Published clinical literature (D'Eredita, 2012); Unpublished clinical case series (Hsu, 2015; Toh C. et al., 2003; Ofo E. et al., 2009; Lalwani A., 2003)
Clinical data from retrospective cohorts and case series were used to demonstrate the safety and efficacy of the SIS material in myringoplasty and tympanoplasty procedures compared to autologous tissue.
myringoplasty; tympanoplasty; SIS material; clinical case series
Safety, efficacy (tympanic membrane closure), and procedure duration
Unpublished clinical data series; Clinical case series; Study Period: 2003-2015
Patients undergoing otologic procedures; Sample Size: 77 total (18, 19, 32, and 8 patients respectively); Number of Sites: 4
Not applicable for this study
Safety (adverse events)
Indications for Use
The Biodesign Otologic Repair Graft is intended for use as an implant to aid in surgical repairs and as an adjunct to aid in the natural healing process in various otologic procedures, including but not limited to myringoplasty and tympanoplasty. The device is supplied sterile and is intended for one-time use.
Device Story
Absorbable multi-layer biomaterial; composed of four layers of lyophilized porcine small intestinal submucosa (SIS) extracellular collagen matrix. Implanted by surgeons during otologic procedures (e.g., myringoplasty, tympanoplasty). Acts as scaffold for host cell infiltration; supports natural tissue repair process; degrades over time. Available as discs (4, 6, 9 mm) or square sheets (2.5x2.5 cm, 5x5 cm). Supplied sterile in Tyvek pouches. Benefits patient by providing structural support for tympanic membrane repair, facilitating healing, and serving as a viable alternative to autologous tissue.
Clinical Evidence
Prospective study (n=404) compared SIS material to temporalis fascia (PTF) in myringoplasty. Results: 97.2% closure rate for SIS vs 94.8% for PTF (follow-up 2-11 years). No adverse reactions observed. Additional unpublished data from four cohorts (n=18, 19, 32, 8) reported no significant adverse events. Bench testing included ISO 10993-1 biocompatibility (genotoxicity, cytotoxicity, sensitization, etc.) and mechanical burst strength testing. Animal studies (chinchilla/mouse) confirmed biocompatibility, efficacy, and degradation profile.
Technological Characteristics
Material: Porcine small intestinal submucosa (SIS) containing collagen types I, III, IV, VI. Form factor: Discs (4-9 mm) and sheets (2.5-5 cm). Thickness: 100-500 μm. Principle: Scaffold for host cell infiltration/remodeling. Sterilization: Ethylene oxide. Shelf life: 18 months. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for patients undergoing otologic surgical procedures, including myringoplasty and tympanoplasty, requiring an implant to aid in surgical repair and natural healing.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{1}------------------------------------------------
Public Health Service
DEPARTMENT OF HEALTH & HUMAN SERVICES - USA
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2015
Cook Biotech Incorprated Katie Molland, Ph.D. Regulatory Affairs Specialist 1425 Innovation Place West Lafayette, IN 47906
Re: K150594
Trade/Device Name: Biodesign Otologic Repair Graft Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, and Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: KHJ Dated: August 14, 2015 Received: August 17, 2015
Dear Dr. Molland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Katie Molland, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours. Deborah L. Falls -5
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150594
Device Name Biodesign(R) Otologic Repair Graft
#### Indications for Use (Describe)
The Biodesign Otologic Repair Graft is intended for use as an implant to aid in surgical repairs and as an adjunct to aid in the natural healing process in various otologic procedures, including but not limited to myringoplasty. The device is supplied sterile and is intended for one-time use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
### March 4, 2015
### Cook Biotech Incorporated
# Biodesign® Otologic Repair Graft
| Manufacturer Name: | Cook Biotech Incorporated<br>1425 Innovation Place<br>West Lafayette, Indiana 47906<br>Telephone: +1 (765) 497-3355<br>FAX: +1 (765) 807-7709 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Perry W. Guinn |
|-------------------|----------------|
|-------------------|----------------|
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Biodesign Otologic Repair Graft |
|-----------------------------|------------------------------------------------------------------------|
| Common Name | Surgical implant polymer material/Surgical<br>adjunct polymer material |
| Classification Regulations: | Class II, 21 CFR §874.3620 (KHJ) |
INDICATIONS FOR USE:
The Biodesign Otologic Repair Graft is intended for use as an implant to aid in surgical repairs and as an adjunct to aid in the natural healing process in various otologic procedures, including but not limited to myringoplasty and tympanoplasty. The device is supplied sterile and is intended for one-time use.
