AUDIFON ARRIVA CIC, IS, IS+, M, S, S+, X TRT, AUDIFON VICO, AUDIFON PRADO, AUDIFON ELIA, CIC, IS, IS+, M, S TRT
K130514 · Audifon-USA, Inc. · KLW · Nov 21, 2013 · Ear, Nose, Throat
Device Facts
Record ID
K130514
Device Name
AUDIFON ARRIVA CIC, IS, IS+, M, S, S+, X TRT, AUDIFON VICO, AUDIFON PRADO, AUDIFON ELIA, CIC, IS, IS+, M, S TRT
Applicant
Audifon-USA, Inc.
Product Code
KLW · Ear, Nose, Throat
Decision Date
Nov 21, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 874.3400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The combined devices are intended for the adult population suffering from a chronic persistent ringing in the ears (Tinnitus), who also need or desire amplification. The amplification suits the needs of a slight to a profound hearing loss (see table below). The products may also be used for masking Tinnitus as part of a Tinnitus Retraining Therapy (TRT) protocol and should be utilized only in consultation with a licensed hearing healthcare professional, who is trained in subsequent rehabilitation therapy, or a qualified audiologist.
Device Story
Digital hearing aids with integrated tinnitus noise generators; designed for Tinnitus Retraining Therapy (TRT). Devices process acoustic input to provide amplification for hearing loss and generate broadband noise for tinnitus masking/distraction. Operated by patients; fitted by hearing healthcare professionals using audifit software and HI-PRO programming interface. Features up to four programmable noise programs; adjustable noise shape (low/high-cut filters) and output levels. Housed in standard In-the-ear (CIC, IS, IS+) or behind-the-ear (S, S+, M, X) form factors. Output levels limited to <80 dB SPL for TRT and <85 dBA per OSHA guidelines; software includes warnings for high-level settings. Benefits include improved hearing and tinnitus management through combined amplification and masking.
Clinical Evidence
Bench testing only. Performance verified per ANSI S3.22-2009 standards. Non-clinical tests demonstrate equivalent acoustic characteristics, frequency range, and safety profile compared to the predicate device.
Technological Characteristics
Digital hearing aid and noise generator; programmable via HI-PRO and audifit software. Form factors: CIC, IS, IS+, M, S, S+, X. Power: 1.4V zinc-air batteries (sizes 10, 312, 13, 675). Connectivity: Wired programming interface. Standards: ANSI S3.22-2009. Features: 4-program capability, adjustable low/high-cut filters, programmable output levels.
Indications for Use
Indicated for adults with chronic persistent tinnitus who also require hearing amplification for slight to profound hearing loss. Contraindicated for use without consultation from a licensed hearing healthcare professional or audiologist trained in rehabilitation therapy.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{1}------------------------------------------------
KB0514
#### 5 510(k) Summary
- 1. Applicant's Name and Address:
audifon USA Inc. 403 Chairman Ct., Suite 1 Debary, Florida 32713 PO BOX 531700 USA
2. Contact Person: Jane E Perrone Phone: 386-6688812
3. Trade or Proprietary Name:
audifon arriva CIC TRT audifon arriva IS TRT audifon arriva IS+ TRT audifon arriva M TRT audifon arriva S TRT audifon arriva S+ TRT audifon arriva X TRT audifon elia CIC TRT audifon elia IS TRT audifon elia IS+ TRT audifon elia M TRT audifon elia S TRT audifon prado CIC TRT audifon prado IS TRT audifon prado IS+ TRT audifon prado M TRT audifon prado S TRT audifon vico CIC TRT audifon vico IS TRT audifon vico IS+ TRT audifon vico M TRT audifon vico S TRT
Hearing Aid, Tinnitus Masker (Regulation Number: 21 CFR 874.3400)
- 4. Device Common Name / Classification Name:
ડ. Product Code: ESD, KLW
- 6. Classification of Device:
Class I for hearing aid Class II for tinnitus masker
- 7. Establishment Registration Number:
3005384855
NOV 2 1 2013
{2}------------------------------------------------
audifon GmbH & Co. KG Werner-von-Siemens.Str. 2 D-99625 Kölleda Germany
9. Market Device with Substantial Equivalence: K091552 audifon switch 8 TRT
Date of Preparation 10.
