The ES2® Awl, ES2® Taps, and ES2® Screwdriver are intended to be used with the ES2® Dilators & ES2® Tap Sleeve during neuromonitoring applications. The neuromonitoring accessory instrument set-up is also intended to be used for bone screw insertion under power.
Device Story
ES2® Neuromonitoring Accessory Instruments (Awl, Taps, Screwdriver) function as surgical nerve stimulators/locators; used in open or percutaneous minimally invasive posterior non-cervical spine surgery. Instruments are cannulated for K-Wire placement; manufactured from surgical grade stainless steel. During surgery, electrodes (clip or probe) attach to the instrument contact zone to deliver electrical stimulation from an external, FDA-cleared neuromonitoring console. The system provides proximity information to the surgeon to assist in nerve avoidance during bone preparation and screw insertion. The screwdriver may be used with or without power (corded or cordless). Output is visual/auditory feedback from the neuromonitoring console, allowing the surgeon to assess nerve proximity and adjust surgical trajectory, thereby reducing the risk of nerve injury.
Clinical Evidence
No human clinical data. Evidence consists of bench testing and a porcine animal study. Bench testing evaluated electrical safety (IEC 60601-1, IEC 60601-1-2) and signal integrity. Porcine study assessed functionality and safety of the powered screw insertion option (corded and cordless) during neuromonitoring. Results demonstrated that the powered configuration does not interfere with neuromonitoring signals and that the device is safe and effective for its intended use.
Technological Characteristics
Instruments: surgical grade stainless steel. Dilators/Tap Sleeves: RADEL®. Design: cannulated for K-wire. Stimulation: electrical via external console/electrodes. Connectivity: connects to external neuromonitoring consoles via clip or probe. Sterilization: reusable, validated for SAL 10^-6. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for use by surgeons to assist in locating spinal nerves by providing proximity information before, during, or after bone preparation and bone screw placement in open and percutaneous minimally invasive posterior surgical approaches of the non-cervical spine.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
### Traditional 510(k) #K140400
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
| Proprietary Name: | ES2® Neuromonitoring Accessory Instruments |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Surgical Nerve Stimulator/Locator |
| Classification Name and Reference: | 21 CFR §874.1820: Surgical Nerve Stimulator/Locator |
| Device Product Code: | ETN |
| Proposed Regulatory Class: | Class II |
| For Information contact: | Soraya King |
| | Regulatory Affairs Specialist |
| | 2 Pearl Court |
| | Allendale, NJ 07401 |
| | Telephone: (201) 760-8296 |
| | Fax: (201) 962-4296 |
| | Email: Soraya.King@Stryker.com |
| Date Summary Prepared: | June 17, 2014 |
| Predicate Devices | NuVasive® NVM5 System - K112718 NuVasive® NVM5 System - K123307 Stryker Spine ES2® Spinal System - K122845 |
| Device Description | The ES2® Neuromonitoring instruments (Awl, Taps, and Screwdriver) are accessory devices to be used with FDA cleared neuromonitoring systems to deliver electrical stimulation to assist in location of the spinal nerves during intraoperative neurological monitoring of the non-cervical spine in open and percutaneous minimally invasive posterior surgical approaches. The instruments are manufactured from surgical grade stainless steel and are provided non-sterile.<br><br>The neuromonitoring application is a surgical option that allows the surgeon to locate the spinal nerves by providing proximity information during bone preparation and placement/insertion of bone screws. The ES2® Awl and ES2® Taps facilitate bone preparation, and the ES2® Screwdriver facilitates bone screw placement/insertion. The surgical accessories are compatible with commercially available FDA cleared neuromonitoring consoles/systems and associated electrodes. The nerves are stimulated using electrodes attached to the subject accessory devices. The neuromonitoring accessory instrument can be used with or without a |
| Indication for Use | The ES2® Awl, ES2® Taps, and ES2® Screwdriver can be<br>used by the surgeon to assist in location of the spinal nerves by<br>providing proximity information before, during or after bone<br>preparation and placement of bone screws. |
| Intended Use | The ES2® Awl, ES2® Taps, and ES2® Screwdriver are<br>intended to be used with the ES2® Dilators & ES2® Tap<br>Sleeve during neuromonitoring applications. The<br>neuromonitoring accessory instrument set-up is also intended<br>to be used for bone screw insertion under power. |
