BIO-LOGIC ABAER CUB WITH INTEGRATED OAE & ABR FUNCTIONS
K021801 · Bio-Logic Systems Corp. · EWO · Jun 27, 2002 · Ear, Nose, Throat
Device Facts
Record ID
K021801
Device Name
BIO-LOGIC ABAER CUB WITH INTEGRATED OAE & ABR FUNCTIONS
Applicant
Bio-Logic Systems Corp.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Jun 27, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bio-logic Otoacoustic Emissions (OAE) and Evoked Potential (EP) product families are indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. This product, the ABaer Cub with Automated OAE and ABR, like it's predicate devices, the AuDX and the ABaer Cub, is especially indicated for use in the screening of infants to determine hearing loss. The Bio-logic OAE and EP Systems can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic OAE and EP families of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
Device Story
Portable, battery-powered hearing screening device; integrates OAE and ABR functions into single unit. Inputs: physiological signals via sensors. Processing: DSP software (same as Sport/AuDX) performs signal recording/analysis; Point Optimized Variance Ratio (POVR) algorithm optimizes signal quality and test speed. Output: screening results displayed on Pocket PC host computer. Used in clinical settings by physicians or trained healthcare professionals. Replaces separate desktop/laptop-based systems with handheld, battery-powered unit; improves portability and ease-of-use. Facilitates diagnosis of auditory disorders in patients unable to perform behavioral tests.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance of predicate devices.
Technological Characteristics
Portable, battery-powered handheld unit. Integrates OAE and ABR screening. Uses DSP software for signal processing. Connectivity: Pocket PC host computer. Software: C++ for Windows Pocket PC. Algorithm: Point Optimized Variance Ratio (POVR) for signal quality optimization.
Indications for Use
Indicated for hearing loss screening in patients of all ages, including infants, children, adults, and geriatric patients. Specifically for individuals where behavioral audiometric testing is unreliable, such as infants, young children, and cognitively impaired or uncooperative adults.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
K 021801
Image /page/0/Picture/1 description: The image shows the logo for Bio-logic Systems Corp. The logo is set against a black background. The words "Bio-logic" are in a large, bold, white font, with the registered trademark symbol to the right of the word "logic". Below "Bio-logic" is the text "SYSTEMS CORP" in a smaller, white font.
One Bio-logic Plaza Mundelein, IL 60060-3700 1-800-323-8326 Fax: 847-949-8615 www.bio-logic.com
JUN 2 7 2002
# SECTION 2: SUMMARY AND CERTIFICATION
## 510(K) SUMMARY
Safety and effectiveness information concerning the ABaer Cub with OAE device modification to the Bio-logic Hearing Screening product family is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
| PREPARED BY: | Bio-logic Systems Corp |
|--------------|------------------------|
| | One Bio-logic Plaza |
| | Mundelein, IL 60060 |
TELEPHONE: (847)-949-5200
CONTACT PERSON: Norman E. Brunner
DATE ON WHICH THE SUMMARY WAS PREPARED: May 24, 2002
NAME OF DEVICE: Bio-logic ABaer Cub with integrated OAE and ABR functions.
COMMON NAME: Audiometer (OAE) and Evoked Response (ABR).
CLASSIFICATION NAME: Audiometer (reference 21CFR 874.1050) and Auditory Evoked Response system (reference 21CFR 882.1900).
PREDICATE DEVICE: The OAE function is substantially equivalent to the Sport/AuDX predicate device (510(k) #K974076) and the ABR function is substantially equivalent to the ABaer Cub predicate device (510(k) #K021215).
## DESCRIPTION OF THE DEVICE:
The Bio-logic Hearing Screening products, including the OAE and ABR test instruments Scout/AuDX and ABaer Cub, are intended to be used for the purpose of testing for hearing loss in patients of all ages, but primarily in infants . The predicate devices referenced above are the latest current products of this type marketed by Bio-logic. This new product, ABaer Cub with OAE, is the integration of both OAE and ABR screening functions into one portable unit, the ABaer Cub, with additional OAE software functions added to the Pocket PC host computer for the ABaer Cub.
