K090720 · Cochlear Bone Anchored Systems AB · LXB · Jun 17, 2009 · Ear, Nose, Throat
Device Facts
Record ID
K090720
Device Name
COCHLEAR BAHA BP100
Applicant
Cochlear Bone Anchored Systems AB
Product Code
LXB · Ear, Nose, Throat
Decision Date
Jun 17, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K090720 · Jun 17, 2009
COCHLEAR BAHA BP100
Cochlear Bone Anchored Systems AB
Published clinical literature (Kompis et al., 2007)
The sponsor utilized published clinical literature reporting on the performance of the predicate device (Baha Divino) in a clinical setting to support the safety and efficacy of the new Baha BP100 device, arguing that bench testing equivalence to the predicate justifies the clinical performance.
Literature review; Clinical performance; Predicate comparison
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Kompis et al., 2007 (Speech understanding in quiet and in noise with the bone-anchored hearing aids Baha Compact and Baha Divino); Clinical study (literature)
Patients with conductive or mixed hearing loss
Not applicable for this study
Speech understanding in quiet and in noise
Indications for Use
The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications: Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL. Bilateral fitting of the BP100 is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies. Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL. Baha for SSD is also indicated for any patient who is indicated for an airconduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Device Story
Baha BP100 is an ear-level sound processor for bone-conduction hearing. It captures sound via microphone; processes signals using digital signal processing (DSP) algorithms; and transmits vibrations to the skull via a titanium implant (osseointegrated) or external headband/softband. Features include multi-band wide-dynamic range compression, automatic noise management, automatic directionality, and active feedback cancellation. Operated by patient or clinician; settings adjusted manually or via computer-based fitting software. Provides three user-selectable programs for different environments (quiet, noise, music). Benefits include improved sound amplification for patients with conductive/mixed hearing loss or single-sided deafness who cannot use air-conduction CROS aids. Includes datalogging, FM/telecoil compatibility, and direct audio input.
Clinical Evidence
Bench testing only. No new clinical data provided. Equivalence to Baha Divino (supported by literature, e.g., Kompis et al., 2007) is established via bench testing showing comparable gain/output performance for the specified patient population (bone-conduction thresholds < 45 dB HL).
Technological Characteristics
External sound processor; digital signal processing; titanium implant (ASTM F67-06); multi-band wide-dynamic range compression; automatic noise management; automatic directionality; active feedback cancellation. Connectivity: FM, telecoil, direct audio input. Fitting: manual or software-based (HI-PRO/NOAH interface). Standards: ISO 13485, ISO 60601-1, ISO 14971, IEC 60118-7, IEC 60188-9, ANSI S3.22.
Indications for Use
Indicated for patients >5 years (implant) or any age (headband/softband) with conductive/mixed hearing loss (bone-conduction thresholds ≤45 dB HL) or single-sided sensorineural deafness (normal hearing in contralateral ear).
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Resound digital 5000 series hearing devices (K984547)
Submission Summary (Full Text)
{1}------------------------------------------------
K090720
Cochlear Americas
510(k) Premarket Notification Cochlear Baha BP100
## 510(k) Summary
## JUN 1 7 2009
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided.
#### Submitted by:
Cochlear Americas 13059 East Peakview Ave. Centennial, CO 80111
#### On behalf of:
Cochlear Bone Anchored Solutions AB Konstruktionsvägen 14 SE-435 33 Mölnlycke Sweden
#### Contact Persons:
Sean Bundy Manager, Regulatory Affairs Phone: 303-524-7139 (office) 303-792-9025 (facsimile) Email: SBundy@cochlear.com
Carol A. Sammeth, Ph.D., CCC-A/FAAA Senior Regulatory/Clinical Specialist Phone: 303-524-6833 (office) Email: CSammeth@cochlear.com
## Date Submission Prepared:
March 2009
#### Device Name:
| Trade or Proprietary Name: | Cochlear Baha® BP100 |
|----------------------------|--------------------------------------------------|
| Common or Usual Name: | Hearing Aid (Bone Conduction) |
| Classification Status: | Class II, 21 CFR §874.3300<br>Product Codes: LXB |
| Panel: | Ear Nose and Throat Specialty Panel |
Panel:
{2}------------------------------------------------
#### Predicate Devices
The predicate devices for the Baha BP100 sound processor are the three currently marketed sound processors in the family of processors available for use with the Baha® auditory osseointegrated implant; namely, the Baha Divino™ (510(k) K042017), Baha IntensoTM (510(k) K081606), and Baha Cordelle II (510(k) K080363). Bench testing shows that the gain/output of the Baha BP100 are comparable to those for the marketed Baha Divino sound processor, and the intended use and technological characteristics of the Baha BP100 are substantially equivalent to the other three marketed sound processors. See Table 2 under Section XI of this submission for a detailed comparison of technological characteristics and features across the Baha family of sound processors used with the auditory osseointegrated implant.
