K123624 · Mimosa Acoustics, Inc. · EWO · Mar 28, 2013 · Ear, Nose, Throat
Device Facts
Record ID
K123624
Device Name
OTOSTAT DPOAE + MEPA MEASUREMENT SYSTEM
Applicant
Mimosa Acoustics, Inc.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Mar 28, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of the OtoStat DPOAE Measurement is to determine the presence of cochlear function. The intended use of the OtoStat MEPA Middle Ear Power Analyzer is to characterize the middle ear status and to assist in diagnosing middle ear pathologies. The intended use of the OtoStat DPOAE+MEPA is to determine the presence of cochlear and middle ear functions and to assist in differential diagnosis between cochlear and middle ear dysfunctions. The OtoStat MEPA system measures various acoustic properties of the ear, namely power reflectance, power absorbance, transmittance, wideband immittance, and equivalent ear canal volume. These measures allow the evaluation of the functional condition of the middle and outer ear. The device is suitable for all populations including new-born infants. The device is to be used by trained personnel only.
Device Story
Handheld audiometric device; measures DPOAE and MEPA. Input: acoustic stimuli delivered via ear probe loudspeakers; complex pressure frequency responses recorded via probe microphone. Operation: DPOAE evaluates cochlear outer-hair-cell activity by detecting cubic distortion tones (2f1-f2); MEPA characterizes middle ear status by measuring power reflectance, absorbance, transmittance, and wideband immittance. Data processed via automated monitoring and signal processing for noise/artifact rejection. Output: plots of distortion levels, noise floors, and acoustic characteristics with pass/refer indicators. Used by trained personnel in clinical settings. Benefits: rapid, objective assessment of cochlear and middle ear function; assists in differential diagnosis of pathologies.
Clinical Evidence
No clinical data. Bench testing only; performance verification and validation conducted by comparing functionality against predicate devices (CUB DIS-DPOAE and HearID MEPA) under various test conditions. Compliance with IEC 60601-1 for safety confirmed.
Technological Characteristics
Handheld audiometric measurement device. Uses acoustic ear probe for stimulus delivery and pressure response recording. DPOAE measures 500-8000 Hz range. MEPA measures power reflectance, absorbance, transmittance, and wideband immittance. Software-based automated data monitoring and signal processing. Complies with IEC 60601-1.
Indications for Use
Indicated for all populations, including newborns, to assess cochlear function via DPOAE and middle/outer ear status via MEPA. Used by trained personnel for differential diagnosis of cochlear and middle ear dysfunctions.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
HearID Wideband Middle Ear Power Analyzer (MEPA) (K053216)
Submission Summary (Full Text)
{0}------------------------------------------------
K123624
### MAR 2 8 2013
2 510(K) SUMMARY
#### 510(k) Summary 2
| Date Prepared: | November 20, 2012 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Mimosa Acoustics, Inc.<br>335 N. Fremont Street<br>Champaign, Illinois 61820<br>(217) 579 9740<br>(217) 579 9741<br>Patricia S. Jeng, Ph.D. |
| Device Information: | |
| Trade Name: | OtoStat DPOAE+MEPA, OtoStat MEPA, Oto-Stat DP, OtoStat DP2000 |
| Common Name: | Distortion Product Otoacoustic Emission(DPOAE)<br>and Wideband power reflectance (WBR), Wideband<br>immittance |
| Classification: | OAE and Auditory Impedance Tester |
| Predicate Devices: | CUBeDIS-DPOAE Measurement System (K981460)<br>HearID Wideband Middle Ear Power Analyzer<br>(MEPA) (K053216) |
### Device Description:
The OtoStat is a small portable handheld audiometric measurement device for measuring either DPOAE (distortion product otoacoustic emission), MEPA (middle ear power analysis), or both, where DPOAE is for evaluating cochlear function and MEPA for middle and outer ear status. Both MEPA and DPOAE measurements present acoustic stimulus into the ear canal via the loudspeakers in the acoustic ear probe, then record the complex pressure frequency responses via the microphone in the same acoustic ear probe. The responses are then analyzed and displayed against relevant normative data with recommended pass/refer indication for the trained user to determine the cochlear and middle ear status and for differential diagnosis.
### Intended Use:
The intended use of the OtoStat DPOAE Measurement is to determine the presence. of cochlear function.
The intended use of the OtoStat MEPA Middle Ear Power Analyzer is to characterize the middle ear status and to assist in diagnosing middle ear pathologies.
The intended use of the OtoStat DPOAE+MEPA is to determine the presence of cochlear and middle ear functions and to assist in differential diagnosis between cochlear and middle ear dysfunctions.
