K130937 · Hansaton Akustik GmbH · KLW · Jan 3, 2014 · Ear, Nose, Throat
Device Facts
Record ID
K130937
Device Name
WAVE 2G, SOUL
Applicant
Hansaton Akustik GmbH
Product Code
KLW · Ear, Nose, Throat
Decision Date
Jan 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hansaton Tinnitus Maskers are air conduction broad band noise generators and hearing aids intended to output noise of sufficient intensity and bandwidth to be used for tinnitus habituation therapy and are suitable for tinnitus masking therapy. Optional, the Hansaton SOUL Tinnitus Maskers provide amplification, intended to be used by those individuals who experience tinnitus and desire amplification. Diagnosis and prescription of the tinnitus maskers must be performed by hearing health specialists (e.g., ENT specialists, audiologists, or hearing system professionals) who are experienced in tinnitus management. The target population is primarily the adult population over 18 years of age. The instruments are not intended for pediatric use.
Device Story
Hansaton WAVE 2G and SOUL are air-conduction broadband noise generators; SOUL includes hearing aid amplification. Available in In-The-Ear (ITE) and Behind-The-Ear (BTE) form factors. Operated by hearing health specialists (audiologists, ENT) for tinnitus habituation and masking therapy. Devices output noise of specific intensity/bandwidth to mask tinnitus; SOUL provides additional amplification for hearing loss. Clinicians program devices to patient needs; patients use devices daily to manage tinnitus symptoms. Benefits include reduced tinnitus perception and improved quality of life through habituation or masking.
Clinical Evidence
Bench testing only. Compliance with IEC 60118 series (electroacoustical characteristics, induction coils, AGC, electrical input, EMC) and ISO 10993 biocompatibility standards. Speech testing performed per Nordic Requirements 7th edition.
Technological Characteristics
Digital circuit; programmable; multiple programs/memories. Available as BTE (X-Mini, Slim) and ITE (Mini Canal, Cymba). Connectivity: SOUL supports wireless communication. Biocompatibility per ISO 10993. Standards: IEC 60118-0, 60118-1, 60118-2, 60118-6, 60118-13.
Indications for Use
Indicated for adults >18 years experiencing tinnitus, with or without hearing loss, requiring tinnitus habituation or masking therapy. Contraindicated for pediatric use.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{0}------------------------------------------------
ﺮ
# 510(k) Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act, Hansaton Akustik GmbH herewith submits a Summary of Safety and Effectiveness.
| Submitter Information: | Hansaton Akustik GmbH<br>Sachsenkamp 5<br>20097 Hamburg<br>Germany<br>Registration Number: 3004597869<br>Owner/Operator Number: 9042953 |
|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Mrs. Kathrin Hesebeck<br>Hansaton Akustik GmbH<br>Sachsenkamp 5<br>20097 Hamburg<br>Phone: +49 40 298 011-146 |
| US Agent (Contact): | Jane A. Flynn<br>209 Johnson Road<br>Morris Plains, NJ 07950 USA<br>Phone: +1 973 290 9866<br>E-mail: jflynnmpi@aol.com |
| Date Prepared: | January 03, 2014 |
| Device(s) Identification:<br>Device Trade Name:<br>Common Name: | WAVE 2G,<br>SOUL<br>Tinnitus Maskers |
| Classification of the device:<br>Device Classification Name:<br>Product Code:<br>Device Classification No.:<br>Panel:<br>Regulatory Status: | Masker, Tinnitus<br>KLW<br>Part 874.3400<br>Ear Nose & Throat<br>Class 2 |
{1}------------------------------------------------
#### Device Description:
The Hansaton Tinnitus Maskers are provided in two versions; the WAVE 2G, a noise generator intended to output noise for tinnitus habituation therapy and tinnitus masking therapy, and SOUL, a noise generator and hearing aid additionally providing amplification. Both, the WAVE 2G and the SOUL, systems are available as 'In The Ear' (ITE) and 'Behind The Ear' (BTE) versions.
