K133352 · Hypersound Health, Inc. · EPF · Feb 12, 2014 · Ear, Nose, Throat
Device Facts
Record ID
K133352
Device Name
HYPERSOUND AUDIO SYSTEM (HSS)
Applicant
Hypersound Health, Inc.
Product Code
EPF · Ear, Nose, Throat
Decision Date
Feb 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HyperSound Group Auditory Trainer is indicated for use as a group auditory trainer or group hearing aid used to communicate simultaneously with one or more listeners with or without hearing loss and with or without the use of hearing aids in order to improve clarity and comprehension of sounds generated from sources such as a microphone, CD/DVD player, TV, Stereo System or other sound generation systems.
Device Story
HyperSound Audio System (HSS) delivers directional sound via ultrasonic carrier waves. Input: audio signals from microphones, TVs, or media players. Processing: amplifier mixes audible sound with ultrasonic carrier; custom emitters project sound in a narrow, beam-like column. Output: audible sound demodulated in air within the beam path. Used in homes, schools, or businesses; operated by users or installed by contractors. No ear-coupling (headphones/earmolds) required. Benefits: improved speech understanding and clarity in noisy environments or at distance by directing sound specifically to the listener's area. Clinical decision-making: assists hearing-impaired users in group communication without traditional hearing aid coupling.
Clinical Evidence
Single-blinded, randomized crossover clinical study (n=10) with mild-to-severe hearing loss subjects. Compared HSS to conventional speakers at 70 dB. Primary endpoints: speech understanding (MRT, CNC, AzBio). Results: Significant gains in speech understanding; median AzBio scores increased from 0.0% to 34.9% (quiet, p=0.008) and 1.8% to 51.6% (noise, p=0.008). Median CNC whole word scores increased from 0.0% to 54.0% (p=0.004).
Technological Characteristics
System includes amplifier and custom ultrasonic emitters. Principle: ultrasonic carrier wave modulation for directional sound projection. Frequency range: 300Hz to 18kHz. Max output: 89dB SPL @ 1m. Power: <15W per channel. Connectivity: wired input from external sound sources. Safety: CE and FCC Part 15 Subpart B compliant; ultrasonic output within OSHA 29 CFR 1910.95 limits.
Indications for Use
Indicated for individuals with or without hearing loss (including mild to severe) to improve sound clarity and comprehension in group settings. Used with sound sources like microphones, TVs, or stereos. No contraindications noted.
Regulatory Classification
Identification
A group hearing aid or group auditory trainer is a hearing aid that is intended for use in communicating simultaneously with one or more listeners having hearing impairment. The device is used with an associated transmitter microphone. It may be either monaural or binaural, and it provides coupling to the ear through either earphones or earmolds. The generic type of device includes three types of applications: hardwire systems, inductance loop systems, and wireless systems.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
{0}------------------------------------------------
### 510(k) Summary HyperSound Health Audio System Model HSS-3000 (HSS) – Group Auditory Trainer 510(k) Number: K133352
#### 1. Submission Sponsor -
HyperSound Health, Inc. 13771 Danielson Street, Ste. L Poway, California 92064 USA Phone: 888-496-8001 Fax: 888-639-2150 Contact: Gus Bock, Management Representative
#### 2. Date Prepared -
October 23, 2013
#### 3. Device Name -
| Trade/Proprietary Name: | HyperSound Audio System (HSS) |
|-----------------------------------|---------------------------------------------|
| Common/Usual Name: | Group Auditory Trainer |
| Classification Name: | Group Auditory Trainer or Group Hearing Aid |
| Classification Regulation: | 874.3320 |
| Classification Panel: | Ear Nose and Throat |
| Product Code : | EPF |
| Device Class: | Class II |
| K Number: | K133352 |
| FDA Establishment Registration #: | 348096 |
#### 4. Predicate Device -
(1) Phonic Ear, Inc. K043090, SE: 12/23/2004
#### 5. Device Description -
The HyperSound Audio System (HSS) utilizes a specially designed emitter (speaker) that delivers beamlike sound that is clear to the listener. The beamed audio content begins to demodulate in the air within six inches of the emitter and results in a cone-like transmission of sound waves that can be directed towards the listener. This device is intended for sale as an over-the-counter product and can be used with any sound source and at any distance from 3 feet to 50 feet or more from the emitters (speakers). Ultrasonic technology has been used for many years in industry and is also utilized in medical devices, such as ultrasonic nebulizers. The ultrasonic sound waves this product is based upon have also used in commercial and localized communications, and they are below the ultrasonic sound safety limits established by OSHA as required by OSHA Noise Standard, 29 CFR 1910.95.
