The GN Otometrics Insert Earphones are accessories to the GN Otometrics Audiometers and Chartr EP Auditory Evoked Response systems. The GN Otometrics Insert Earphones devices perform as the means for delivering auditory stimulus to the ears of the patient under test. The interface to the ears of the patient is provided by means of disposable foam eartips, designed to fit infants, children, and adults.
Device Story
Insert earphones function as transducers converting electrical signals from audiometers or auditory evoked response systems into acoustic stimuli; delivered to patient ears via disposable foam eartips. Device comprises electrical transmission cable, passive analog filter for response shaping, speaker, acoustic transmission path (tubes/cavities for resonance smoothing), and transducer case. Used in clinical settings by audiologists or technicians to facilitate hearing assessment or evoked potential testing. Output allows healthcare providers to measure patient auditory response; aids in diagnosis of hearing status. Benefits include precise stimulus delivery and patient comfort across all age groups.
Clinical Evidence
Bench testing only. Performance verified through non-clinical acoustic and electrical parameter testing in accordance with ANSI/ASA S 3.6-2004, Specification for Audiometers.
Technological Characteristics
Passive transducer system. Components: shielded electrical cable, passive analog filter, speaker, acoustic transmission path (stainless steel tubes, rubber molded parts, silicon tubing), and transducer case. Connectivity: wired connection to audiometer/stimulator. Disposable foam eartips. Complies with ANSI/ASA S 3.6-2004.
Indications for Use
Indicated for infants, children, and adults undergoing auditory testing or evoked response procedures to deliver auditory stimuli to the ear.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
# 510(k) Submission (21 CFR 807.90(e)) for GN Otometrics Insert Earphones
## Section 5:
.
## JUL 1 4 2008
## 510(k) Summary
| Date: | April 30, 2008 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | GN Otometrics<br>125 E. Commerce Drive<br>Schaumburg, IL 60173<br>USA |
| Telephone:<br>Fax: | 847-534-2150 #6932<br>847-534-2153 |
| Contact Person: | Daniel Sansonetti<br>Manager of Research & Development<br>dsansonetti@gnotometrics.com |
| Name of Device: | OtoInsert |
| Common Name: | Insert Earphones for auditory stimulus delivery |
| Classification Name: | Accessories to devices with classification:<br>Audiometer (per 21 CFR section 874.1050)<br>and<br>Accessories to devices with classification:<br>Stimulator, Auditory, Evoked Response.<br>(per 21 CFR section 882.1900) |
| Predicate Device: | Etymotic Research ER-3 Insert Earphones, 510(k)<br>#K930003 |
.
:
: 上一篇:
-
.
{1}------------------------------------------------
#### Description of the Device:
The GN Otometrics Insert Earphones are transducers that convert electrical stimulus, provided by GN Otometrics Audiometers and GN Otometrics Auditory Evoked Response Stimulators, into acoustic stimulus, which is then coupled to the patient's ears. They are comprised of five sections: 1.) Electrical Transmission Path, 2.) Electrical Filter, 3.) Speaker, 4.) Acoustic Transmission Path, and 5.) Transducer Case.
The Electrical Transmission Path consists of a shielded cable. The cable is connected to the Audiometer or Auditory Evoked Response Stimulator, by means of a connector. The other end of the cable is attached to the left and right Electrical Filter Sections.
Each Electrical Filter section (left and right) consists of a passive analog filter to provide electrical pre-emphasis response shaping to the stimulus prior to reaching the Speakers.
Each Speaker (left and right) converts the electrical stimulus into an acoustic stimulus. The acoustic stimulus is delivered to the patient's ear by means of the Acoustic Transmission Path.
Each Acoustic Transmission Path (left and right) consists of a front acoustic transmission path and a back acoustic transmission path. The back acoustic transmission path consists of a series of tubes and cavities which are configured to create two resonant acoustic circuits. These resonant circuits are used to smooth the acoustic output at the eartip, by reducing the primary and secondary resonant peaks caused by the impedance mismatch at both ends of the front transmission path. The front acoustic transmission path consists of silicon tubing, tube nipple, and an eartip. The tube nipple provides acoustic and mechanical connection to the disposable eartip. The disposable eartips interface to the patient's ear and deliver the acoustic stimulus.
