The Mapping Suction Probe is a surgical instrument that allows the surgeon to remove secretions and test surgical tissue with nerve stimulation simultaneously and with the same instrument is intended for use only by a licensed physician and in conjunction with compatible nerve locator monitor systems.
Device Story
Mapping Suction Probe is a single-use, disposable surgical instrument combining a suction lumen and a monopolar stimulation probe. Used by surgeons and neurophysiologists in the OR to aspirate fluids/tissue during tumor resection while simultaneously identifying nervous structures to prevent damage. The device connects to external suction equipment via tubing and to external neuromonitoring systems via a standard connector cable. A counter electrode is placed at the resection margin to complete the circuit. The probe shaft is partially insulated with PTFE to ensure localized stimulation at the non-insulated tip. The device provides the surgeon with real-time feedback on nerve proximity through the connected monitor, facilitating safer surgical navigation and tissue preservation.
Clinical Evidence
No clinical data was performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993 series), electrical conductivity, dielectric strength, suction performance (vacuum/flow rate), mechanical stability (pull-out/bending stress), and shelf-life testing.
Technological Characteristics
Stainless-steel shaft, partially insulated with PTFE. Monopolar stimulation principle. Dimensions: 9cm or 13cm length; 6fg or 9fg lumen. Connectivity: DIN 42802 touch-proof connector for neuromonitoring; standard suction tube interface. Sterilization: Ethylene Oxide (ISO 11135). No software or internal power source.
Indications for Use
Indicated for use by licensed physicians during surgical procedures to remove secretions and simultaneously test surgical tissue via nerve stimulation in conjunction with compatible nerve locator monitor systems.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
November 11, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
inomed Medizintechnik GmbH Shuofei Cheng Regulatory Affairs Manager Im Hausgruen 29 Emmendingen. Baden-Wuerttemberg 79312 Germany
Re: K212164
Trade/Device Name: Mapping Suction Probe Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: September 29, 2022 Received: October 3, 2022
Dear Ms. Cheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Patrick Antkowiak -S
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212164
#### Device Name
Mapping Suction Probe 3/90 monopolar (Art. No. 525650), Mapping Suction Probe 3/130 monopolar (Art. No. 525651), Mapping Suction Probe 2/90 monopolar (Art. No. 525655), Mapping Suction Probe 2/130 monopolar (Art. No. 525656)
#### Indications for Use (Describe)
The Mapping Suction Probe is a surgical instrument that allows the surgeon to remove secretions and test surgical tissue with nerve stimulation simultaneously and with the same instrument is intended for use only by a licensed physician and in conjunction with compatible nerve locator monitor systems.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submission<br>Date: | 08 July 2021 | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------------------------------------------------------------|
| 510(k) Holder: | inomed Medizintechnik GmbH<br>Im Hausgrün 29<br>79312 Emmendingen, Germany | | |
| Contact Person: | Shuofei Cheng Ph.D.<br>Phone: +49 7641 9414 849<br>Email: S.Cheng@inomed.com | | |
| Manufacturing<br>Site: | inomed Medizintechnik GmbH<br>Im Hausgrün 29<br>79312 Emmendingen, Germany | | |
| Name of Device: | Mapping Suction Probe | | |
| Common and<br>Classification<br>Name: | Suction Stimulator Probe | | |
| Classification<br>Name: | Surgical nerve stimulator/locator (21 CFR 874.1820) | | |
| Regulatory<br>Class: | II | | |
| Product Code: | ETN | | |
| Regulation<br>Medical<br>Specialty: | Surgical nerve stimulator/locator | | |
| | | Predicate<br>510(k)<br>Number | Predicate Manufacturer/ Model |
| | Predicate Device: | K110712 | Neurovision™ Medical<br>Products, Inc.™./ DryTouch<br>Suction Stimulator Probe |
| Device<br>Description: | The Mapping Suction Probe is a surgical instrument which combines surgical suction instrument and a monopolar stimulation probe in one instrument | | |
a
{4}------------------------------------------------
| The stimulation function of the Mapping Suction Probe is used for | | | |
|---------------------------------------------------------------------------|--|--|--|
| monopolar stimulation during intraoperative surgical interventions and | | | |
| is connected to electrical stimulators of neuromonitoring devices outside | | | |
| the sterile field via an appropriate integrated connection cable. To | | | |
| ensure a correct monopolar stimulation, a counter electrode is also | | | |
| included in the sterile package which must be placed at the resection | | | |
| marqin. | | | |
The suction function of the Mapping Suction Probe is used for aspiration during tumor resection and is connected to suction devices via a tube, which is not part of the delivered package.