| Summary of the Technological<br>Characteristics | The Stryker Spine ES2® Neuromonitoring Accessory Instruments are<br>substantially equivalent to the predicate devices in terms of design,<br>function, principals of operation, technological characteristics, and<br>indications and intended uses. As compared to the NuVasive®<br>NVM5® predicates, the ES2® Awl, ES2®Taps, and ES2®<br>Screwdriver are cannulated to allow for K-Wire placement during<br>percutaneous minimally invasive surgical approaches. The predicate<br>NuVasive® NVM5® instruments and the ES2® Neuromonitoring<br>Accessories are used with dilators and/or a tap sleeve to provide<br>insulation. The electrical signal for both the NuVasive® NVM5®<br>and ES2® instruments are supplied via electrodes (clip or probe)<br>that are attached to the neuromonitoring contact zone of the awls,<br>taps, or screwdrivers.<br><br>In addition, the ES2® Screwdriver can be used with or without<br>power for bone screw insertion. As FDA cleared in 510(k)<br>#K122845, the power supply can be corded or cordless. Bench<br>testing demonstrated that the ES2® Screwdriver can deliver safe and<br>effective neuromonitoring signals using the non-powered or<br>powered (corded and cordless) bone screw placement/insertion<br>options. |
# Section 008: 510(k) Summary
-
{1}------------------------------------------------
Traditional 510(k) #K140400
,
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
{2}------------------------------------------------
raditional 510(k) #K140400
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
# OMPARISON OF ES2® NEUROMONITORING ACCESSORY INSTRUMENT AND THE PREDICATE DEVICE FOR ISSOCIETY
| Characteristic | Subject Device:<br>ES2® Neuromonitoring<br>Accessory Instruments | Predicate Device:<br>NuVasive® NVM5® System | Substantial Equivalence |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Neuromonitoring Accessories<br>Instruments | Awl, Taps, and Screwdriver | Taps and Screwdriver | Yes - The ES2® and predicate<br>system use the same types of<br>accessory instruments for<br>neuromonitoring applications during<br>bone preparation/pilot hole starter<br>and bone screw placement. The<br>ES2® Neuromonitoring Accessory<br>Instruments are employed and<br>assembled in a similar manner with<br>comparable set-up configurations to<br>deliver electrical signals. Testing<br>confirmed that the powered ES2®<br>Screwdriver configuration does not<br>interfere with the neuromonitoring<br>signals and is safe and effective. |
| Use of Dilators | Dilators or Tap Sleeve | Dilators | Yes |
| Compatible with Common<br>Neuromonitoring Consoles &<br>Software | Compatible with FDA cleared<br>neuromonitoring systems to<br>include the NuVasive® NVM5®<br>System | NuVasive® NVM5® System | Yes |
| Connection to Neuromonitoring<br>unit | Clip or Probe (based on<br>Neuromonitoring System used) | Clip | Yes - the use of a clip or probe is<br>dependent on the neuromonitoring<br>console and/or surgeon's<br>preference. The clip or probe are<br>attached to the accessory instrument<br>on the exposed instrument contact<br>area above the dilator or tap sleeve<br>for both the ES2® and predicate<br>devices |
Page 3 of 9
{3}------------------------------------------------
| 400<br>0<br>P<br>- |
|--------------------|
| ਮ<br>:47 |
| 1<br>ਮ |
| ી(<br>l |
| ్ర |
| aditional<br>Tre |
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
.
| Materials | Indication for Use | Unknown |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Awl, Taps, & Screwdriver:<br>Surgical Grade Stainless Steel<br>Dilators & Tap Sleeve: RADEL®<br><br>The ES2® Awls, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and percutaneous posterior surgical approaches of the non-cervical spine. | The NVM5® System is a medical device that is intended for intraoperative neurophysiologic monitoring during spinal surgery. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NVM5® provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), transcranial motor evoked potential (TceMEP) or somatosensory evoked potential (SSEP) responses of nerves. XLIF (Detection) - The XLIF (Detection) function allows the surgeon to locate and evaluate spinal nerves, and is used as a nerve avoidance tool. Basic & Dynamic Screw Test – The Screw Test functions allow the surgeon to locate and evaluate the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws. Free Run EMG – The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions. Twitch Test (Train of Four) – The Twitch Test Function allows the | Yes – Instruments composed of well-characterized materials and accepted biocompatible materials.<br><br>Yes – the indications of the ES2® Neuromonitoring Accessory Instruments are a subset of the indications of the predicate device. This does not result in a new or different intended use as compared to the predicate. The subject and predicate instruments are employed in the same manner when used as a tool to assist the surgeon in locating spinal nerves before, during, or after bone preparation and bone screw placement for open and percutaneous surgical approaches. |
Page 4 of 9
.
:
·
.
