MODIFICATION TO OTOACOUSTIC EMISSIONS (OAE0 TEST INSTRUMENT

K964132 · Bio-Logic Systems Corp. · EWO · May 13, 1997 · Ear, Nose, Throat

Device Facts

Record IDK964132
Device NameMODIFICATION TO OTOACOUSTIC EMISSIONS (OAE0 TEST INSTRUMENT
ApplicantBio-Logic Systems Corp.
Product CodeEWO · Ear, Nose, Throat
Decision DateMay 13, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1050
Device ClassClass 2
AttributesPediatric

Intended Use

The intended use of the current Scout Otoacoustic Emissions Test Instrument is to measure or determine cochlear function either by measuring and recording transient (click evoked) otoacoustic emissions or by measuring and recording distortion product otoacoustic emissions utilizing continuous pure tones. The intended use of this new software feature is essentially the same as that of the predicate device. It provides for an additional test methodology which measures DPOAE threshold instead of the more-routine DPOAE amplitude test. The threshold test (also called the Input/Output (I/O) function) involves varying the stimulus intensity levels while keeping the frequencies constant. Both test methods can be effectively used to differentiate normal and hearing- impaired subjects, but the amplitude test is usually preferred because it usually requires a shorter testing time. However, in cases where the amplitude test results in a DP-gram with a low DP response (close to the noise floor) at specific frequencies, the I/O function test can be used to determine if there is an improved response at other stimulus frequencies. Therefore, the I/O function test offers the potential to improve the identification of patients with normal hearing.

Device Story

Device measures cochlear function via transient or distortion product otoacoustic emissions (DPOAE). Input: acoustic signals via Etymotic ER-10C probe. Operation: software controls stimulus generation; original device keeps intensity constant while varying frequency; new feature (I/O function) keeps frequency constant while varying stimulus intensity to determine DPOAE threshold. Output: DPOAE amplitude/threshold data displayed on PC. Used in clinical settings by audiologists/technicians. Output assists in identifying hearing impairment by assessing outer hair cell integrity. Benefit: improved identification of normal hearing in cases where routine amplitude tests show low response near noise floor.

Clinical Evidence

Bench testing only. System verified and validated in accordance with IEEE Standards for Software Engineering and internal development procedures to ensure performance matches specifications.

Technological Characteristics

PC-controlled OAE test instrument. Uses Etymotic ER-10C ear probe. No direct electrical patient connection. Software-controlled stimulus generation for transient and distortion product emissions. Software developed per IEEE standards.

Indications for Use

Indicated for adults, children, and infants to assess outer hair cell function in the cochlea via DPOAE threshold or amplitude measurements to differentiate normal and hearing-impaired subjects.