The device is suitable for all populations including new-born infants. The device is to be used by trained personnel only.
## Technical Characteristics:
Mimosa Acoustics OtoStat DPOAE and CUB®DIS-DPOAE Measurement System (K981460) have the same technical design and use the same ear probe. The Oto
{1}------------------------------------------------
Stat DPOAE is designed to provide stimuli for rapid clinical measurement and evaluation of odd order distortion product tones generated by the cochlea, and measured in the external ear canal. These tones are known as distortion-product otoacoustic emissions (DPOAEs). The system measures 4, 6, or 12 audiometric octave frequencies (or other frequencies, as desired), between 500 through 8000 Hz with no known artifacts, a low system-noise floor, and a high signal-to-noise ratio. The presence in the ear of a cubic (2f1-f2) distortion tone that is detectable above the system noise level indicates the presence of healthy outer-hair-cell activity in the cochlea. The distortion product level and the corresponding noise floor level are plotted together with recommended pass/refer indicator.
Mimosa Acoustics OtoStat middle-ear power analyzer (MEPA) and HearID middle-ear power analyzer (MEPA3) (K053216) have the same technical design and use the same ear probe. With the ear probe in the patient's ear canal,the MEPA software generated acoustic stimulus is delivered into the external ear canal. Some of the sound is absorbed by the middle ear and the remainder is reflected back into the ear canal. The percentage of reflected power to incident power is defined as power reflectance (IT'). Different middle-ear conditions affect the amount of power that is reflected or absorbed in each frequency band. MEPA measures and plots reflectance, as well as other related acoustic characteristics of the ear, over a wide frequency range, aiding the diagnosis of middle-ear disorders. The OtoStat MEPA instrumentation employs computer-generated stimuli, automated data monitoring, and advanced signal processing for noise and artifact rejection.
### Nonclinical Performance Evaluations:
Following the design control procedure, the performance verification and validation of the OtoStat DPOAE and MEPA Measurement System was conducted by comparing their functionality against CUB DIS-DIS-DPOAE and HearID MEPA Measurement Systems, the predicate devices, for various test conditions. The test results demonstrate substantial equivalence in its effectiveness.
Bench mark evaluations were performed in accordance to the recognized consensus standards. The OtoStat has been demonstrated to comply with Medical Electrical Equipment, Part 1: General Requirements for Safety (IEC 60601-1).
Conclusion: The OtoStat DPOAE+MEPA as a modification to the predicate devices (the original devices) uses the same or identical technology and has the same intended use as the cited predicate devices, and is therefore substantially equivalent (21 CFR 807.92(a)(3)). The nonclinical performance evaluations have demonstrated it is as safe and effectiveness as the original and predicate devices.
2 - 2
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a bird-like figure with three curved lines.
Public Health Service
March 28, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Mimosa Acoustics, Inc. % Patricia S. Jeng, Ph.D. 335 N. Fremont Street Champaign, IL 61820
Re: K123624
Trade/Device Name: OtoStat MEPA + DPOAE Measurement System Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO, ETY Dated: February 25, 2013 Received: February 27, 2013
Dear Dr. Jeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Dr. Patricia S. Jeng
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
# EricADMann
for Malvina B. Eydelman, M.D.
Director Division of Ophthalmic and Ear, Nose
and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### l Indications for Use Statement
### Indications for Use
K123624 510(k) Number (if known):
Device Name: QtoStat DPOAE + MEPA Measurement System
Indications for Use:
The intended use of the OtoStat DPOAE Measurement is to determine the presence of cochlear function.
The intended use of the OtoStat MEPA Middle Ear Power Analyzer is to characterize the middle ear status and to assist in diagnosing middle ear pathologies.
The intended use of the OtoStat DPOAE+MEPA is to determine the presence of cochlear and middle ear functions and to assist in differential diagnosis between cochlear and middle ear dysfunctions.
The OtoStat MEPA system measures various acoustic properties of the ear, namely power reflectance, power absorbance, transmittance, wideband immittance, and equivalent ear canal volume. These measures allow the evaluation of the functional condition of the middle and outer ear.
The device is suitable for all populations including new-born infants.
AND/OR
The device is to be used by trained personnel only.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Digitally signed by Shuchen Peng |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| DN: ou=U.S. Government, ou=HHS, o=FDA, ou=People, cn=Shuchen Peng, 0.9.2342.19200300.100.1.1=1300399622<br>Date: 2013.03.28 10:21:42 -04'00 |
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number: K123624
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.