IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identify nerves and spinal nerve roots and to assess nerve function. IOM Stimulator Probes disposable surgical stimulators are indicated for tissue dissection and stimulation of cranial and peripheral nerves for location and identification during surgery, including spinal nerve roots.
Device Story
IOM Stimulator Probes are disposable, sterile surgical instruments used by surgeons during intraoperative neurological monitoring. The device consists of a stainless steel tip (AISI 316L or 304) for patient contact, a biocompatible plastic handle, and a flexible lead wire with a 'touch-proof' safety connector. The probe is connected to an external electrical stimulator; it delivers electrical current to tissue to identify cranial and peripheral nerves and spinal nerve roots and to assess nerve function. The device is used in an operating room setting. By facilitating nerve location and identification, the probe assists the surgeon in preserving nerve integrity during surgical procedures. The device is single-use and sterilized via Ethylene Oxide.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility, dielectric strength, packaging integrity, and sterilization validation.
Technological Characteristics
Materials: Stainless steel (AISI 316L/304) tips, biocompatible polystyrene handles, medical PVC cable insulation, PTFE or heat-shrinkable tube insulation. Connectivity: Touch-proof safety connector (DIN 42802). Sterilization: Ethylene Oxide. Form factor: Monopolar, bipolar, and tripolar configurations with varying tip geometries (ball, hook, minifork).
Indications for Use
Indicated for tissue dissection and stimulation of cranial and peripheral nerves for location and identification during surgery, including spinal nerve roots, in patients undergoing intraoperative neurological monitoring.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting movement or connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 21, 2014
Giorgio Facco Quality Assurance and Regulatory Affairs Spes Medica s.r.1. Via Europa – Zona Industriale Battipaglia (SA), Italy 84091
Re: K133348
Trade/Device Name: IOM Stimulator Probes Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: II Product Code: ETN Dated: October 9, 2014 Received: October 23, 2014
Dear Mr. Facco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K133348
Device Name IOM Stimulator Probes
#### Indications for Use (Describe)
IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identitY nerves and spinal nerve roots and to assess nerve function. 10M Stimulator Probes disposable surgical stimulators are indicated for tissue dissection and stimulation of cranial and peripheral nerves for location and identification during surgery, including spinal nerve roots.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------|
| | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Spes Medica. The logo features a globe on the left side and the words "spes medica" in a stylized font on the right side. Below the words "spes medica" is the tagline "the medical accessories source" in a smaller font.
## 510 (k) Summary of safety and effectiveness
#### SUBMITTER INFORMATION
- A. Company Name:
- B. Company Address:
- C. Company Phone: Company Fax:
- D. Contact Person:
Spes Medica s.r.l. Via Europa - Zona industriale Battipaglia (SA) – Italy 84091 +39 010 39 03 43 +39 010 30 72 345 Giorgio Facco Quality Assurance Spes Medica s.r.l. e-mail: quality@spesmedica.com November 04th 2014
E. Date Summary Prepared:
#### DEVICE IDENTIFICATION
| A. Device name: | IOM Stimulator Probe |
|----------------------------|-------------------------------------------------------|
| B. Trade/Proprietary name: | Stim Probe |
| C. Classification name: | Surgical nerve stimulator/locator (21 CFR §874.1820) |
| D. Product code: | ETN |
#### LEGALLY MARKETED DEVICES (PREDICATE DEVICES)
| Predicate device | Device Classification<br>Name | 510(k) Applicant | 510(k) number |
|-----------------------------------------------------|-------------------------------|-----------------------------------------------------|---------------|
| RHYTHMLINK Monopolar<br>stimulating instrument | Stimulator Nerve | RHYTHMLINK<br>INTERNATIONAL, LLC | K112435 |
| Drytouch Suction Stimulator probe | Stimulator Nerve | NEUROVISION MEDICAL<br>PRODUCTS, INC. | K110712 |
| XIAN FRIENDSHIP MEDICAL nerve<br>stimulator probes | Stimulator Nerve | XIAN FRIENDSHIP<br>MEDICAL ELECTRONICS<br>CO., LTD. | K112426 |
| SURGICAL STIMULATORS | Stimulator Nerve | TECHNOMED EUROPE | K110422 |
| STIMULUS/DISSECTION<br>INSTRUMENTS, BALL-TIP PROBES | Stimulator Nerve | MEDTRONIC XOMED,<br>INC | K031003 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for "spes medica". The logo features a stylized globe on the left side, with the company name "spes medica" written in a cursive font to the right of the globe. Below the company name, there is a tagline that reads "the medical accessories source" in a smaller font. The overall design is simple and professional, suggesting a company that is involved in the medical accessories industry.
