AUDIOSCREENER OAE+ABR

K024205 · Everest Biomedical Instruments Co. · EWO · Mar 7, 2003 · Ear, Nose, Throat

Device Facts

Record IDK024205
Device NameAUDIOSCREENER OAE+ABR
ApplicantEverest Biomedical Instruments Co.
Product CodeEWO · Ear, Nose, Throat
Decision DateMar 7, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1050
Device ClassClass 2
AttributesPediatric

Indications for Use

The Audioscreener OAE+ABR may be used for patients of all ages, from newborn infants through adults. The Distortion Product Otoacoustic Emissions, Transient Evoked Otoacoustic Emissions, and Auditory Brainstem Response tests are indicated for use in screening individuals for hearing loss for whom behavioral audiometric responses are deemed to be unreliable, such as in infants, young children, and uncooperative or cognitively impaired adults.

Device Story

The Audioscreener OAE+ABR is a diagnostic instrument for evaluating hearing function. It performs Distortion Product Otoacoustic Emissions (DPOAE), Transient Evoked Otoacoustic Emissions (TEOAE), and Auditory Brainstem Response (ABR) testing. The device processes auditory stimuli and physiological responses to screen for hearing loss. It is intended for use by clinicians in clinical settings. The device provides objective hearing screening data, assisting healthcare providers in identifying hearing impairment in patients unable to provide reliable behavioral responses. This facilitates early intervention and management of hearing loss.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

The device is an audiometer and evoked response auditory stimulator. It complies with UL 2601-1, CSA 22.2 No. 601.1, IEC 60601-1-2, ANSI S3.6, IEC 60601-2-40, IEC 645-3, and IEC 60601-2-26. It utilizes software to perform DPOAE, TEOAE, and ABR testing.

Indications for Use

Indicated for screening hearing loss in patients of all ages (newborns to adults) where behavioral audiometry is unreliable, including infants, young children, and uncooperative or cognitively impaired adults.

Regulatory Classification

Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO 24205 ## Premarket Notification [510(k)] Summary MAR 0 7 2003 SUBMITTED BY: Everest Biomedical Instruments Co. 16690 Swingley Ridge Rd. Suite 140 Chesterfield. MO 63017 V: 636-519-7770 F: 636-519-7991 CONTACT PERSON: Randall J. Krohn DATE OF PREPARATION December 19, 2002 TRADE NAME: AUDIOscreener OAE+ABR COMMON NAME: Otoacoustic Emissions Test Instrument / Evoked Response Auditory Stimulator CLASSIFICATION NAME: Audiometer (per 21 CFR section 874.1050) Panel 77 Product Code EWO Evoked Response Auditory Stimulator (per 21 CFR section 882.1900) Panel 84 Product Code GWJ PREDICATE DEVICES: Kedly Audioscreener OAE+ABR (K001058) ## DEVICE DESCRIPTION: The Audioscreener OAE+ABR is a Otoacoustic Emissions and Audiotory Brainstem Response testing device to be used in the evaluation of hearing function. This device is essentially the Audioscreener OAE+ABR unit with additional software required to perform a Transient Evoked OAE in addition to the Distortion Product OAE test. INTENDED USE: The Audioscreener OAE+ABR may be used for patients of all ages, from newborn infants through adults. The Distortion Product Otoacoustic Emissions, Transient Evoked Otoacoustic Emissions, and Auditory Brainstem Response tests are indicated for use in screening individuals for hearing loss for whom behavioral audiometric responses are deemed to be unreliable, such as in infants, young children, and uncooperative or cognitively impaired adults. {1}------------------------------------------------ ## TECHNOLOGICAL CHARACTERISTICS: The Audioscreener OAE+ABR is similar in its intended use to predicate devices and existing methodologies. In addition, the device complies with the following safety and performance standards as applicable to its classification: UL 2601-1 Medical Electrical Equipment Part 1: General Requirements for Safety 18 ed. 1997-10 CSA 22.2 No. 601.1 Medical Electrical Equipment Part 1: General Requirements for Safety-M90 IEC60601-1-2 Medical Electrical Equipment Part 1: General Requirements for Safety 2. Collateral Standard: Electromagnetic Compatibility - Requirements and Tests 1st ed. 1993-04 ANSI S3.6 Specification for Audiometers (sections 4-10 as applicable) 1996 IEC 60601-2-40 Particular Requirements for the Safety of Electromyographs and Evoked Response Equipment (1998-02) FDA Electroencephalograph Devices Guidance for 510(k) Content Draft Document Version 1.0 November 3, 1997 IEC 645-3 Audiometers Part 3: Auditory Test Signals of Short Duration for Audiometric and Neuro-Otological Purposes (1994-10) IEC 60601-2-26 Particular Requirements for the Safety of Electroencephalographs (1994-04) {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle head facing right, composed of three curved lines that suggest depth and dimension. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. MAR 0 7 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Everest Biomedical Instruments Co. c/o Randall J. Krohn VP Engineering - Neuro Products 16690 Swingley Ridge Rd Suite 140 Chesterfield, MO 63017 Re: K024205 Trade/Device Name: AUDIOscreener OAE+ABR Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: December 19, 2002 Received: December 20, 2002 Dear Mr. Krohn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Randall J. Krohn This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, A. halyi Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510(k) Number (if known): Audioscreener OAE+ABR Device Name: Indications for Use: The Audioscreener OAE+ABR may be used for patients of all ages, from newborn infants through adults. The Distortion Product Otoacoustic Emissions, Transient Evoked Otoacoustic Emissions, and Auditory Brainstem Response tests are indicated for use in screening individuals for hearing loss for whom behavioral audiometric responses are deemed to be unreliable, such as in infants, young children, and uncooperative or cognitively impaired adults. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE) Veena Abrol (Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises 510(k) Number K024305 | Prescription Use | | |----------------------|--| | (Per 21 CFR 801.109) | | (Optional Format 3-10-98)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...