K982642 · Fischer-Zoth Audiologic Systems, Inc. · EWO · Oct 21, 1998 · Ear, Nose, Throat
Device Facts
Record ID
K982642
Device Name
ECHO-SCREEN
Applicant
Fischer-Zoth Audiologic Systems, Inc.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Oct 21, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Echo-Screen Otoacoustic Emissions Test Instrument is a hand-held instrument with an attached probe (that is placed in the outer ear canal) which emits sound that evokes otoacoustic emissions. A microphone in the probe picks up the emissions if any are present, and the device then analyzes the emissions and mathematically separates noise from cochlear emissions. The presence of these emissions indicates cochlear function in adults, infants and children. The instrument requires no active participation from the patient.
Device Story
Handheld, battery-powered screening system for detecting cochlear dysfunction via Otoacoustic Emissions (OAE). Device emits click-bursts (40-100Hz) into ear canal via miniature probe; microphone records cochlear response. Automated evaluation software processes signals to separate noise from emissions; provides 'PASS' indication based on signal statistics. Used by trained personnel in medical or school environments. Output includes evaluation results, stimulus stability, and artifact rate. Facilitates early identification of hearing disorders; does not prescribe or fit assistive devices.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by compliance with electrical safety standards (IEC601-1, UL2601-1, VDE750) and verification of automated signal processing algorithms for OAE detection.
Technological Characteristics
Handheld device (21.5x10x5.4cm, 600g); 6V rechargeable battery. Probe contains one speaker and one microphone. Stimulus: 40-100Hz click-bursts; bandwidth 1.4kHz-4kHz. Max output 85dB SPL, software-limited. Connectivity: None during measurement. Software: C-based automated evaluation algorithm. Standards: ISO 9001, EN46001, IEC601-1, UL2601-1.
Indications for Use
Indicated for the detection of cochlear dysfunction in neonates, infants, children, and adults. Contraindications: None stated.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
Predicate Devices
Grason Stadler Inc. model GS170 AudioPath Screener (K974237)
{1}------------------------------------------------
K982642
OCT 21 1998
'510 (k) Summary' Summary and safety effectiveness
Submitter's Name & Address: Fischer-Zoth Audiologic Systems 11309 South Brandon Park Dr. 3 Sandy, Utah, 84092
Contact person & Telephone: Mr. Michael Soerensen (801) 253 0704
May 5, 1998 Data summary prepared:
Device name: Classification Name - Audiometer Common/Usual Name - Screening Audiometer Proprietary name - Echo-Screen
Predicate Device: Grason Stadler Inc. model GS170 AudioPath Screener (ref. 510(k) #K974237) and Otodynamics LTD. model ILO88 (ref. 510(k) #K962995,
Device description , intended Use & Effectiveness:
The Fischer-Zoth Audiologic System model 'Echo-Screen' is based upon Otoacoustic Emissions and it is similar to the Grason Stadler Inc. GSI70 AudioPath Screener and the Otodynamics OAE Analyser, model ILO88. __ ____________________________________________________________________________________________________________________________________
The purpose of the device is the reliable and earliest possible detection of hearing disorders due to cochlear dysfunction. It can be used beginning from neonates to adults.
The 'echo-screen' is a handheld, automated OAE-based screeming-system which is easy to use . The measurement flow is menu guided and the evaluation is based upon signal statistics.
It is intended to be used by trained personnel in a medical or school environment. The echo-screen is not intended for use in prescribing or fitting assisstive listening devices such as hearing aids or cochlea implants.
Technological Characterisitics: see also Attachment A
- Probe: miniature probe easy to apply even for premature infants - Hardware:
- **Hardware:**
Hand held device : size 21,5x10x5,4cm , weight 600g. Battery powered : 6V/1000mAh, rechargeable Stimulus Rate: 40 ... 100Hz Signal Bandwidth : 1,4KHz to 4KHz
- Software: echo-screen evaluation SW
Automated stimulus adjustment and online control
Signal statistical criterion for 'PASS' condition
Additional hints to measurement quality (stimulus stability and artefact rate)
Safety:
Battery powered, no connection to mains required Sound Pressure level: 70-85 dB SPL maximum
{2}------------------------------------------------
Safety:
The 'Echo-Screen' Screening System is designed to provide electrical safety to the patient as well as the user. The system is designed to meet the following standards related electrical safety: IEC601-1, UL2601 and VDE750 (German MedGv).
