The Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes. The Titan Impedance System measures various acoustic properties of the ear, namely power reflectance, power absorption, transmittance, reflectance group delay, complex acoustic impedance and admittance, and equivalent ear canal volume. These measures allow the evaluation of the functional condition of the middle and outer ear. The Titan Impedance System is suitable for all populations including new-born infants. The Titan Impedance System is to be used by trained personnel only such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
Device Story
Handheld audiometric device (Titan) with cradle and PC software (Titan Suite/IMP440); used by audiologists/ENTs in clinical settings. Device inserts probe into ear canal; presents calibrated tones/clicks (250-8000Hz); records intensity/phase of reflected sound. Measures acoustic properties: power reflectance, absorption, transmittance, impedance, admittance, and ear canal volume. Includes artifact rejection system to remove noise-contaminated responses. Data transmitted via Bluetooth to PC for analysis. Provides normative data for guidance; displays tympanograms and acoustic reflex measurements. Assists clinicians in diagnosing otologic disorders by evaluating middle ear mobility and functional status.
Clinical Evidence
No clinical data performed. Evidence based on bench testing confirming conformance to international standards (IEC 60645-5/ANSI S3.39) and comparison to predicate devices. Literature cited for normative data and measurement validation.
Technological Characteristics
Handheld diagnostic acoustic impedance instrument; Type 1 (IEC 60645-5/ANSI S3.39). Connectivity: Bluetooth to PC. Power: Battery/Cradle. Stimuli: Chirp/Click. Calibration: IEC 60645-3. Features artifact rejection, normative data overlays, and customizable measurement protocols.
Indications for Use
Indicated for diagnostic hearing evaluations and assessment of middle/outer ear functional condition in all populations, including newborns. Features tympanometry and acoustic reflexes. Contraindicated for users without necessary training/education.
Regulatory Classification
Identification
An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.
Special Controls
*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
HEARID WIDEBAND MIDDLE EAR POWER ANALYZER (K053216)
Submission Summary (Full Text)
{0}------------------------------------------------
Titan, IMP440 WBT
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
### Titan IMP440 WBT
Interacoustics A/S
Drejervaenget 8
SUBMITTER INFORMATION Company Name Company Address
NOV 2-9 2012
Company Phone Company Fax Company e-mail Contact Person
Assens DK-5610 Denmark +45 6371 3555 +45 6371 3522 info@interacoustics.com Erik Nielsen. Director, Quality and Regulatory Affairs June. 21 2012
Date Summary Prepared
DEVICE IDENTIFICATION Trade Name Common Name Classification Name
Product Code Panel Device Class
Titan Audiometric equipment. Tester, Auditory Impedance / Evoked response auditory stimulator ETY / GWJ Ear Nose & Throat / Neurology Class II
SUBSTANTIAL EQUIVALENCE (TITAN WITH IMP440)
Predicate Device Manufacturer 510(k) No. Date Cleared
Titan with IMP440 Interacoustics K083861 04/09/2006
SUBSTANTIAL EQUIVALENCE (MIMOSA HEARID)
Predicate Device Manufacturer 510(k) No. Date Cleared
HEARID WIDEBAND MIDDLE EAR POWER ANALYZER MIMOSA ACOUSTICS, INC K053216 02/10/2006
Image /page/0/Picture/18 description: The image is a logo for Interacoustics. The logo consists of a stylized globe on the left and the word "Interacoustics" in a sans-serif font on the right. A registered trademark symbol is located to the right of the word. The logo is white on a black background.
{1}------------------------------------------------
### Description of device
The instrument is audiometric equipment used for measuring aural acoustic impedance and admittance (According to ANSI S3.39).
The Titan is a platform with multi functions depending of licensing. The license function for this submission is IMP440 (IMP) for measurement of aural acoustic impedance and admittance.
The Titan IMP test is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The IMP test function is already cleared by FDA 510(k) premarket notification K083861 (04/09/2009)
The instrument is a diagnostic/clinical instrument (as defined in IEC60645-5), but provides normative data for guidance.
Titan consists of a handheld unit named Titan, Titan Cradle (cleared by FDA 510(k) premarket notification K083861) and PC software (Titan Suite/IMP440). The measurements are controlled by the handheld unit.
A connection box (shoulder box) enables different types of accessories to be connected to the platform
The Titan platform can connect to PC software via a Bluetooth connection.
### Indications for Use
The Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes.
The Titan Impedance System measures various acoustic properties of the ear, namely power reflectance, power absorption, transmittance, reflectance group delay, complex acoustic impedance and admittance, and equivalent ear canal volume. These measures allow the evaluation of the functional condition of the middle and outer ear.
The Titan Impedance System is suitable for all populations including new-born infants.
