PORTAREM-3000, PFS-6000

K964340 · Starkey Laboratories, Inc. · EWO · Jan 10, 1997 · Ear, Nose, Throat

Device Facts

Record IDK964340
Device NamePORTAREM-3000, PFS-6000
ApplicantStarkey Laboratories, Inc.
Product CodeEWO · Ear, Nose, Throat
Decision DateJan 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1050
Device ClassClass 2

Indications for Use

The portaREM-3000, PFS-6000 is a portable hearing aid dispensing system which allows the Hearing Health Care Professional to test hearing aids to IEC/ANSI standards. It also allows for Real Ear testing, and In Situ Audiometry via the use of the probe microphone, and offers a 2cc coupler conversion in which the 2cc coupler response desired can be determined from the audiogram. Standard puretone measurement is also available. With the PFS software, which will be standard with PFS-6000 and optional for the pR-3000, prescriptive formulas, matrix and hearing aid selection, order form printing, along with other options described in the user manual are available.

Device Story

Portable hearing aid dispensing system for clinical use by hearing health care professionals. Inputs: probe microphone signals (Real Ear/In Situ), puretone audiometry, and hearing aid test chamber measurements. System performs IEC/ANSI standard testing, intermodulation distortion analysis, and 2cc coupler conversions. PFS software provides prescriptive fitting formulas (NAL, POGO I/II, 1/2 Gain, 1/3 Gain, Berger), matrix selection, and order printing. Output: visual data on 4.7" LCD/external monitor, hard copy via built-in thermoprinter, and hearing aid performance metrics. Benefits: enables precise hearing aid calibration, fitting, and verification of performance against prescriptive targets.

Clinical Evidence

Bench testing only; system complies with IEC 601 and FCC Part 15 Class B standards. No clinical trial data provided.

Technological Characteristics

2-channel system; 4.7" 640x480 color LCD; built-in thermoprinter; 3.5" floppy drive; 2x RS-232 ports; 6W loudspeaker (50-15,000 Hz). Measurement range: 42-15,600 Hz; dynamic range 130 dB; accuracy ±0.2 dB. Electrical THD < 0.01%. Power: 110/220V AC. Complies with IEC 601 and FCC Part 15 Class B.

Indications for Use

Indicated for hearing health care professionals to perform hearing aid testing, real ear measurements, in situ audiometry, and hearing aid fitting/selection for patients requiring hearing aid calibration and adjustment.

Regulatory Classification

Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

Submission Summary (Full Text)

{1} JAN 10 1997 K964340 # SAFETY AND EFFECTIVENESS SUMMARY portaREM-3000, PFS-6000 NAME OF DEVICE: Rastronics portaREM-3000, Starkey PFS-6000 TYPE OF DEVICE: Hearing aid calibrator and analysis system INTENDED USE: The portaREM-3000, PFS-6000 is a portable hearing aid dispensing system which allows the Hearing Health Care Professional to test hearing aids to IEC/ANSI standards. It also allows for Real Ear testing, and In Situ Audiometry via the use of the probe microphone, and offers a 2cc coupler conversion in which the 2cc coupler response desired can be determined from the audiogram. Standard puretone measurement is also available. With the PFS software, which will be standard with PFS-6000 and optional for the pR-3000, prescriptive formulas, matrix and hearing aid selection, order form printing, along with other options described in the user manual are available. FEATURES: Easy to use menu-driven software, connections for color monitor and external printer, complete 2cc coupler test according to IEC/ANSI, intermodulation distortion measurement, composite signals, Ear Canal Resonance (REUR), Insertion gain measurements (REIR), In Situ response (REAR), In Situ audiometry, Real Ear 2cc coupler conversion (NAL), Insertion gain with Prescription Rules and Articulation Index, Fitting prescriptions: NAL, POGO I, POGO II, 1/2 Gain, 1/3 Gain, and Berger. There is a 3.5 inch disk drive, built in Thermoprinter for hard copy, test chamber for hearing aid measurement, 4.7" color LCD monitor, battery simulator, two RS-232 serial port interfaces, real ear probe mic. ASSEMBLY: Electronic circuitry all complies to IEC 601 and FCC part 15 class B. Display is 4.7" LCD 640*480 resolution, color monitor, Thermoprinter. Loudspeaker min. 6W in 4 Ohms, freq. range 50 Hz-15,000 Hz. All electrical components are of a standard, off the shelf variety. # TECHNICAL CHARACTERISTICS: Technical specifications complies to IEC 601 and FCC part 15 class B, 2 channels, LCD monitor, Thermoprinter, single 3.5 in floppy drive, measurement circuits freq. range: 42 - 15600 Hz, dynamic range: 130 dB, dynamic resolution: 0.25 dB, accuracy: ±0.2dB, loudspeaker output: min. 6W in 4 Ω, freq. range: 50 - 15000 Hz, electrical THD: &lt; 0.01% MDD compliance will be achieved by Jan. 1997. CONTROLS: The keyboard is a fully functional PC style keyboard. POWER: 110 or 220 V ac; 50-60 Hz; at 220 V AC: 250 mA max., 110 V AC: 500 mA max. 10/28//96
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...