The ANM T30 CR® Tinnitus Therapy System is intended to provide relief from the disturbance of tinnitus while using the system, and with regular use (over several months) may provide relief to the patient when not using the system. The ANM T30 CR® Stimulator is a patient device for home use. The ANM T30 CR® Stimulator is fitted and programmed by a qualified healthcare professional familiar with tinnitus treatment. Stimulation needs to be applied for several hours a day over a period of weeks or months. The target population is adults over 18 years of age.
Device Story
ANM T30 CR® Tinnitus Therapy System provides acoustic tinnitus relief. System comprises: Programmer (PC with analysis software), Patient Console, Stimulator (wearable device), and Earphones. Healthcare professional uses Programmer/Software to measure tinnitus, adjust therapeutic stimulation signals, and program Stimulator. Patient uses Console to adjust frequency/loudness during setup; uses Stimulator at home to apply pre-programmed acoustic tones via earphones. Rechargeable battery-powered Stimulator generates signals. Regular use (hours/day over weeks/months) intended to provide relief during and after treatment. Output is software-controlled (default 80 dB(A), up to 115 dB(A) with professional consent). Benefits include symptom relief when combined with counseling. Device requires professional fitting; patient must avoid use during activities requiring full attention.
Clinical Evidence
No clinical data provided. Substantial equivalence established via nonclinical verification and validation testing, including software requirement specifications, risk analysis, and electrical safety testing (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Digital tinnitus masker. Components: Programmer, Console, Stimulator, Earphones. Stimulator dimensions: 1.85" x 1.65" x 0.91". Energy: 2.4V (2x Ni-MH rechargeable button cells). Frequency range: 90 Hz - 13 kHz. Output: 80 dB(A) default, max 115 dB(A). Biocompatible patient contact materials (ISO-10993). Electrical safety: IEC 60601-1, IEC 60601-1-2. Software-controlled signal generation.
Indications for Use
Indicated for adults over 18 years of age experiencing tinnitus disturbance. Used for relief during system use and potentially after regular use over several months. Requires fitting and programming by a qualified healthcare professional. Contraindicated for use during activities requiring full attention (e.g., driving).
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{0}------------------------------------------------
K112752
DEC 1 6 2011
#### 510(k) Summary 5
| Company Name:<br>Address: | ANM Adaptive Neuromodulation GmbH<br>Im Mediapark 6d<br>50670 Cologne<br>Germany |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Telephone:<br>FAX: | +49 (0)221 454-6300 Ext. 6340<br>+49 (0)221 454-6302 |
| Contact person: | Dr. Ingrid Rohm<br>Quality & Regulatory Manager |
| Date summary prepared: | September 9, 2011 |
| Classification Name/<br>Common Name: | Tinnitus Masker |
| Trade Name: | ANM T30 CR® Tinnitus Therapy System |
| Product Code: | KLW |
| Device Classification: | Class II |
| Predicate Devices: | Neuromonics Tinnitus Treatment System, K043274<br>(Neuromonics PTY Ltd.)<br>TinniTech ANMP System, K030791 (Tinnitech Ltd.) |
## Device Description:
The ANM T30 CR® is a computer-controlled system for tinnitus therapy is provided to the patient by individualized, ambient acoustic tones. The system consists of four components:
ANM T30 CR® Programmer (computer) including the tinnitus analysis software ANM T30 CR® SW: The ANM T30 CR® Software is used to measure the tinnitus, adjust the individual therapeutic stimulation signals and program the patient device (ANM T30 CR® Stimulator).
## ANM T30 CR® Patient Console:
The ANM T30 CR® Patient Console is connected to the ANM T30 CR® Programmer and is used by the patient to adjust the parameters specific to his/her tinnitus (frequency and loudness) during tinnitus measurement and stimulation signal adjustment.
# ANM T30 CR® Stimulator (patient device):
The stimulator is used by the patient for the application of the therapeutic stimulation signal. It has a rechargeable battery for the energy supply and electronic components to generate the stimulation signal.
{1}------------------------------------------------
## ANM T30 CR® Earphones:
The ANM T30 CR® Earphones are to be connected either to the ANMT30 CR® Patient Console to measure the tinnitus and adjust the stimulation signal or to the ANM T30 CR® Stimulator to apply the stimulation signal.