The Titan Impedance System is to be used by trained personnel only such as audiologists, ENT surgeons, doctors, healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
Section 05 p. 3 - 11
Image /page/1/Picture/16 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left and the word "Interacoustics" in bold, white letters on a black background. The globe icon has lines running across it, and the word "Interacoustics" is written in a sans-serif font.
leading diagnostic solutions
{2}------------------------------------------------
### Technological Characteristics
Titan consists of a handheld unit named Titan, Titan Cradle and PC software (Titan Suite, Titan modules/ Titan Applications). The measurements are controlled by the handheld unit. A license system makes it possible within each configuration to select which functionality the user wants to be incorporated in the system.
A connection box (shoulder box) enables different types of accessories to be connected to the platform (depending on the module) The Titan platform can connect to PC software via a Bluetooth connection.
For measuring acoustic properties of the middle ear a probe is placed into the ear canal. The probe presents a calibrated tone or click sound into the ear canal and by recording the intensity and phase of the remaining tone or click in the ear canal it can be derived how much of the sound is absorbed by or transferred into the ear. Most common measures are:
- A tympanogram, in which the compliance/absorbance is measured as function of the presented frequencies and pressure in the ear canal;
- An acoustic reflex, in which the compliance/absorbance is measured over time as function of the presented frequencies while an acoustic activator is presented that above certain intensity is expected to cause a contraction of the stapedius muscle which in return is expected to cause a change in the compliance/absorbance of the ear.
The levels of the tonal and click sounds are to be calibrated in an artificial ear according to peak to peak equivalent signal level principles in IEC 60645-3
Titan with IMP440 WBT is IEC 60645-5/ANSI S3.39, Type 1 (Diagnostic/Clinical) acoustic impedance instrument
Section 05 p. 4 - 11
Image /page/2/Picture/11 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left and the word "Interacoustics" in bold, white letters on a black background. A registered trademark symbol is present to the right of the word.
leading diagnostic solutions
{3}------------------------------------------------
| Description | Titan with IMP440 | Titan with IMP440 WBT | Equivalence |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes. | The Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes.<br><br>The Titan Impedance System measures various acoustic properties of the ear, namely power reflectance, power absorption, transmittance, reflectance group delay, complex acoustic impedance and admittance, and equivalent ear canal volume. These measures allow the evaluation of the functional condition of the middle and outer ear.<br><br>The Titan Impedance System is suitable for all populations including new-born infants.<br><br>The Titan Impedance System is to be used by trained personnel only such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted. | Same<br><br>The first paragraph for Titan IMP440 WBT is identical with the previous version of the Titan with IMP440<br><br>The addition of subsequent paragraphs is the subject of this 510(k) |
| Target population | The devices are suitable for all populations including new-born infants | The devices are suitable for all populations including new-born infants | Same |
| Intended user | The devices are to be used by trained personnel only | The devices are to be used by trained personnel only | Same |
## Comparison table for Titan with IMP440 and Titan IMP440 WBT
## Section 05 p. 5 - 11
leading diagnostic solutions
Inte
{4}------------------------------------------------
| Type | HearID (Predicate) | Titan IMP440 WBT | Comments |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Audiometric equipment | Audiometric equipment | Same |
| Indications for use | The intended use of the<br>HearID-wbMEPA Middle Ear<br>Power Analyzer is to<br>characterize the middle ear<br>status and to assist in<br>diagnosing middle ear<br>pathologies.<br>The HearID-wbMEPA system<br>measures various acoustic<br>properties of the ear, namely<br>power reflectance, power<br>absorption, transmittance,<br>reflectance group delay,<br>complex acoustic impedance<br>and admittance, and<br>equivalent ear canal volume.<br>These measures allow the<br>evaluation of the functional<br>condition of the middle and<br>outer ear. The devices are<br>suitable for all populations<br>including new-born infants.<br>The devices are to be used by<br>trained personnel only.<br>The HearID system comes in<br>two versions: HearID-b-<br>wbMEPA, and HearID e-<br>wbMEPA, where the difference<br>is the variation of the<br>hardware platform. | The Titan Impedance System<br>is an electroacoustic test<br>instrument that produces<br>controlled levels of test tones<br>and signals intended for use<br>in conduction diagnostic<br>hearing evaluations and<br>assisting in the diagnosis of<br>possible otologic disorders. It<br>features tympanometry and<br>acoustic reflexes.<br><br>The Titan Impedance System<br>measures various acoustic<br>properties of the ear, namely<br>power reflectance, power<br>absorption, transmittance,<br>reflectance group delay,<br>complex acoustic impedance<br>and admittance, and<br>equivalent ear canal volume.