{4}------------------------------------------------
raditional 510(k) #K140400
Stryker Spinc ES2® Spinal System Neuromonitoring Accessory Instruments
.
| | | surgeon to assess moderate<br>degrees of neuromuscular block<br>in effect by evaluating muscle<br>contraction following a train of<br>four stimulation pulses.<br>● TCMEP – Transcranial<br>stimulation techniques for motor<br>evoked potentials are used to<br>assess for acute dysfunction in<br>axonal conduction of the<br>corticospinal tract. The TCMEP<br>function provides an adjunctive<br>method to allow the surgeon to<br>monitor spinal cord motor<br>pathway integrity during<br>procedures with a risk of<br>surgically induced motor injury.<br>● SSEP – The SSEP function<br>allows the surgeon to assess<br>sensory spinal cord function in<br>surgical procedures during which<br>the spinal cord is at risk.<br>● Remote Reader – The Remote<br>Reader function provides real<br>time remote access to the<br>NVM5® System for monitoring<br>physician outside of the operating<br>room.<br>● The Guidance function is<br>intended as an aid for use in either<br>open or percutaneous pedicle<br>cannulation producedure in the<br>lumbar spine of adult patients,<br>and when used in conjunction<br>with radiographic imaging<br>anatomy for the creation of a<br>cannulation trajectory for bone<br>screw placement. |
|---------------|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | Instruments provided as reusable | Yes<br>As selected for individual accessories. |
Page 5 of 9
{5}------------------------------------------------
Fraditional 510(k) #K140400
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
| | non-sterile devices with validated<br>sterilization parameters to assure a<br>SAL of 10-6 | and validated to assure a SAL of 10-6 | |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Surgical Approach | Open or Percutaneous/Minimally<br>Invasive | Open or Percutaneous/Minimally<br>Invasive | Yes |
| Electromagnetic Compatibility &<br>Electrical Safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC 61000-3-2<br>IEC 61000-3-3 | IEC 60601-1-2 | Yes |
| Min. exposed surface area during<br>tissue stimulation | 0.53mm² | unknown | Yes - The test data demonstrated<br>that the design of the subject device,<br>the exposed surface area during<br>tissue stimulation, and<br>neuromonitoring instrument set-<br>up/configurations do not impact or<br>interfere with the electrical<br>signaling and are comparable to the<br>predicate devices. |
Page 6 of 9
,
.
.
)
:
{6}------------------------------------------------
# raditional 510(k) #K140400
।
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
:
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
:
・
·
-
:
.
{7}------------------------------------------------
| Summary of the Performance Data | This 510(k) pre-market notification seeks expanded indications for<br>the ES2® Awl, ES2® Taps, and ES2® Screwdriver to be used as<br>accessory instruments during neuromonitoring applications. No<br>design modifications, or changes in the materials of construction<br>were made to the ES2® Awl, ES2® Taps, or ES2® Screwdriver, as<br>previously presented in 510(k) #K122845, to facilitate<br>neuromonitoring. All instruments are manufactured from surgical<br>grade stainless steel. The materials of construction have been<br>well characterized and shown to have stable chemical and<br>mechanical properties that are not affected by aging or storage<br>conditions.<br><br>Performance testing was performed to demonstrate that the subject<br>devices are substantially equivalent to the identified predicate<br>instruments in terms of design, performance and intended use. The<br>ES2® Neuromonitoring Accessory Instruments were tested for<br>electrical safety in accordance with IEC 60601-1-2 for EMC and<br>Safety. The instruments were also evaluated as per IEC 60601-1. A<br>porcine animal study was conducted to assess the performance,<br>functionality, and safety of the ES2® Neuromonitoring Accessory<br>Instruments and implants utilizing the powered screw insertion<br>option, corded and cordless. Bench test data and assessment<br>confirmed substantial equivalence to the NuVasive® NVM5®<br>predicates, and the safety and efficacy of the devices.<br><br>Laboratory tests were conducted in compliance with applicable<br>Good Laboratory Practices (GLP) requirements stipulated in 21 CFR<br>Part 58. |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the information and comparison tables included in this<br>510(k), the ES2® Neuromonitoring Accessory Instruments are<br>substantially equivalent to the NuVasive® NVM5® predicates. The<br>instruments are employed in the same manner, have similar intended |
| | and indications for use, principles of operation, and technological<br>characteristics. The ES2® subject devices met all required<br>acceptance criteria and did not create new safety or efficacy<br>concerns. |
{8}------------------------------------------------
### Traditional 510(k) #K140400
:
·
:
・
1
Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments
ن
r
.
.
:
!
バ
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains a logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is in black and white.
Public Health Service
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
July 1, 2014
Stryker Spine Sorava King Regulatory Affairs Specialist 2 Pearl Court Allendale. NJ 07401
Re: K140400
Trade/Device Name: ES2® Neuromonitoring Accessory Instruments Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: April 1. 2014 Received: April 2. 2014
Dear Ms. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I'DA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{10}------------------------------------------------
Page 2 - Ms. Soraya King
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Felipe Aguel -S
Carlos L. Peña, Ph.D., M.S. for
Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) KI40400
### Device Name
ES2® Neuromonitoring Accessory Instruments: ES20 Awl, ES20 Taps, and ES2® Screwdriver
Indications for Use (Describe)
The ES2® Awl, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and perculaneous minimally invasive posterior surgical approaches of the non-cervical spine,
### Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Felipe Aquel -S
Date: 2014.07.01 14:20:03 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.