#### Intended Use:
The Hansaton Tinnitus Maskers are air conduction broad band noise generators and hearing aids intended to output noise of sufficient intensity and bandwidth to be used for tinnitus habituation therapy and are suitable for tinnitus masking therapy. Optional, the Hansaton SOUL Tinnitus Maskers provide amplification, intended to be used by those individuals who experience tinnitus and desire amplification. Diagnosis and prescription of the tinnitus maskers must be performed by health specialists (e.g., ENT specialists, audiologists, or hearing system professionals) who are experienced in tinnitus management.
The target population is primarily the adult population over 18 years of age. The instruments are not intended for pediatric use.
#### Predicate devices:
| 1.<br>Device Trade Name:<br>Applicant:<br>510(k) No.: | TCI Combi<br>Siemens Hearing Instruments<br>K003558 | | Hansaton<br>WAVE 2G | Hansaton SOUL | TCI Combi<br>K003558 | TCI<br>K003559 | Custom TCI Combi<br>K011366 | Widexlink in<br>clear series<br>hearing aids<br>K101699 |
|-------------------------------------------------------|------------------------------------------------------------|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 2.<br>Device Trade Name:<br>Applicant:<br>510(k) No.: | TCI<br>Siemens Hearing Instruments<br>K003559 | Intended Use | The Hansaton<br>Tinnitus<br>Maskers are air<br>conduction<br>broad band<br>noise generators<br>and hearing aids<br>intended to<br>output noise of<br>sufficient<br>intensity and<br>bandwidth to be<br>used for tinnitus<br>habituation<br>therapy and are<br>suitable for<br>tinnitus masking<br>therapy.<br>Diagnosis and<br>prescription of<br>the tinnitus<br>maskers must<br>be performed by<br>hearing health<br>specialists (e.g.,<br>ENT specialists,<br>audiologists, or<br>hearing system<br>professionals)<br>who are<br>experienced in<br>tinnitus<br>management.<br>The target<br>population is<br>primarily the<br>adult population<br>over 18 years of<br>age. The<br>instruments are<br>not intended for<br>pediatric use. | The Hansaton<br>Tinnitus Maskers<br>are air<br>conduction broad<br>band noise<br>generators and<br>hearing aids<br>intended to<br>output noise of<br>sufficient<br>intensity and<br>bandwidth to be<br>used for tinnitus<br>habituation<br>therapy and are<br>suitable for<br>tinnitus masking<br>therapy.<br>Optional, the<br>Hansaton SOUL<br>Tinnitus Maskers<br>provide<br>amplification,<br>intended to be<br>used by those<br>individuals who<br>experience<br>tinnitus and<br>desire<br>amplification.<br>Diagnosis and<br>prescription of<br>the tinnitus<br>maskers must be<br>performed by<br>hearing health<br>specialists (e.g.,<br>ENT specialists,<br>audiologists, or<br>hearing system<br>professionals)<br>who are<br>experienced in<br>tinnitus<br>management.<br>The target<br>population is<br>primarily the<br>adult population<br>over 18 years of<br>age. The<br>instruments are<br>not intended for<br>pediatric use. | Mask tinnitus as<br>part of tinnitus<br>management<br>program<br>Provide<br>amplification for<br>compensation of<br>hearing loss | Mask tinnitus<br>as part of<br>tinnitus<br>management<br>program | Mask tinnitus as part<br>of tinnitus<br>management<br>program<br>Provide amplification<br>for compensation of<br>hearing loss | Mask tinnitus as<br>part of tinnitus<br>management<br>program |
