← Product Code [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ) · K152747

# DSG Threaded Drill System (K152747)

_Spineguard S.A. · PDQ · Jun 10, 2016 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ/K152747

## Device Facts

- **Applicant:** Spineguard S.A.
- **Product Code:** [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ.md)
- **Decision Date:** Jun 10, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1820
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The DSG™ Threaded Drill System is indicated for use during pedicle screw pilot hole drilling to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the probe and may indicate contact of the tip with soft tissues and possible vertebration. The DSG™ Threaded Drill System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine. DSG™ Threaded Drill System is also indicated for use with fluoroscopic guidance in percutaneous (MIS) surgical approaches to the spine. The DSG™ Threaded Drill System also is specifically indicated for use in intraoperative electromyographic ("EMG") surveillance to assist in the location and evaluation of spinal nerves during surgery of the spine, by administration of low voltage electrical energy to tissues and nerves at the operative site, and EMG monitoring of muscle with those nerves.

## Device Story

DSG Threaded Drill System is a surgical instrument for pedicle screw pilot hole preparation. Device features a sensor at the distal 3mm of the shaft that measures electrical impedance of surrounding tissue. Input signals are processed to provide real-time visual and audible feedback to the surgeon; varying pitch and cadence indicate tissue density changes, helping identify potential cortical wall breaches. Device also performs intraoperative EMG surveillance by delivering low-voltage electrical energy to tissues and monitoring associated muscle groups. Used in OR by surgeons during open or percutaneous spinal procedures, often with fluoroscopic guidance. Output allows surgeons to adjust drilling trajectory, potentially reducing risk of nerve injury or vertebral cortex perforation. System includes a T-handle with ratchet, stainless steel shaft, and ceramic insulator.

## Clinical Evidence

Bench testing only. Mechanical bending tests confirmed sensor characteristics identical to predicate. Detection performance of new pin lengths verified as equivalent. Durability, cleaning, and sterilization validation performed for reusable components.

## Technological Characteristics

Materials: 316L Stainless Steel (ASTM F138), 304/316L/17-4PH Stainless Steel (ASTM F899), Zirconium Oxide ceramic insulator. Sensing: Electrical impedance at distal 3mm tip. Energy: Lithium-Ion battery. Form: Cannulated shaft with 25mm threaded portion and cutting flutes; T-handle with ratchet. Connectivity: Standalone. Sterilization: Validated for reusable components. Software: Firmware on circuit board with capacitors, resistors, and diodes.

## Regulatory Identification

A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.

## Predicate Devices

- Cannulated PediGuard Nerve Detector ([K143159](/device/K143159.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three figures representing health, services, and human aspects.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 10, 2016

SpineGuard, S.A. c/o Dr. John J. Smith Hogan Lovells US LLP 555 13th Street, NW Washington, DC 20004

Re: K152747

Trade/Device Name: SpineGuard DSGTM Threaded Drill System Regulation Number: 21 C.F.R. §874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: II Product Code: PDQ Dated: May 9, 2016 Received: May 9, 2016

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J.
Heetderks -A

Digitally signed by William J. Heetderks -A
DN: c=US, o=U.S. Government, ou=HHS, ou=NI
ou=People,
0.9.2342.19200300.100.1.1=0010149848,
cn=William J. Heetderks -A
Date: 2016061013-50-34-04'00'

for

Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K152747

Device Name DSGTM Threaded Drill System

### Indications for Use (Describe)

The DSG™ Threaded Drill System is indicated for use during pedicle screw pilot hole drilling to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the probe and may indicate contact of the tip with soft tissues and possible vertebration. The DSG™ Threaded Drill System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine. DSG™ Threaded Drill System is also indicated for use with fluoroscopic guidance in percutaneous (MIS) surgical approaches to the spine.

The DSG™ Threaded Drill System also is specifically indicated for use in intraoperative electromyographic ("EMG") surveillance to assist in the location and evaluation of spinal nerves during surgery of the spine, by administration of low voltage electrical energy to tissues and nerves at the operative site, and EMG monitoring of muscle with those nerves.

| Type of Use ( <i>Select one or both, as applicable</i> )                                                      |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div align="center">Prescription Use (Part 21 CFR 801 Subpart D) <span style="font-size: 2em;">☑</span></div> | <div align="center">Over-The-Counter Use (21 CFR 801 Subpart C) <span style="font-size: 2em;">☐</span></div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

{3}------------------------------------------------

### 510(k) SUMMARY

### SpineGuard, S.A.'s DSG™ Threaded Drill System (K152747)

### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

SpineGuard, S.A. 10 cours Louis Lumière 94300 Vincennes - France France Phone: +(33) 1 45 18 45 19

- Contact Person: Stephane Bette Chief Technology Officer SpineGuard, S.A. 1388 Sutter Street Suite 510 San Francisco, CA 94109 Phone + 1 415 512 2504
Date Prepared: June 7, 2016

### Name of Device

SpineGuard DSG™ Threaded Drill System

#### Common or Usual Name

Nerve Stimulator

#### Classification Name

21 C.F.R. §874.1820 Surgical Nerve Stimulator/Locator, PDQ

#### Predicate Devices

SpineGuard S.A., Cannulated PediGuard Nerve Detector (K143159)

#### Purpose of the 510(k) notice

The DSG™ Threaded Drill System is a modification to the cleared PediGuard System intended to add additional options for the surgeon to create pilot holes for pedicles screw systems.

