Self-Fitting Air-Conduction Hearing Aid

Page Type
Product Code
Definition
A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.
Physical State
The self-fitting air-conduction hearing aid consists of the following components that are worn on the body, typically at ear level: microphone, receiver, and signal processor. In addition, the device will have a user interface for adjusting the hearing aid.
Technical Method
A self-fitting air-conduction hearing aid incorporates technology, including software, used by the listener to configure parameters, settings, and listening modes to meet the user's listening needs.
Target Area
The self-fitting air-conduction hearing aid provides amplification to the ear.
Regulation Medical Specialty
Ear, Nose, Throat
Review Panel
Ear, Nose, Throat
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
874.3325
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 874.3325 Self-fitting air-conduction hearing aid

§ 874.3325 Self-fitting air-conduction hearing aid.

(a) Identification. A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical data must evaluate the effectiveness of the self-fitting strategy.

(2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested.

(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.

(4) Software verification, validation, and hazard analysis must be performed.

(5) If the device incorporates wireless technology:

(i) Performance testing must validate safety of exposure to non-ionizing radiation;

(ii) Performance data must validate wireless technology functions; and

(iii) Labeling must specify instructions, warnings, and information relating to wireless technology and human exposure to non-ionizing radiation.

(6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.

(7) Patient labeling must include the following:

(i) Information on how a patient can self-identify as a candidate for the device;

(ii) Information about when to seek professional help;

(iii) A warning about using hearing protection in loud environments;

(iv) A warning about staying alert to sounds around the user of the device;

(v) Technical information about the device, including information about EMC; and

(vi) Information on how to correctly use and maintain the device.

[84 FR 57612, Oct. 28, 2019]

Self-Fitting Air-Conduction Hearing Aid

Page Type
Product Code
Definition
A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.
Physical State
The self-fitting air-conduction hearing aid consists of the following components that are worn on the body, typically at ear level: microphone, receiver, and signal processor. In addition, the device will have a user interface for adjusting the hearing aid.
Technical Method
A self-fitting air-conduction hearing aid incorporates technology, including software, used by the listener to configure parameters, settings, and listening modes to meet the user's listening needs.
Target Area
The self-fitting air-conduction hearing aid provides amplification to the ear.
Regulation Medical Specialty
Ear, Nose, Throat
Review Panel
Ear, Nose, Throat
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
874.3325
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 874.3325 Self-fitting air-conduction hearing aid

§ 874.3325 Self-fitting air-conduction hearing aid.

(a) Identification. A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical data must evaluate the effectiveness of the self-fitting strategy.

(2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested.

(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.

(4) Software verification, validation, and hazard analysis must be performed.

(5) If the device incorporates wireless technology:

(i) Performance testing must validate safety of exposure to non-ionizing radiation;

(ii) Performance data must validate wireless technology functions; and

(iii) Labeling must specify instructions, warnings, and information relating to wireless technology and human exposure to non-ionizing radiation.

(6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.

(7) Patient labeling must include the following:

(i) Information on how a patient can self-identify as a candidate for the device;

(ii) Information about when to seek professional help;

(iii) A warning about using hearing protection in loud environments;

(iv) A warning about staying alert to sounds around the user of the device;

(v) Technical information about the device, including information about EMC; and

(vi) Information on how to correctly use and maintain the device.

[84 FR 57612, Oct. 28, 2019]