Patient contact parts are manufactured from materials that meet the ISO-10993 biocompatibility requirement contact with intact skin, as tested by their suppliers. The ANM T30 CR® Tinnitus Therapy System meets the applicable general electrical safety requirements of IEC 60601 1, and the requirements of IEC 60601-1-2.
## Intended Use:
The ANM T30 CR® Tinnitus Therapy System is intended to provide relief from the disturbance of tinnitus while using the system, and with regular use (over several months) may provide relief to the patient when not using the system.
The ANM T30 CR® Stimulator is a patient device for home use.
The ANM T30 CR® Stimulator is fitted and programmed by a qualified healthcare professional familiar with tinnitus treatment.
Stimulation needs to be applied for several hours a day over a period of weeks or months. The target population is adults over 18 years of age.
Table 5.1: Comparison to Predicate Device:
| | ANM T30 CR® System | Neuromonics Tinnitus Treatment System | TinniTech ANMP System |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The ANM T30 CR®<br>Tinnitus Therapy<br>System is intended to<br>provide relief from the<br>disturbance of tinnitus<br>while using the system,<br>and with regular use<br>(over several months)<br>may provide relief to<br>the patient when not<br>using the system.<br>The ANM T30 CR®<br>Stimulator is a patient<br>device for home use.<br>The ANM T30 CR®<br>Stimulator is fitted and<br>programmed by a<br>qualified healthcare<br>professional familiar<br>with tinnitus<br>treatment.<br>Stimulation needs to be<br>applied for several<br>hours a day over a<br>period of weeks or<br>months.<br>The target population is<br>adults over 18 years of<br>age. | The Neuromonics Tinnitus<br>Treatment System is<br>intended to provide relief<br>from the disturbance of<br>tinnitus while using the<br>system, and with regular<br>use (over several months)<br>may provide relief to the<br>patient when not using the<br>system.<br>The Neuromonics Tinnitus<br>Treatment System is<br>intended to interact and<br>intermittently interact with<br>the patients tinnitus as part<br>of a tinnitus management<br>program. The addition of<br>the music is to aid the<br>promotion of relaxation<br>during the tinnitus<br>interaction process.<br>The initial hearing and<br>tinnitus tests are conducted<br>by a qualified audiologist<br>familiar with the treatment<br>of tinnitus; the subsequent<br>management of the<br>treatment is carried out by<br>an appropriate healthcare | The ANMP System is<br>intended to completely<br>mask and intermittently<br>mask tinnitus as part of a<br>tinnitus management<br>program. The addition of<br>the music is to aid the<br>promotion of relaxation<br>during the tinnitus masking<br>process.<br>The initial hearing and<br>tinnitus tests are conducted<br>by a qualified audiologist<br>familiar with the treatment<br>of tinnitus; the subsequent<br>management of the<br>treatment is carried out by<br>an appropriate healthcare<br>professional.<br>The target population for<br>the device is adults (18<br>years and over) who<br>present with tinnitus that<br>may or may not be<br>accompanied with hearing<br>loss at the higher<br>frequencies and who are<br>participating in a tinnitus |
| | ANM T30 CR® System | Neuromonics Tinnitus Treatment System | TinniTech ANMP System |
| | | professional. | management program. |
| | | The target population for the device is adults (18 years and over) who present with tinnitus that may or may not be accompanied with hearing loss at the higher frequencies and who are participating in a tinnitus management program. | |
| Physical dimensions | ANM T30 CR® Stimulator (1.85" x 1.65" x 0.91") with preprogrammed auditory stimuli delivered through earphones | iPod-like device (4.06" x 2.32" x 0.71") with prerecorded audio stimuli delivered through headphones | The Philips eXanium 401 mini MP3 mini CD player with earphones is supplied by TinniTech<br>Size:<br>3.6" x 4.7" x 1.1" (LxHxD) |
| Maximum Output Characteristics | Maximum output is 80 dB(A) (software-controlled).<br>This maximum can be exceeded only with the consent of a qualified healthcare professional to 115 dB(A). As this maximum output exceeds the OSHA (Standard 1910.95) 8 hour time weighted average for the occupational workplace, warnings are included in the User Manual | Fixed at 80 dB SPL | The maximum output from the earphones could exceed 85 dBA, the OSHA (Standard 29CFR 1910.95) 8 hour time weighted average for the occupational workplace, and therefore a warning about setting the player volume setting at a comfortable (safe) level is included in the User's Manual and on the player |