#### DEVICE DESCRIPTION
IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identify nerves and spinal nerve roots and to assess nerve function. The probes are supplied sterile and are for single use only. The probes are connected to an electrical stimulator using a flexible lead wire(s) and a 'touch-proof' safety connector(s) on the distal end.
All IOM Stimulator Probes are sterilized by Ethylene Oxide and single use only.
The IOM Stimulator Probes are composed by the part in contact with the patient that is medical grade stainless steels, then there is the handle part and cable in biocompatible plastic; finally the cable has a 'touch-proof' safety connectors.
#### INTENDED USE
IOM Stimulator Probes are used by the surgeon as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring, to identify nerves and spinal nerve roots and to assess nerve function. IOM Stimulator Probes disposable surgical stimulators are indicated for tissue dissection and stimulation of cranial and peripheral nerves for location and identification during surgery, including spinal nerve roots.
{5}------------------------------------------------
IOM Stimulator Probes
### SUBSTANTIAL EQUIVALENCE
The IOM Stimulator Probes are similar in interials, packaging and other technological characteristics to the predicate devices.
| Comparison Chart IOM Stimulator Probes | | | | | | | | | | | | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------------|----------------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------|-----------------------------|---------------|---------------|--------------------------------------------------------------------|
| Predicate<br>device | 510(k)<br>Applicant | 510(k)<br>number | Intent for use | Tip<br>Diam.<br>mm | Shaft<br>Length<br>mm* | Material<br>In contact<br>with patient | Touch<br>proof | Bend Angle | Wire Length<br>cm | Single<br>Use | Steri<br>liz. | Ref.<br>Document |
| RHYTHMLINK<br>Monopolar<br>stimulating<br>instrument<br>(Legally<br>Marketed) | RHYTHMLINK<br>INTERNATION<br>AL, LLC | K112435 | Rhythmlink<br>International<br>Monopolar-<br>Stimulating<br>Instruments Is<br>indicated for<br>stimulation of cranial<br>and peripheral motor<br>nerves for location<br>and identification<br>during surgery;<br>including spinal<br>nerve roots. The<br>Rhythmlink<br>International<br>Monopolar -<br>Stimulating<br>Instruments is sterile<br>and for single use . | 3<br>2.3<br>1.6<br>0.75<br>Stim<br>Surface<br>mm³ | 100<br>160 | Stainless<br>steel | YES | NO | 150<br>250 | YES | EtO | 004_<br>Section 09<br>Rhythmlin<br>k |
| Drytouch<br>Suction<br>Stimulator<br>probe<br>(Legally<br>Marketed) | NEUROVISION<br>MEDICAL<br>PRODUCTS,<br>INC. | K110712 | The Suction<br>Stimulator Probe is a<br>dedicated manual<br>surgical instrument<br>that allows the<br>surgeon to clear<br>secretions and test<br>surgical tissue with<br>nerve stimulation at<br>the same time and<br>with the same<br>instrument. It is<br>intended for use only<br>by a licensed<br>physician and in<br>conjunction with the<br>Neurovision SE<br>(Nerveäna) Nerve<br>locator Monitor<br>System. | 2<br>Stim<br>Surface<br>mm³<br>8 | 130<br>260 | Stainless<br>steel | YES | NO | 150 | YES | EtO | 003_<br>Section 09<br>Neurovisio<br>n |
| XIAN<br>FRIENDSHIP<br>MEDICAL<br>nerve<br>stimulator<br>probes<br>(Legally<br>Marketed) | XIAN<br>FRIENDSHIP<br>MEDICAL<br>ELECTRONICS<br>CO., LTD. | K112426 | Xian Friendship<br>Disposable Nerve<br>Stimulator Probes is<br>used to perform<br>localized stimulation<br>of neural tissue and<br>to locate, identify<br>and monitor cranial<br>motor nerves,<br>peripheral nerves<br>and spinal nerve<br>roots during surgery. | 2.3<br>1.3<br>0.8<br>1.2<br>Stim<br>Surface<br>mm³ | 150÷22<br>0<br>80÷150<br>80÷120<br>70÷120 | Stainless<br>steel | YES | 90° (double<br>Hook Probe)<br>100° (double<br>and triple<br>Hook Probe)<br>180° (double<br>and triple<br>Hook Probe) | Information<br>not findable | YES | EtO | 006_<br>Section 09<br>Xian<br>Friendship<br>Medical<br>Electronics |
