KARTUSH TM PATCHER

K964484 · Xomed, Inc. · KHJ · Jan 24, 1997 · Ear, Nose, Throat

Device Facts

Record IDK964484
Device NameKARTUSH TM PATCHER
ApplicantXomed, Inc.
Product CodeKHJ · Ear, Nose, Throat
Decision DateJan 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3620
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Kartush TM Patcher is intended for use as a patch to aid coverage of a tympanic membrane perforation in a dry ear.

Device Story

Kartush TM Patcher; silicone elastomer implant; double-flange collar-button design; no lumen/through-hole. Used by ENT surgeons to cover tympanic membrane (TM) perforations. Device placed in dry ear; flanges conform to TM surface to secure patch over perforation; central shaft provides stability to prevent displacement. Trimmed to size by surgeon. Silicone material biocompatible; similar to existing tympanostomy tubes and nasal septal buttons. Benefits: provides secure coverage of perforation; restores middle ear protection; less invasive than surgical repair. Potential risks: infection, premature extrusion, perforation widening, or granulomatosis reaction.

Clinical Evidence

No clinical data provided. Substantial equivalence based on design, material, and intended use comparison to existing ENT silicone implants.

Technological Characteristics

Silicone elastomer; double-flange collar-button configuration; non-porous (no lumen); soft/pliable for folding and trimming; biocompatible material.

Indications for Use

Indicated for patients with tympanic membrane perforations in a dry ear requiring coverage to aid healing.

Regulatory Classification

Identification

Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.

Predicate Devices

Submission Summary (Full Text)

{0} K964484 # 510(k) Summary ## 1.0 Date Prepared November 6, 1996 JAN 24 1997 ## 2.0 Submitter (Contact) David Timlin Xomed Surgical Products Jacksonville, FL (904) 279-7532 ## 3.0 Device Name Proprietary Name: **Kartush TM Patcher** Common Name: Tympanic Membrane (TM) patch Classification Name: Not Classified ## 5.0 Device Classification The Kartush TM Patcher has not yet been classified. It is similar to a number of current ENT devices: a tympanostomy tube, in that both are intended to be implanted in the tympanic membrane; a nasal septal button, in that both are intended to cover a perforation in mucosal lined tissue; and silicone sheeting, as both are silicone implant material for ENT repairs and both are used for patching TM perforations. | Tympanostomy Tube | Procode 77 ETD | Class II; 21CFR 874.3880 | Tier 2 | | --- | --- | --- | --- | | Nasal Septal Button | Procode 77 LFB | Class II; | Tier 1 | | Silicone sheeting | Procode 77 KHJ | Class II; 21 CFR 874.3620 | Tier 2 | ## 6.0 Device Description The Kartush TM Patcher is comprised of a double flange design, using silicone elastomer as the biocompatible material. The configuration is similar to a collar button style tympanostomy tube, with the exception that no through-hole or lumen is present. A variety of sizes or diameters of the TM Patcher will be available for approximating different size perforations. Alternately, the TM Patcher may be trimmed to size by the surgeon to fit the existing perforation. The flange thickness is the same as silicone sheeting currently used to treat TM perforations. The soft, pliable flange, which can conform to the surface of the TM, also easily folds for removal of the device as needed. As with nasal septal buttons, the silicone elastomer material is the same as used in the manufacture of Xomed tympanostomy tubes. ## 7.0 Intended Use Kartush TM Patcher is intended for use as a patch to aid coverage of a tympanic membrane perforation in a dry ear. 0022 {1} 0023 ## 8.0 Substantial Equivalence The Kartush TM Patcher described in this notification has the same intended use as the currently used silicone sheeting. But silicone sheets, although commonly used, are not designed specifically for TM patching. The TM Patcher, designed and intended specifically for TM patching, draws on the design benefits of the other predicate devices, the tympanostomy tube and the nasal septal button, that have proven to be safe as a TM implant and effective as a nasal perforation patch, respectively. The TM patch functions in the same regard as the silicone sheeting, but the central shaft and inner flange of the TM Patcher provide added security to help reduce displacement. A more secure patch helps to restore the protection of the middle ear effected by the tympanic membrane. Although it is expected that there could be adverse effects similar to the predicate devices, such as infection, premature extrusion, widening or persistent perforation, or granulomatosis reaction; the TM Patcher does not raise any new issues of safety or effectiveness and certainly involves less risk than that associated with a surgical repair.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...