The ANMP System is intended to completely mask and intermittently mask tinnitus as part of a tinnitus management program. The addition of the music is to aid the promotion of relaxation during the tinnitus masking process. The initial hearing and tinnitus tests are conducted by a qualified audiologist familiar with the treatment of tinnitus: the subsequent management of the treatment is carried out by an appropriate healthcare professional.
Device Story
System delivers spectrally adapted relaxation music and wide-spectrum noise to mask tinnitus. Input: patient audiogram used to customize sound frequency response. Operation: pre-recorded MP3 files on mini CDs played via portable stereo CD player with earphones. Two modes: continuous masking (CD 1) and intermittent masking (CD 2) to facilitate habituation. Used in home setting under direction of healthcare professional (audiologist/otolaryngologist). Output: audible sound delivered to patient ears. Benefit: tinnitus relief and relaxation promotion. Healthcare provider uses audiogram to tailor sound output; patient controls volume. Safety: includes warnings regarding volume levels and situational awareness.
Clinical Evidence
No clinical data provided. Bench testing only; performance specifications of the MP3 player were tested to ensure suitability for delivery of ANMP treatment sounds.
Technological Characteristics
System consists of pre-recorded MP3 mini CDs and a portable stereo CD player (Philips eXpanium 401). Materials: standard plastic and diecast metals. Energy: 1.5V AA battery or 110 VAC adapter. Output: max 5mW per channel; max output 85 dBA. Connectivity: standalone. Software: pre-recorded digital audio files.
Indications for Use
Indicated for adults (18 years and over) with tinnitus, with or without high-frequency hearing loss, participating in a tinnitus management program. Contraindicated for patients without prior medical evaluation by a physician to rule out treatable underlying diseases.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{0}------------------------------------------------
KO30791
APR 1 7 2003
## 510(k) SUMMARY
# TinniTech ANMP System
| Applicant: | TinniTech Ltd<br>Level 2, 55 Market Street<br>Sydney NSW 2000<br>Australia<br>Phone: +61 2 9283 0601<br>Fax: +61 2 9283 0602 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Date of<br>Submission: | 28 February 2003 |
| Manufacturing Site: | ResMed Ltd.<br>97 Waterloo Road<br>North Ryde NSW 2113<br>Australia |
| Establishment<br>Registration<br>Number: | 8020057 |
| Distributor: | Mr R West, Go West, Inc.<br>2502 Tournament Court<br>Castle Rock, CO 80104<br>Phone/Fax: (303) 663-8089 |
| Contact Name for<br>Submission: | Lachlan James<br>(contact address as per Applicant) |
| Trade or<br>Proprietary Name | ANMP (Acoustic Neuro Modulation Protocol) |
| Common or Usual<br>Names | Tinnitus Masker |
| Classification Name<br>(FDA) | KLW : Masker Tinnitus |
| Classification | Class II device |
Comments of the comments of
{1}------------------------------------------------
#### Indications for Use:
The TinniTech ANMP System comprises two (2) mini compact disks that have been prerecorded with selected relaxation music and other sounds spectrally adapted to suit the particular patient's hearing thresholds. The device is indicated to mask and intermittently mask tinnitus as part of a tinnitus management program. The addition of the music is to aid the promotion of relaxation during the tinnitus masking process.
The spectral adaptation of the sound is determined by the patient's audiogram measured by a qualified audiologist; the patient's tinnitus therapy is managed by an appropriate healthcare professional. The target population for the device is adults (18 years of age and over) who present with tinnitus that may or may not be accompanied with hearing loss at the higher frequencies.
#### Intended use
The ANMP System is intended to completely mask and intermittently mask tinnitus as part of a tinnitus management program. The addition of the music is to aid the promotion of relaxation during the tinnitus masking process. The initial hearing and tinnitus tests are conducted by a qualified audiologist familiar with the treatment of tinnitus: the subsequent management of the treatment is carried out by an appropriate healthcare professional.
#### Target population
The target population for the device is adults (18 years and over) who present with tinnitus that may or may not be accompanied with hearing loss at the higher frequencies and who are participating in a tinnitus management program.
#### Device description
The TinniTech ANMP system consists of:
One CD pre-recorded in the MP3 format with selected relaxation music and superimposed low amplitude, wide spectrum noise that is spectrally adapted to compensate for the individual patient's higher frequencies hearing loss. This CD is intended for the complete masking of tinnitus.