{1}------------------------------------------------
The device consists of an amplifier, software, and an emitter (speaker). No ear plugs or head phones are required to use this device. The device is distant from and does not touch the intended patient/user.,
#### 6. Indications for Use -
The HyperSound Group Auditory Trainer is indicated for use as a group auditory trainer or group hearing aid used to communicate simultaneously with one or more listeners with or without hearing loss and with or without the use of hearing aids in order to improve clarity and comprehension of sounds generated from sources such as a microphone, CD/DVD player, TV, Stereo System or other sound generation systems.
#### 7. Technological Characteristics and Substantial Equivalence -
The following table compares the HyperSound Audio System to the Predicate Devices with respect to intended use, technological characteristics and principles of operation. These comparisons provide more detailed information regarding the basis for the determination of substantial equivalence.
The HyperSound Audio System (HSS) shares the same basic technology as standard consumer amplifiers and speakers as well as those speakers used in certain group auditory trainers, group hearing aids and assistive listening devices. Similar and substantially equivalent to K043090 (Easy Listener 2 FM) and other sound field amplification systems, HSS employs reproducing audio in the air and not through any coupling to the ear through either earphones or ear molds. There is no contact with the body. HSS is also substantially equivalent to various types of approved Group Hearing Aids or Group Auditory Trainers that are directly connected or coupled to the listener with similar results.
| Device<br>Proprietary Name | Submitted Device<br>HyperSound Audio System | Easy Listener 2 FM |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Being submitted | K043090 |
| Device Classification<br>Type and<br>Classification Name | Class II<br>Group hearing aid or auditory trainer | Class II<br>Group hearing aid or auditory trainer |
| Manufacturer | HyperSound Health, Inc. | Phonic Ear, Inc. |
| Product Code | EPF | EPF |
| Regulation Number | 874.3320 | 874.3320 |
| Device | Submitted Device | |
| Proprietary Name | HyperSound Audio System | Easy Listener 2 FM |
| Intended Use | The HyperSound Audio System is intended to be<br>used to improve the clarity of sound in the<br>home, business, school or other institutional<br>environment by individuals or groups with or<br>without hearing loss and with or without hearing<br>aids. The device is intended to overcome any<br>negative effects of distance, poor room<br>acoustics and background noise to improve<br>clarity (speech understanding and speech<br>discrimination) for those with normal hearing or<br>the hearing impaired with mild to severe hearing<br>loss without earphones or ear- molds. | Sound Field wireless amplification systems are<br>intended for use as group auditory trainers used to<br>communicate simultaneously with one or more<br>listeners with normal hearing or hearing impairments.<br>The device is used with an associated transmitter<br>microphone and speakers systems and does not<br>provide coupling to the ear through either earphones<br>or ear molds. The use of FM systems also reduces the<br>detrimental effects caused by the distance between<br>the speaker or sound source and the hearing impaired<br>listener. |
| Indications for Use | The HyperSound Group Auditory Trainer is<br>indicated for use as a group auditory trainer or<br>group hearing aid used to communicate<br>simultaneously with one or more listeners with<br>or without hearing loss and with or without<br>hearing aids in order to improve clarity and<br>comprehension of sounds generated from<br>sources such as a microphone, CD/DVD player,<br>TV, Stereo System or other sound generation<br>systems. | Sound Field wireless amplification systems are<br>intended for use as group auditory trainers used to<br>communicate simultaneously with one or more<br>listeners with normal hearing or hearing impairments.<br>The device is used with an associated transmitter<br>microphone and speakers systems and does not<br>provide coupling to the ear through either earphones<br>or ear molds.<br>These systems use FE and/or wireless technology<br>which include receivers, transmitters, and strategic<br>placement of high quality speaker systems and can be<br>used with or without BTE hearing aids nad/or personal<br>FM devices. The talker's voice is mildly amplified and<br>dispersed throughout the room to assist listeners,<br>regardless of seating location, to consistently hear<br>whatever the talker is saying. |
