NIM Surgeon Control Probe, Prass Tip, NIM Surgeon Control Probe, 1mm Ball Tip, NIM Surgeon Control Probe, 1mm Ball Tip, 35cm Long
K213246 · Medtronic Xomed, Inc. · ETN · Mar 21, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K213246
Device Name
NIM Surgeon Control Probe, Prass Tip, NIM Surgeon Control Probe, 1mm Ball Tip, NIM Surgeon Control Probe, 1mm Ball Tip, 35cm Long
Applicant
Medtronic Xomed, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Mar 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The NIM™ surgeon control probes are intended for use as intraoperative nerve stimulators.
Device Story
NIM™ Surgeon Control Probes are single-use, sterile, handheld surgical instruments used to deliver electrical stimulation to cranial and peripheral motor nerves. Operated by surgeons in an OR setting, these probes connect to a patient interface to facilitate nerve location and identification. The device features integrated controls allowing the surgeon to adjust stimulation current and key functions directly from the surgical site. By providing controlled electrical pulses, the probe helps surgeons map nerve pathways, potentially reducing the risk of nerve damage during procedures. The device is available in Prass-tip or Ball-tip configurations.
Clinical Evidence
Bench testing only. Testing included electromechanical, dimensional, and visual design performance; sterilization validation (ISO 11135:2014); electrical safety/EMC (IEC 60601); biocompatibility (ISO 10993-1:2018); shelf-life stability; and packaging/distribution testing (ISTA 3A, ASTM D4169).
Technological Characteristics
Materials: 303 Stainless Steel with Xylamed or PTFE coating. Principle: Electrical nerve stimulation. Form factor: Handheld probe with Prass-tip or 1mm Ball-tip. Connectivity: Wired connection to patient interface/console. Sterilization: Ethylene oxide (ISO 11135:2014).
Indications for Use
Indicated for stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots. Contraindicated for use with paralyzing anesthetic agents that may reduce or eliminate patient EMG response.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
March 21, 2022
Medtronic Xomed, Inc. Alexandra Oliver Senior Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville. Florida 32216
Re: K213246
Trade/Device Name: NIM™ Surgeon Control Probes Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: February 23, 2022 Received: February 25, 2022
### Dear Alexandra Oliver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213246
Device Name NIM™ Surgeon Control Probes
Indications for Use (Describe)
The NIM™ Surgeon Control Probes are intended to stimulate cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| I. SUBMITTER | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Company (submitter): | Medtronic Xomed, Inc.<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216 USA<br>Telephone Number: (904) 296-9600 |
| Date Prepared: | March 15, 2022 |
| Contact Person: | Alexandra Oliver<br>Senior Regulatory Affairs Specialist<br>Telephone: (904) 332-8936<br>Email Address:<br>Alexandra.j.oliver@medtronic.com |
## II. DEVICE
| Proprietary (Trade) Name: | NIM™ Surgeon Control Probes |
|---------------------------|-----------------------------------|
| Common Name: | Stimulator, nerve |
| Regulation Name: | Surgical nerve stimulator/locator |
| Regulation Number: | 21 CFR 874.1820 |
| Product Code: | ETN |
| Classification: | II |
| Panel: | 77 (Ear, Nose, & Throat) |
#### III. PREDICATE DEVICE(s)
The NIM™ Surgeon Control Probes are substantially equivalent in intended use and technological characteristics to the following predicate devices:
| 510(k) Number | 510(k) / Device Name | 510(k) Clearance Date |
|---------------|-------------------------------------|-----------------------|
| K873964 | Nicolet Disposable<br>Prass Probe | November 20, 1987 |
| K992869 | Xomed Monopolar<br>Stimulator Probe | October 07, 1999 |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
#### Device Description
The NIMTM surgeon control probes carry stimulation current from the patient interface to the patient. It also enables the user to adjust stimulation current and key functions from the surgical site. All probes are single use devices.
#### Intended Use
The NIM™ surgeon control probes are intended for use as intraoperative nerve stimulators.
#### V. INDICATIONS FOR USE
#### Indications for Use
The NIM™ surgeon control probes are intended to stimulate cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.