| | | | Xian Friendship<br>Disposable Nerve<br>Stimulator Probes<br>are sterile (EtO),<br>single-patient-use<br>device. | $0.5÷100$ | | | | | | | | |
| SURGICAL<br>STIMULATORS<br>(Legally<br>Marketed) | TECHNOMED<br>EUROPE | K110422 | The Technomed<br>Europe Kartush<br>disposable surgical<br>stimulators are<br>indicated for tissue<br>dissection and<br>stimulation of cranial<br>and peripheral motor<br>nerves for location<br>and identification<br>during surgery,<br>including spinal<br>nerve roots. | 0.5<br>Stim<br>Surface<br>mm3<br>0.125 | 38,70,1<br>00 | Stainless<br>steel | YES | NO | Information<br>not findable | YES | EtO | 005_<br>Section 09<br>Technome<br>d Europe |
| STIMULUS/DIS<br>SECTION<br>INSTRUMENTS<br>, BALL-TIP<br>PROBES<br>(Legally<br>Marketed) | MEDTRONIC<br>XOMED, INC | K031003 | The Stimulus-<br>Dissection<br>Instruments are<br>indicated for tissue<br>dissection and<br>stimulation of cranial<br>and peripheral motor<br>nerves for location<br>and identification<br>during surgery,<br>including spinal<br>nerve roots. | 0.5<br>1.3<br>1<br>2.3<br>Stim<br>Surface<br>mm3<br>$0.125÷12$ | / | Stainless<br>steel | YES | NO | Information<br>not findable | YES | EtO | 002_<br>Section 09<br>Medtronic |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for "spes medica the medical accessories source". The logo features a globe on the left side and the company name on the right side. The text is in a stylized font.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for "spes medica the medical accessories source". The logo features a globe on the left side and the company name on the right side. The globe is a wireframe design, and the company name is written in a cursive font. The text "the medical accessories source" is written in a smaller font below the company name.
IOM Stimulator Probes
*The information about Handle length in not avaible for any models of any predicates device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Spes Medica. The logo features a globe on the left side and the company name "spes medica" on the right side. Below the company name, it says "the medical accessories source" in a smaller font.
IOM Stimulator Probes
# IOM Stimulator Probes Comparison
| Spes Medica Product<br>Name and Drawing | Comparison | Spes Medica Description | Predicate device Photo |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Neural stimulator probe<br>insulated monopolar<br>Image: Neural stimulator probe insulated monopolar | The materials is the same,<br>stainless steel, and the probe<br>diameter is in the same range:<br>1.2 mm for Spes Medica and<br>1.5 mm for Predicate Device<br>(PD). The shaft length is<br>longer in the PD compared to<br>Spes Medica. The cable length<br>is the same 250 cm (PD has<br>150 cm too). Both the<br>products as Touch proof<br>connection . Both products<br>are sterilized with EtO . | The probe is a monopolar stimulator probe, It is made by<br>stainless steel AISI 316L, It is insulated with termosrhinkable<br>tube with an insulating thickness of 0,25 mm (KBM100). The<br>part in contact with the patient is the tip of the probe (AISI<br>316L), the diameter is 1,2 mm and the exposure tip (uncoated<br>surface) is 2 mm . The probe length is 45 mm.<br>The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and<br>lead wire. After the polystyrene there is black<br>thermosrhinkable (18 mm) tube to protect the cable. The cable<br>insulation is made by medical PVC (diameter 1,2 mm). The<br>connector is a coloured T.P. DIN 42802 . The device is EO<br>sterilized and for only single use | Image: Standard Monopolar Stimulating Probe<br>Standard Monopolar Stimulating Probe<br>Rhythmlink |