Regulatory Classification

Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K964132 BIO-LOGIC MAY 13 1997 One Bio-logic Plaza Mundelein, Illinois 60060-3700 1-800-323-8326 Fax: 847-949-8615 # SECTION 2: SUMMARY AND CERTIFICATION 510(K) SUMMARY SAFETY AND EFFECTIVENESS SUMMARY Safety and effectiveness information concerning this modification to the Scout OAE device is summarized below. Because this is not a CLASS III device, the special certification defined for this section is not required. PREPARED BY: Bio-logic Systems Corp One Bio-logic Plaza Mundelein, IL 60060 TELEPHONE: (847)-949-5200 CONTACT PERSON: Norman E. Brunner DATE ON WHICH THE SUMMARY WAS PREPARED: October 7, 1996 NAME OF DEVICE: Modification and Software Feature Addition to the Scout Otoacoustic Emissions (OAE) Test Instrument. COMMON NAME: Otoacoustic Emissions Test Instrument CLASSIFICATION NAME: Audiometer (per CFR 874.1050). PREDICATE DEVICE: Bio-logic Scout OAE Instrument (K944735) Page 7 {1} REVISED STATEMENT OF INTENDED USE FOR 510(k) #K964132 INTENDED USE: The intended use of the current Scout Otoacoustic Emissions Test Instrument is to measure or determine cochlear function either by measuring and recording transient (click evoked) otoacoustic emissions or by measuring and recording distortion product otoacoustic emissions utilizing continuous pure tones. The intended use of this new software feature is essentially the same as that of the predicate device. It provides for an additional test methodology which measures DPOAE threshold instead of the more-routine DPOAE amplitude test. The threshold test (also called the Input/Output (I/O) function) involves varying the stimulus intensity levels while keeping the frequencies constant. Both test methods can be effectively used to differentiate normal and hearing- impaired subjects, but the amplitude test is usually preferred because it usually requires a shorter testing time. However, in cases where the amplitude test results in a DP-gram with a low DP response (close to the noise floor) at specific frequencies, the I/O function test can be used to determine if there is an improved response at other stimulus frequencies. Therefore, the I/O function test offers the potential to improve the identification of patients with normal hearing. DPOAE threshold is the lowest stimulus intensity level which produces a DPOAE response amplitude that is distinguishable from the level of system distortion and from the noise floor. DPOAE threshold is expressed in much of the published literature as the lowest intensity of the F2 stimulus that generates a clear DPOAE response. DPOAE threshold does not equate to and should not be confused with audiometric or auditory threshold. Whereas auditory threshold measures assess the integrity of the entire auditory system from outer ear to cortex, DPOAE threshold measures are limited to assessment of outer hair cell function in the cochlea only. (This statement replaces the text on pages 8 and 11 of the original 510(k) submission.) {2} Modification and Feature Addition to Bio-logic Scout Otoacoustic Emissions (OAE) Test Instrument Page 3 ## PATIENT POPULATION: Adults, children and infants. ## SAFETY AND EFFECTIVENESS: No changes have been made to the hardware design of the Scout product line. Therefore, the Scout product which incorporates this new software feature does not present any safety issues which have not already been addressed in the predicate Scout 510(k) notifications. Also, the Scout software modules which are used to control the hardware have not been changed. This new feature only changes the sequence in which tones of varying intensity and frequency are generated when conducting a test. Instead of keeping stimulus intensity constant and varying frequency, the frequencies are kept constant and the stimulus intensity is varied. Therefore, this new feature does not affect the safety of the Scout product. To establish the safety and effectiveness of the software which controls the Scout OAE Test Instrument with the addition of the Input/Output feature, the system was validated in accordance with the IEEE Standards for Software Engineering, as well as Bio-logic internal software development policies and procedures modeled after the IEEE Standards. Changes to the standard Scout PC computer program necessary to generate new test sequences and the presentation of this data were all developed and tested as specified in these procedures. The system, for which this application is submitted, was verified and validated; it was found to perform in accordance with specifications. Page 9 {3} Modification and Feature Addition to Bio-logic Scout Otoacoustic Emissions (OAE) Test Instrument Page 4 The following comparison is provided as a summary of technological characteristics relative to the predicate device Scout. This is to demonstrate that the addition of this new Input/Output software feature creates no significant differences which would adversely affect product safety and effectiveness. | Parameter for comparison. | Similarity or Difference. | | --- | --- | | Intended Use | No differences. | | Population | No differences. | | Hardware Configuration | No differences. | | Computer Control Software | Only differences are changes to Scout software required to support the new Input/Output feature addition. There are no differences in software test algorithms. | | Ear Probe System | No differences. The system still uses the Etymotic ER-10C probe. | | Performance | The time to perform a DPOAE I/O (threshold) test is somewhat longer than that required for the more routine DPOAE amplitude test. However, the I/O test can provide additional information which the routine test cannot easily provide, which can, in some cases, be clinically useful. | | Safety Characteristics | No difference. There is no direct electrical connection to the patient in any Bio-logic OAE system. | Page 10
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...