K122496 · Bio-Logic Systems Corp. · EWO · Nov 1, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K122496
Device Name
AUDX PRO OTOACOUSTIC EMISSIONS MEASUREMENT SYSTEM
Applicant
Bio-Logic Systems Corp.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Nov 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The AuDX Pro Otoacoustic Emissions Measurement System delivers controlled acoustic signals in the ear canal and measures the resulting evoked otoacoustic emissions (OAEs) that are generated by the outer hair cells of the inner ear. The AuDX Pro device performs transient evoked otoacoustic emissions (TEOAE) and distortion product otoacoustic emissions (DPOAE) tests. AuDX Pro is indicated for use by trained health care professionals (audiologists, physicians) and personnel (nurses, technicians) who are trained to operate the device under their supervision to perform otoacoustic emissions testing to assess cochlear function. The device can be used for patients of all ages, from newborn infants through adults, to include geriatric patients. The otoacoustic emissions test is especially indicated for use in testing individuals for whom behavioral results are deemed unreliable, such as infants, young children, and cognitively impaired adults.
Device Story
Handheld, battery-operated OAE measurement system; uses Bio-logic ear probe with miniature receivers and microphone. Device delivers controlled acoustic stimuli to ear canal; records evoked emissions from outer hair cells. System performs TEOAE and DPOAE tests; collects/averages data until parameters met. Operated by trained healthcare professionals/supervised personnel in clinical settings. Features LCD for real-time data display, pass/refer recommendations, and graphic review. Data can be exported to PC-based databases (P&TI) or hearing tracking systems (HATS) for analysis. Provides objective assessment of cochlear function; aids clinical decision-making for patients unable to provide reliable behavioral responses.
Clinical Evidence
Bench testing only. Verification and validation performed to ensure device meets performance specifications and demonstrates equivalence to predicate devices. No clinical data presented.
Technological Characteristics
Handheld, battery-operated (rechargeable Li-ion). Uses Bio-logic OAE ear probe (miniature receivers/microphone). Connectivity via RS-232/USB to PC. Standards: IEC 60601-1, IEC 60601-1-2, CISPR 11. Software: Firmware written in assembly code for stimulus control/data collection. Biocompatibility: Ear tips per ISO 10993-1 (cytotoxicity, sensitization, irritation).
Indications for Use
Indicated for patients of all ages (newborns to geriatric) to assess cochlear function via TEOAE and DPOAE testing. Especially indicated for individuals where behavioral audiometric testing is unreliable, such as infants, young children, and cognitively impaired adults.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
Predicate Devices
AuDX Otoacoustic Emissions Measurement System with AuDX I/O Function (K111618)
{0}------------------------------------------------
K122496
NOV
1 2012
### 510(k) SUMMARY
Manufacturer's Name:
Natus Medical Incorporated One Bio-logic Plaza Mundelein, IL 60060
Corresponding Official:
Telephone Number: Fax Number:
Timothy Karlovsky Quality and Regulatory Assurance Manager Natus Medical Incorporated One Bio-logic Plaza Mundelein, IL 60060 847.573.5430 847.949.8615
Summary Date:
October 23, 2012
Trade Name:
and Number:
Classification Name
Common or Usual Name: Audiometer
Audiometer 21 CFR 874.1050, Product Code: EWO
AuDX Pro Otoacoustic Emissions Measurement System
Predicate Devices:
K111618 AuDX Otoacoustic Emissions Measurement System with AuDX I/O Function
K112247 ABaer with ABaer I/O
K072033 Otoport
Device Description:
The AuDX Pro Otoacoustic Emissions Measurement System is a handheld battery operated device that performs otoacoustic emissions tests. Connection to the patient's ear is via the Bio-logic ear probe. Using a combination of hardware and software, the system produces a controlled acoustic signal in the ear canal and measures the resulting evoked emissions that are generated by the outer hair cells of the inner ear associated with normal cochlear function. The stimuli are presented via miniature receivers and the acoustic response in the external ear canal is recorded via a miniature microphone, all embedded in the Bio-logic OAE probe. The system collects and averages data samples until specified measurement parameters are achieved. The AuDX Pro system includes optional software selections
{1}------------------------------------------------
distinguished by the AuDX Pro, AuDX Pro II, and AuDX Pro Plus naming designations.