November 07, 2013
{3}------------------------------------------------
### Indications for Use
The combined devices are intended for the adult population suffering from a chronic persistent ringing in the ears (Tinnitus), who also need or desire amplification. The amplification suits the needs of a of a slight to a profound hearing loss (see table below). The products may also be used for masking Tinnitus as part of a Tinnitus Retraining Therapy (TRT) protocol and should be utilized only in consultation with a licensed hearing healthcare professional, who is trained in subsequent rehabilitation therapy, or a qualified audiologist.
| Device Name | Hearing loss level |
|----------------|---------------------------|
| arriva CIC TRT | slight to moderate severe |
| arriva IS TRT | slight to moderate severe |
| arriva IS+ TRT | slight to moderate severe |
| arriva M TRT | slight to severe |
| arriva S TRT | slight to severe |
| arriva S+ TRT | slight to severe |
| arriva X TRT | mild to profound |
| elia CIC TRT | slight to moderate severe |
| elia IS TRT | slight to moderate severe |
| elia IS+ TRT | slight to severe |
| elia M TRT | mild to profound |
| elia S TRT | slight to severe |
| prado CIC TRT | slight to moderate severe |
| prado IS TRT | slight to moderate severe |
| prado IS+ TRT | slight to moderate severe |
| prado M TRT | mild to profound |
| prado S TRT | slight to severe |
| vico CIC TRT | slight to moderate severe |
| vico IS TRT | slight to moderate severe |
| vico IS+ TRT | slight to severe |
| vico M TRT | mild to profound |
| vico S TRT | slight to severe |
### Description of Device
The above mentioned TRT devices are digital noise generators and hearing aids which were developed to be used in a tinnitus retraining therapy. These products have up to four different programs, which can be programmed in shape and level to fit the individual users needs. The programming can be done with a standard HI-PRO and the audifit software. Within the software the amplification of the combi-masker can be fitted to the individual needs. The noise can be adjusted in shape with low- and high-cut filters and in the output level. It is housed in a standard In-the-ear instrument housing (CIC, IS+ housing) or in a standard behind-the-ear instrument housing (S, S+, M and X housing).
{4}------------------------------------------------
#### Comparison Information to Predicate Device
The mentioned devices are substantially equivalent to the audifon switch 8 TRT (K091552). The mentioned devices and the audifon switch 8 TRT are fully digital noiser, with programmable noises. Within the program the level and the shape of the noise can be adjusted. Also the mentioned devices and the switch 8 TRT provide an additional amplification and can be programmed with the fitting software and a standard HI-PRO programming box.
The non-clinical performance data which were measured according to official standards (ANSI S3.22-2009) verify that the devices have a similar effectiveness as the predicate device, For TRT therapy only low sound levels below 80 dB SPL are needed. So for an effective TRT system levels above this are not needed. Also according the OSHA (29CFR 1910.95) output levels should not exceed 85 dBA. Therefore a warning in the software will occur that the higher levels should not be used or only in case of a hearing loss. So the lower maximum output has no influence on the effectiveness of the devices.
The frequency range provides an equivalent white noise with the same sound quality.
In conclusion the non-clinical tests demonstrate that the submitted devices are as safe, as effective, and perform as well as the predicate device.
The submission for tinnitus masker relies on a special control that is defined in section 874.3400. The special controls are identical applied as with the predicate device and supports the substantial equivalence:
The following tables compare the submitted devices and the audifon switch 8 TRT.