Unique to the new Baha BP100 sound processor is that it can be programmed with fitting software, as well as manually adjusted like the predicate devices. The programming software and interface (described in Section XV of this submission) is substantially equivalent to other programming software that has been used for fitting of hearing devices that have digital signal processing algorithms. The fitting software, used on a standard computer, interfaces with the sound processor via cables to a standard HI-PRO box or NOAH-link, and has the same intended use and works in the same manner as that of many programmable hearing aids and devices. Predicate devices proposed for the fitting software system are 510(k) number K973880 for the Sonix hearing aid fitting and programming system, 510(k) number K931372 for the Resound® portable prescriptive programming system, and 510(k) number K942749 for the HI-PRO/NOAH universal hearing instrument by Madsen Electronics.
Also unique to this new Baha sound processor is the ability to measure bone-conduction thresholds directly via pure tones generated by the sound processor. The intended use and process to accomplish these measurements is also substantially equivalent to a procedure used by a previous hearing device cleared for marketing under the 510(k) process. Specifically, the predicate for obtaining data to assist in more accurate fittings by obtaining measurements for pure tones produced by the hearing device can be found in 510(k) number K984547 for the LGOB test using the Resound® digital 5000 series hearing devices. Although that approach uses suprathreshold measures to determine perceived loudness and the BP100 direct bone conduction measurements use threshold-level stimuli, the production of pure tones through the hearing device for standard audiometric measurements to be used in a prescription has the same intended use, function, and technology as the predicate.
#### Device Description
The Baha implant system works by combining a sound processor with an abutment and a small titanium implant placed in the skull behind the ear. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the
{3}------------------------------------------------
Cochlear Americas
510(k) Premarket Notification Cochlear Baha BP100
implant. Thus, the titanium implant becomes one with the bone, allowing high-quality amplified and processed sound to be conducted via the skull bone directly to a cochlea with residual functionality. Baha sound processors can also be fitted as external (non-implanted) bone-conduction vibrators with either the Baha Headband or Baha Softband as an alternative to a conventional bone-conduction hearing aid.
The Baha BP100 will be the newest addition to a current family of three marketed sound processors for use with the Baha auditory osseointegrated implant, or with the Baha Headband/Softband setup. In its initial release, it will provide a moderate gain ear-level sound processor that will serve as a premium-feature option to the currently marketed Baha Divino. At a later date, the BP100 platform may be used to provide a range of programmable products to meet end-user needs.
#### Intended Use
The Baha system is indicated for patients who have conductive or mixed hearing loss, and who can still benefit from sound amplification. Patients with bilaterally symmetric conductive or mixed hearing loss may be implanted bilaterally. The Baha system is also indicated for patients with sensorineural deafness in one ear and normal hearing in the other ear (i.e. single-sided deafness or "SSD"), and patients who are candidates for an airconduction contralateral routing of signals (AC CROS) hearing aid but who for some reason cannot or will not wear an AC CROS device.
For the Baha implant, patients must be > 5 years old (per 510(k) K984162). They also must (either by themselves or with the aid of others) be able to maintain hygiene of the abutment/skin interface of the Baha. Finally, they should have sufficient bone volume and bone quality to support successful fixture placement. For patients not meeting these criteria for implantation, the Baha sound processor can be fit externally (non-implanted) with the Baha Headband or Baha Softband for transcutaneous bone-conduction stimulation.
The BP100 offers gain/output that is comparable to that supplied by the currently marked Baha Divino sound processor. Thus, it will be marketed for use for conductive or mixed hearing loss patients who have average bone-conduction thresholds up to < 45 dB HL (across 0.5, 1, 2, and 3 kHz), the same Indications For Use as cleared for marketing of the Baha Divino.
#### Technological Characteristics
Baha BP100 is an external sound processor that utilizes digital signal processing with a number of advanced automatic features, for use with the Baha auditory osseointegrated implant, or with the Baha Headband/Softband. It has substantially equivalent technology, and intended use, as the other three currently marketed sound processors for the Baha system (Baha Divino, Baha Intenso, Baha Cordelle II). The external sound processors of the
{4}------------------------------------------------
Cochlear Americas ·
Baha system differ in style, signal processing capability, features, and degree of available gain and output, but they are all interchangeable in that they all snap onto the abutment of the Baha auditory osseointegrated implant or are affixed to the skull using a Baha Headband/Softband.
Choice of Baha sound processor(s) depends on the individual needs and desires of a given patient. The Intenso is often chosen for patients who have greater gain needs than can be met with the Divino (i.e. those with mixed losses and poorer bone-conduction hearing thresholds), but who want the cosmetic discretion of wearing an ear-level device rather than the bulkier (body-worn) but most powerful Cordelle II. The BP100 will offer a "premium" alternative to the Baha Divino because it offers more advanced signal processing and available features, and can be programmed either manually via button pushing or with software (while the previously marketed sound processors are only able to be manually. adjusted). Specifically, Baha BP100 provides advanced capabilities including multi-band wide-dynamic range compression, automatic background noise management, automatic directionality, and active feedback cancellation. Three user-selectable programs are available to customize the amplification for different environments such as listening in quiet, in noise, or to music (or, alternately, one program can be dedicated to accessory use via the DAI input port). It also will provide a new design with high cosmetic appeal, and a number of other features to meet end-user needs, such as FM unit and telecoil compatibility, direct audio input via an adaptor, tamperproofing for children, and usage datalogging. Use of the (recommended, but optional) software fitting capability will provide greater fitting accuracy, especially for difficult-to-fit patients.
