The Madsen Zodiac (Type 1096) is an auditory impedance tester that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. This device is also used to measure the acoustic reflex threshold and decay testing as well as eustachian tube function testing for intact and perforated tympanic membranes.
Device Story
Madsen Zodiac 1096 is an auditory impedance tester for middle ear evaluation. It delivers acoustic probe signals and air pressure stimuli (-600 to +400 daPa) to the external auditory canal. The device measures and graphs tympanic membrane mobility, acoustic reflex thresholds, reflex decay, and eustachian tube function. It operates as a standalone unit (via built-in keypad/display) or as a PC-based system using OTOsuite Immittance software. Used in hospitals, clinics, and hearing aid dispensing centers by audiologists and ENTs. Output includes graphical representations of admittance and numerical test results, which assist clinicians in diagnosing middle ear pathologies. Benefits include objective assessment of middle ear function and reflex pathways.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including compliance with ANSI/AAMI ES60601-1, IEC 60601-1-2, EN/IEC 60645-5, and ANSI S3.39. Biocompatibility was evaluated per ISO 10993. Software validation and risk analysis were performed.
Technological Characteristics
Auditory impedance tester; Type BF applied parts. Materials evaluated for biocompatibility (ISO 10993). Connectivity: USB 2.0, standalone or PC-based. Power: External supply (AFM60US24). Probe types: Quick Check, Classic, Comfort. Standards: EN 60601-1, EN 60645-5, ANSI S3.39. Software: C# based.
Indications for Use
Indicated for use by audiologists, ENTs, and healthcare professionals to test hearing in infants, children, and adults. Used to evaluate middle ear function, acoustic reflex thresholds, reflex decay, and eustachian tube function in patients with intact or perforated tympanic membranes.
Regulatory Classification
Identification
An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.
Special Controls
*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other. The profiles are stylized with flowing lines, giving the impression of movement or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2016
Gn Otometrics Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct. Naples, FL 34114
Re: K161707
Trade/Device Name: Madsen Zodiac Regulation Number: 21 CFR 874.1090 Regulation Name: Auditory Impedance Tester Regulatory Class: Class II Product Code: ETY Dated: August 4, 2016 Received: August 9, 2016
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161707
Device Name Madsen Zodiac Model 1096 (SA or PC)
#### Indications for Use (Describe)
The Madsen Zodiac (Type 1096) is an auditory impedance tester that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. This device is also used to measure the acoustic reflex threshold and decay testing as well as eustachian tube function testing for intact and perforated tympanic membranes.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary, 510(k) K161707 Submitter: GN Otometrics A/S Hoerskaetten 9 Taastrup, DENMARK DK-2630 Registration number: 9612197 Tel : +45 45 75 55 55 Fax: +45 45 75 55 59 Contact: Asif Muhammad, Global Director RA/QA Date Prepared: June 16, 2016
| 1. | Identification of the Device: | 510(k) Number K161707 |
|----|-------------------------------|-------------------------------|
| | Proprietary-Trade Name: | Madsen Zodiac 1096 (SA or PC) |
| | Common Name: | Tester, Auditory Impedance |
| | Classification Name: | Auditory impedance tester. |
| | Product Code: | ETY |
| | Regulation Number | 874.1090 |
- 2. Equivalent legally marketed device: 510(K) Number. K033645
| Proprietary-Trade Name: | OTOflex 100 Type 1012, GN Otometrics A/S. |
|-------------------------|-------------------------------------------|
| Common Name: | Tester, Auditory Impedance |
| Classification Name: | Auditory impedance tester. |
| Product Code: | ETY |
| Regulation Number | 874.1090 |
- 3. Description of the Device: The MADSEN Zodiac Type 1096 device is designed to measure and evaluate the acoustic impedance/admittance of the human ear by means of acoustic probe signals at different frequencies and with temporal characteristics. MADSEN Zodiac is a compact device for immittance testing. You can operate Zodiac as a stand-alone unit using the built-in keypad and display, or, if you are using the OTOsuite Immittance software module, you can operate Zodiac using the PC's keyboard and mouse with the OTOsuite Immittance module acting as the display. MADSEN Zodiac supports the following probes:
- The hand-held Quick Check probe
- The two diagnostic probe types, Classic and Comfort
- A contralateral insert phone or TDH-39.
Supported tests: Depending on the configuration, Zodiac supports the following tests and functionalities:
- Tympanometry
- · Reflex Screening
- Reflex Threshold
- · Reflex Decay
• ETF-I (Eustachian Tube Function - Intact) • ETF-P (Eustachian Tube Function -Perforated)
{4}------------------------------------------------
- Admittance Recording (multiple uses, e.g. patulous Eustachian Tube evaluation, acoustic reflexes with external stimulus)
- Manual Tympanometry
This is a MODIFIED version of our previous model, the predicate device.
- 4. Indications for Use (intended use): The Madsen Zodiac (Type 1096) is an auditory impedance tester that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. This device is also used to measure the acoustic reflex threshold and decay testing as well as eustachian tube function testing for intact and perforated tympanic membranes. (Prescription Use).