{5}------------------------------------------------
| | audifon arriva CIC TRT<br>audifon arriva IS TRT<br>audifon arriva IS+ TRT<br>audifon arriva M TRT<br>audifon arriva S TRT<br>audifon arriva S+ TRT<br>audifon arriva X TRT | switch 8 TRT |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | The combined devices are intended for the adult<br>population suffering from a chronic persistent<br>ringing in the ears (Tinnitus), who also need or<br>desire amplification. The amplification suits the<br>needs of a slight to a profound hearing loss. The<br>products may also be used for masking Tinnitus<br>as part of a Tinnitus Retraining Therapy (TRT)<br>protocol and should be utilized only in<br>consultation with a licensed hearing healthcare<br>professional, who is trained in subsequent<br>rehabilitation therapy, or a qualified audiologist. | The device is addressed to the adult population<br>with a chronic persistent ringing in the ears<br>(Tinnitus), who also need or desire amplification.<br>The amplification suits the needs of a mild to a<br>moderate hearing loss. It may be used for<br>masking tinnitus as part of tinnitus management<br>program that is prescribed by a licensed hearing<br>healthcare professional, who is trained in<br>subsequent rehabilitation therapy, or a qualified<br>audiologist. Therefore it generates a broadband<br>noise with sufficient bandwidth and intensity and<br>is applied on the ear. |
| Operation /<br>Mechanism | Uses broadband noise; Manages tinnitus through<br>masking and distraction<br>Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>Yes (M, S, S+ X)<br>Optional (CIC, IS, IS+) | Uses broadband noise; Manages tinnitus through<br>masking and distraction<br>Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>No |
| | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones |
| Where Used | May be used anywhere | May be used anywhere |
| Physical<br>Description | Standard In-the-ear instrument housing<br>(CIC, IS and IS+ housing)<br>Standard behind-the-ear instrument housing<br>(M, S, S+ and X housing) | Standard receiver-in-the-ear instrument housing |
| | RMS Output Characteristics: | RMS Output Characteristics: |
| Maximum<br>Output | White noise:<br>70 - 74 dB SPL | White noise:<br>100 dB SPL |
| Characteristics | frequency range:<br>200 - 8000 Hz | frequency range:<br>200 - 6000 Hz |
| Power Source | standard 10 zinc air 1,4V hearing aid battery<br>(CIC housing)<br>standard 312 zinc air 1,4V hearing aid battery<br>(IS and S housing)<br>standard 13 zinc air 1,4V hearing aid battery<br>(IS+, M and S+ housing)<br>standard 675 zinc air 1,4V hearing aid battery<br>(X housing) | Uses standard 312 zinc air 1.4V hearing aid<br>battery |
| Quality<br>Assurance<br>Standard | ANSI S3.22-2009 to ensure proper functioning of<br>HA | ANSI S3.22-2009 to ensure proper functioning of<br>HA |
.
{6}------------------------------------------------
| | audifon elia CIC TRT<br>audifon elia IS TRT<br>audifon elia IS+ TRT<br>audifon elia M TRT<br>audifon elia S TRT | switch 8 TRT |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | The combined devices are intended for the adult<br>population suffering from a chronic persistent<br>ringing in the ears (Tinnitus), who also need or<br>desire amplification. The amplification suits the<br>needs of a slight to a profound hearing loss. The<br>products may also be used for masking Tinnitus<br>as part of a Tinnitus Retraining Therapy (TRT)<br>protocol and should be utilized<br>only<br>in<br>consultation with a licensed hearing healthcare<br>professional, who is trained in subsequent<br>rehabilitation therapy, or a qualified audiologist. | The device is addressed to the adult population<br>with a chronic persistent ringing in the ears<br>(Tinnitus), who also need or desire amplification.<br>The amplification suits the needs of a mild to a<br>moderate hearing loss. It may be used for<br>masking tinnitus as part of tinnitus management<br>program that is prescribed by a licensed hearing<br>healthcare professional, who is trained in<br>subsequent rehabilitation therapy, or a qualified<br>audiologist. Therefore it generates a broadband<br>noise with sufficient bandwidth and intensity and<br>is applied on the ear. |
| | Uses broadband noise; Manages tinnitus through<br>masking and distraction | Uses broadband noise; Manages tinnitus through<br>masking and distraction |
| Operation /<br>Mechanism | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>Yes (M, S)<br>Optional (IS, IS+)<br>No (CIC) | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>No |
| | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones |
| Where Used | May be used anywhere | May be used anywhere |
| Physical<br>Description | Standard In-the-ear instrument housing<br>(CIC, IS and IS+ housing)<br>Standard behind-the-ear instrument housing<br>(M, and S housing) | Standard receiver-in-the-ear instrument housing |
| Maximum<br>Output | RMS Output Characteristics: | RMS Output Characteristics: |
| Characteristics | White noise:<br>71 - 72 dB SPL | White noise:<br>100 dB SPL |
| | frequency range:<br>200 - 8000 Hz | frequency range:<br>200 - 6000 Hz |
| Power Source | standard 10 zinc air 1,4V hearing aid battery<br>(CIC housing)<br>standard 312 zinc air 1,4V hearing aid battery<br>(IS and S housing)<br>standard 13 zinc air 1,4V hearing aid battery<br>(IS+, and M housing) | Uses standard 312 zinc air 1.4V hearing aid<br>battery |
| Quality<br>Assurance<br>Standard | ANSI S3.22-2009 to ensure proper functioning of<br>HA | ANSI S3.22-2009 to ensure proper functioning of<br>HA |