| Neural stimulator probe<br>bipolar hook<br>Image: Neural stimulator probe bipolar hook | The materials is the same,<br>stainless steel, and the probe<br>diameter is 1.2 mm for PD and<br>0.65 mm for Spes Medica.<br>The shaft length is longer in<br>the PD (70÷120mm)<br>compared to Spes Medica (20<br>mm). Both the products as<br>Touch proof connection . Both<br>products are sterilized with<br>EtO . | The probe is a bipolar hook stimulator probe, It is made by<br>Stainless steel AISI 304, It is PTFE insulated with insulating<br>thikness 0,010-0,015 mm (Duraskin). The part in contact with<br>the patient are the tips of the probe (AISI 304), the diameter is<br>0,65 mm and the exposure tip (uncoated surface) is 6 mm with<br>a radius R 6 mm. The distance between the two probes is 9 mm<br>(center-center). The probes length are 20 mm. The body of the<br>probe is made by coloured polystyrene, It contains the<br>connections (Tin soldering) between probes and lead wires.<br>After the polystyrene there is black thermosrhinkable (18 mm)<br>tube to protect the cables. The intertwisted cables insulation is<br>made by medical PVC (diameter 1,2 mm). The connector is a<br>coloured T.P. DIN 42802 . The device is EO sterilized and for<br>only single use | Image: Disposable Double Hook Nerve Stimulator Probe-180 degree<br>Disposable Double Hook Nerve Stimulator Probe-180 degree<br>Xian Friendship Medical Electronics |
| Neural stimulator probe<br>tripolar hook<br>Image: Neural stimulator probe tripolar hook | The materials is the same,<br>stainless steel, and the probe<br>diameter is 1.2 mm for PD and | The probe is a bipolar hook stimulator probe, It is made by<br>Stainless steel AISI 304, It is PTFE insulated with insulating<br>thikness 0,010-0,015 mm (Duraskin). The part in contact with | Image: Disposable Triple Hook Nerve Stimulator Probe-180 degree<br>Disposable Triple Hook Nerve Stimulator Probe-180 degree |
| | 0.65 mm for Spes Medica.<br>The shaft length is longer in<br>the<br>PD (70÷120mm)<br>compared to Spes Medica (20<br>mm). Both the products as<br>Touch proof connection. Both<br>products are sterilized with<br>EtO. | the patient are the tips of the probe (AISI 304), the diameter is<br>0,65 mm and the exposure tip (uncoated surface) is 6mm with<br>a radius R 6 mm. The distance between the three probes is 4,5<br>mm (center-center each). The probes length are 20 mm. The<br>body of the probe is made by coloured polystyrene, It contains<br>the connections (Tin soldering) between probes and lead wires.<br>After the polystyrene there is black thermoshrinkable (18 mm)<br>tube to protect the cables. The intertwisted cables insulation is<br>made by medical PVC (diameter 1,2 mm). The connector is a<br>coloured T.P. DIN 42802. The device is EO sterilized and for<br>only single use | Xian Friendship Medical Electronics |
| Neural stimulator probe<br>monopolar ball | The materials is the same,<br>stainless steel, and the probe<br>diameter is 3 mm for PD and<br>2 mm for Spes Medica. The<br>shaft length is longer in the<br>PD (100 mm) compared to<br>Spes Medica (30 mm). The<br>cable length is the same 250<br>cm (PD has 150 cm too). Both<br>the products as Touch proof<br>connection. Both products<br>are sterilized with EtO. | The probe is a monopolar ball stimulator probe, It is made by<br>stainless steel AISI 316L, It is insulated with thermoshrinkable<br>tube with an insulating thickness 0,25 mm (KBM100). The part<br>in contact with the patient is the tip of the probe (AISI 316L),<br>the diameter is 1,7 mm, the ball (exposure tip - uncoated<br>surface) has a diameter of 2 mm. The probe length is 30 mm.<br>The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and lead<br>wire. After the polystyrene there is black<br>thermoshrinkable (18 mm) tube to protect the cable. The cable<br>insulation is made by medical PVC (diameter 1,2 mm). The<br>connector is a coloured T.P. DIN 42802.<br>The device is EO sterilized and for only single use | Disposable Ball Tip Direct Nerve Stimulator Probe<br>Xian Friendship Medical Electronics |
| Neural stimulator probe<br>monopolar ball | The materials is the same,<br>stainless steel, and the probe<br>diameter is 3 mm for PD and<br>2 mm for Spes Medica. The<br>shaft length is longer in the<br>PD (100 mm) compared to<br>Spes Medica (45 mm). The<br>cable length is the same 250<br>cm (PD has 150 cm too). Both | The probe is a monopolar ball stimulator probe, It is made by<br>stainless steel AISI 316L, It is insulated with thermoshrinkable<br>tube with an insulating thickness 0,25 mm (KBM100). The part<br>in contact with the patient is the tip of the probe (AISI 316L),<br>the diameter is 1,7 mm, the ball (exposure tip - uncoated<br>surface) has a diameter of 2 mm. The probe length is 45 mm.<br>The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and lead<br>wire. After the polystyrene there is black<br>thermoshrinkable | Disposable Ball Tip Direct Nerve Stimulator Probe |