The Mapping Suction Probe is a surgical instrument that allows the Indications for surgeon to remove secretions and test surgical tissue with nerve Use: stimulation simultaneously and with the same instrument.
> The instrument is intended for use only by a licensed physician and in conjunction with compatible nerve locator monitor systems.
Comparison of Technological Characteristics with the Predicate Device:
| Characteristic | Subject device | Predicate device |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System | Mapping Suction Probe | DryTouch Suction Stimulator Probe |
| Manufacturer | inomed Medizintechnik GmbH | Neurovision Medical Produtcts, Inc. |
| 510(k)Number | K212164 | K110712 |
| Picture | Image: Mapping Suction Probe | Image: DryTouch Suction Stimulator Probe |
| Indications<br>for use | The Mapping Suction Probe is a surgical<br>instrument that allows the surgeon to<br>remove secretions and test surgical tissue<br>with nerve stimulation simultaneously and<br>with the same instrument.<br>The instrument is intended for use only by a<br>licensed physician and in conjunction with<br>compatible nerve locator monitor systems." | The Suction Stimulator Probe is a dedicated<br>manual surgical instrument that allows the<br>surgeon to clear secretions and test surgical<br>tissue with nerve stimulation at the same<br>time and with the same instrument.<br>It is intended for use only by a licensed<br>physician and in conjunction with the<br>Neurovision SE (Nerve; na) Nerve locator<br>Monitor System. |
| Contraindicati<br>ons | The use of muscle relaxants is<br>contraindicated, as these cause significant | The use of paralytic anaesthetics will cause<br>abnormal EMG performance. Advise the |
| | reductions in the evoked muscle potentials. | anaesthesia provider of this<br>contraindication. |
| User group | Surgeon and Neurophysiologist | Licensed physician |
| Risk<br>management | Risk Management according to standard:<br>ISO 14971: 2019 | Risk Management according to standard:<br>ISO 14971: 2007 |
| | • The device is a disposable surgical<br>instrument | • The device is a disposable surgical<br>instrument |
| | • The product combines a surgical<br>suction instrument and a<br>monopolar stimulation probe<br>simultaneously | • The product combines a surgical<br>suction instrument and a<br>monopolar stimulation probe<br>simultaneously |
| | • Monopolar stimulation on the non-<br>insulated tip of the suction probe<br>component to localize and identify<br>nervous structures and to prevent<br>their damage during surgery. | • Monopolar stimulation on the non-<br>insulated tip of the suction probe<br>component to localize and identify<br>nervous structures and to prevent<br>their damage during surgery. |
| Operating<br>Principle | • To ensure a specific stimulation at<br>the point of interest, the residual<br>metal part of the probe's shaft is<br>insulated. | • To ensure a specific stimulation at<br>the point of interest the residual<br>metal part of the probe's shaft is<br>insulated. |
| | • The connector on the end of the<br>suction probe component is for<br>mechanical connection to suction<br>devices via suction tubes. | • The connector on the end of the<br>suction probe component is for<br>mechanical connection to suction<br>devices via suction tubes. |
| | • Depending on the case, the<br>suction function is used for<br>aspiration during tumor resection<br>or to clear the surgical field from<br>fluids for better visualization. | • Depending on the case the suction<br>function is used for aspiration<br>during tumor resection or to clear<br>the surgical field from fluids for<br>better visualization. |
| Materials | The instrument consists of a stainless-<br>steel shaft, partially insulated with<br>biocompatible PTFE and ending with an<br>oblong hole suction. | The instrument consists of a stainless-<br>steel shaft, partially insulated with<br>biocompatible PTFE and ending with a<br>ball- tip suction. |
| Tissue<br>contact | The medical device comes in direct tissue<br>contact. | The medical device comes in direct tissue<br>contact. |
| Shelf life | 3 years | 1 year |
