K230853 · Nvision Biomedical Technologies, Inc. · ETN · Oct 6, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K230853
Device Name
EARP Nerve Cuff Electrode
Applicant
Nvision Biomedical Technologies, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Oct 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The EARP Nerve Cuff Electrode is used to perform localized stimulation of neural tissue and to locate, identify, and monitor spinal nerve roots during surgery.
Device Story
EARP Nerve Cuff Electrode is a sterile, single-use surgical accessory for intraoperative neuromonitoring. It functions as a passive conduit for electrical signals between external, commercially available neuromonitoring systems and target neural tissue. The device features a distal C-shaped cuff with platinum contacts for direct nerve engagement and proximal 1.5mm safety socket connectors for equipment interface. During surgery, a clinician (e.g., surgeon or neurophysiologist) positions the cuff around spinal nerve roots to facilitate localized stimulation or monitoring. The device does not generate or record signals independently; it relies on the host neuromonitoring system. By enabling precise nerve identification and monitoring, the device assists the surgeon in avoiding nerve damage, potentially improving surgical outcomes. It is intended for use in an operating room environment.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional analysis, tensile/flexural testing, electrical safety (high potential/leakage), and biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity). All tests passed.
Technological Characteristics
Bipolar nerve cuff electrode. Materials: Platinum contacts, polyimide insulation, silicone cuff, stainless steel lead wire, FEP lead insulation. Dimensions: 2mm strip contact, 1.8m lead length, 260mm applicator. Connectivity: 1.5mm safety socket for external neuromonitoring systems. Sterilization: Sterile, single-use. Standards: ISO 14708-1 (electrical safety), IEC 60601-1 (lead wire performance), ISO 10993 (biocompatibility).
Indications for Use
Indicated for use in patients undergoing surgery requiring localized stimulation, identification, and monitoring of spinal nerve roots.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
Ad-Tech Medical Cueva Cranial Nerve Electrode (K944061)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date October 6, 2023. The text is written in a clear, sans-serif font. The date is likely part of a document or calendar entry. The text is horizontally aligned and centered.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nvision Biomedical Technologies, Inc. % Jeffrey Brittan Vice President of Product Realization Watershed Idea Foundry 1815 Aston Ave.. Suite 106 Carlsbad, California 92008
Re: K230853
Trade/Device Name: EARP Nerve Cuff Electrode Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: September 6, 2023 Received: September 7, 2023
Dear Jeffrey Brittan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Patrick Antkowiak -S
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name EARP Nerve Cuff Electrode
#### Indications for Use (Describe)
The EARP Nerve Cuff Electrode is used to perform localized stimulation of neural tissue and to locate, identify, and monitor spinal nerve roots during surgery.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
DATE PREPARED October 4, 2023
## MANUFACTURER
Nvision Biomedical Technologies, Inc. 4590 Lockhill Selma San Antonio, TX 78249, USA Telephone: (210) 545-3713 Fax: (866) 764-1139 Official Contact: Diana Langham, Director of Regulatory and Corporate Compliance
#### REPRESENTATIVE/CONSULTANT
Jeffrey Brittan Vice President of Product Realization Watershed Idea Foundry Telephone: (714) 287-6780 Email: jeffbrittan(a)watershedideas.com
### PROPRIETARY NAME OF SUBJECT DEVICE
EARP Nerve Cuff Electrode
COMMON NAME Nerve Stimulator
#### DEVICE CLASSIFICATION
Surgical nerve stimulator/locator (Classification Regulations: 21 CFR 874.1820, Product Codes: ETN, Class: II)
PREMARKET REVIEW Ear, Nose, & Throat
#### INDICATIONS FOR USE
The EARP Nerve Cuff Electrode is used to perform localized stimulation of neural tissue and to locate, identify, and monitor spinal nerve roots during surgery.
#### DEVICE DESCRIPTION
The EARP Nerve Cuff Electrode conducts electrical signal as a component of intraoperative neuromonitoring. The EARP Nerve Cuff Electrode is used with commercially available neuromonitoring systems and does not stimulate or record signal itself. The standard connectors at the proximal end of the EARP Nerve Cuff Electrode interface with the neuromonitoring equipment and the distal cuff contacts the target tissue. The EARP Nerve Cuff Electrode is provided sterile packaged and is for single use only.
{5}------------------------------------------------
K230853 Page 2 of 3
#### PREDICATE DEVICE IDENTIFICATION
The subject EARP Nerve Cuff Electrode is substantially equivalent to the following predicates:
| 510(k) Number | Manufacturer & Predicate Device Name | Predicate |
|---------------|-----------------------------------------------|------------|
| K103128 | Cadwell Lab Disposable Stimulator Probes | Primary |
| K944061 | Ad-Tech Medical Cueva Cranial Nerve Electrode | Additional |
#### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for nerve stimulators or electrodes. The EARP Nerve Cuff Electrode was evaluated to ensure that it performed as intended and supported substantial equivalence to the predicate devices. Testing and analysis included:
- Analysis of dimensional characteristics, materials, function, and intended use .