| | the products as Touch proof<br>connection . Both products<br>are sterilized with EtO . | thermosrhinkable (18 mm) tube to protect the cable. The cable<br>insulation is made by medical PVC (diameter 1,2 mm). The<br>connector is a coloured T.P. DIN 42802 . The device is EO<br>sterilized and for only single use | |
| Neural stimulator probe<br>bipolar balls<br>Image: [Neural stimulator probe bipolar balls] | The materials is the same,<br>stainless steel, and the probe<br>diameter is the same for PD<br>and Spes Medica (2 mm). The<br>shaft length is longer in the<br>PD ( $70\div120$ mm) compared to<br>Spes Medica (30 mm).<br>Both the products as Touch<br>proof connection . Both<br>products are sterilized with<br>EtO . | The probe is a bipolar balls stimulator probe, It is made by<br>stainless steel AISI 316L, It is insulated with termosrhinkable<br>tube with an insulating thickness 0,25 mm (KBM100). The part<br>in contact with the patient is the tip of the probe (AISI 316L),<br>the diameter is 1,7 mm, the balls (exposure tips - uncoated<br>surfaces) have a diameter of 2 mm . The distance between the<br>two probes is 6,5 mm (center-center). The probe length is 30<br>mm.<br>The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and<br>lead wire. After the polystyrene there is black<br>thermosrhinkable (18 mm) tube to protect the cables. The<br>cable insulation is made by medical PVC (diameter 1,2 mm).<br>The connector is a coloured T.P. DIN 42802 .<br>The device is EO sterilized and for only single use | Image: [Disposable Double Ball Tip Direct Nerve Stimulator Probe]<br>Xian Friendship Medical Electronics |
| Neural stimulator probe<br>bipolar concentric<br>Image: [Neural stimulator probe bipolar concentric] | The materials is the same,<br>stainless steel, and the probe<br>diameter is 0.8 mm for PD and<br>0.32 mm for Spes Medica<br>The shaft length is longer in<br>the PD ( $80\div150$ mm)<br>compared to Spes Medica (45 | The probe is a bipolar concentric stimulator probe, It is made<br>by stainless steel AISI 304 canula with internal sensor in<br>stainless steel AISI 304, It is insulated with PTFE with an<br>insulating thickness 0,010-0,015 mm (Duraskin). The part in<br>contact with the patient is AISI 304, the diameter of canula is<br>1,5 mm, the sensor tip has a diameter of 0,32 mm and a<br>length of 0.3 mm. The probe length is 45mm. | Image: [Disposable Concentric Direct Nerve Stimulator Probe]<br>Xian Friendship Medical Electronics |
| | mm).<br>Both the products as Touch<br>proof connection . Both<br>products are sterilized with<br>EtO . | The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and<br>lead wire. After the polystyrene there is black<br>thermosrhinkable (18 mm) tube to protect the cables. The<br>cable insulation is made by medical PVC (diameter 1,2 mm).<br>The connector is a coloured T.P. DIN 42802.<br>The device is EO sterilized and for only single use | Image: Xian Friendship Medical Electronics |
| | | | Xian Friendship Medical Electronics |
| Image: Neural stimulator probe bipolar minifork | The materials is the same,<br>stainless steel, and the probe<br>diameter is 1.3 mm for PD and<br>0.65 mm for Spes Medica.<br><br>Both the products as Touch<br>proof connection . Both<br>products are sterilized with<br>EtO . | The probe is a bipolar minifork stimulator probe, It is made by<br>stainless steel AISI 304, It is insulated with PTFE with an<br>insulating thickness0,010-0,015 mm (Duraskin). The diameter<br>are 0,65 mm, the exposure tips (uncoated surface) have a<br>length of 2 mm . The distance between the two probes is 1,5<br>mm (center-center). The probe length is 30 mm.