<br>These measures allow the<br>evaluation of the functional<br>condition of the middle and<br>outer ear.<br><br>The Titan Impedance System<br>is suitable for all populations<br>including new-born infants.<br><br>The Titan Impedance System<br>is to be used by trained<br>personnel only such as<br>audiologists, ENT surgeons,<br>doctors, hearing healthcare<br>professionals or personnel<br>with a similar level of<br>education. The device should<br>not be used without the<br>necessary knowledge and<br>training to understand its use<br>and how results should be<br>interpreted. | Same<br><br>The first paragraph for<br>Titan IMP440 WBT is<br>identical with the<br>previous version of the<br>Titan with IMP440<br><br>Paragraph 2, 3 and 4 for<br>Titan IMP440 WBT are<br>identical with paragraph<br>2 for the predicate<br>device.<br><br>The 3rd paragraph for the<br>predicate device is just a<br>listing of variants of<br>devices and has been<br>omitted in the<br>comparison |
| Target population | The devices are suitable for all<br>populations including new-<br>born infants | The devices are suitable for<br>all populations including new-<br>born infants | Same |
| Intended user | The devices are to be used by<br>trained personnel only. | The devices are to be used by<br>trained personnel only. | Same |
| Safety standard | IEC 60601-1 (assumed) | IEC 60601-1 | Same. |
| | HearID (Predicate) | Titan IMP440 WBT | Comments |
| EMC standard | IEC 60601-1-2 (assumed) | IEC 60601-1-2 | predicate device complies<br>as it at least is CE marked<br>Same.<br>It is assumed that the<br>predicate device complies<br>as it at least is CE marked |
| Performance standard | Unknown | IEC 60645-5 / ANSI S3.39 | The Titan IMP440 WBT<br>complies with current<br>state of the art standards.<br>It is unknown if the<br>predicate device<br>complies. |
| Frequency Range | 200-6000 Hz | 250-8000Hz | Substantial equivalent<br>Titan system use better<br>transducers so the<br>frequency range can be<br>expanded to 8kHz. We<br>trust that this does not<br>influence the safety or<br>effectiveness of the<br>system |
| Intensity Range | 0 - 80dB SPL | 90/94 dB peSPL =<br>70/74 dB SPL | Substantial equivalent<br>The intensity range is<br>fixed but inside the range<br>as the HeardID. The<br>variable range for<br>HeardID is used for pure<br>tone stimulation and this<br>is not relevant for the<br>Titan WBT. We trust that<br>when the range is inside<br>the predicate device we<br>can declare equivalence<br>for safety and<br>effectiveness. |
| Sample time | 0.1 - 10 seconds per point | 21, 5 Hz (Click sample time) | These parameters are not<br>comparable, as the range<br>for the predicate device is<br>for pure tone, and the<br>fixed sample time for<br>Titan is fixed for the<br>broadband click |
| Stimuli | Chirp<br>Pure Tone | Chirp (Frequency linearized<br>click)<br>Not relevant | Substantial equivalent |
| Artefact rejection | Present | Present | Substantial equivalent.<br>Titan IMP440 WBT has an<br>artefact rejection system |
| | HearID (Predicate) | Titan IMP440 WBT | Comments |
| | | | to remove responses contaminated with noise or noise artefacts. HearID has equivalent artefact rejection system to minimize noisy measurements. |
| Real time display of signal and noise | Present | No real time presentation. | Different. The Titan IMP440 WBT does not have real time display of signal and noise. Titan IMP440 WBT has a real time display of measurement data during acquisition. We trust that such "quality indicator" is irrelevant because the measurement time is so short that an indicator cannot be monitored with any benefit. After the test has finished, accepted and rejected sample ratio can be viewed. |
| Normative data | Present | Present | The functionality is substantial equivalent. Both systems have normative data present in the user interface to indicate normal middle ear function. Normative data are acquired individually for each system and is based on different databases. The Titan IMP440 WBT normative data is based on [1], [2]. The HearID norm data is based on other data. |
| Reflectance Area Index for enhanced diagnosis | Present | Present | Substantial equivalent. Titan IMP440 WBT uses a shaded area indicating normative region in either 10/90 percentiles or 5/95 percentiles for evaluating measurement results. Identical with normative data above |
| | HearID (Predicate) | Titan IMP440 WBT | Comments |
| measurement<br>protocols. | Present | Present | Users can change the<br>sequence of<br>measurements to have<br>test protocols fit for<br>diagnostic and clinical<br>purposes. |
| Customizable display<br>parameters. | Present | Present | Substantial equivalent.<br>The user can change how<br>many and how graphs are<br>presented on the display.<br>Graphs may differ but the<br>intended purpose is the<br>same. |
| Otitis media with<br>effusion examples | Present | Present | The functionality is<br>substantial equivalent.<br>For Titan IMP440 WBT<br>these are sketched<br>examples of curve<br>characteristics based on<br>measurements [4] |
| Measurements | | | |
| Power Reflectance | Present | Present | Substantial equivalent.<br>Power Reflectance is<br>equal to 1 - Power<br>Absorbance.<br>Power absorbance is<br>plotted in Titan IMP440<br>WBT.<br>Refer to Power<br>Absorbance |
| Power Absorbance | Present | Present…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.