The test consists of measuring and recording transient (clickevoked) otoacoustic emissions (TEOAE) or distortion product otoacoustic emissions (DPOAE) utilizing pure tones. The same ear probe is used for both types of tests. For transient otoacoustic emissions (TEOAEs), the reproducibility and the difference value between the TEOAE and the noise floor amplitudes appear on the LCD display on the front of the AuDX Pro unit. For distortion product otoacoustic emissions (DPOAEs), the DPOAE and noise floor amplitudes appear on the LCD. A pass or refer result is assigned at the end of the test based on a comparison of the patient's OAE response to normalized data. Additionally, a graphic display of the data is presented on the LCD allowing users to view and/or analyze the results.
The data collected by the AuDX system can be sent to a host computer where it can be saved or placed in a database. Optional Natus software programs are available for that function. Natus software programs provide a mechanism for users to define their own specific test protocols and download them to the AuDX Pro device.
The AuDX Pro Otoacoustic Emissions Measurement System expands upon the capabilities of the AuDX Otoacoustic Emissions Measurement System with AuDX I/O Function [K111618]. The AuDX Pro System provides the following enhancements over the AuDX System:
- Protocol Setup utilities to load expanded frequency . protocols to the AuDX Pro:
- Options to upload data to a PC based Patient and Test . Information Database (P&TI);
- . Options to graphically review data from a PC based Hearing Assessment & Tracking System (HATS) software utility linked with the P&TI Database; and.
- . A Graphical User Interface (GUI) on an enhanced LCD screen for testing, patient information, and data review.
Intended Use:
The AuDX Pro Otoacoustic Emissions Measurement System is indicated for use when it is necessary for a trained health care professional to assess cochlear function. The device can be used for patients of all ages, from newborn infants through adults, to include geriatric patients. The otoacoustic emissions test is especially indicated for use in testing individuals for whom behavioral responses are deemed
{2}------------------------------------------------
unreliable, such as infants, young children, and cognitively impaired adults.
The AuDX Pro Otoacoustic Emissions Measurement System performs transient evoked otoacoustic emissions (TEOAE) and distortion product otoacoustic emissions (DPOAE) tests. Using a combination of hardware and software, the AuDX system produces a controlled acoustic signal in the ear canal and measures the resulting evoked emissions that are generated by the inner ear as a result of normal cochlear function. The stimuli are presented via miniature receivers and the acoustic response in the external ear canal is recorded via a miniature microphone, all embedded in the Bio-logic OAE probe. The system collects and averages data samples until specified measurement parameters are achieved. For transient evoked otoacoustic emissions (TEOAEs), the reproducibility and the difference value between the TEOAE and the noise floor amplitudes are calculated and presented to the user. For distortion product otoacoustic emissions (DPOAEs), the DP and noise floor amplitudes are calculated and presented to the user. A pass or refer recommendation is assigned automatically at the end of the test based on user defined custom protocols or default test protocols and measured OAE test parameters. The AuDX Pro Otoacoustic Emissions Measurement System is equivalent to the devices cleared under K111618. K112247, and K072033.
Nonclinical Tests:
Technological Characteristics:
> Design verification and validation were performed to assure that the AuDX Pro meets its performance specifications and demonstrates equivalence to the specified predicate devices.
The verification and validation summary report and risk analysis documentation provided in this 510(k) support the conclusion that the AuDX Pro Otoacoustic Emissions Measurement System is safe and effective.
{3}------------------------------------------------
# Comparison Table
:
.