#### Non-Clinical Tests
Although Cochlear is not declaring conformity to these standards, note that the following standards (or relevant elements of these standards) were used as guidelines in the manufacturing and development process for the Baha BP100 implant:
- . ISO 13485 (Medical devices. Quality management systems. Requirements for regulatory purposes)
- ISO 60601-1 (Medical electrical equipment Part 1: General requirements for basic . safety and essential performance)
- ISO 11137-1 (Sterilization health care products. Requirements for validation and . routine control. Radiation sterilization)
- ISO 14971 (Medical devices. Application of risk management to medical devices) .
- . ISO 5832-2 (Implants for surgery - Metallic materials. Part 2: Unalloyed titanium)
- ISO 10993-1 (Biological evaluation of medical devices. Part 1: Evaluation and . testing)
- ISO 17664 (Sterilization of medical devices Information to be provided by the . manufacturer for the processing of resterilizable medical devices)
{5}------------------------------------------------
- . EN 10088-1 (Stainless steels. List of stainless steels)
- EN 552 (Sterilization of medical devices. Validation and routine control of . sterilization by irradiation)
- . EN 556-1 (Sterilization of medical devices) Requirements for medical devices to be designated "sterile". Requirements for terminally sterilized medical devices)
- . EN 868-1 (Packaging materials and systems for medical devices which are to be sterilized. Part 1: General requirements and tests methods, and Part 5: Heat and selfsealable pouches and reels of paper and plastic film construction. Requirements and tests methods)
- EN1174-1 (Sterilization of medical devices. Estimation of the population of micro-. organisms on product. Requirements)
- . EN 980 (Graphic symbols for use in the labeling of medical devices)
- EN 1041 (Information supplied by the manufacturer with medical devices) .
- . ASTM F67-06 (Standard specification for unalloyed titanium for surgical implant applications).
- IEC 60118-7 (Hearing aids: measurements of performance characteristics of hearing . aids for quality inspection for delivery purposes)
- IEC 60188-9 (Hearing aids: methods of measurements of characteristics of hearing . aids with bone vibrator output)
- ANSI S3.22 (Specification of hearing aid characteristics) .
#### Clinical Performance Data
Published data in the literature support the safety and efficacy of the Baha Divino for patients with average bone-conduction thresholds of < 45 dB HL (e.g. Kompis et al., 2007). Because bench testing shows that the Baha BP100 is essentially equivalent (or better) than the Baha Divino in terms of gain/output, we believe that further clinical performance data are unnecessary to prove that the BP100 can successfully fit the same patient population; i.e. those with bone-conduction thresholds averaged across 500, 1000, 2000, and 3000 Hz of < 45 dB HL.
#### Conclusions
The design and function of the Baha BP100 are substantially equivalent to previous Baha® sound processors cleared by the 510(k) process. The optional software fitting system for the Baha BP100 is substantially equivalent to that routinely used to fit conventional airconduction hearing aids that use digital signal processing, including direct bone conduction (DBC) measurements, and the manual fitting approach for the BP100 is substantially equivalent to that used with currently marketed Baha sound processors. In addition, testing of the Baha BP100, including inspectional, functional and environmental tests, verify that the device meets key requirements in the design specification. Bench testing supports the proposed Indications For Use, since gain/output are equivalent to the Baha Divino which is currently marketed under 510(k) K042017 for the same Indications For Use.
{6}------------------------------------------------
Cochlear Americas
.
.
.
510(k) Premarket Notification Cochlear Baha BP100
.
#### Reference:
い
·
:
Kompis, M, Krebs, M, Hausler, R. (2007). Speech understanding in quiet and in noise with the bone-anchored hearing aids Baha Compact and Baha Divino. Acta Otolaryngol; 127(8): 829-835.
.
.
. . . . .
{7}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes extending upwards and to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# 2009 7: 1 NUN
Cochlear Americas c/o Mr. Sean Bundy Manager, Regulatory Affairs 13059 East Peakview Ave. Centennial, CO 80111
Re: K090720
Trade/Device Name: Cochlear Baha® BP 100 Regulation Number: 21 CFR 874.3300 . Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: May 20, 2009 Received: May 21, 2009
Dear Mr. Bundy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
#### Page 2 - Mr. Sean Bundy
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known): K090720
Device Name: Cochlear Baha® BP100
#### Indications for Use Statement:
The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications:
- Patients who have a conductive or mixed hearing loss and can still benefit from . sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
- Bilateral fitting of the BP100 is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
- Baha for SSD is also indicated for any patient who is indicated for an airconduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Prescription Use (21 CFR 801 Subpart C) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kurent boken
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices
510(k) Number KD90720
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.