- Tensile and flexural testing (to verify mechanical integrity, continuity, isolation, and . visual appearance)
- Electrical safety testing (high potential and electrical leakage per ISO 14708-1)
- Electrode lead wire performance per 21 CFR 898.12 (IEC 60601-1)
- Biocompatibility testing including cytotoxicity, sensitization, irritation/intracutaneous . reactivity, acute systemic toxicity, and material-mediated pyrogenicity:
| Endpoint | Test Description | Conclusion | Pass/Fail |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Cytotoxicity<br>(ISO 10993-5) | Elution Method<br>(1X MEM extraction vehicle) | The test article extract showed no evidence of<br>causing cell lysis or toxicity. The test article<br>extract met the requirements of the test since the<br>grade was less than a grade 2 (mild reactivity). | Pass |
| Sensitization<br>(ISO 10993-10) | Maximization Sensitization Study<br>(Polar and non-polar extraction<br>vehicles) | The test article extracts showed no evidence of<br>causing delayed dermal contact sensitization in<br>the guinea pig. The test article was not<br>considered a sensitizer in the guinea pig<br>maximization test. | Pass |
| Irritation<br>(ISO 10993-23) | Intracutaneous Study<br>(Polar and non-polar extraction<br>vehicles) | The test article met the requirements of the test<br>since the difference between each test article<br>extract overall mean score and corresponding<br>control extract overall mean score was 0.0 and<br>0.1 for the SC and SO test article extracts,<br>respectively. | Pass |
| Acute Systemic<br>Toxicity<br>(ISO 10993-11) | Systemic Toxicity Study<br>(Polar and non-polar extraction<br>vehicles) | There was no mortality or evidence of systemic<br>toxicity from the extracts injected into mice.<br>Each test article extract met the requirements of<br>the study. | Pass |
| Material-Mediated<br>Pyrogenicity<br>(USP<151>,<br>ISO 10993-11) | Pyrogen Study - Material Mediated<br>(Sterile non-pyrogenic saline<br>extraction vehicle) | The total rise of rabbit temperatures during the 3<br>hour observation period was within acceptable<br>USP requirements. The test article met the<br>requirements for the absence of pyrogens. | Pass |
{6}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Subject Device | Predicate Devices | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| | EARP Nerve Cuff Electrode | Cadwell Lab | Ad-Tech Medical |
| | | Disposable Stimulator Probes | Cueva Cranial Nerve Electrode |
| Indications<br>for Use | The EARP Nerve Cuff Electrode is<br>used to perform localized stimulation<br>of neural tissue and to locate,<br>identify, and monitor spinal nerve<br>roots during surgery. | Cadwell Disposable Stimulator Probe<br>is used to perform localized<br>stimulation of neural tissue and to<br>locate, identify and monitor cranial<br>motor nerves, peripheral nerves and<br>spinal nerve roots during surgery. | The Cueva Cranial Nerve Electrode is<br>intended for use to monitor cranial<br>nerves during skull base type<br>surgeries. |
| Principle of<br>Operation | Conducts signal between<br>neuromonitoring equipment and<br>nerve (interfaces w/ commercially<br>available neuromonitoring systems) | Conducts signal between<br>neuromonitoring equipment and<br>nerve (interfaces w/ commercially<br>available neuromonitoring systems) | Conducts signal between<br>neuromonitoring equipment and<br>nerve (interfaces w/ commercially<br>available neuromonitoring systems) |
| Product Codes<br>/Reg# | ETN (21 CFR 874.1820)<br>Stimulator, Nerve | ETN (21 CFR 874.1820)<br>Stimulator, Nerve | GZL (21 CFR 882.1330)<br>Electrode, Depth |
| Duration of Use | Intraoperative<br>(Single use) | Intraoperative<br>(Single use) | Intraoperative<br>(Single use) |
| Implanted | No<br>(Removed after use) | No<br>(Removed after use) | No<br>(Removed after use) |
| Tissue<br>Engagement | Direct nerve contact<br>(C-shaped cuff) | Direct nerve contact<br>(Forked tip) | Direct nerve contact<br>(C-shaped cuff) |
| Tip Contact<br>Exposure | 2mm<br>(Strip) | 2mm<br>(Balls) | ~2mm<br>(Strip) |
| Electrode Exposed<br>Surface Area | $15.4 - 42.6mm^2$ | $~25.1 - 33.2mm^2$ | $~10.4 - 14.5mm^2$ |
| Bipolar or<br>Monopolar? | Bipolar | Bipolar and monopolar versions | Monopolar |
| Electrode Contact<br>Material | Conductive metal<br>(Platinum) | Conductive metal<br>(Stainless Steel) | Conductive metal<br>(Platinum) |
| Electrode<br>Insulation | Polymer<br>(Polyimide) | Polymer<br>(PTFE) | Polymer |
| Cuff Material | Polymer<br>(Silicone) | N/A | Polymer<br>(Silicone) |
| Lead<br>Length | 1.8m | 2.0m | 1.8m |
| Lead Wire<br>Material | Conductive metal<br>(Stainless steel) | Conductive metal<br>(Tin plated copper) | Conductive metal |
| Lead Wire<br>Insulation | Polymer<br>(FEP) | Polymer<br>(PVC) | Polymer |
| Connector Style | 1.5mm safety socket | 1.5mm safety socket | 1.5mm safety socket |
| Applicator Tool<br>Length | 260mm | 190 to 330mm (handle & shaft) | 250mm |
| Provided Sterile? | Yes | Yes | Yes |
#### CONCLUSION
Based on testing and analysis of device characteristics, it can be concluded that the subject device does not raise new issues of safety compared to the predicate devices. The equivalent indications for use, technological characteristics, and performance characteristics for the EARP Nerve Cuff Electrode are assessed to be substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.