One CD pre-recorded in the MP3 format with selected relaxation music that is spectrally adapted to compensate for the individual patient's higher frequencies hearing loss however this CD has been recorded such that the dynamic characteristics of the sound allow the tinnitus to intermittently break through during the quieter passages of the recorded music.
{2}------------------------------------------------
A commercially available, battery powered, stereo, MP3 mini compact disc player with earphones with a performance specification that TinniTech has tested and shown to be suitable for the delivery of the ANMP treatment sounds or waveforms. At this time Tinnitech will supply the Phillips eXpanium 401™ MP3 mini compact disc player. Other suitable players will be offered as TinniTech tests and approves them for the ANMP System.
#### Substantial equivalence
TenniTech is claiming substantial equivalence to two devices:
Equivalence is claimed to the Dynamic Tinnitus Mitigation System, DTM-6 manufactured by Petroff Audio Technologies Inc. 510k number; K974501
Equivalence is also claimed to the Custom TCI Instrument manufactured by Siemens Hearing Instruments, 510k number; K011364.
#### K974501
The Dynamic Tinnitus Mitigation System, DTM-6 manufactured by Petroff Audio Technologies Inc. 510k number; K974501 is a four (4) CD set system capable of being played on any commercial CD player with headphones or speakers. CD#1 provides digital tinnitus masking sounds only, CDs #2, #3 and #4 provide digital tinnitus masking sounds plus relaxation messages, alpha rhythms (a gentle musical sound generated on electronic musical instruments), music and/or nature sounds. CDs #2. #3 and #4 add the indication of promoting relaxation during the tinnitus masking process to the indication of providing 'temporary relief of tinnitus symptoms'.
Like the predicate device K974501 the TinniTech ANMP sounds are recorded on mini CDs (in the MP3 format) capable of being played on any one of a number of good quality, portable MP3 mini disc players, the patient using appropriate earphones to listen to the sound recorded on the discs.
The TinniTech ANMP, like CDs #2. #3 and #4 of the predicate device includes on CD1 digital tinnitus masking sounds together with selected musical sounds to promote relaxation during the tinnitus continuous masking process.
#### K011364
The Custom TCI Instrument manufactured by Siemens Hearing Instruments, 510k number: K011364. is a fully digital. low level noise generator. The device is programmable, with selected noise and output levels and is indicated for use in the delivery of tinnitus habituation and tinnitus masking therapies.
The output noise in the Custom TCI Instrument can be custom-tailored to the user's individual requirements.
{3}------------------------------------------------
Like the predicate device, K011364, the TinniTech ANMP custom-tailors the sounds on both CD 1 and CD 2 to suit the individual patients' hearing requirements. This is achieved analysing the patient's audiogram and boosting the amplitude of those sounds or tones where the patient has been shown to have a reduced hearing threshold.
The TinniTech ANMP also includes on CD 2 specially selected music with a dynamic characteristic that allows the patient's tinnitus sound to intermittently break through the masking effect of the music. This function, like the predicate device K011364, assists in enhancing the patient's habituation to the tinnitus sound.
{4}------------------------------------------------
| Characteristic | New Device | Predicate Device<br>K974501 | Predicate Device<br>K011364 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Continuously and<br>intermittently mask<br>tinnitus as part of a<br>tinnitus management<br>program with masking<br>noise and selected<br>music to promote the<br>relaxation of the<br>patient during the<br>delivery of the<br>Tinnitech ANMP<br>therapy. | Temporary relief of<br>tinnitus symptoms and<br>the promotion of<br>relaxation during the<br>tinnitus masking<br>process. | Continuously mask<br>and partially mask<br>tinnitus as part of a<br>tinnitus management<br>program. |
| Target Population | Adults (18 years and<br>over), both with and<br>without high<br>frequency loss, with<br>tinnitus who are<br>participating in a<br>tinnitus management<br>program. | All tinnitus sufferers,<br>including that<br>percentage of severe<br>sufferers who are "un-<br>maskable" through the<br>application of a<br>masking sound. | Adults and children<br>(over 5 years of age)<br>with tinnitus who are<br>participating in a<br>tinnitus |
| Audio signal<br>technology | Digital | Digital | Digital |