<br>The body of the probe is made by coloured polystyrene, It<br>contains the connection (Tin soldering) between probe and<br>lead wire. After the polystyrene there is black<br>thermosrhinkable (18 mm) tube to protect the cables. The<br>cable insulation is made by medical PVC (diameter 1,2 mm).<br>The connector is a coloured T.P. DIN 42802<br>The device is EO sterilized and for only single use | Image: Kartush Side-by-Side Bipolar Stimulator Probe<br><br>Medtronic |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo for "spes medica, the medical accessories source". The logo includes a stylized globe on the left side. The text is in a serif font and is slightly stylized.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Spes Medica. The logo features a globe on the left side and the words "spes medica" on the right side. Below the words is the phrase "the medical accessories source" in a smaller font.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Spes Medica. The logo consists of a globe on the left and the words "spes medica" on the right. Below the words is the phrase "the medical accessories source".
{12}------------------------------------------------
Image /page/12/Picture/2 description: The image shows a logo for "spes medica". The logo features a globe on the left side and the company name on the right side. Below the company name is the text "the medical accessories source".
| Neural stimulator probe insulated monopolar | | |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Rhythmlink – Standard Monopolar<br>Stimulating Probe |
| Stimulating Surface Area (mm2) | 7.5 mm² | 4.7 mm² |
| Tip Diameter (mm) | 1.2 mm | 0.75 mm |
| Shaft Length (mm) | 45 mm | 100 mm |
| Lead Length (mm) | 250 cm | 250 cm and 150 cm |
| Bend Angle (degrees) | 0° | 0° |
| Discussion | The differences between the two devices are shaft length and tip diameter.<br>Shaft is insulated and the length is only an ease of use. The tip diameter is<br>bigger in Spes Medica device, the stimulating area is larger than the PD,<br>this doesn't change in dangerous way current densities. So these two<br>differences not affect the safety and effectiveness as compared to the<br>predicate. | |
| Neural stimulator probe bipolar hook | | |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Xian Friendship Medical Electronics –<br>Disposable Double Hook Nerve<br>Stimulator Probe |
| Stimulating Surface Area (mm2) | 30.6 mm2 | 37.7 mm2 |
| Tip Diameter (mm) | 0.65 mm | 1.2 mm |
| Shaft Length (mm) | 20 mm | 70-120 mm |
| Lead Length (mm) | 250 cm | Information not available |
| Bend Angle (degrees) | 180° | 180° |
| Discussion | The differences between the two devices are shaft length and tip diameter.<br>Shaft is insulated and the length is only an ease of use. The tip diameter is<br>bigger in PD, but the stimulating area doesn't change in dangerous way<br>current densities. So these two differences not affect the safety and<br>effectiveness as compared to the predicate. | |
| Neural stimulator probe tripolar hook | | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Xian Friendship Medical Electronics –<br>Disposable Triple Hook Nerve<br>Stimulator Probe |
| Stimulating Surface Area (mm²) | 30.6 mm² | 37.7 mm² |
| Tip Diameter (mm) | 0.65 mm | 1.2 mm |
| Shaft Length (mm) | 20 mm | 70-120 mm |
| Lead Length (mm) | 250 cm | Information not avaible |
| Bend Angle (degrees) | 180° | 180° |
| Discussion | The differences between the two devices are shaft length and tip diameter.<br>Shaft is insulated and the length has only an ease of use. The tip diameter is<br>bigger in PD device, stimulating area is smaller than the PD, but the<br>stimulating area doesn't change in dangerous way current densities.<br>So these two differences not affect the safety and effectiveness as<br>compared to the predicate. | |
{13}------------------------------------------------
Image /page/13/Picture/2 description: The image shows the logo for "spes medica". The logo consists of a globe on the left and the company name on the right. Below the company name is the tagline "the medical accessories source".