#### Indication for Use

The DSG™ Threaded Drill System is intended to be used for the preparation of pedicle screw holes. The DSG™ Threaded Drill System is indicated for use during pedicle screw pilot hole drilling to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the probe and may indicate contact of the tip with soft tissues and possible vertebral cortex perforation. The DSG™ Threaded Drill System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine. The DSG™ Threaded Drill System is also indicated for use with fluoroscopic guidance in percutaneous (MIS) surgical approaches to the spine.

The DSG™ Threaded Drill System also is specifically indicated for use in intraoperative electromyographic ("EMG") surveillance to assist in the location and evaluation of spinal 

{4}------------------------------------------------

nerves during surgery of the spine, by administration of low voltage electrical energy to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves.

# Technological Characteristics

The DSG™ Threaded Drill System is designed to prepare pedicle screw holes. The sensor at the distal 3mm of the shaft detects the electrical impedance of the tissue immediately surrounding the sensor and provides real-time audio and visual information to the surgeon to assist the surgeon to determine the relative density of the tissue at the tip of the shaft. A skilled surgeon can interpret the varying pitch and cadence of the feedback to determine potential breaches of the cortical wall. The first 6mm at the instrument are very similar to the cleared PediGuard models, with a pointy awl type design to penetrate the bone. After that, a threaded portion of 25mm in length, featuring cutting flutes, allows for an easy insertion in bone. The threads allow for a smooth and progressive penetration in bone that results in steady and consistent readings of the impedance of the tissue, making the determination of potential breaches of the cortical wall easy for the surgeon.

### Performance Data

The SpineGuard DSG™ Threaded Drill System was tested mechanically (bending tests of the shaft) to show the sensor characteristics were identical to the predicate devices. The detection performance of the new length of pins was also demonstrated to be equivalent to the predicates. In addition, the device durability was tested. Cleaning and sterilization validation demonstrated that the reusable components can be adequately cleaned and sterilized prior to reuse.

The SpineGuard DSG™ Threaded Drill System complies with the following recognized standards:

- . ASTM F138-13, Standard specification for wrought 18 chromium-14 nickel-2.5 molybdenum stainless steel bar and wire for surgical implants (2013).
- . ASTM F899-12b, Standard specification for wrought stainless steel for surgical instruments (2012).
- . IEC 60601-1 Medical Electrical Equipment - Part 1: General requirements for safety, Amendment 1 (2007).
- IEC 60601-1-4, Medical Electrical Equipment Part 1-4: General requirements for safety, . Collateral Standard: Programmable electrical medical systems, Edition 1.1
- . ISO 10993-5, Biological Evaluation of Medical Devices: Tests for in vitro cytotoxicity (2010).
- . ISO 10993-7, Biological evaluation of medical devices: Ethylene Oxide sterilization residuals (2008).
- . ISO 10993-10, Biological Evaluation of Medical Devices: Tests for irritation and sensitization (2010).
- ISO 10993-11, Biological Evaluation of Medical Devices: Tests for systemic toxicity (2009).
- ISO 11607-1, Packaging for terminally sterilized medical devices: Requirements for materials, sterile barrier systems and packaging systems (2006).
- ISO 11607-2, Packaging for terminally sterilized medical devices: Validation requirements for . forming, sealing and assembly processes (2006).
- . ISO 11737-1, Sterilization of medical devices - Microbiological methods: Determination of a population of microorganisms on products (2006).
- . ISO 11737-2, Sterilization of medical devices - Microbiological methods: Tests of sterility performed in the definition, validation and maintenance of a sterilization process (2010).

{5}------------------------------------------------

- ASTM D4169, Standard practice for performance testing of shipping containers and systems . (2014).
- ASTM F88, Standard test method for seal strength of flexible barrier materials (2009).
- ASTM F1929, Standard test method for detecting seal leaks in porous medical packaging by . dve penetration.
- . ASTM F1980, Standard guide for accelerated aging of sterile medical device packages (2011).
- . ISO 11135-1, Sterilization of health care products - Ethylene oxide: Requirements for the development, validation and routine control of a sterilization process for medical devices (2014).
- ISO 17665-1, Sterilization of health care products Moist heat Part 1: Requirements for . the development, validation and routine control of a sterilization process for medical devices (2006).