| | AuDX with AuDX I/O<br>Function | ABaer with ABaer I/O<br>Function | Otodynamics Otoport | AuDX Pro |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate 510(k)<br>number | K111618 | K112247 | K072033 | New device |
| Intended Use | To measure or<br>determine cochlear<br>function via<br>performing transient<br>evoked otoacoustic<br>emissions (TEOAE),<br>distortion product<br>otoacoustic emissions<br>(DPOAE) and<br>determining DPOAE<br>Input / Output (I/O)<br>functions. | Performs an<br>automated auditory<br>evoked response<br>(ABaer) screening<br>and/or an automated<br>otoacoustic emissions<br>(AOAE) screening.<br>The device performs<br>TEOAE and DPOAE<br>tests and determining<br>DPOAE Input / Output<br>(I/O) functions. | Can perform automated<br>TEOAE and DPOAE<br>based measurements<br>via an ear probe,<br>including I/O<br>functionality. | To measure or<br>determine cochlear<br>function via<br>performing transient<br>evoked otoacoustic<br>emissions (TEOAE)<br>and distortion product<br>otoacoustic emissions<br>(DPOAE). |
| User | Trained health care<br>professionals | Same | Same | Same |
| Patient Population | Newborn infants<br>through adults, to and<br>including geriatric<br>patients. Especially for<br>use in testing<br>individuals where<br>behavioral audiometric<br>results are unreliable. | Same | Same | Same |
| Anatomical sites | External ear canal | External ear canal<br>for OAE and for<br>ABR testing; scalp,<br>ear and other skin<br>sites proper for<br>surface electrode.<br>placement. | External ear canal | External ear canal |
| Noninvasive | Yes | Yes | Yes | Yes |
| Physical Form | Standalone handheld<br>system that can be<br>connected to a PC. | PC based system<br>with handheld size<br>external data<br>acquisition box. | Standalone handheld<br>system. | Standalone handheld<br>system that can be<br>connected to a PC. |
| DPOAE | Yes | Yes | Yes | Yes |
| I/O function | Yes | Yes | Yes - Otoport Advance | No |
| TEOAE | Yes | Yes | Yes | Yes |
| ABR | No | Yes | No | No |
| Can provide<br>Pass/Refer<br>recommendation | Yes | Yes | Yes | Yes |
| Background noise<br>check during test | Yes | Yes | Yes | Yes |
| Measurement<br>based stopping<br>criteria | Yes | Yes | Yes | Yes |
| Performs in-the-<br>ear calibration | Yes | Yes | Yes | Yes |
| User customization<br>of test protocols | Yes - User has<br>control over DPOAE<br>I/O function test<br>parameters, but not<br>automated DPOAE or<br>TEOAE testing. | Yes - User has<br>control over DPOAE<br>I/O function test<br>parameters, but not<br>automated DPOAE or<br>TEOAE testing. | Yes, Can create, edit<br>and save user's own<br>test protocols and<br>pass/refer criteria for<br>automated DPOAE,<br>TEOAE and DPOAE I/O<br>testing. | Yes, Can create, edit<br>and save user's own<br>test protocols and<br>pass/refer criteria for<br>automated DPOAE<br>and TEOAE testing. |
| | AuDX with AuDX I/O<br>Function | ABaer with ABaer I/O<br>Function | Otodynamics Otoport | AuDX Pro |
| Predicate 510(k)<br>number | K111618 | K112247 | K072033 | New device |
| Fundamental<br>Scientific<br>Technology | Records data in the<br>form of acoustic<br>signals, i.e. OAE with<br>the use of an ear<br>probe. The probe<br>houses miniature<br>receivers to deliver a<br>stimulus and a<br>microphone to records<br>stimulus and the<br>resulting evoked<br>OAEs. The test<br>consists of either<br>measuring and<br>recording TEOAEs or<br>DPOAEs. Sensed<br>signals are amplified,<br>filtered and averaged<br>to improve the signal<br>quality and signal-to-<br>noise ratio. The<br>resulting recorded<br>measurements are<br>automatically<br>displayed and can be<br>evaluated for OAE<br>signals. I/O software<br>feature to the DPOAE<br>test suite provides<br>ability to allow the<br>user to change the<br>stimulus levels at<br>constant stimuli<br>frequencies and plot<br>the corresponding<br>stimulus amplitude vs<br>DPOAE level<br>functions to evaluate<br>the input/output<br>relations of the<br>generated DPOAEs at<br>specific frequency<br>regions as defined by<br>the user. | Same as K111618<br>with respect to OAE<br>testing (TEOAE,<br>DPOAE and DPOAE<br>I/O). Additionally, the<br>device performs<br>screening by<br>recording and<br>analyzing Auditory<br>Brainstem Responses<br>(ABR). | Can screen for cochlear<br>function using either<br>Transient Evoked or<br>Distortion Product<br>Otoacoustic Emissions<br>(TE or DPOAEs). It<br>uses an ear probe to<br>deliver stimulus and to<br>record evoked OAEs.<br>Recorded signals are<br>evaluated automatically<br>using signal analysis to<br>provide robust<br>indications of cochlear<br>function and high<br>immunity to extraneous<br>noise. Additionally, the<br>system includes I/O<br>functionality. | Same as K111618,<br>K112247 and<br>K072033 with respect<br>to TEOAE and<br>DPOAE. |