| 3.<br>Device Trade Name:<br>Applicant:<br>510(k) No.: | Custom TCI Combi<br>Siemens Hearing Instruments<br>K011366 | Target<br>population | Individuals who<br>experience<br>tinnitus. The<br>target population<br>is primarily the<br>adult population<br>over 18 years of | Individuals who<br>experience<br>tinnitus and<br>desire<br>amplification.<br>The target<br>population is<br>primarily the | Adults and children<br>(≥ 5 years) with<br>tinnitus and<br>hearing loss that<br>are participating in<br>a tinnitus<br>management | Adults and<br>children (≥ 5<br>years) with<br>tinnitus that are<br>participating in<br>a tinnitus<br>management | Adults and children<br>(≥ 5 years) with<br>tinnitus that are<br>participating in a<br>tinnitus management<br>program | Individuals with a<br>full range of<br>hearing loss<br>severity (from<br>slight (16 to 25 dB<br>HL) to profound<br>(90+ dB HL)) and |
য :
WIDEXLINK IN CLEAR SERIES HEARING AIDS Device Trade Name: Applicant: Office Research in Clinical Amplification 510(k) No.: K101699
{2}------------------------------------------------
{3}------------------------------------------------
| | Hansaton<br>WAVE 2G | Hansaton SOUL | TCI Combi<br>K003558 | TCI<br>K003559 | Custom TCI Combi<br>K011366 | Widexlink in<br>clear series<br>hearing aids<br>K101699 |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | instruments are<br>not intended for<br>pediatric use. | adult population<br>over 18 years of<br>age. The<br>instruments are<br>not intended for<br>pediatric use. | | | | configurations. |
| Circuit type | Digital | Digital | Digital | Digital | Digital | Digital |
| Programmable | Yes | Yes | Yes | Yes | Yes | Yes |
| Multiple<br>programs/<br>memories | Yes<br>(except Cymba) | Yes<br>(except Cymba) | Yes | No | Yes | Yes |
| Available noises | Four | Four | One | Four | One | n/a |
| Volume control | Yes | Yes | Yes | Yes | Yes | Yes |
| Number of<br>channels | Eight | (G/AGC)<br>Business: 12/12<br>Economy:<br>8/4 | Four | n/a | n/a | n/a |
| Physical<br>description | Available as<br>standard behind-<br>the-ear (X-Mini<br>and Slim) and<br>in-the-ear<br>solutions (Mini<br>Canal and<br>Cymba):<br>Mini Canal is<br>custom made | Available as<br>standard behind-<br>the-ear (X-Mini<br>and Slim) and in-<br>the-ear solutions<br>(Mini Canal and<br>Cymba);<br>Mini Canal is<br>custom made | Standard behind-<br>the-ear instrument<br>housing | Standard<br>behind-the-ear<br>instrument<br>housing | Custom product,<br>available as in-the-<br>ear, half shell, and<br>in-the-canal shell<br>styles | n/a |
| RMS output<br>characteristics<br>(Noiser) | Mini Canal<br>80 dB<br><br>Slim<br>80/90 dB<br><br>X-Mini<br>(S-receiver)<br>70 dB<br><br>X- Mini<br>(M-receiver)<br>80 dB | Mini Canal: 75<br>dB<br><br>Slim:<br>75/85 dB<br>(Mini tube/<br>earhook)<br><br>X-Mini<br>(S-receiver):<br>65 dB<br><br>X-Mini<br>(M-receiver):<br>75 dB<br><br>X-Mini<br>(P-receiver)<br>85 dB | 102 dB Broadband<br>noise | White noise<br>71 dB SPL<br><br>Pink noise<br>69 dB SPL<br><br>Speech noise<br>69 dB SPL<br><br>High-tone<br>noise<br>76 dB SPL | ITE:<br>89 dB<br>Broadband noise<br><br>In-the-canal and half<br>shell:<br>84 dB Broadband<br>noise | n/a |
| RMS output<br>characteristics<br>(Hearing aid<br>amplifier) | n/a | (ANSI S3.22-<br>2003)<br>HF-Average<br>OSPL 90<br><br>Mini Canal: | 105 dB HF-<br>Average OSPL 90<br>(ANSI S3.22-1996) | n/a | ITE:<br>105 dB<br>HF-Average OSPL<br>90 (ANSI S3.22-<br>1996) (110/40/03<br>matrix) | n/a |
| | Hansaton WAVE 2G | Hansaton SOUL | TCI Combi<br>K003558 | TCI<br>K003559 | Custom TCI Combi<br>K011366 | Widexlink in clear series<br>hearing aids<br>K101699 |
| | | Slim:<br>116 dB<br><br>X-Mini<br>(S-receiver):<br>101 dB<br><br>X-Mini<br>(M-receiver):<br>114 dB<br><br>X-Mini<br>(P-receiver)<br>116 dB | | | In-the-canal and half<br>shell:<br>107 dB<br>HF-Average OSPL<br>90 (ANSI S3.22-<br>1996) (110/35/03<br>matrix) | |