# Substantial Equivalence

The SpineGuard DSG™ Threaded Drill System has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared PediGuard System. The minor differences in the DSG™ Threaded Drill System's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the SpineGuard DSG™ Threaded Drill System is substantially equivalent to the cleared PediGuard System.

|                                       | Subject DSG™ Threaded Drill<br>System (K152747)                                                                                                                                                                                                                                                                                                                                                                                                                  | Cannulated PediGuard® models<br>(K143159)                                                                                                                                                                                                                                                                                          |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use / Indications<br>for Use | The DSG™ Threaded Drill System<br>is intended to be used for the<br>preparation of pedicle screw holes.<br>The DSG™ Threaded Drill System<br>is indicated for use during pedicle<br>screw pilot hole drilling to provide<br>feedback to the surgeon via visual<br>and audible alerts that indicate a<br>change in impedance at the tip of<br>the probe and may indicate<br>contact of the tip with soft tissues<br>and possible vertebral cortex<br>perforation. | The PediGuard® is indicated for<br>use during pedicle screw pilot hole<br>drilling to provide feedback to the<br>surgeon via visual and audible<br>alerts that indicate a change in<br>impedance at the tip of the probe<br>and may indicate contact of the tip<br>with soft tissues and possible<br>vertebral cortex perforation. |
|                                       | The DSG™ Threaded Drill System<br>is indicated for use in both open<br>and percutaneous (MIS) surgical<br>approaches to the spine. The<br>DSG™ Threaded Drill System is<br>also indicated for use with<br>fluoroscopic guidance in<br>percutaneous (MIS) surgical<br>approaches to the spine.                                                                                                                                                                    | The PediGuard System is<br>indicated for use in both open and<br>percutaneous (MIS) surgical<br>approaches to the spine.<br>PediGuard is also indicated for<br>use with fluoroscopic guidance in<br>percutaneous (MIS) surgical<br>approaches to the spine.                                                                        |
|                                       | The DSG™ Threaded Drill System<br>also is specifically indicated for<br>use in intraoperative<br>electromyographic ("EMG")                                                                                                                                                                                                                                                                                                                                       | The PediGuard also is specifically<br>indicated for use in intraoperative<br>electromyographic ("EMG")<br>surveillance to assist in the                                                                                                                                                                                            |

{6}------------------------------------------------

|                        | surveillance to assist in the<br>location and evaluation of spinal<br>nerves during surgery of the spine,<br>by administration of low voltage<br>electrical energy to tissues and<br>nerves at the operative site, and<br>EMG monitoring of muscle groups<br>associated with those nerves. | location and evaluation of spinal<br>nerves during surgery of the<br>spine, by administration of low<br>voltage electrical energy to tissues<br>and nerves at the operative site,<br>and EMG monitoring of muscle<br>groups associated with those<br>nerves. |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Handle Shape           | T-Handle with Ratchet                                                                                                                                                                                                                                                                      | T-Handle                                                                                                                                                                                                                                                     |
| Components             | Stainless Steel shaft, plastic<br>handle, ceramic insulator, modular<br>with removable handle, ratchet<br>and reusable shaft                                                                                                                                                               | Stainless Steel shaft, plastic<br>handle, ceramic insulator, modular<br>with removable handle                                                                                                                                                                |
| Shaft Material         | Inner electrode: 316L Stainless<br>Steel (ASTM F138) Outer<br>electrode and/or shaft:<br>304 Stainless Steel (ASTM F899),<br>316L Stainless Steel and/or 17-<br>4PH (ASTM F899); Ceramic<br>Zirconium Oxide Insulator                                                                      | Inner electrode: 316L Stainless<br>Steel (ASTM F138) Outer<br>electrode and/or shaft:<br>304 Stainless Steel (ASTM F899),<br>316L Stainless Steel and/or 17-<br>4PH (ASTM F899); Ceramic<br>Zirconium Oxide Insulator                                        |
| Safety Features        | Device cannot be turned off until<br>battery exhausted. Prevents<br>reuse of device.                                                                                                                                                                                                       | Device cannot be turned off until<br>battery exhausted. Prevents reuse<br>of device.                                                                                                                                                                         |
| Power Source           | Lithium-Ion Battery                                                                                                                                                                                                                                                                        | Lithium-Ion Battery                                                                                                                                                                                                                                          |
| Sterility              | Sterile/Non-sterile                                                                                                                                                                                                                                                                        | Sterile                                                                                                                                                                                                                                                      |
| Single Use or Reusable | Single-use (T-Handle and Active<br>Stylet)<br>Re-usable (Ratcheting Handle,<br>Threaded Drill Shaft and Sleeve)                                                                                                                                                                            | Single Use                                                                                                                                                                                                                                                   |
| Distal Shaft Shape     | Straight (cannulated) and active<br>stylet                                                                                                                                                                                                                                                 | Straight (cannulated) with<br>removable inner starter stylet<br>(optional) and active stylet                                                                                                                                                                 |
| Dimensions             | 1.7mm inner diameter (hole)<br>cannulated metal shaft with<br>flanges (threads) and cutting<br>edges; Thread outer diameters of<br>4.0, 5.5 and 8.0mm.                                                                                                                                     | Cannulated, inner diameter:<br>2.5mm; Tapered shaft outer<br>diameter from 3 to 4mm.                                                                                                                                                                         |
| Circuit Board          | Capacitors, Resistors and Diodes<br>- Firmware (programmable chip)<br>on circuit board                                                                                                                                                                                                     | Capacitors, Resistors and Diodes<br>- Firmware (programmable chip)<br>on circuit board                                                                                                                                                                       |

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ/K152747](https://fda.innolitics.com/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ/K152747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