{1}------------------------------------------------
Snecial 510(k): ABaer Cub with OAE
The ABaer Cub predicate device performs Evoked Potential screening, recording and analysis functions, provides one channel of data recording, and includes the Point Optimized Variance Ratio (POVR) algorithm for optimizing signal quality, implementing the screening function and enhancing speed of test completion. This new ABaer Cub with OAE device performs these same functions in the same ways, but has the added capability of performing the OAE tests of the Sport/AuDX predicate device, all within the same hardware package as the ABaer Cub.
The Sport/AuDX predicate device can be used in either a stand-alone (AuDX) or host-connected (Sport) mode. The same DSP software is used in both types of box. The primary difference between the two products is that the Sport requires the use of a host computer for user control information and to store and display the data returned from the Sport box. For the AuDX, all user control and display is right at the AuDX box, making this a fully portable test instrument but with limited control and display capability. However, after an AuDX test is completed, it can be connected to a host computer to download test data for display on the host computer screen if desired. This same Sport/AuDX DSP software functionality has been incorporated into the ABaer Cub box. Much of the Sport host computer display software has been incorporated into the ABaer Cub Pocket PC (host computer), resulting in a completely battery-powered, portable hand-held test instrument capable of performing both the Automated OAE and ABR tests. The Sport OAE predicate device uses a desktop or laptop computer instead of the Pocket PC. The electronic hardware is the same as the ABaer Cub hardware used in that predicate device. Because both the Pocket PC and the ABaer Cub unit are battery powered, this significantly improves the portability and usability of the device over that of the Sport OAE predicate device. The ABaer Cub software for control of this device is a combination of the Sport and ABaer Cub predicate device software, with some GUI changes necessary in the Sport OAE software to modify the graphics display for use with the smaller Pocket PC LCD screen. The host software for the Sport/AuDX OAE predicate device is written in the C++ programming language for the Windows 95/98/ME operating system, whereas the host software for the ABaer Cub is written in the C++ programming language for the Windows Pocket PC operating system. Together, these software additions implement the same functionality and perform the same intended use as both predicate devices, but with improved portability and ease-of-use.
### INTENDED USE:
The Bio-logic Otoacoustic Emissions (OAE) and Evoked Potential (EP) product families are indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders.
This product, the ABaer Cub with Automated OAE and ABR, like it's predicate devices, the AuDX and the ABaer Cub, is especially indicated for use in the screening of infants to determine hearing loss.
The Bio-logic OAE and EP Systems can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic OAE and EP families of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 7 2002
Bio-logic Systems Corp. c/o Norman E. Brunner Vice-President of Research & Development One Bio-logic Plaza Mundelein, IL 60060-3700
Re: K021801
Trade/Device Name: ABaer Cub with Automated OAE and ABR Regulation Number: 21 CFR 874.1050 & 21 CFR 882.1900 Regulation Name: Audiometer & Auditory Evoked Response System Regulatory Class: Class II Product Code: EWO & GWJ Dated: May 31, 2002 Received: June 3, 2002
Dear Mr. Brunner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Norman E. Brunner
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
Not Assigned 510(k) Number (if known):
Device Name: ABaer Cub with Automated OAE and ABR.
#### Indications For Use:
The Bio-logic Otoacoustic Emissions (OAE) and Evoked Potential (EP) product families are indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders.
This product, the ABaer Cub with Automated OAE and ABR, like it's predicate devices, the AuDX and the ABaer Cub, is especially indicated for use in the screening of infants to determine hearing loss.
The Bio-logic QAE and EP Systems can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Biologic OAE and EP families of products is to be performed under the prescription and supervision of other trained health care professional.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ooms K. Kane, M.D.
Prescription Use
(Per 21 CFR 801.109) X
OR
Over-The-Counter Use _
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.