| | audifon prado CIC TRT<br>audifon prado IS TRT<br>audifon prado IS+ TRT<br>audifon prado M TRT<br>audifon prado S TRT | switch 8 TRT |
| Indications For<br>Use | The combined devices are intended for the adult<br>population suffering from a chronic persistent<br>ringing in the ears (Tinnitus), who also need or<br>desire amplification. The amplification suits the<br>needs of a slight to a profound hearing loss. The<br>products may also be used for masking Tinnitus<br>as part of a Tinnitus Retraining Therapy (TRT)<br>protocol and should be utilized only in<br>consultation with a licensed hearing healthcare<br>professional, who is trained in subsequent<br>rehabilitation therapy, or a qualified audiologist. | The device is addressed to the adult population<br>with a chronic persistent ringing in the ears<br>(Tinnitus), who also need or desire amplification.<br>The amplification suits the needs of a mild to a<br>moderate hearing loss. It may be used for<br>masking tinnitus as part of tinnitus management<br>program that is prescribed by a licensed hearing<br>healthcare professional, who is trained in<br>subsequent rehabilitation therapy, or a qualified<br>audiologist. Therefore it generates a broadband<br>noise with sufficient bandwidth and intensity and<br>is applied on the ear. |
| | Uses broadband noise; Manages tinnitus through<br>masking and distraction | Uses broadband noise; Manages tinnitus through<br>masking and distraction |
| Operation /<br>Mechanism | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>Yes (M, S)<br>Optional (IS, IS+)<br>No (CIC) | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>No |
| | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones | white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones |
| Where Used | May be used anywhere | May be used anywhere |
| Physical<br>Description | Standard In-the-ear instrument housing<br>(CIC, IS and IS+ housing)<br>Standard behind-the-ear instrument housing<br>(M, and S housing) | Standard receiver-in-the-ear instrument housing |
| Maximum | RMS Output Characteristics: | RMS Output Characteristics: |
| Output<br>Characteristics | White noise:<br>71 - 72 dB SPL | White noise:<br>100 dB SPL |
| | frequency range:<br>200 - 8000 Hz | frequency range:<br>200 - 6000 Hz |
| Power Source | standard 10 zinc air 1,4V hearing aid battery<br>(CIC housing)<br>standard 312 zinc air 1,4V hearing aid battery<br>(IS and S housing)<br>standard 13 zinc air 1,4V hearing aid battery<br>(IS+, and M housing) | Uses standard 312 zinc air 1.4V hearing aid<br>battery |
| Quality<br>Assurance<br>Standard | ANSI S3.22-2009 to ensure proper functioning of<br>HA | ANSI S3.22-2009 to ensure proper functioning of<br>HA |
| | audifon vico CIC TRT<br>audifon vico IS TRT<br>audifon vico IS+ TRT<br>audifon vico M TRT<br>audifon vico S TRT | switch 8 TRT |
| Indications For<br>Use | The combined devices are intended for the adult<br>population suffering from a chronic persistent<br>ringing in the ears (Tinnitus), who also need or<br>desire amplification. The amplification suits the<br>needs of a slight to a porfound hearing loss. The<br>products may also be used for masking Tinnitus<br>as part of a Tinnitus Retraining Therapy (TRT)<br>protocol and should be utilized only in<br>consultation with a licensed hearing healthcare<br>professional, who is trained in subsequent<br>rehabilitation therapy, or a qualified audiologist. | The device is addressed to the adult population<br>with a chronic persistent ringing in the ears<br>(Tinnitus), who also need or desire amplification.<br>The amplification suits the needs of a mild to a<br>moderate hearing loss. It may be used for<br>masking tinnitus as part of tinnitus management<br>program that is prescribed by a licensed hearing<br>healthcare professional, who is trained in<br>subsequent rehabilitation therapy, or a qualified<br>audiologist. Therefore it generates a broadband<br>noise with sufficient bandwidth and intensity and<br>is applied on the ear. |
| Operation /<br>Mechanism | Uses broadband noise; Manages tinnitus through<br>masking and distraction | Uses broadband noise; Manages tinnitus through<br>masking and distraction |
| | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>Yes (M, S)<br>Optional (CIC, IS, IS+)<br><br>white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones | Circuit type:<br>Digital<br>Programmable:<br>Yes<br>Available noises:<br>Four<br>Volume control:<br>No<br><br>white-noise is adjustable<br>noise level is programmable<br>adjustable Low Battery Indicator<br>programmable Program Switch Tones |
| Where Used | May be used anywhere | May be used anywhere |
| Physical<br>Description | Standard In-the-ear instrument housing<br>(CIC, IS and IS+ housing)<br><br>Standard behind-the-ear instrument housing<br>(M and S housing) | Standard receiver-in-the-ear instrument housing |
| RMS Output Characteristics: | | RMS Output Characteristics: |
| Maximum<br>Output<br>Characteristics | White noise:<br>107 - 113 dB SPL | White noise:<br>100 dB SPL |
| | frequency range:<br>200 - 8000 Hz | frequency range:<br>200 - 6000 Hz |
| Power Source | standard 10 zinc air 1,4V hearing aid battery<br>(CIC housing)<br>standard 312 zinc air 1,4V hearing aid battery<br>(IS and S housing)<br>standard 13 zinc air 1,4V hearing aid battery<br>(IS+, and M housing) | Uses standard 312 zinc air 1.4V hearing aid<br>battery |
| Quality<br>Assurance<br>Standard | ANSI S3.22-2009 to ensure proper functioning of<br>HA | ANSI S3.22-2009 to ensure proper functioning of<br>HA |
·
{7}------------------------------------------------
.