Each Transducer Case (left and right) houses an Electrical Filter, Speaker, and back acoustic transmission path. Each Transducer case provides means of handling and labeling and means to secure the Electrical Transmission Path (cable).
#### Intended Use of the Device:
The GN Otometrics Insert Earphones are accessories to the GN Otometrics Audiometers and GN Otometrics Auditory Evoked Response systems. The GN Otometrics Insert Earphone devices perform as the means for delivering auditory stimulus to the ears of the patient under test. The interface to the ears of the patient is provided by means of disposable foam eartips, designed to fit infants, children, and adults.
{2}------------------------------------------------
## 510(k) Submission (21 CFR 807.90(e)) for GN Otometrics Insert Earphones
#### Comparison to Predicate Device:
The GN Otometrics Insert Earphones were designed to be a direct replacement to the Etymotic Research ER-3 Insert Earphones (510(k) #K930003). The GN Ottometrics Insert Earphones share the same technological characteristics as the Etymotic Research ER-3 Insert Earphones (510(k) #K930003). Both insert earphones are comprised of the same sections: 1.) Electrical Transmission Path, 2.) Electrical Filter, 3.) Spcaker, 4.) Acoustic Transmission Path, and 5.) Transducer Case.
The Electrical Transmission Path in both the GN Otometrics and Etymotic Insert Earphones are comprised of a cable with a connector on each end. The only differences are cable manufacturer, cable color, durometer, nominal dimensions, and connector type.
The Electrical Filter sections of both the GN Otometrics and Etymotic Insert Earphones are identical with respect to component values and configuration. The only difference is component manufacturers.
The Speaker in both the GN Otometrics and Etymotic Insert Earphones are from the same manufacturer and are the same model number.
The Acoustic Transmission Path in both the GN Otometrics and Etymotic Insert Earphones are identical with respect to nominal dimensions and configuration. Both the GN Otometrics and Etymotic Insert Earphones have a front and back acoustic transmission path. The only difference in the back transmission path is the material and construction method. The Etymotic Insert Earphones use PVC tubing to create the acoustic resonant circuits, while the GN Otometrics uses stainless steel tubes inseted into a rubber molded part. The end result of both construction methods is the same. The front transmission path is identical with respect to materials, dimensions, and counting means to the patient. Both the GN Otometrics and Etymotic Insert Earphones use the same disposable eartips from the same manufacturer.
The Transducer Case in both the GN Otometrics and Etymotic Insert Earphones are similar with exception to shape and color.
The intended use of both the GN Otometrics and Etymotic Insert Earphones is the same. Both the GN Otometrics and Etymotic Insert Earphones are intended to been a transducers to convert electrical stimulus, delivered by an Audiometer or Evoked Potential stimulator, into acoustic stimulus, which is then delivered to the patient's ear by means of a disposable Eartip.
### Substantial Equivalence Performance Metrics:
Substantial equivalence to the Etymotic Research ER-3 Insert Earphones (510(k) #K930003) was based on non-clinical performance testing of the acoustic and electrical parameters as specified in the ANSI/ASA S 3.6-2004, Specification for Audiometers.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, symbolizing the department's mission to protect and promote the health and well-being of Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## UUL 1 4 2008
GN Otometrics c/o Daniel Sansonetti Manager of Research & Development 125 E. Commerce Drive Schaumburg, IL 60173
Re: K081234
> Trade/Device Name: GN Otometrics Insert Earphones Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: June 27, 2008 Received: June 27, 2008
Dear Mr. Sansonetti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced aboye and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egerkins, mD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K081234
#### 510(k) Submission (21 CFR 807.90(e)) for GN Otometrics Insert Earphones
#### Section 4:
#### Indications for Use
510(k) Number (if known): 1/208 /234
Device Name: GN Otometrics Insert Earphones
Indications for Use:
The GN Otometrics Insert Earphones are accessories to the GN Otometrics Audiometers and Chartr EP Auditory Evoked Response systems. The GN Otometrics Insert Earphones devices perform as the means for delivering auditory stimulus to the ears of the patient under test. The interface to the ears of the patient is provided by means of disposable foam eartips, designed to fit infants, children, and adults.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laura Sohn
n of Ophthalmic Far and Throat Devises
510(k) Number K081234
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.