| PREDICATE DEVICES: | EpiFilm® Otologic Lamina, K982870 |
|--------------------|-----------------------------------|
| | MeroGel®Otologic Pack, K001148 |
### DEVICE DESCRIPTION:
The Biodesign Otologic Repair Graft is an absorbable multi-layer biomaterial composed of four layers of laminated extracellular collagen matrix derived from porcine small intestinal submucosa (SIS). The SIS material is lyophilized and then punched into the desired shape. The device is available in 4 mm, 6 mm and 9 mm diameter discs, as well as 2.5 x 2.5 cm and 5 x 5 cm square sheets. Upon implantation, the Biodesign Otologic Repair Graft is infiltrated by the host cells and acts as a scaffold for these cells during the body's natural repair process.
Additionally, the circular configurations of the device are packaged in a dried state and supplied sterile in a tray inside a sealed Tyvek® pouch. The square
{5}------------------------------------------------
configurations of the device are also packaged sterile in a dried state inside a sealed Tyvek® pouch.
## EQUIVALENCE TO MARKETED DEVICES
The Biodesign Otologic Repair Graft is similar with respect to intended use, materials (naturally occurring constituents of the extracellular matrix) and technological characteristics of the predicate devices in terms of section 510(k) substantial equivalence. Substantial equivalence is supported by biocompatibility testing (conducted in accordance to ISO 10993-1 standards), mechanical, pre-clinical and clinical testing.
## Biocompatibility testing
The following biocompatibility tests were performed on sterilized SIS devices which are identical in composition to the Biodesign Otologic Repair Graft (according to the ISO 10993-1 standard):
- Genotoxicity ●
- Direct contact in vitro hemolysis ●
- Cytotoxicity ●
- Muscle implantation ●
- Acute intracutaneous reactivity ●
- ISO sensitization
- Acute systemic toxicity ●
- Pyrogenicity
- LAL endotoxins ●
- Subchronic systemic toxicity
The results of these tests provide evidence that the Biodesign Otologic Repair Graft meets the biocompatibility requirements of the ISO standard.
## Mechanical Testing
The Biodesign Otologic Repair Graft material was tested for burst strength and the results compared with its predicates.
The results of this mechanical test provide evidence that the Biodesign Otologic Repair Graft has adequate mechanical strength for its application.
## Animal Testing
The SIS material that comprises the Biodesign Otologic Repair Graft was tested in animal studies that included an efficacy study using a chinchilla model and an implant study using a mouse model to characterize cellular response and device degradation. The efficacy study, which compared SIS repair with autologous tissue repair, suggested that SIS was a viable alternative to autologous tissue for tympanic
{6}------------------------------------------------
membrane perforation repair. Additionally, the mouse implant study compared the Biodesign Otologic Repair Graft against the MeroGel Otologic Pack and showed that the subject device performed similarly to the predicate in terms of device degradation, and non-inflammatory host responses. These animal studies provide evidence that the Biodesign Otologic Repair Graft is biocompatible and safe for its application.
## Clinical Testing
Prospective data was collected on the use of the SIS material (labeled as Surgisis), the same material that comprises the Biodesign Otologic Repair Graft (D'Eredita, 2012, abstract). In this 404 patient study, the SIS material was used in 217 myringoplasty procedures and compared to 215 temporalis fascia (PTF) repairs performed by the same surgeon. Follow- up was from 2-11 years (average 7.7 years) (data from manuscript submitted by invitation to the International Journal of Pediatric Otorhinolaryngology (February 5, 2013) by D'Eredita). Data analysis included safety, efficacy and procedure duration. No adverse reactions were observed with either type of repair. Stable tympanic membrane closures were seen in 212/217 (97.2%) of SIS repairs compared to 204/215 (94.8%) of PTF procedures. The difference in procedural times between the two (2) arms was not statistically significant.