| Available<br>noises/sounds | Pre-adapted to the<br>patient's hearing<br>characteristics,<br>a wide selection of<br>musical sounds<br>incorporating Tinnitus<br>Masking noise (20Hz<br>- 20KHz) digitally<br>recorded on mini<br>compact discs in MP3<br>format. | Proprietary dynamic<br>formats of sounds<br>incorporating Dynamic<br>Tinnitus Mitigation<br>technology (20Hz -<br>20KHz) recorded on<br>digital storage means<br>that are played at<br>audible levels. | Programmable noise<br>spectrum with pre-<br>selection of four noise<br>types digitally stored<br>in solid state memory. |
| Medium | Two mini CDs (but<br>could be expanded by<br>providing a variety of<br>music on additional<br>discs). | Four CDs/audio tapes | Solid state memory. |
| Characteristic | New Device | Predicate Device<br>K974501 | Predicate Device<br>K011364 |
| Operation - continued | | | |
| Volume control | Yes, user controlled<br>with warning in the<br>User Instruction<br>Manual stating "To<br>prevent the possible<br>damage to your<br>hearing the volume<br>setting of the disc<br>player should not be<br>set at levels where<br>you are uncomfortable<br>with the sound". | Yes, user controlled<br>with a note stating<br>that the volume<br>should always be set<br>at a low level when<br>first trying any CD. | Yes, user controlled<br>within limits pre-<br>programmed by the<br>managing health<br>professional. |
| Distribution | To be sold via direct<br>and indirect channels<br>involving an<br>appropriately qualified<br>healthcare<br>professional. | Currently sold and<br>supported via mail<br>order. | The sale and fitting of<br>the Siemens Custon<br>TCI will only be<br>conducted through a<br>Hearing Healthcare<br>Professional, such as<br>an audiologist or<br>otolaryngologist. |
| Components | The components of<br>the complete<br>TinniTech ANMP<br>system include:<br>Sound files on<br>storage means<br>(digital mini CDs)A high quality,<br>digital MP3 mini<br>CD player, of<br>normal<br>commercial<br>supplyUser's Manual | The components of<br>the complete<br>predicate device<br>system include:<br>Sound files on<br>storage means<br>(digital CDs or<br>Audio tapes)A digital CD or<br>Audio Tape<br>player, either<br>portable or fixed<br>of normal<br>commercial<br>supply or pre-<br>existing in the<br>home or car.User's Manual (called<br>Workbook) | The Siemens Custom<br>TCI is a fully digital,<br>programmable, low<br>level noise generator<br>that is available in a<br>full in-the-ear, half<br>shell, in-the-canal, or<br>helix shell.<br>User's Manual. |
| Characteristic | New Device | Predicate Device<br>K974501 | Predicate Device<br>K011364 |
| Audio player device. | The Philips eXpanium<br>401 mini MP3 mini CD<br>player with earphones<br>is supplied by<br>TinniTech<br><br>Size<br>3.6"L x 4.7" H x<br>1.1" D<br>Weight: 200 g (7.1<br>ounces)<br>Battery: 1 x 1.5V AA<br>Accessories:<br>Earphone | A standard Walkman<br>type digital player or<br>equivalent player is a<br>commonly used<br>listening/delivery<br>device.<br><br>The Sony model<br>CDFV-15 is<br>recommended.<br><br>Size:<br>5.5" x 5.0" x 7/8"<br>Weight: 105 g<br>(3.6 ounces)<br>Battery: 2 x 1.5V AA<br>Accessories:<br>headphones | Not applicable. |
| Energy delivered | The supplied mini disc<br>player, the Philips<br>eXpanium 401 has a<br>maximum power<br>output per channel of<br>5mW.<br>The maximum output<br>from the earphones<br>could exceed 85 dBA,<br>the OSHA (Standard<br>29CFR 1910.95) 8<br>hour time weighted<br>average for the<br>occupational<br>workplace, and<br>therefore a warning<br>about setting the<br>player volume setting<br>at a comfortable<br>(safe) level is included<br>in the User's Manual<br>and on the player | No Maximum<br>stipulated or<br>controlled. A home<br>stereo system could<br>be capable of more<br>than 200 Watts<br>output. | Maximum output is<br>86dBA which exceeds<br>the OSHA (Standard<br>29CFR 1910.95) 8<br>hour time weighted<br>average for the<br>occupational<br>workplace therefore<br>warnings are included<br>in the User's Manual. |
| Energy used | The supplied mini disc<br>player, the Philips<br>eXpanium 401<br>operates from either<br>one AA cell or a 110<br>VAC mains adapter<br>provided with the<br>player. | The audio player will<br>utilise mains voltage<br>(1110 VAC) or battery<br>voltage (typically up<br>to 9 VDC) | Small hearing aid type<br>battery. |
| Characteristic | New Device | Predicate Device<br>K974501 | Predicate Device<br>K011364 |
| Transducer<br>(headphones type) | Use the earphones<br>provided with the mini<br>disc player. | Recommended to use<br>either Radio Shack<br>headphones PRO-35<br>or<br>Sony premium<br>headphones or<br>earbuds. | The Siemens Custom<br>TCI is an "in-the-ear"<br>device and is<br>available in four<br>versions:<br>full in-the-ear, half<br>shell, in the canal, or<br>helix shell. |