| Overall Design | A microphone, CD/DVD or other sound source is<br>routed to an amplifier, which mixes audible<br>sound with an ultrasonic carrier. The output is<br>sent to specially designed emitters (speakers),<br>which deliver the waves to produce sound into<br>the air through a conical beam. The waves begin<br>to demodulate into high quality sound waves<br>within six inches of the emitter and continue to<br>travel within a conical beam to the desired area.<br>The sound is clearly delivered to the listener at<br>normal volume levels free with reduced external<br>noise. The HSS System may be portable or<br>installed permanently by a sound contractor. | The device consists of FM or IR transmitters and<br>microphone used by a speaker to send auditory signals<br>to FM or IR receiver/amplifier which processes the<br>voice signals and broadcasts the signals through one or<br>more loudspeakers. These systems may be installed by<br>a sound contractor. Sound is reproduced by the<br>speaker or speaker systems to recreate the sound at a<br>slightly increased volume level. |
| Device<br>Proprietary Name | Submitted Device<br>HyperSound Audio System | Easy Listener 2 FM |
| How it Works | HSS electronically mixes audible sound with an<br>ultrasonic carrier. These audible tones are<br>projected within a beam of silent energy heard<br>by those in the targeted area. Unlike a<br>conventional speaker, sound is not created<br>omni-directionally at the emitter (speaker)<br>surface but is created along and within a highly<br>directional air column. Sound is only heard if a<br>listener's head is within the beam or the beam<br>hits a reflective surface whereupon sound is<br>created at the point of reflection. This sound,<br>which is created in the air, can be directed to<br>nearly any desired point in the listening<br>environment, controlling the direction of sound. | The system works essentially as a monaural or stereo<br>system for the processing of sound input through the<br>amplifier for delivery to the speakers placed throughout<br>the audience. Sound is not directional and outside noise<br>may comingle with.the sound delivered. FM or IR are<br>used in lieu of cables, but serve the same purpose. |
| Physical dimensions | Amplifier 6x6x1.5 inches, individual or<br>multiple emitters: 12.5x6.5x1 inches | Varies based on type of FM transmitter, receiver<br>and speaker(s) employed |
| Maximum audio<br>output<br>characteristics | 89dB SPL @ 1m | 90dB SPL @ 1m |
| Frequency range | Generally 300Hz to 18kHz (may be used with off-<br>shelf consumer powered woofer where<br>improved bass (lower frequencies) is desired) | Generally 50Hz to 20kHz depending on speaker size<br>type |
| Energy used | < 15 watts per channel | Generally 30W per speaker |
| Transducer type | Custom emitter (speaker) | Conventional speaker |
| Safety | HyperSound delivers audio through the air<br>within a conical beam and without any earphone<br>or ear mold to deliver sound free of external<br>sounds or sources of distortion.<br>Users are instructed to not aim an emitter<br>(speaker) directly in to the ear canal of an<br>individual at less than 3 feet distance.<br>The amplifier is CE compliant, FCC compliant<br>(Part 15, Subpart B) and approved for Electrical<br>Safety by Nemko CCL (Nationally Recognized) | There are no known contraindications associated<br>with the use of sound field group amplification<br>systems.<br>Many case studies, experiments, and trials exist that<br>document the benefits and effectiveness of speech<br>perception from sound field amplification.<br>Amplifier and speakers meet FCC and Consumer<br>Electrical Safety requirements. |
...