{8}------------------------------------------------
## Conclusion
- . The devices have similar acoustic characteristics as the predicate device.
- The devices are similar in style (ITE or BTE) as the predicate device. .
- The devices are similar in material as the predicate device. .
- . The devices are similar in intended use as the predicate device
- The devices have the same targeted population as the predicate device .
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/8/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 21, 2013
Audifon USA, Inc. c/o Ms. Jane E. Perrone Vice President of U.S. Operations 403 Chairman Ct. Suite 1 Debary, FL 32713
Re: K130514
Trade/Device Name: TRT products arriva, vico, prado and elia Regulation Number: 21 CFR 874.3400 Regulation Name: Hearing Aid, Tinnitus Masker Regulatory Class: Class II (for tinnitus masker) Product Code: KLW, ESD Dated: August 19, 2013 Received: August 23, 2013
Dear Ms. Perrone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
Page 2 - Ms. Jane E. Perrone
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## Indications for Use
510(k) Number (if known):
K130514
Device Name:
audifon arriva CIC TRT audifon arriva IS TRT audifon arriva IS+ TRT audifon arriva M TRT audifon arriva S TRT audifon arriva S+ TRT audifon arriva X TRT audifon elia CIC TRT audifon elia IS TRT audifon elia IS+ TRT audifon elia M TRT audifon elia S TRT audifon prado CIC TRT audifon prado IS TRT audifon prado IS+ TRT audifon prado M TRT audifon prado S TRT audifon vico CIC TRT audifon vico IS TRT audifon vico IS+ TRT audifon vico M TRT audifon vico S TRT
(hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator) (hearing aid with TRT noise generator)
Indications For Use:
The combined devices are intended for the adult population suffering from a chronic persistent ringing in the ears (Tinnitus), who also need or desire amplification. The amplification suits the needs of a slight to a profound hearing loss (see table below). The products may also be used for masking Tinnitus as part of a Tinnitus Retraining Therapy (TRT) protocol and should be utilized only in consultation with a licensed hearing healthcare professional, who is trained in subsequent rehabilitation therapy, or a qualified audiologist.
Page 1 of 2
{12}------------------------------------------------
| Device Name | Hearing loss level |
|----------------|---------------------------|
| arriva CIC TRT | slight to moderate severe |
| arriva IS TRT | slight to moderate severe |
| arriva IS+ TRT | slight to moderate severe |
| arriva M TRT | slight to severe |
| arriva S TRT | slight to severe |
| arriva S+ TRT | slight to severe |
| arriva X TRT | mild to profound |
| elia CIC TRT | slight to moderate severe |
| elia IS TRT | slight to moderate severe |
| elia IS+ TRT | slight to severe |
| elia M TRT | mild to profound |
| elia S TRT | slight to severe |
| prado CIC TRT | slight to moderate severe |
| prado IS TRT | slight to moderate severe |
| prado IS+ TRT | slight to moderate severe |
| prado M TRT | mild to profound |
| prado S TRT | slight to severe |
| vico CIC TRT | slight to moderate severe |
| vico IS TRT | slight to moderate severe |
| vico IS+ TRT | slight to severe |
| vico M TRT | mild to profound |
| vico S TRT | slight to severe |
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Vasant Dasika -S 2013.11.19 16:56:53 -05'00'
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.