K102862 · Neurovision Medical Products, Inc. · ETN · Dec 21, 2010 · Ear, Nose, Throat
Device Facts
Record ID
K102862
Device Name
NEUROVISION SE (NERVEANA)
Applicant
Neurovision Medical Products, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Dec 21, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Neurovision Nerve Locator/Monitor is an electronic device consisting of a surgical nerve stimulator and an evoked EMG monitor with integrated detecting and warning capability. This device is intended for use in surgical procedures where motor nerves are at risk to assist the surgeon in locating these nerves. This device is intended for use only by qualified, trained medical practitioners who perform operative surgery and who fully understand that this device is only an adjuvant to proper surgical technique and good surgical judgment. The EMGView software is an accessory to the Neurovision Nerve Locator/Monitor. When the dedicated USB output of the EMG unit is connected to a personal computer, the PC installed software provides status monitoring data, troubleshooting data, real-time EMG waveforms, and digital backup of all data obtained during clinical use of the device. The real-time display of EMG waveforms is intended to assist the surgeon with nerve integrity monitoring.
Device Story
EMGView Software is a PC-based accessory for the Neurovision Nerve Locator/Monitor. It receives real-time EMG data from the Nerve Locator/Monitor via a dedicated serial USB connection. The software functions as a read-only display and documentation tool; it does not control or modify the Nerve Locator/Monitor. It provides surgeons with real-time EMG waveform visualization, status monitoring, troubleshooting data, and digital data backup. Used in surgical settings, it assists surgeons in nerve location and integrity monitoring by providing visual feedback and postoperative review capabilities. It serves as an adjuvant to surgical technique and judgment.
Clinical Evidence
Bench testing only. Verification was performed using the Nerveana bench testing performance as a baseline. Validation confirmed that clinical use of the software conforms to user needs.
Technological Characteristics
PC-installed software accessory; connects via dedicated serial USB to the Neurovision Nerve Locator/Monitor. Provides real-time EMG waveform display, threshold audio alarms, and database/playback functionality. Read-only architecture; does not control the primary device.
Indications for Use
Indicated for use by qualified, trained medical practitioners performing operative surgery where motor nerves are at risk, to assist in nerve location and integrity monitoring.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text on a white background. The text includes the alphanumeric string "K102862" and "P. 1 of 3". The handwriting appears to be in black ink and is slightly slanted.
Image /page/0/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, sans-serif font. Below that, the word "medical" is in a smaller, sans-serif font, with two horizontal lines above it.
Traditional 510k - EMGView Software
DEC 2 1 2010
## 510(k) Summarv
نر
In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:
Date: 12/10/10
Submitter: Christine Vergely Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 Telephone: (866) 815-6999 Fax: (314) 310-4579
Trade or Proprietary Name: Common or Usual Name: Classification Name:
EMGView Software EMG software for Windows® PC Surgical nerve stimulator/locator (21 CFR §874.1820)
Predicate Device:
Neurovision Nerve Locator/Monitor (K954601)
Device Description:
The EMGView Software is a Windows PC operating system software dedicated to viewing and documentation of the output of the Neurovision (Nerveäna) Nerve Locator/Monitor. The software is dedicated to the EMG device by a serial USB connection. The EMGView software is a read-only device and is thus incapable of modifying or controlling the Nerveäna Device.
Intended Use:
The EMGView software is an accessory to the Neurovision Nerve Locator/Monitor. When the dedicated USB output of the EMG unit is connected to a personal computer, the PC installed software provides status monitoring data, troubleshooting data, realtime EMG waveforms, and digital backup of all data obtained during clinical use of the device.
Indications for Use:
The Neurovision Nerve Locator/Monitor is an electronic device consisting of a surgical nerve stimulator and an evoked EMG monitor with integrated detecting and warning capability. This device is intended for use in surgical procedures where motor nerves are at risk to assist the surgeon in locating these nerves. This device is intended for
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text on a white background. The text includes the alphanumeric string "k102862" and the phrase "p. 2 of 3". The text is slightly rotated, giving it a dynamic appearance.
Image /page/1/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Below the word "neurovision" are three horizontal lines with a small EKG waveform in the middle. To the right of the horizontal lines is the word "MEDICAL" in bold, black letters.
Traditional 510k - EMGView Software
use only by qualified, trained medical practioners who perform operative surgery and who fully understand that this device is only an adjuvant to proper surgical technique and good surgical judgment.
The EMGView software is an accessory to the Neurovision Nerve Locator/Monitor. When the dedicated USB output of the EMG unit is connected to a personal computer, the PC installed software provides status monitoring data, troubleshooting data, realtime EMG waveforms, and digital backup of all data obtained during clinical use of the device. The real-time display of EMG waveforms is intended to assist the surgeon with nerve integrity monitoring.