| Sterilization | • EO validation is compliant with<br>ISO 11135:2014 + A1:2018<br>• residuals analysis is compliant with<br>ISO 10993-7:2008<br>• The Sterility Assurance Level is<br>10-6<br>• Device is not labeled non-pyrogenic. | • EO validation is compliant with<br>ISO 11135-1:2007<br>• EO residuals analysis is compliant<br>with ISO 10993-7:2008<br>• The Sterility Assurance Level is 10-6<br>• Device is not labeled non-<br>pyrogenic. |
| Product length | • 9 cm<br>• 13 cm | • 13 cm |
| Inner<br>diameter<br>(Lumen) | 9fg 6fg | 6fg |
| Insulated tip<br>length | 2mm | 2mm |
| Accessories | 1 Subdermal Needle Electrode<br>(black lead wire) 3.0m wire, 20 mm<br>Needle-0.45mm gauge Connection Cable (red) 3.0m cable<br>length | 2 Neurovision™ Subdermal Needle<br>Electrodes (white and green lead<br>wire) 2.5m wire, 12 mm Needle-0.4<br>mm gauge, 27G 2 Alcohol Wipes |
| Site of<br>application | No specific site | No specific site |
| Duration of<br>use | < 30 days | < 30 days |
| Single use or<br>reusable | Single use | Single use |
| Provided sterile | Yes | Yes |
| Connector<br>standard | DIN 42802 | DIN 42802 |
| Connector | Touch Proof | Touch Proof |
{5}------------------------------------------------
## inomed fi
www.inomed.com
### K212164
# 510(k) Summary Mapping Suction Probe
FM 3-1
Page 3 of 7
Stand 1.1 vom 2017-11-02
{6}------------------------------------------------
### Performance Data:
Biocompatibility:
The Mapping Suction Probe and Needle Electrode component were tested according to the following standards:
- -ISO 10993-1:2018
- ISO 10993-5:2009 -
- ISO 10993-7:2008 .
- -ISO 10993-17:2002
- ISO 10993-18:2005 -
| Test | Test Performed | Results and Conclusion |
|--------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | Cell Growth<br>Analysis via XTT-Staining | The results showed no reduction of cell proliferation<br>and/or cell viability. With the highest extract<br>concentration (100 %) the dehydrogenase activity was<br>101 %.<br>Microscopically, no inhibition of cell growth and no cell<br>lysis were observed at all extract concentrations<br>used. |
{7}------------------------------------------------
| | | Conclusion: The item under test is considered non-<br>cytotoxic. |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensitization | Guinea Pig<br>Maximization Test<br>with polar and non-<br>polar extract | The sensitization rate after application of the test item<br>extract was 0%.<br>Conclusion: The item under test is considered to have<br>no sensitizing properties. |
| Irritation /<br>Intracutaneous<br>Reactivity | Irritation Test<br>(Intracutaneous<br>Reactivity) in the<br>Rabbit with polar<br>and non-polar<br>extract | The primary irritation index of the polar and the nonpolar<br>test item extracts compared to the controls were 0.<br>Conclusion: The item under test is classified as not<br>irritant. |
| Acute<br>Systemic<br>Toxicity | Acute Systemic<br>Toxicity in the<br>Mouse with polar<br>and non-polar<br>extract | No compound-related mortalities and no other signs of<br>toxicity were recorded within 72 hours post-dose for any<br>of the animals.<br>Conclusion: The item under test showed no acute<br>systemic toxic characteristics. |
| Material<br>Mediated<br>Pyrogenicity | Rabbit Pyrogen<br>Test according to<br>USP | Temperature increase for each rabbit was not > 0.50°C.<br>Conclusion: The materials of the item under test are<br>considered non pyrogenic. |
| Hemolysis | Direct and indirect<br>contact test under<br>static conditions<br>according to ASTM<br>F 756-17 | The data show that the haemolytic indices of the blood<br>substrate supernatants with direct (0.11%) and indirect<br>(0.08%) contact to the test item were below 2% and<br>therefore the material is classified as non-haemolytic.<br>Conclusion: The item under test does not induce<br>haemolysis. |
| Bacterial<br>Endotoxins | Limulus-<br>Amoebocyte-Lysate<br>(LAL) Test - Kinetic<br>Turbidimetric Assay<br>(KTA) | The endotoxin content of each test items was < 1.55<br>EU/test item.<br>Conclusion: The test items are non-pyrogenic under<br>consideration of the acceptance criterion (< 2.15<br>EU/device) |
The tests are conducted on the Device: '525651 Mapping Suction Probe 3/130, monopolar'.