- 5. Technological Characteristics. This device has the same indications for use as the predicate device and employs similar technology to accomplish the same tasks. Modifications to the cleared device are not substantial and do not change the operating principles of the device. The intended use and fundamental technological characteristics remain the same as the predicate device and modifications do not affect the safety or effectiveness of the device.
| Characteristic | K033645, OTOflex 100 Type 1012,<br>GN Otometrics A/S. | K161707 Madsen Zodiac 1096<br>GN Otometrics A/S |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The OTOflex 100 Type 1012 is an<br>auditory impedance tester that is<br>intended to change the air pressure in<br>the external auditory canal and measure<br>and graph the mobility characteristics of<br>the tympanic membrane to evaluate the<br>functional condition of the middle ear.<br>This device is also used to measure the<br>acoustic reflex threshold and decay<br>testing as well as eustachian tube<br>function testing for intact and<br>perforated tympanic membranes.<br>(Prescription Use). | The Madsen Zodiac (Type 1096) is an<br>auditory impedance tester that is<br>intended to change the air pressure in<br>the external auditory canal and measure<br>and graph the mobility characteristics of<br>the tympanic membrane to evaluate the<br>functional condition of the middle ear.<br>This device is also used to measure the<br>acoustic reflex threshold and decay<br>testing as well as eustachian tube<br>function testing for intact and<br>perforated tympanic membranes.<br>(Prescription Use). (SAME) |
| User population | Audiologists, ENTs and other health<br>care professionals in testing the<br>hearing of infants, children and<br>adults. | Same as predicate |
| Distribution | ENT Doctors, Audiologists and<br>professional Hearing Aid dispensers. | Same as predicate |
| Energy delivered | Acoustic stimuli:<br>Air pressure stimuli -600daPa to<br>+400daPa | Same as predicate |
| Characteristic | K033645, OTOflex 100 Type 1012,<br>GN Otometrics A/S. | K161707 Madsen Zodiac 1096<br>GN Otometrics A/S |
| Where used | Hospitals, private clinics and Hearing<br>Aid dispensers | Same as predicate |
| Instructions | The predicate devices functionality is<br>covered by a user's manual. | Same as predicate |
| Compliance<br>measuring system<br>Probe tones | 226 Hz at 85dB SPL $\pm$ 1.5 dB<br>1000Hz at 75dB SPL $\pm$ 1.5 dB<br>THD: < 3% in 2 cc<br>Frequency accuracy: $\pm$ 0.5%<br>Range: 0.1 ml to 8.0 ml $\pm$ 5% or 0.1 ml,<br>whichever is greater | 226 Hz at 85 dB SPL $\pm$ 3 dB<br>678 Hz at 72 dB SPL $\pm$ 3 dB<br>800 Hz at 70.5 dB SPL $\pm$ 3 dB<br>1000 Hz at 69 dB SPL $\pm$ 3 dB<br>THD: < 1% in 2 cc<br>Frequency accuracy: $\pm$ 0.5%<br>Range: 0.2 ml to 5.0 ml $\pm$ 5% or 0.05 ml<br>which ever is greater, 5 ml to 8.0 ml $\pm$<br>15% |
| Acoustic reflex<br>Contralateral<br>Stimulation | Pure tones: 500 Hz, 1000 Hz, 2000<br>Hz, 3000 Hz, 4000 Hz<br>Frequency accuracy: $\pm$ 0.5%<br>Noise White Noise according to IEC<br>61027<br>Low Pass 400 to 1600 Hz.<br>High Pass 1600 to 4000 Hz.<br>Roll off >12 dB/Octave.<br>Range at: BBN, LPN at 50 to 100 dB<br>SPL $\pm$ 3 dB, HPN at 50 to 95 dB SPL $\pm$ 3<br>dB<br>Step size dB 1, 2, 5, 10 dB<br><i>E-A-RTONE</i> ® 3A:<br>Range at: 500 Hz at 50 to 105 dB HL $\pm$<br>3 dB<br>1000 Hz at 50 to 120 dB HL $\pm$ 3 dB<br>2000 Hz at 50 to 115 dB HL $\pm$ 3 dB<br>3000 Hz at 50 to 105 dB HL $\pm$ 3 dB<br>4000 Hz at 50 to 110 dB HL $\pm$ 3 dB<br>THD: < 3% in 2 cc (measured 5 dB<br>below max output) | Same as predicate<br>Same as predicate<br>Broad Band Noise according to IEC<br>60645-5<br>Low Pass: TDH-39: 250 - 1600 Hz,<br>Insert: 400 - 1600 Hz.<br>High Pass: TDH-39: 1600 - 6000 Hz,<br>Insert: 1600 - 4000 Hz.<br>Same as predicate<br>Range: BBN, LPN, HPN at 50 to 110<br>dB SPL $\pm$ 3 dB<br>Same as predicate<br>Contralateral insert phone:<br>Range: 500 Hz at 50 to 115 dB HL $\pm$ 3<br>dB<br>Same as predicate<br>2000 Hz at 50 to 120 dB HL $\pm$ 3 dB<br>Frequency not used<br>4000 Hz at 50 to 120 dB HL $\pm$ 3 dB<br>THD: < 5% for levels below 110 dB HL,<br>< 10% for levels above 110 dB HL<br>Contralateral TDH-39 phone:<br>500 Hz at 50 to 115 dB HL $\pm$ 3 dB<br>1000 Hz at 50 to 120 dB HL $\pm$ 3 dB<br>2000 Hz at 50 to 115 dB HL $\pm$ 3 dB |