| Volume control<br>range | Programmable:<br>OFF, 8 dB,<br>16 dB, 32 dB | Programmable:<br>OFF, 8 dB,<br>16 dB, 32 dB | Programmable:<br>OFF, 8 dB, 16 dB,<br>32 dB | Programmable:<br>OFF, 8 dB,<br>16 dB, 32 dB | Programmable:<br>OFF, 8 dB, 16 dB,<br>32 dB | n/a |
| Wireless<br>communication<br>of pair of<br>hearing aids or<br>device<br>accessories | No | Yes | No | No | No | Yes |
,
{4}------------------------------------------------
The Hansaton WAVE 2G and SOUL Tinnitus Maskers are considered substantial equivalent to the Siemens TCI Combi (K003558), Siemens TCI (K003559), Siemens Custom TCI Combi (K011366), and Office Research in Clinical Amplification's WIDEXLINK (K101699). There is no significant difference in intended use or technology.
### Summary of performance testing:
The Hansaton WAVE 2G and SOUL tinnitus maskers have been developed and tested in accordance with hearing aids specific standard IEC 60118 series.
- IEC 60118-0: Hearing aids Measurement of electroacoustical characteristics .
- IEC 60118-1: Hearing aids Hearing aids with induction pick-up coil input .
- IEC 60118-2: Hearing aids -Hearing aids with automatic gain control circuits .
- IEC 60118-6: Hearing aids -Characteristics of electrical input circuits for hearing aids .
- IEC 60118-13: Hearing aids -- Electromagnetic compatibility .
All patient contacting materials have been evaluated in terms of biocompatibility in accordance with the ISO 10993 standard series.
Additionally a speech test was performed in accordance with the Nordic Requirements 7th edition appendix A.
{5}------------------------------------------------
### Conclusion:
Hansaton Akustik GmbH believes that the Hansaton Tinnitus Maskers WAVE 2G and SOUL are substantially equivalent to the currently legally marketed devices. They do not introduce new indications for use, have the same technological characteristics and do not introduce new potential hazards or safety risks.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
N
Public Health Service
DEPARTMENT OF HEALTH & HUMAN SER
January 3, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hansaton Akustik GmbH c/o Mr. Nick Burmester Regulatory Affairs & Project Manager, Prosystem AG Beim Strohhause 17 Hamburg, Germany 20097
Re: K130937
Trade/Device Name: WAVE 2G, SOUL Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: November 18, 2013 Received: December 2, 2013
Dear Mr. Burmester:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{7}------------------------------------------------
Page 2 - Mr. Nick Burmester
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bradley S. Cunningham -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) number (if known): K130937
Device Name:
WAVE 2G, SOUL Tinnitus Masker
Indications for Use:
The Hansaton Tinnitus Maskers are air conduction broad band noise generators and hearing aids intended to output noise of sufficient intensity and bandwidth to be used for tinnitus habituation therapy and are suitable for tinnitus masking therapy. Optional, the Hansaton SOUL Tinnitus Maskers provide amplification, intended to be used by those individuals who experience tinnitus and desire amplification. Diagnosis and prescription of the tinnitus maskers must be performed by hearing health specialists (e.g., ENT specialists, audiologists, or hearing system professionals) who are experienced in tinnitus management.
The target population is primarily the adult population over 18 years of age. The instruments are not intended for pediatric use.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Srinivas Nandkumar -S
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.