| Frequency Range | 90 Hz - 13 kHz | 100 Hz - 12.5 kHz | 20 Hz - 20 KHz |
| Audio Signal Technology | Digital | Digital | Digital |
| Medium | Flash Memory | SD Memory Card | Two mini CDs |
| Energy used | 2.4V, 2 x rechargeable Ni-MH button, 1.2V | Lithium polymer, output 6V | The supplied mini disc player, the Philips eXpanium 401 operates from either one AA cell or a 110 VAC mains adapter provided with the player |
| Transducer (headphones type) | Use only earphones supplied with the ANM T30 CR® Tinnitus Therapy System. The | Use only earphones supplied with the Neuromonics device. Other earphones may not | Use the earphones provided with the mini disc player |
| | ANM T30 CR® System | Neuromonics Tinnitus<br>Treatment System | TinniTech ANMP System |
| | stimulation sequence<br>program has been<br>acoustically aligned<br>specifically for use with<br>these earphones | sounds that are essential<br>for treatment | |
| Where used | Home use under the<br>supervision of a<br>qualified healthcare<br>professional | Home use under the<br>supervision of a qualified<br>clinician | Home use under the<br>management of an<br>appropriate qualified<br>healthcare professional |
| Safety | Avoid using the ANM<br>T30 CR® Tinnitus<br>Therapy System while<br>driving or other<br>activities that require<br>the patient's full<br>attention | Avoid using the<br>Neuromonics device while<br>driving, cycling, or while<br>performing any other<br>activity that requires the<br>patient's full attention | The ANMP therapy should<br>never be undertaken when<br>tinnitus masking sounds<br>might prevent the patient<br>from hearing cues or<br>warnings of likely harm or<br>danger |
{2}------------------------------------------------
.
{3}------------------------------------------------
## Summary of performance data:
Nonclinical verification and validation testing activities were conducted to establish the performance, functionality and reliability characteristics of the device. Testing involved verification of software requirement specifications from risk analysis. The testing shows that the ANM T30 CR® System is as safe, as effective and performs as well as or better than the predicate devices identified in the above Comparison Table.
## Risks to health
With a maximum output of 115 dB(A) the ANM T30 CR® Stimulator could exceed the occupational workplace OSHA standard 29 CFR 1910.95. To avoid possible hearing damage the health care professional only can program the stimulator to allow for acoustic signals above 80 dB(A) for use by tinnitus patients who suffer from impaired hearing.
The User's Manual carries precautions to reduce the acoustic volume or stop the treatment if the auditory stimulation volume is uncomfortable. It also states that stimulation should not been done when undisturbed attention is required.
## Hearing Healthcare Professional Diagnosis
The sale and fitting of the ANM T30 CR® system will only be conducted through a Hearing Healthcare Professional.
# Benefits
Supported by appropriate counselling and/or tinnitus therapy, relief from tinnitus symptoms may be experienced by using the ANM T30 CR® Stimulator.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ANM Adaptive Neuromodulation GmbH c/o Dr. Ingrid Rohm Quality & Regulatory Manager Im Mediapark 6d 50670 Cologne Germany
DEC 1 6 2011
Re: K112752
Trade/Device Name: ANM T30 CR® Tinnitus Therapy System Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: September 9, 2011 Received: September 21, 2011
Dear Dr. Rohm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. Ingrid Rohm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Melvina R. Evdelman, M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K//2+52
#### Indications for Use 4
510(k) Number (if known):
ANM T30 CR® Tinnitus Therapy System Device Name:
Indications For Use:
The ANM T30 CR® Tinnitus Therapy System is intended to provide relief from the disturbance of tinnitus while using the system, and with regular use (over several months) may provide relief to the patient when not using the system.
The ANM T30 CR® Stimulator is a patient device for home use.
The ANM T30 CR® Stimulator is fitted and programmed by a qualified healthcare professional familiar with tinnitus treatment.
Stimulation needs to be applied for several hours a day over a period of weeks or months. The target population is adults over 18 years of age.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Anne K. Kane Ph.D.
hsion Sign-Off) ision Sign-On)
sion of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.