This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
Device Story
NeuralMAS is a multichannel intraoperative monitor used by surgeons to locate and map motor nerves. The system inputs include electrical stimulus (1-15 mA, monophasic square pulse) applied via a monopolar probe and mechanomyographic (MMG) signals captured by skin-surface sensors from muscles innervated by the target nerve. The device transforms these inputs by detecting mechanical muscle contractions resulting from nerve stimulation. The system consists of a touchscreen PC control unit, a patient module, and various probes/sensors. It provides real-time feedback to the surgeon, alerting them when a nerve is activated, thereby assisting in nerve identification and trauma minimization during surgery. The device is intended for use in an operating room environment.
Clinical Evidence
No clinical data. Bench testing and animal study (sheep) performed. Animal study compared NeuralMAS to a predicate device during simulated surgery, demonstrating positive and overall percent agreement in ascertaining spinal nerve status and location.
Technological Characteristics
Multichannel (10-channel) intraoperative monitor. Uses skin-surface MMG sensors (electromechanical) vs. subdermal needle electrodes. Stimulator: constant electrical current (1-15 mA), monophasic square pulse, 100µs pulse duration, 2Hz rate. Control unit: touchscreen PC. Connectivity: disposable probes, stimulators, sensors, electrodes. Sterilization: not specified.
Indications for Use
Indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2 in patients undergoing surgical procedures where motor nerves are at risk.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
# 510(K) SUMMARY
## 2 4 2010 SEP
Page .
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### Submitter's Identification: 1.
Christopher Wybo Director of Engineering and Regulatory Innovative Surgical Solutions, LLC 1002 Hoffman Ave. Royal Oak, MI 48067 Tel: (248) 416-0740 Fax: (800) 683-8144
# Date Summary Prepared: September 24, 2010
### Name of the Device: 2.
| Trade or Proprietary Name: | NeuralMAS™ |
|----------------------------|-----------------------------------|
| Classification Name: | Surgical Nerve Stimulator/Locator |
| Device Class: | Class II |
| Classification: | 21 CFR §874.1820 |
| Product Code: | ETN |
### Common or Usual Name: 3.
(MMG) Mechanomyographic System, Intraoperative Avoidance Mapping and Nerve Monitor/Stimulator, Nerve Locator/Stimulator
#### Predicate Device Information: 4.
The subject device NeuralMAS™ is substantially equivalent to one or more of the following predicate devices:
- NIM-Spine, Medtronic Xomed, Jacksonville, Florida K031510
- Rhythmlink, Columbia, South Carolina K072736
- Axon Systems, Hauppauge, New York KO62996
- UniPlex Nanoline Cannula, Pajunk GMBH, Geisingen, Germany K000722
{1}------------------------------------------------
Page L
#### 5. Device Description:
The NeuralMAS™ system is a multichannel intraoperative monitor for use during surgeries in which a motor nerve is at risk. The NeuralMAS™ system records mechanomyographic (MMG) signals from muscles innervated by the affected nerve, which may originate from operator applied electrical stimulus or from direct or indirect mechanical stimulus occurring during the course of surgery. The monitor will assist early nerve identification by providing the surgeon with a tool to help locate and identify the particular nerve at risk to minimize trauma by alerting the surgeon when a particular nerve has been activated.
The NeuralMAS™ system consists of a reusable Patient Module, a Control Unit comprised of a touch-screen PC and an assortment of disposable conductive probes, stimulators, sensors, electrodes and electrode leads.
#### Intended Use: 6.
This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
#### Comparison to Predicate Devices: 7.
The subject device has indications for use which are substantially equivalent to the predicate device, is composed of the same or equivalent materials as one or more commercially marketed devices, has the same or equivalent design features as the predicate device, is substantially equivalent in its method of use, and has functional characteristics which are the same or equivalent to those of one or more of the predicate devices. Due to the equivalency of indications for use, materials of composition, design features, method of use, and functional characteristics, the device raises no new safety or effectiveness issues.
NeuralMAS™ Nerve Mapping & Avoidance System
{2}------------------------------------------------
A
Page 3
# Table 1: Overview & Indications
| Subject Device: | Predicate Device: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISS NeuralMAS | Medtronic NIM-Spine #K031510 |
| System Overview | System Overview |
| Uses low level electrical current to stimulate<br>nerves sufficient to produce a detectable<br>mechanical signal in contracting muscle using<br>skin-surface mounted mechanomyographic<br>(MMG) sensors. | Uses electrical current varying between<br>low and high levels to induce current flow<br>in stimulated nerves sufficient to produce a<br>detectable electrical signal in contracting<br>muscle using EMG needle electrodes<br>placed in the muscle. |
# Table 2: Physical Attributes
| Attribute | ISS NeuralMAS | Medtronic NIM-Spine<br>#K031510 |
|-----------------------------------------|--------------------------|---------------------------------------------------------------------------|
| Control Module: PC/Display | | |
| Type: | Touchscreen | Touchscreen |
| Number of Channels: | 10 | 8 |
| Channel<br>Enable/Disable<br>Controls: | Touchscreen controlled | Dedicated function touchpads for<br>independent channel<br>enable/disable |
| Event Threshold<br>Control and Display: | Touchscreen controlled | Adjustable graduated touchbar<br>with voltage threshold display |
| Stimulator | | |
| Type Constant: | Electrical Current (mA) | Electrical Current (mA) |
| Range: | 1-15 mA | 1-200 mA, max 540V compliant |
| Control: | Digitally controlled | Digitally controlled |
| Waveform: | Monophasic, square pulse | Monophasic, square pulse |
| Pulse Duration: | 100µs | 50, 100, 150, 200 or 250µs |
.
{3}------------------------------------------------
| Rate: | 2Hz | Switch selectable: 1, 5, or 10Hz |
|-------------------|-----------------------------|-----------------------------------------|
| Stimulus Probe: | Monopolar | Monopolar or bipolar |
| Activation: | Touch screen or hand switch | Two step touch screen or hand<br>switch |
| Monitoring Sensor | | |
| Function: | Mechanomyographic (MMG) | Electromyographic (EMG) |
| Type: | Electromechanical | Subdermal needle electrodes |
| Attachment Site: | Skin surface | Subdermal |
# Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as 8. follows:
A comparative performance evaluation was conducted in the sheep to assess statistical agreement between the subject device and the predicate. Results for positive, and overall percent agreement have established that 1) the principle functions of the NeuralMAS™ system are effective, such that spinal nerve status and location may be ascertained and monitored, and 2) the data acquired by the subject device during simulated surgical use in the animal model correlate well with those acquired simultaneously by a predicate device.
## Discussion of Clinical Tests Performed: ರು
Not Applicable
### 10. Conclusions:
The conclusions drawn from our non-clinical testing (to include our animal testing) of the subject device demonstrates that the subject device is as safe, as effective and performs as well as the legally marketed predicate device. The subject device is substantially equivalent in function to the legally marketed predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized bird in flight, composed of three curved lines, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Innovative Surgical Solutions, LLC c/o Ms. Susan Goldstein-Falk MDI Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, New York 10021
Re: K100992
Trade/Device Name: NeuralMAS Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: August 27, 2010 Received: September 1, 2010
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
SEP 24 2010
{5}------------------------------------------------
Page 2 - Ms. Susan Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robinm.k
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K100992
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
SEP 2 4 2010
Device Name: NeuralMAS
Indications For Use:
This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
100992
Prescription Use
(Per 21 CFR 801.109)
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.