TULA Tube Delivery System

K171239 · Tusker Medical · ETD · Jun 28, 2017 · Ear, Nose, Throat

Device Facts

Record IDK171239
Device NameTULA Tube Delivery System
ApplicantTusker Medical
Product CodeETD · Ear, Nose, Throat
Decision DateJun 28, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 874.3880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Tula Tube Delivery System is intended to provide a means to create a myringotomy with insertion of a preloaded Grommet type tympanostomy tube.

Device Story

TULA Tube Delivery System (TDS) is a handheld device for performing myringotomy and tympanostomy tube insertion. Device consists of four concentric components: cutter, dilator, shield, and stabilizer. Operation involves button-controlled activation to penetrate the tympanic membrane, create the myringotomy, dilate the incision, and deploy a preloaded Grommet-style tympanostomy tube. Used by clinicians in a clinical setting. The device automates the mechanical steps of tube placement, reducing procedural complexity compared to manual methods. Benefits include precise, controlled tube delivery and simplified myringotomy creation.

Clinical Evidence

Bench testing only. Testing included tube deployment verification, deployment time, hearing safety, loudness discomfort level, and diametral interference. Biocompatibility testing performed per ISO 10993-1 and FDA G95-1. Sterilization validated per ISO 11135:2014.

Technological Characteristics

Handheld delivery system with four concentric components (cutter, dilator, shield, stabilizer). Mechanical actuation via button control. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1 compliant. No software or electronic components described.

Indications for Use

Indicated for patients requiring myringotomy and insertion of a preloaded Grommet type tympanostomy tube.

Regulatory Classification

Identification

A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to form a single, abstract figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993 - 0002 June 28, 2017 Tusker Medical Pavan Sethi Director. Regulatory Affairs 155 Jefferson Drive Suite 200 Menlo Park, California 94025 Re: K171239 Trade/Device Name: TULA Tube Delivery System Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy Tube Regulatory Class: Class II Product Code: ETD Dated: June 1. 2017 Received: June 2, 2017 Dear Pavan Sethi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {1}------------------------------------------------ (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Eric A. Mann -S for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Tusker Medical. The logo features a stylized origami elephant in brown and white, followed by the word "TUSKER" in bold, black letters. Below "TUSKER" is the word "MEDICAL" in a smaller, gray font. Underneath the logo is the text "TULA Tube Delivery System". ## INDICATIONS FOR USE | 510(k) Number (if known): | K171239 | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name: | TULA® Tube Delivery System | | Common Name: | Tympanostomy Tube and Tympanostomy Tube Delivery System | | Indications For Use: | The Tula Tube Delivery System is intended to provide a means to create a<br>myringotomy with insertion of a preloaded Grommet type tympanostomy<br>tube. | Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "Special 510(k) Premarket Notification". The text is written in a simple, sans-serif font and is centered on the image. The text is likely a title or heading for a document or presentation related to medical device regulation. ## 510(k) SUMMARY | Sponsor/Submitter: | Tusker Medical<br>155 Jefferson Drive, Suite 200<br>Menlo Park, California 94025 | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Pavan Sethi, Ph.D.<br>Director, Regulatory Affairs<br>Phone: (408) 513-7529<br>Fax: (650) 223-6769 | | Date of Submission: | April 26, 2017 | | Device Trade Name: | TULA® Tube Delivery System | | Common Name: | Tympanostomy Tube and Tympanostomy Tube Delivery System | | Device Classification: | Class II | | Regulation Number: | 21 CFR 874.3880 | | Classification Name: | Tube, Tympanostomy | | Product Code: | ETD | | Predicate Device: | Tympanostomy Tube Delivery System (K103595) | | Device Description: | The Tula Tube Delivery System (TDS) is a device that penetrates the<br>tympanic membrane and inserts the Grommet type tympanostomy tube with<br>a button controlled activation. It is intended to provide a means to create a<br>myringotomy with insertion of a preloaded grommet type tympanostomy<br>tube. | | Indications for Use: | The Tula Tube Delivery System is intended to provide a means to create a<br>myringotomy with insertion of a preloaded Grommet type tympanostomy<br>tube. | | Technological<br>Characteristics: | Tula Tube Delivery System is composed of four concentric components that<br>cut, dilate, shield and stabilize. The cutter creates the myringotomy. The<br>dilator opens the incision while the shield introduces the tympanostomy tube<br>through the myringotomy. The subject device is identical to the predicate<br>device in method of operation, materials and intended use. The aspects of the<br>device that control the tympanostomy tube delivery (Cutter, Dilator, Shield,<br>Plunger, Clear Tip, Outer Tube) are unchanged between the subject and<br>predicate device. | | | The main differences in technological characteristics are: the modified device<br>will be pre-loaded with a Grommet style tympanostomy tube instead of a<br>Paparella style tympanostomy tube, and the Tube Delivery System is slightly<br>modified to accommodate and implant a Grommet type tympanostomy tube.<br>These differences in technological characteristics do not raise different<br>questions of safety and effectiveness than the predicate and do not render this<br>device Not Substantially Equivalent (NSE). | | Performance Data: | Bench verification testing was conducted to verify that the modified device<br>meets the design inputs and intended performance characteristics. The<br>testing included tube deployment test, tube deployment time upon actuation,<br>hearing safety test, loudness discomfort level and diameteral interference. | | | Sterilization of the modified TULA Tube Delivery System is validated per<br>ISO 11135:2014 - "Sterilization of Health Care Products-Ethylene Oxide:<br>Requirements for Development, Validation, and Routine Control of a<br>Sterilization Process for Medical Devices". The method used for sterilization<br>validation is the overkill (half-cycle approach) in a fixed chamber method per<br>Annex B of ISO 11135:2014. | | | Biocompatibility evaluation and testing conducted gave acceptable<br>results per ISO 10993-1 and FDA Bluebook Memorandum G95-1. | | Summary of<br>Substantial<br>Equivalence: | The subject device TULA Tube Delivery System is substantially equivalent<br>to the predicate device in indication for use, performance, fundamental<br>scientific technology, safety and effectiveness. | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for Tusker Medical. The logo features a stylized origami elephant in shades of gray, followed by the text "TUSKER" in bold, black letters. Below "TUSKER" is the word "MEDICAL" in a smaller, gray font. Underneath the logo is the text "TULA Tube Delivery System" in a serif font.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...