{2}------------------------------------------------
ﺑ
{3}------------------------------------------------
{4}------------------------------------------------
#### 8. Non-Clinical Testing
Comparable systems using the same technology as the HyperSound Audio System have been in commercial use since 2001 for general audio use primarily commercial advertising systems. The manufacturer has submitted the applicable report to the FDA pursuant to the Radiation Control for Health and Safety Act of 1968 (Title 21, CFR, sub-chapter J) as it pertains to ultrasonic devices. Laboratory bench testing, comparative testing and verifications related to device performance, production of sound volume over distance, has demonstrated the HyperSound Audio System consistently and safely delivers clear and understandable sounds, letters and words to the patient or to the listener at distances from 3 feet to 50 feet or more with sound levels of up to 76.5 dB in normal listening and sound reproduction environments.
Both sound pressure level measurement and sound beam plotting measurements were performed to demonstrate equivalent output at various distances. Beam plotting demonstrated focused delivery of frequencies important to speech discrimination in a targeted beam important to augment hearing for selected users in home, classroom or other group settings without the need for headphones or segregation.
Equivalency testing to a predicate device was performed by an independent laboratory pursuant to ANSI S3.2-2009-American National Standard Method for Measuring the Intelligibility of Speech over Communication Systems. Testing of the HSS system demonstrated a mean ANSI S3.2-2099 MRT (Modified Rhyme Test) score of 91% and the predicate device had a mean score of 79.2% demonstrating more than equivalency in word list discrimination amongst normal subjects.
#### 9. Clinical Testing -
A clinical study was conducted by the California Hearing and Balance Center in La Jolla, CA under a formal, written, Institutional Review Board (IRB) approved protocol. The clinical study was performed in a controlled audiology laboratory as a single-blinded, randomized cross over clinical study using adult subjects with hearing loss ranging from mild to severe degree. Testing was based upon speech tests (such as Modified Rhyme Test (MRT) and Consonant-Nucleus-Consonant or CNC in quiet and AzBio in noise) to assess the effectiveness of the HSS System compared to conventional speakers at comparable volume levels.
Ten adult patients with mild to severe hearing loss with pure-tone average (PTA) of > 30 dB and word discrimination scores of < 80% in both ears were tested at one meter distance from the audio source. Significant gains in speech understanding were associated with the HSS versus conventional speaker for all test conditions at 70 dB. Median AZBio scores increased from 0.0% to 34.9% (p=0.008) in quiet, and from 1.8% to 51.6% in noise (p=0.008). Median CNC whole word test scores increased from 0.0% to 54.0% (p=0.004) and median phoneme test scores from 4.0% to 63.4% (p=0.004).
Data and results demonstrated improvement in sound clarity over conventional speakers at 70 dB, including in background noise, in those with mild to severe hearing loss.
{5}------------------------------------------------
#### 10. Conclusion -
By definition, a subject device (new device) is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device or the subject device has the same intended use and different technological characteristics that subject device can be demonstrated to be substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
The HyperSound Audio System meets these criteria, and as designed and manufactured, is considered to be substantially equivalent to the referenced predicate devices based on testing and clinical evaluation.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2014
HyperSound Health, Inc. c/o Mr. James Barnes 13771 Danielson Street, Ste. L Poway, CA 92064
Re: K133352
Trade/Device Name: HyperSound Audio System (HSS) Regulation Number: 21 CFR 874.3320 Regulation Name: Group Auditory Trainer or Group Hearing Aid Regulatory Class: Class II Product Code: EPF Dated: January 15, 2014 Received: January 16, 2014
Dear Mr. Barnes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. James Barnes ………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
:
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
## Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### Indications for Use Statement
#### HyperSound Health Audio System Model HSS-3000 (HSS) - Group Auditory Trainer 510(k) Number: K133352
Device Name: HyperSound Health Audio System Model HSS-3000 (HSS) - Group Auditory Trainer
Indications for Use:
The HyperSound Group Auditory Trainer is indicated for use as a group avditory trainer or group hearing aid used to communicate simultaneously with one or more listeners with or without hearing loss and with or without the use of hearing aids in order to improve clarity and comprehension of sounds generated from sources such as a microphone, CD/DVD player, TV, Stereo System or other sound generation systems.
_ {Part 21 CFR 801 Subpart D} Prescription Use _ AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Cherish R. Giusto -A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.