| Neural stimulator probe monopolar ball | | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Xian Friendship Medical Electronics -<br>Disposable Ball Tip Direct Nerve<br>Stimulator Probe |
| Stimulating Surface Area (mm2) | 50.2 mm2 | 66 mm2 |
| Tip Diameter (mm) | 2 mm | 2.3 mm |
| Shaft Length (mm) | 30 mm and 45 mm | 70-150 mm |
| Lead Length (mm) | 250 cm | Information not available |
| Bend Angle (degrees) | 0° | 0° |
| Discussion | The differences between the two devices are shaft length and tip diameter.<br>Shaft is insulated and the length has only an ease of use. The tip diameter is<br>bigger in Spes Medica device stimulating area is smaller than the PD, but<br>the stimulating area doesn't change in dangerous way current densities.<br>So these two differences not affect the safety and effectiveness as<br>compared to the predicate | |
| Neural stimulator probe bipolar balls | | |
|---------------------------------------|-----------------------------|------------------------------------------------------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Xian Friendship Medical Electronics –<br>Disposable Double Ball Tip Direct<br>Nerve Stimulator Probe |
| Stimulating Surface Area (mm²) | 51 mm² | 51 mm² |
| Tip Diameter (mm) | 2 mm | 2 mm |
| Shaft Length (mm) | 30 mm | 70-150 mm |
| Lead Length (mm) | 250 cm | Information not available |
| Bend Angle (degrees) | 0° | 0° |
| Neural stimulator probe bipolar concentric | | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Specification | Proposed Spes Medical Probe | Xian Friendship Medical Electronics –<br>Disposable Concentric Direct Nerve<br>Stimulator Probe |
| Stimulating Surface Area (mm²) | 211 mm² | 201-377 mm² |
| Tip Diameter (mm) | 1.5 mm | 0.8 mm |
| Shaft Length (mm) | 45 mm | 80-150 mm |
| Lead Length (mm) | 250 mm | Information not available |
| Bend Angle (degrees) | 0° | 0° |
| Discussion | The differences between the two devices are shaft length and tip diameter.<br>Shaft is insulated and the length has only an ease of use. The tip diameter is<br>bigger in Spes Medica device and shaft length is bigger in the PD. At the<br>end the Spes Medica stimulating area is in the range of the PD stimulating<br>area. So these two differences not affect the safety and effectiveness as<br>compared to the predicate. | |
| Neural stimulator probe bipolar minifork |
|------------------------------------------|
|------------------------------------------|
| Specification | Proposed Spes Medical Probe |
|---------------|----------------------------------|
| | Medtronic – Kartush Side-by-Side |
{14}------------------------------------------------
Image /page/14/Picture/2 description: The image shows the logo for Spes Medica. The logo features a globe on the left side and the company name "spes medica" on the right side. Below the company name, it says "the medical accessories source" in a smaller font.
| | | Bipolar Stimulator Probe |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Stimulating Surface Area (mm2) | 9.4 mm2 | 8 mm2 |
| Tip Diameter (mm) | 1.5 mm | 1.3 mm |
| Shaft Length (mm) | 45 mm | Information not avaible |
| Lead Length (mm) | 250 mm | Information not avaible |
| Bend Angle (degrees) | 0° | 0° |
| Discussion | The differences between the two devices is the tip diameter that is bigger in Spes Medica device. At the end the stimulating area for the two devices is larger than the PD, this doesn't change in dangerous way the current. So this difference not affect the safety and effectiveness as compared to the predicate. | |
| Nonclinical testing performed |
|-------------------------------|
| Biocompatibility |
| Ethylene Oxide Sterilization |
| Dielectric strength |
| Packaging |
| Cleaning Room |
#### CONCLUSION
Comparing these tables and each Spes Medica IOM Stimulator Probes models with PD models we can say that the safety and effectiveness of our device as compared to legally marketed predicate devices. The conclusion of Spes Medica s.r.l. is that the IOM Stimulator Probes are safe and effective as the predicate devices and introduce no new questions concerning safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.