| Characteristic | K033645, OTOflex 100 Type 1012,<br>GN Otometrics A/S. | K161707 Madsen Zodiac 1096<br>GN Otometrics A/S |
| | | 4000 Hz at 50 to 115 dB HL ± 3 dB |
| | | < 2.5 % for levels below 110 dB HL |
| | | < 5 % for levels above 110 dB HL |
| Ipsilateral<br>Stimulation | Tone: 500 Hz, 1000 Hz, 2000 Hz, 3000<br>Hz, 4000 Hz | Tone: 500 Hz, 1000 Hz, 2000 Hz, 4000<br>Hz |
| | Frequency accuracy: ± 0.5% | Same as predicate |
| | Noise White Noise according to IEC<br>61027 | Broad Band Noise according to IEC<br>60645-5 |
| | Low Pass 400 to 1600 Hz | Same as predicate |
| | High Pass 1600 to 4000 Hz | Same as predicate |
| | Roll off >12 dB/Octave | Same as predicate |
| | Step size dB: 1, 2, 5, 10 dB | Same as predicate |
| | Range at: 500 Hz at 50 to 105 dB HL ±<br>3 dB | Same as predicate |
| | 1000 Hz at 50 to 120 dB HL ± 3 dB | 1000 Hz at 50 to 110 dB HL ± 3 dB |
| | 2000 Hz at 50 to 115 dB HL ± 3 dB | 2000 Hz at 50 to 110 dB HL ± 3 dB |
| | 3000 Hz at 50 to 105 dB HL ± 3 dB | Frequency not used |
| | 4000 Hz at 50 to 110 dB HL ± 3 dB | 4000 Hz at 50 to 100 dB HL ± 3 dB |
| | THD: < 3% in 2 cc (measured 5 dB<br>below max output) | THD: < 5% for levels below 110 dB HL,<br>< 10% for levels above 110 dB HL |
| Air pressure<br>system | Range: Normal +200 to -400 daPa/s.<br>Extended +400 to -600 daPa/s | Same as predicate |
| | Pressure sweep rate: 50, 100, 200,<br>400 daPa/s, A.F.A.P | Pressure sweep rate: 50, 100, 200,<br>400, 600 daPa/s |
| | Pressure accuracy: ±10% or ±10 daPa,<br>whichever is greatest | Same as predicate |
| | Pump measure direction: Positive to<br>negative or negative to positive | Same as predicate |
| | Safety: Separate safety +530 daPa<br>and -730 daPa. ±70 daPa | Same as predicate |
| | Software safety +450 daPa and -650<br>daPa. ±70 daPa. | Same as predicate |
| | | Manual control of pressure |
| Graph units | Unit of admittance graph Y-axis: ml,<br>cc, mmho, µl | Same as predicate |
| | Unit of graph X-axis: daPa, sec | Same as predicate |
| Device display | Graphic 128x128 dots | Display: 7 inch, 15:9 WVGA<br>Resolution: 800 x 480 pixel |
| | | |
| | | Light indication on probes |
| Light indications | | |
| | | |
| | Interface | Wireless Bluetooth data transfer to |
| Characteristic | K033645, OTOflex 100 Type 1012, GN Otometrics A/S. | K161707 Madsen Zodiac 1096 GN Otometrics A/S |
| | PC | Compatible: USB 2.0 |
| Operating environment | Temperature: +15°C to +35°C (59°F to +95°F)<br>Rel. humidity: 30 to 90 %, non-condensing<br>Warm-up time: < 2 min.<br>Air pressure: 600 hPa to 1060 hPa<br>Operation at temperatures below -20°C or above +60°C may cause permanent damage. | Same as predicate |
| Storing and handling | Temperature: -20°C to +60°C (-4°F to +140°F)<br>Rel. humidity: < 90 %, non-condensing<br>Air pressure: 500 hPa to 1060 hPa | Same as predicate |
| Dimensions (HxWxD) | MADSEN OTOflex 100 (HxWxD): 20 cm x 4.9 cm x 7.8 cm (7.9" x 1.9" x 3.0")<br>Charger unit (HxWxD): 18 cm x 4.9 cm x 7.8 cm (6.9" x 1.9" x 3.0") | Stand-alone version: 190 mm x 248 mm x 261 mm (7.5" x 9.8" x 10.3")<br>PC-based version: 100 mm x 240 mm x 240 mm (3.9" x 9.4" x 9.4")<br>Quick Check probe: 28 mm x 22 mm x 100 mm (1.1" x 0.9" x 3.9")<br>Diagnostic probe: 10 mm x 10 mm x 25 mm (0.4" x 0.4" x 1.0")…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.