The echo-screen is a stand alone-system. It is battery powered with a 6V battery. During operation there are no connections to the mains. There are no conductive connections to the patient or to the user during measurement.
To prevent excessive tone levels within the ear, the echo-screen system continiously controls the level of the output tone or click-burst. The max. output level is controlled via Software to 85dB SPL (sound pressure level) and the electronic hardware is build in such a way, that the speakers within the probe are incapable of producing enough decibels to permanently damage the ear.
The system will also be certified according to the following standards:
- ISO 9001 (EN29001) Quality Management System
- EN46001 Class I or Class IIa device
- Group 1 (medical equipment, conducted and radiated emissions) - EN 25011
- IEC801-3 Radiated Electromagnetic Field Susceptibility
- Conforms the provisions of Eruopean council Directive 93/42EEC - CE Mark concerning medical devices.
## ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Summary of Effectiveness:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The 'echo-screen' is a portable OAE based screening system.
Its aim is to detect and track hearing loss and abnormal cochlear conditions as early as possible in the child's life. The automated test makes it easy to use by trained personnel in a medical or clinical environment.
{3}------------------------------------------------
## Attachement A
i
------
۰ -----
| Technological<br>Specification | GSI70 | echo-screen |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power source | Powered by rechargeable<br>battery | Powered by rechargeable<br>battery |
| Battery Low<br>Indication | Battery low indicator | Battery low indicator |
| Safety Compliance | IEC 601-1 and UL2601-1 | IEC601-1 and UL2601-1 |
| Computer interface | IRDA communication | none, during measurement |
| Audiometry | Tones are presented to a<br>miniature speaker within the<br>probe which perceived is<br>indicated by the patient using a<br>response switch. Tone quality<br>and frequency specified per<br>ANSIS3.6 | Tones (clicks) are presented to a<br>miniature speaker within the<br>probe and controlled via SW.<br>The response from the cochlea is<br>received and registrated.<br>Click tone quality and amplitude is<br>controlled via SW and HW and is<br>limited to a max. output level<br>85dB SPL. |
| OAE probe design | Two speakers and a<br>microphone present and<br>record response from<br>cochlea of ear. | One speaker and a microphone<br>present and record response<br>from cochlea of ear. |
| SW | GSI 60 SW | Echo-screen Automated Evaluation SW |
| Supporting SW | Windows based SW that<br>shows result., allows export<br>of data, and updating of<br>patient ID | C-based SW, that shows evaluation results,<br>artefact rate, stimulus stability.<br>After measurement export of test-result<br>and patient data are possible |
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name encircling an emblem. The emblem is a stylized image of an eagle with three overlapping profiles, suggesting a sense of community or collaboration. The overall design is simple and professional, reflecting the department's role in public health and human services.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 21 1998
David G. Wood, Esq. Fischer-Zoth Audiologic System, Inc. 11309 Brandon Park Drive Sandy, Utah 84092
Re:
K982642 Echo-Screen Dated: July 27, 1998 Received: July 29, 1998 Regulatory class: II 21 CFR 874.1050/Procode: 77 EWO
Dear Mr. Wood:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we-have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). . You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your-device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the programs on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odff/dsmaldsmam.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510(k) Number (if known): | K982642 |
|---------------------------|-------------|
| Device Name: | ECHO-SCREEN |
| Indications For Use: | |
The Echo-Screen Otoacoustic Emissions Test Instrument is a hand-held instrument with
an attached probe (that is placed in the outer ear canal) which emits sound that evokes
otoacoustic emissions. A microphone in the probe picks up the emissions if any are
present, and the device then analyzes the emissions and mathematically separates noise
from cochlear emissions. The presence of these emissions indicates cochlear function in
adults, infants and children. The instrument requires no active participation from the
patient.
Otoacoustic emissions (OAEs) are low-level audio-frequency sounds that are generated by
the outer hair cells of the cochlea in response to sound input. The presence of OAEs is an
indication of cochlear function. If the Echo-Screen signals a "pass" reading as it analyzes
the presence of OAEs, the clinician can determine that the cochlea is operating normally.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982642
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.