RESOUND REAL EAR LOUDNESS MAPPING (RELM) SYSTEM
Device Facts
| Record ID | K960229 |
|---|---|
| Device Name | RESOUND REAL EAR LOUDNESS MAPPING (RELM) SYSTEM |
| Applicant | Resound Corp. |
| Product Code | ETW · Ear, Nose, Throat |
| Decision Date | Mar 7, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.3310 |
| Device Class | Class 2 |
Indications for Use
The ReSound® Real Ear Loudness Mapping (RELM) System is designed to measure loudness growth in hearing-impaired subjects in sound field, instead of with insert phones, in the unaided and aided conditions, through the use of subjective and objective measures, thus integrating both electroacoustic and psychoacoustic test capabilities.
Device Story
System measures loudness growth in hearing-impaired subjects; utilizes sound field speaker system and real ear probe microphone; inputs include subjective patient responses and objective sound pressure level measurements in ear canal; PC software processes data to generate frequency intensity curves; integrates electroacoustic and psychoacoustic testing; calculates target response curves for hearing device fitting; enables fine-tuning of hearing device parameters to patient loudness targets; used in clinical settings by hearing healthcare professionals; output displayed on PC screen; assists clinicians in optimizing hearing aid performance.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through comparison of functional capabilities, measurement methods, and data processing logic against predicate audiometric diagnostic and hearing aid analysis systems.
Technological Characteristics
System components: proprietary PC software, sound field speaker system, real ear probe microphone. Operates via PC-based signal generation and data acquisition. Displays frequency intensity curves. Software runs under Noah operating system. No specific material standards or sterilization methods cited.
Indications for Use
Indicated for hearing-impaired subjects requiring hearing device fitting and fine-tuning of hearing device parameters based on loudness growth measurements.
Regulatory Classification
Identification
A hearing aid calibrator and analysis system is an electronic reference device intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid, master hearing aid, group hearing aid or group auditory trainer. The device consists of an acoustic complex of known cavity volume, a sound level meter, a microphone, oscillators, frequency counters, microphone amplifiers, a distoration analyzer, a chart recorder, and a hearing aid test box.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
Predicate Devices
- ReSound® Portable Prescriptive Programming (P³) System (K912669)
- Frye Electronics, Inc., Type 6400 Real Ear Hearing Aid Analyzer (K872242)