PICCOLO K-AMP VC+, PICCOLO P-A1, PICCOLO P-AO AND PICCOLO ATC
Device Facts
| Record ID | K964451 |
|---|---|
| Device Name | PICCOLO K-AMP VC+, PICCOLO P-A1, PICCOLO P-AO AND PICCOLO ATC |
| Applicant | Rexton, Inc. |
| Product Code | ESD · Ear, Nose, Throat |
| Decision Date | Nov 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.3300 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Indications for Use
To amplify and transmit sound to the ear.
Device Story
Piccolo series (K-amp VC+, P-A1, ATC, P-AO) BTE hearing aids; compact housing compatible with glasses; utilizes standard hearing aid components; powered by size 13 battery. Features include AGC-I (Automatic Gain Control-Input), active low-frequency tone control, and 3-position switch (Off-Telecoil-Microphone). Operated by patient; frequency response adjusted by clinician based on individual audiogram. Benefits include improved battery life and reduced distortion via modern receiver technology.
Clinical Evidence
No clinical data; bench testing only per ANSI S3.22-1987 specifications.
Technological Characteristics
BTE hearing aid; standard components; size 13 battery; AGC-I; active low-frequency tone control; 3-position switch (Off-Telecoil-Microphone). Complies with ANSI S3.22-1987.
Indications for Use
Indicated for individuals with hearing impairment requiring sound amplification, with frequency response customized based on individual audiograms.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.