| Where used | Home use under the<br>management of an<br>appropriately qualified<br>healthcare<br>professional | Home use | Home use |
| Safety | The ANMP therapy<br>should never be<br>undertaken when the<br>tinnitus masking<br>sounds might prevent<br>the patient from<br>hearing cues or<br>warnings of likely<br>harm or danger. | External low-level<br>noise may not be<br>heard due to the fact<br>that the soundwaves<br>are played at audible<br>levels to suppress<br>Tinnitus | |
| Materials | Typical audio product<br>materials such as<br>plastic and diecast<br>metals. | Typical audio product<br>materials such as<br>plastic and diecast<br>metals. | Typical hearing ail<br>product materials<br>such as plastics. |
| Performance | The TinniTech ANMP<br>system enables the<br>user to determine<br>whether the delivered<br>therapy assists in the<br>management of their<br>tinnitus. | The predicate device<br>enables the user to<br>determine whether the<br>sound therapy assists<br>them manage their<br>tinnitus | The predicate device<br>enables the end user<br>to determine whether<br>the sound therapy<br>assists them to<br>suppress Tinnitus. |
| Instructions | The TinniTech ANMP<br>System comes with a<br>User's Guide | The predicate device<br>comes with a 23 page<br>Workbook | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Treatment management and diagnosis
The TinniTech ANMP System is for home use under the direction of an appropriately qualified healthcare professional such as an otolaryngologist, an audiologist, or a licensed hearing aid dealer.
Patients should receive a medical evaluation by a licensed physician (preferably a physician who specializes in diseases of the ear) to rule out medically or surgically treatable diseases for which tinnitus is a symptom before proceeding with the TinniTech ANMP treatment.
#### Risks and warnings for safe use
The maximum output could exceed the occupational workplace OSHA standard 29CFR 1910.95 of 85 dBA. To avoid possible hearing damage the CD player and the User's Manual carry warnings stating that the sounds on the discs should not be played at uncomfortable volume levels.
The CD player carries a label warning against use of the ANMP system if such use prevents the user from hearing sounds warning of danger.
#### Benefits
Relief from the effects of tinnitus may be provided by the TinniTech ANMP system when used within a tinnitus management program.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol with three parallel lines curving upwards, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 1 7 2003
TinniTech Ltd c/o Lachlan S. James Level 2, 55 Market Street Sydney NSW 2000 Australia
Re: K030791
Trade/Device Name: ANMP (Acoustic Neuro Modulation Protocol) Regulation Number: 21 CFR §874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: February 28, 2003 Received: March 12, 2003
Dear Mr. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 - Lachlan S. James
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. halyi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
### 3.2 Indications for Use
510(k) Number (if known): Device Name: TinniTech ANMP System:
KO30791
Indications for Use:
The TinniTech ANMP System comprises two (2) mini compact disks and a portable, stereo, MP3 mini compact disc player with earphones. The discs have been prerecorded with selected relaxation music and other sounds spectrally adjusted to suit the particular patient's spectral hearing thresholds as shown by their audiogram. The sounds on the discs are reproduced by the mini compact disc player and delivered to the ears by the earphones supplied with the player. The device is indicated for adult (18 years and over) tinnitus sufferers who may or may not suffer higher frequencies hearing loss and are participating in a tinnitus management program.
The TinniTech ANMP System is for home use under the direction of an appropriately qualified healthcare professional such as an otolaryngologist, an audiologist, or a licensed hearing aid dealer.
Patients should receive a medical evaluation by a licensed physician (preferably a physician who specializes in diseases of the ear) to rule out medically or surqically treatable diseases for which tinnitus is a symptom before proceeding with non-medical tinnitus management.
The Tinnitech ANMP System may be used to completely mask and intermittently mask tinnitus.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | X |
|---------------------------------------|---|
|---------------------------------------|---|
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|

(Division Sign-Off)
Division of Ophthalmic Ear, Nose and Throat Devises
| 510(k) Number | K030791 |
|---------------|---------|
|---------------|---------|
Page __ of __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.