>Sweden |
| CONTACT PERSON: | Karolin Isberg |
| TELEPHONE NUMBER: | +46 31 748 6153 |
| FAX NUMBER: | +46 31 687 756 |
| E-MAIL: | kai@oticonmedical.se |
| DATE OF SUBMISSION: | April 6, 2009 |
#### 1. Identification of device
Proprietary Name: Ponto Pro Common Name: Hearing Aid, Bone Conduction Classification Status: Class II per regulations 21 CFR § 874.3300 Product Codes: LXB
#### 2. Equivalent devices
Oticon Medical believes that the Ponto Pro, regarding intended use, function and procedure, is substantially equivalent to the OBC cleared in 510(k) K082108. has substantially equivalent fitting technology as air conduction hearing aids with digital sound processing (exempt from 510(k)) and has substantially equivalent accessories for direct audio input as the Baha® Divino cleared in 510(k) K042017.
#### 3. Description of the Device
The Ponto Pro is a bone conduction-type hearing aid. The Ponto Processor is connected to an implant with a skin penetrating abutment which as been surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The Ponto Pro sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. The Ponto Pro can alternatively be connected to head band accessories, to function as a conventional bone conductor. Using a computer based fitting system the Ponto Pro can be adjusted to the patient's individual hearing requirements. The sound processor is intended to work with the Oticon Medical bone anchored implant system or the Baha® Abutment snap coupling from Cochlear BAS (ref no. 90410, 90305, 90434, 90480). In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha Intenso"" Baha Cordelle II, Baha BP100; ref no. 90500, 90510, 90511, 90502, 90512, 90503, 90513, 90730, 90731, 90732, 90733, HBC 400-0, HCB 401-0, HCB 402-0, 91301, 91302, 91303, 91304, 91305).
{1}------------------------------------------------
### 4. Intended use
The Ponto Pro sound processor is intended for improvement of hearing for patients with conductive and mixed hearing losses, bilateral fitting and single sided deafness.
### ડ. Technological characteristics, comparison to predicate device
| Characteristics | Ponto Pro | OBC | BAHA Divino | Air conduction<br>Hearing Aids<br>with digital<br>sound processing | S/Eq |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------|
| Design | Bone conduction<br>sound processor<br>connected to an<br>implant which has<br>been surgically<br>placed in the bone<br>behind the ear | Bone conduction<br>sound processor<br>connected to an<br>implant which has<br>been surgically<br>placed in the bone<br>behind the ear | Bone conduction<br>sound processor<br>connected to an<br>implant which has<br>been surgically<br>placed in the bone<br>behind the ear | Air conduction<br>sound processor<br>placed behind the<br>ear or in the ear<br>canal. | Yes |
| Intended use | Improvement of<br>hearing for<br>patients with<br>conductive and<br>mixed hearing<br>losses, bilateral<br>fitting and single<br>sided deafness | Improvement of<br>hearing for<br>patients with<br>conductive and<br>mixed hearing<br>losses, bilateral<br>fitting and single<br>sided deafness | Improvement of<br>hearing for<br>patients with<br>conductive and<br>mixed hearing<br>losses, bilateral<br>fitting and single<br>sided deafness | Improvement of<br>hearing | Yes |
| Material | Sound processor<br>coupling: PEEK<br>Implant: Titanium | Sound processor<br>coupling: PEEK<br>Implant: Titanium | Sound processor<br>coupling: PEEK<br>Implant: Titanium | Multiple | Yes |
| Power<br>requirement | Zinc-air battery | Zinc-air battery | Zinc-air battery | Zinc-air battery | Yes |
| Max gain | 33dB | 33dB | 33dB | N/A | Yes |
| Frequency<br>response | 125 Hz-8 kHz | 125 Hz - 8 kHz | 125 Hz - 8 kHz | N/A | Yes |
| Sound processing | Digital | Digital | Digital | Digital | Yes |
| Patient fitting | Individual<br>adjustment to<br>patient audiogram<br>and needs by a<br>computer based<br>fitting system used<br>by the audiologist | Individual<br>adjustment to<br>patient audiogram<br>and needs by a<br>computer based<br>fitting system at<br>the manufacturer<br>on request from<br>the audiologist. | Individual<br>adjustment to<br>patient audiogram<br>and needs by<br>potentiometers<br>adjusted by the<br>audiologist using a<br>screwdriver. | Individual<br>adjustment to<br>patient audiogram<br>and needs by a<br>computer based<br>fitting system used<br>by the audiologist | Yes |
| Accessories | Various<br>accessories for<br>direct audio input<br>(DAI) | Various<br>accessories for<br>direct audio input<br>(DAI) | Various<br>accessories for<br>direct audio input<br>(DAI) | N/A | Yes |
| Manufacturer | Oticon Medical<br>AB | Oticon Medical<br>AB | Entific Medical<br>Systems AB | N/A | |
| K-number | No number yet | K082108 | K042017 | Exempt from<br>510(k) | |
### Comnarison table
#### 6. Discussion of testing
Testing of the Ponto Pro include functional testing and software validation. These tests verify that the Ponto Pro is functionally equivalent to the OBC, why we have come to the conclusion that further testing will not raise new issues of safety and efficacy.
### 7. Conclusion
Based on the comparison to the predicate devices, the Oticon Medical Ponto Pro, is substantially equivalent to devices already on the market, both cleared by and exempt from the 510(k) process, and presents no new concerns about safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Oticon Medical AB C/O Karolin Isberg Quality and Regulatory Affairs Manager Ekomomiv. 2 SE-436 33 Askim Sweden
JUL 2 9 2009
Re: K090996
Trade/Device Name: Ponto Pro Bone Anchored Sound Processor with Accessories Regulation Number: 21 CFR 874,3300 Regulation Name: Hearing Aid, Bone Conduction Regulatory Class: Il Product Code: LXB Dated: July 10, 2009 Received: July 14, 2009
Dear Ms. Isberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address. http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Ra. An. no/p
Malvina B. Eydelman, M.I Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Kojo 996
## Indications for Use
510(k) Number (if known): K090996
Device Name: Ponto Pro
Indications For Use:
The Ponto Pro is intended for the following patients and indications;
- Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2 and 3 kHz).
- I Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
- l Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than 20 dB HL (measured at 0.5. 1, 2 and 3 kHz).
The placement of a bone anchored implant is contraindicated for patients below the age of 5.
The Ponto Pro sound processor is intended to be connected to the Oticon Medical bone anchored implant system or to the Baha® Abutment snap coupling from Cochlear BAS. In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha IntensoTM, Baha Cordelle II, Baha BP100),
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jams K. Kane ph.D
Page 1 of 1
(Division Sign-Off) vision of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K090996
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.