The following tests were performed for the Needle Electrode component.
| Test | Test Performed | Results and Conclusion |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | Cell Growth<br>Analysis via XTT-<br>Staining | The results showed no relevant reduction (- relevant<br>meaning > 30 % reduction -) of cell proliferation<br>and/or cell viability. With the highest extract<br>concentration (100 %) the dehydrogenase activity was<br>reduced to 89 %.<br>Microscopically, slightly reduced cell growth was<br>observable from 66.7 % extract concentration<br>onwards. Slight cell lysis was observable only in the<br>undiluted extract.<br>Conclusion: The item under test is considered non-<br>cytotoxic. |
| Sensitization | Guinea Pig<br>Maximization Test<br>with polar and non-<br>polar extract | The sensitization rate after application of the test item<br>extract was 0%.<br>Conclusion: The item under test is considered to have<br>no sensitizing properties. |
| Irritation /<br>Intracutaneous<br>Reactivity | Irritation Test<br>(Intracutaneous<br>Reactivity) in the<br>Rabbit with polar<br>and non-polar<br>extract | The primary irritation index of the polar and the nonpolar<br>test item extracts compared to the controls were 0.<br>Conclusion: The item under test is classified as not<br>irritant. |
{8}------------------------------------------------
| Acute<br>Systemic<br>Toxicity | Acute Systemic<br>Toxicity in the<br>Mouse with polar<br>and non-polar<br>extract | No compound-related mortalities and no other signs of<br>toxicity were recorded within 72 hours post-dose for any<br>of the animals.<br>Conclusion: The item under test showed no acute<br>systemic toxic characteristics. |
|--------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material<br>Mediated<br>Pyrogenicity | Rabbit Pyrogen<br>Test according to<br>USP | Temperature increase for each rabbit was not > 0.50°C.<br>Conclusion: The materials of the item under test are<br>considered non pyrogenic. |
| Bacterial<br>Endotoxins | Limulus-<br>Amoebocyte-Lysate<br>(LAL) Test — Kinetic<br>Turbidimetric Assay<br>(KTA) | The endotoxin content of the test items was 0.6300,<br>0.7800 and 0.6600 EU/test item.<br>Conclusion: The test items are non-pyrogenic under<br>consideration of the acceptance criterion (< 20.0<br>EU/device) |
Test results indicate that the Mapping Suction Probe and the Needle Electrode component comply with the applicable standards and demonstrated the safety of subject device.
The Mapping Suction Probe does not contain any kind of software, Software: and therefore, this section does not apply to it.
The Mapping Suction Probe is a device which is not electrically-Electrical Safety: powered, and therefore, this section does not apply to it.
Electromagnetic The Mapping Suction Probe is a device which is not electrically-Compatibility: powered, and therefore, this section does not apply to it.
The Mapping Suction Probe was tested for performance in Performance Testing - Bench accordance with internal requirements and the following standards:
- IEC 60601-1:2005/AMD1:2012 -
- IEC 60601-1-2:2014 -
- IEC 60601-2-40:2016 -
- . Electrical Tests: Electrical Conductivity, Dielectric Strength Functional Tests: Impedance Measurement, Stimulation Current Test Dimension Test: Dimension of Components
- Suction Performance Tests: Reached Vacuum in Air, . Volumetric Flow Rate
- . Mechanical Suction Tests: Deformation Test, Leakage Test
- Mechanical Tests: Pull-out Force Test, Distractive Force Test, . Mechanical Stability Test, Bending Stress Test
- Insulation Test Shelf-Life Tests
The tests were conducted on the subject device Mapping Suction Probe and/or its component.
{9}------------------------------------------------
Performance Clinical testing was not performed for the Mapping Suction Probe. Testing – Clinical
- Conclusions The testing and assessments performed demonstrate that the subject device performs comparable to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.