Technological Characteristics:
The EMGView software is a PC installed software accessory to the Neurovision Nerve Locator/Monitor.
| | Subject Device | Predicate device (after approved modification |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Neurovision (Nerveäna) with<br>EMGView Software | Nerveäna (post 2006) K954601 |
| Surgical Nerve<br>Locator/Monitor | Yes | Yes |
| Device data<br>output | Dedicated serial USB to PC | Dedicated serial USB to PC |
| Indications for<br>Use | The Neurovision Nerve Locator/Monitor is an<br>electronic device consisting of a surgical nerve<br>stimulator and an evoked EMG monitor with<br>integrated detecting and warning capability.<br>This device is intended for use in surgical<br>procedures where motor nerves are at risk to<br>assist the surgeon in locating these nerves.<br>This device is intended for use only by qualified,<br>trained medical practioners who perform<br>operative surgery and who fully understand that<br>this device is only an adjuvant to proper<br>surgical technique and good surgical judgment.<br>The EMGView software is an accessory to the<br>Neurovision Nerve Locator/Monitor. When the<br>dedicated USB output of the EMG unit is<br>connected to a personal computer, the PC | The Neurovision Nerve Locator/Monitor is an<br>electronic device consisting of a surgical nerve<br>stimulator and an evoked EMG monitor with<br>integrated detecting and warning capability. This<br>device is intended for use in surgical procedures<br>where motor nerves are at risk to assist the<br>surgeon in locating these nerves. This device is<br>intended for use only by qualified, trained medical<br>practioners who perform operative surgery and who fully understand that this device is only an<br>adjuvant to proper surgical technique and good surgical judgment. |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a label or identifier. The text includes 'k102862' and 'P. 3 of 3', suggesting it might be a page number or a reference code within a larger document. The handwriting is clear, but the text is oriented diagonally in the image.
Image /page/2/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word is a black line with a small EKG-like waveform in the middle, followed by the word "MEDICAL" in smaller, bold, black letters.
## Traditional 510k - EMGView Software
| | installed software provides status monitoring<br>data, troubleshooting data, real-time EMG<br>waveforms, and digital backup of all data<br>obtained during clinical use of the device. The<br>real-time display of EMG waveforms is intended<br>to assist the surgeon with nerve integrity<br>monitoring. | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Software | Dedicated <i>EMGView</i> Program | N/A |
| EMG Waveform<br>Display | PC Based Accessory | N/A |
| PC based<br>threshold audio<br>alarm | PC Based Accessory | N/A |
| Database and<br>Playback<br>capability | PC Based Accessory | N/A |
## Safety and Efficacy
The EMGView Software (v4.31, current version) requirements were verified using the Nerveana bench testing performance as a baseline. The software was validated by confirming that its clinical use conforms to user needs.
The Nerveäna device with EMGView software is as safe and effective as the Nerveäna device alone in locating and monitoring motor nerves during surgery:
- The EMGView software, being a read-only device accessory, does not affect the . performance of the Nerveäna device.
- · No surgeon input is required for operation of the Nerveäna accessory.
- · The EMGView software provides additional troubleshooting tools for the surgeon to verify proper Nerveäna operation and electrode position.
- · The Database Feature of the Nerveäna provides focus by the surgeon on the issue of proper nerve location and preservation as well as providing postoperative review data that would otherwise be unobtainable in answering any questions about the relationship of nerve monitoring to the surgical outcome.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Neurovision Medical Products, Inc. c/o Ms. Christine Vergely Regulatory Manager 2225 Sperry Avenue, Suite 1000 Ventura, CA 93003
DEC212010
Re: K102862
Trade/Device Name: EMGView Software (v4.31) Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: December 10, 2010 Received: December 13, 2010
Dear Ms. Vergely:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
R. A. Rosenbaum Jr.
Malvina B. Eydelman, Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K102862
DEC 2 1 2010
Device Name: EMGView Software
Indications for Use:
The Neurovision Nerve Locator/Monitor is an electronic device consisting of a surgical nerve stimulator and an evoked EMG monitor with integrated detecting and warning capability. This device is intended for use in surqical procedures where motor nerves are at risk to assist the surgeon in locating these nerves. This device is intended for use only by qualified, trained medical practitioners who perform operative surgery and who fully understand that this device is only an adjuvant to proper surgical technique and good surgical judgment.
The EMGView software is an accessory to the Neurovision Nerve Locator/Monitor. When the dedicated USB output of the EMG unit is connected to a personal computer, the PC installed software provides status monitoring data, troubleshooting data, real-time EMG waveforms, and digital backup of all data obtained during clinical use of the device. The real-time display of EMG waveforms is intended to assist the surgeon with nerve integrity monitoring.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------|--|
| Division of Ophthalmic, Neurological and Ear, | |
| Nose and Throat Devices | |
| 510(k) Number | K102862 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.