#### VI. SUBSTANTIAL EQUIVALENCE
#### Substantial Equivalence/Device Comparison (Subject Device(s) to Predicate Devices)
| Feature/Attribute | Surgeon Control<br>Probes<br>(Subject Device(s)) | Nicolet Disposable Prass<br>Probe / K873964<br>(Predicate Device) | Xomed Monopolar<br>Stimulator Probe /<br>K992869<br>(Predicate Device) |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | ETN | ETN | ETN |
| Regulation<br>Number | 21 CFR 874.1820 | 21 CFR 874.1820 | 21 CFR 874.1820 |
| Regulation<br>Description | Surgical nerve<br>stimulator/locator | Surgical nerve<br>stimulator/locator | Surgical nerve<br>stimulator/locator |
| Classification | Class II | Class II | Class II |
| Common Name | Nerve Stimulator | Nerve Stimulator | Nerve Stimulator |
| Device Description | The NIM™ surgeon<br>control probes carry<br>stimulation current from<br>the patient interface to<br>the patient. It also<br>enables the user to adjust<br>stimulation current and<br>key functions from the<br>surgical site. All probes<br>are single use devices. | The Prass Probes carry<br>stimulation current from<br>the console, via the<br>Patient Interface, to the<br>patient [...] The<br>incrementing prass probes<br>provide the surgeon with<br>the means to adjust the<br>stimulation current at the<br>surgical site. | The Monopolar Ball-Tip<br>Stimulating Probes<br>carry stimulation current<br>from the console, via the<br>Patient Interface, to the<br>patient [...] The<br>incrementing probes<br>provide the surgeon<br>with the means to adjust<br>the stimulation current<br>at the surgical site. |
| Intended Use | The NIM™ surgeon<br>control probes are<br>intended for use as<br>intraoperative nerve<br>stimulators | This device is intended<br>for use as an<br>intraoperative motor<br>nerve stimulator with the<br>Nerve Integrity Monitor<br>(NIM™) | To stimulate cranial and<br>peripheral motor nerves<br>for location and<br>identification during<br>surgery, including spinal<br>nerve roots |
| Indications for Use | The NIM™ surgeon<br>control probes are<br>intended to stimulate | This device is indicated<br>for intraoperative motor<br>nerve monitoring | The Ball-Tip Monopolar<br>Stimulating Probe is<br>intended to stimulate |
| Feature/Attribute | Surgeon Control<br>Probes<br>(Subject Device(s)) | Nicolet Disposable Prass<br>Probe / K873964<br>(Predicate Device) | Xomed Monopolar<br>Stimulator Probe /<br>K992869<br>(Predicate Device) |
| | cranial and peripheral<br>motor nerves for location<br>and identification during<br>surgery, including spinal<br>nerve roots | cranial and peripheral<br>motor nerves for location<br>and identification during<br>surgery, including spinal<br>nerve roots | cranial and peripheral<br>motor nerves for<br>location and<br>identification during<br>surgery, including spinal<br>nerve roots |
| Contraindications | The NIM™ surgeon<br>control probes are<br>contraindicated for use<br>with paralyzing<br>anesthetic agents when<br>monitoring a motor<br>nerve as these may<br>reduce or eliminate the<br>patient's EMG response. | The NIM 3.0 is<br>contraindicated for use<br>with paralyzing anesthetic<br>agents that will<br>significantly reduce, if<br>not completely eliminate,<br>EMG responses to direct<br>or passive nerve<br>stimulation. | The NIM 3.0 is<br>contraindicated for use<br>with paralyzing<br>anesthetic agents that<br>will significantly<br>reduce, if not<br>completely eliminate.<br>EMG responses to direct<br>or passive nerve<br>stimulation. |
| Operating<br>Principle | Electrical stimulation | Electrical stimulation | Electrical stimulation |
| Design | Offered in Prass-tip or<br>Ball-tip configurations | Prass-tip configuration | Ball-tip configuration |
| Probe tip (exposed<br>area) tissue<br>contact surface<br>geometry | Probe Prass Tip: Flat<br>•<br>Probe Ball Tip:<br>•<br>1mm ball | Flat (Prass) | 1mm ball |
| Probe working<br>length | Probe Prass: 3.76"<br>•<br>[9.6cm]<br>Probe Ball: 3.76"<br>•<br>[9.6cm] | 3.76" [9.6cm] | 3.76" [9.6cm] |
| Patient-contacting<br>Materials | 303 Stainless Steel with<br>Xylamed coating<br>insulation or Teflon<br>Polytetrafluoroethylene<br>(PTFE) coating<br>insulation | 303 Stainless Steel with<br>Xylamed coating<br>insulation | 303 Stainless Steel with<br>Teflon<br>Polytetrafluoroethylene<br>(PTFE) coating<br>insulation |
| Patient contact | Direct | Direct | Direct |
| Biocompatible | Yes | Not available | Yes |
| Sterile | Yes | Yes | Yes |
| Single-Use<br>Disposable | Yes | Yes | Yes |
| Duration of Use | Limited (<24 hours) | Limited (<24 hours) | Limited (≤24 hours) |
{5}------------------------------------------------
#### Substantial Equivalence Discussion
A comparison of technological characteristics was provided in the submission to establish substantial equivalence. Both the subject device(s) and the predicate device(s) share similar characteristics (design, materials, operating principle, energy source, and performance). The similar characteristics do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
#### Performance Testing Discussion
Design performance testing was completed to ensure the functionality and intended use of the NIM™ Surgeon Control Probes was met in accordance with external standards, and device specifications via pre-defined acceptance criteria. Bench engineering test samples were subjected to simulated real-life conditions during functional testing to establish baseline data and accelerated aging data. The testing included in this submission provides objective evidence through passing results that key technological characteristics such as the ones listed below are proven to be safe and effective. This performance testing was used to support substantial equivalence, proving the subject device(s) are as safe and effective as their predicate devices.
- . Electromechanical, dimensional, and visual design performance
- Sterilization validation to ISO 11135:2014
- Electrical safety & EMC testing to IEC 60601
- Biocompatibility testing and risk analysis to ISO 10993-1:2018
- Stability testing of proposed shelf life ●
- Packaging performance of environmental conditioning to ISTA 3A and distribution simulation to ASTM D4169
### VIII. CONCLUSION
### Conclusion
Utilizing FDA's Guidance for Industry and Food and Drug Administration Staff, "Format for Traditional and Abbreviated 510(k)s" issued on September 13, 2019, a comparison of key performance characteristics demonstrates that the subject device(s), the NIM™ Surgeon Control Probes do not raise different questions of safety and effectiveness compared to the predicate device(s). The NIM™ Surgeon Control Probes are proven to be as safe and effective as their legally marketed predicate devices. The design performance testing has also demonstrated that the NIM™ Surgeon Control Probes perform as well as the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.