Additional unpublished data are available in which the device was implanted in:
- a) 18 patients (Hsu, DuPage Medical Group, 2015);
- b) 19 patients (Toh C. et al., Birmingham Heartland Hospital, UK,2003);
- c) 32 patients (Ofo E. et al.,North West London Hospital, UK, 2009); and
- d) 8 patients (Lalwani A. San Francisco, CA, COSM 2003).
No significant adverse events were reported.
Results of these clinical studies show that the Biodesign Otologic Repair Graft is safe and effective for its intended use.
# SUBSTANTIAL EQUIVALENCE
Table 1 below provides a comparison of the subject device and its predicates.
{7}------------------------------------------------
| Device | Biodesign Otologic Repair<br>Graft (subject) | EpiFilm Otologic Lamina<br>(Predicate) | MeroGel™ Otologic Pack<br>(Predicate) |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Cook Biotech Incorporated | Xomed Surgical Products | Medtronic Xomed |
| 510(k) Number | Not assigned | K982870 | K001148 |
| Intended Use | The Biodesign Otologic Repair<br>Graft is intended for use as an<br>implant to aid in surgical<br>repairs and as an adjunct to aid<br>in the natural healing process<br>in various otologic procedures,<br>including but not limited to<br>myringoplasty and<br>tympanoplasty. | Intended for use as an implant to<br>aid in surgical repairs and as an<br>adjunct to aid in the natural<br>healing process in various<br>otologic procedures. EpiFilm<br>Otologic Lamina is indicated for<br>use in myringoplastic and<br>tympanoplastic surgical<br>procedures. | MeroGel™ Otologic Pack is a<br>space occupying dressing and/or<br>stent intended to separate<br>mucosal surfaces, help control<br>minimal bleeding and act as an<br>adjunct to aid in the natural<br>healing process. The device is<br>indicated for use in the middle<br>ear and external canal following<br>canalplasty, myringoplasty,<br>tympanoplasty, and stapes and<br>mastoid surgery. |
| Material | Small intestinal submucosa<br>(SIS)<br>Primarily Types I, III, IV and<br>VI collagen (constituents of the<br>extracellular matrix) | HYAFF® (ester of hyaluronic<br>acid, a naturally occurring<br>constituent of the extracellular<br>matrix.) | HYAFF® (ester of hyaluronic<br>acid, a naturally occurring<br>constituent of the extracellular<br>matrix.) |
| Method of action | Has a scaffold structure which<br>is infiltrated with host cells,<br>forms gel as the process of<br>remodeling occurs | Has micro-perforation providing<br>permeable surface and acts as a<br>scaffold for migrating host cells | Hygroscopic, forms gelatinous<br>mass in contact with fluids |
| Dimensions | 4 mm, 6mm, 9 mm diameter<br>2.5 x 2.5 cm<br>5 x 5 cm | 8 mm diameter (EpiDisc)*<br>2.5 cm x 2.5 cm | 1 cm x 5 cm, 4 cm x 4 cm |
| Thickness | 100 μm to 500 μm | NA | 340 μm† |
| Sterilization | Ethylene oxide | Gamma irradiation | Gamma irradiation |
| Shelf life | 18 months | NA | 48 months |
# Table 1 – Substantial Equivalence Comparison
NA – Not available
†N=2
*EpiDisc and EpiFilm Otologic Lamina are the same material and sold under the same 510(k). EpiDisc is a smaller sized device than EpiFilm
CONCLUSION: The biocompatibility, pre-clinical and clinical tests performed on the Biodesign Otologic Repair Graft show that the device is substantially equivalent to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.