| Safety<br>Characteristics | No direct electrical<br>connection to the<br>patient. | Same | Same | Same |
| Energy delivered<br>to the patient | Audible acoustic<br>stimulus to the<br>subject's auditory<br>system. | Same | Same | Same |
| Energy delivered<br>to the patient | Audible acoustic<br>stimulus to the<br>subject's auditory<br>system. | Same | Same | Same |
| Energy delivered<br>to the patient | Audible acoustic stimulus<br>to the subject's auditory | Same | Same | Same |
| | AuDX with AuDX I/O, | ABaer with ABaer I/O | Otodynamics Otoport | AuDX Pro |
| Predicate 510(k)<br>number | Function<br>K111618 | Function<br>K112247 | K072033 | New device |
| Energy delivered<br>to the patient | Audible acoustic<br>stimulus to the<br>subject's auditory<br>system. | Same | Same | Same |
| Mechanical safety | No moving<br>mechanical parts in<br>contact with the<br>subject | Same | Same | Same |
| Electrical Safety<br>and Performance<br>Standards | CISPR 11, IEC 60601-1,<br>IEC 60601-1-1,<br>IEC 60601-1-2,<br>IEC 610003-2,<br>IEC 610003-3,<br>IEC 610004-2,<br>IEC 610004-3,<br>IEC 610004-4,<br>IEC 610004-5,<br>IEC 610004-6,<br>IEC 610004-8,<br>IEC 610004-11,<br>IEC 60101-1-4 | CISPR 11, IEC 60601-1,<br>IEC 60601-1-1,<br>IEC 60601-1-2,<br>IEC 610003-2,<br>IEC 610003-3,<br>IEC 610004-2,<br>IEC 610004-3,<br>IEC 610004-4,<br>IEC 610004-5,<br>IEC 610004-6,<br>IEC 610004-8,<br>IEC 610004-11,<br>IEC 60101-1-4 | BS EN 60601-1,<br>BS EN 60601-1-1,<br>BS EN 60601-1-2,<br>BS EN 60601-1-4 | CISPR 11,<br>IEC 60601-1 + Coor. 2,<br>IEC 60601-1-2 + A1,<br>IEC 610003-2,<br>IEC 610003-3 + A1 +<br>IEC 610004-2,<br>IEC 610004-3,<br>IEC 610004-4,<br>IEC 610004-5,<br>IEC 610004-6,<br>IEC 610004-8,<br>IEC 610004-11,<br>IEC 60101-1-4 |
| Battery powered<br>hardware (Battery<br>type) | Yes (Rechargeable<br>Lithium Ion Battery) | No (6 V DC) | Yes (Rechargeable Lithium<br>polymer) | Same as K111618 |
| Data display | Information displayed<br>on the AuDX screen<br>and the PC monitor<br>screen for the AuDX<br>I/O | Information displayed<br>on computer monitor<br>screen. | Information displayed<br>on the Otoport device<br>screen. | Information displayed<br>on the device's LCD<br>screen and can be<br>displayed on a<br>computer monitor<br>screen. |
| Built in keys for<br>user interface on<br>the collection box | Yes; 5-keys | No | Yes | Yes, 7-keys |
| Computer<br>compatibility | IBM compatible PC. | Same | Same | Same |
| Operating System | Windows for host PC<br>for AuDX I/O and<br>utilities. No OS for the<br>AuDX hardware<br>device. | Windows® 2000,<br>Windows® XP,<br>Windows® 7 | Windows 2000/XP/Vista<br>for PC based utilities | No OS for the AuDX<br>hardware device. |
| Communication with<br>PC | RS-232 and USB | RS-232 and USB | USB | RS-232 and USB |
| System Design | With respect to<br>TEOAE and DPOAE,<br>the main stimulus<br>generation and data<br>acquisition hardware<br>connected to OAE<br>probe. For the AuDX<br>I/O functionality, the<br>main stimulus<br>generation and data<br>acquisition hardware<br>connected to OAE<br>probe and a personal<br>computer hosting<br>testing software, data<br>display and user<br>interface. | Main stimulus<br>generation and data<br>acquisition hardware<br>connected to patient<br>cables, transducers<br>and personal<br>computer hosting<br>testing software, data<br>display and user<br>interface. | Main stimulus<br>generation and data<br>acquisition hardware<br>connected to OAE<br>probe and hosting the<br>testing software, data<br>display and user<br>interface functionality<br>without a need to<br>connect to PC. | Main stimulus<br>generation and data<br>acquisition hardware<br>connected to OAE<br>probe. |
| | AuDX with AuDX I/O.<br>Function | ABaer with ABaer I/O<br>Function…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.