Tympanic Membrane Direct Contact Hearing Aid

Page Type
Product Code
Definition
A tympanic membrane contact hearing aid is intended to compensate for impaired hearing by transmitting amplified sound by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.
Physical State
A tympanic membrane contact hearing aid consists of an externally worn audio processor, and a custom fitted tympanic membrane receiver/transducer.
Technical Method
A tympanic membrane contact hearing aid is a device worn by the patient that uses an external audio processor to transmit amplified sound by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane
Target Area
A tympanic membrane contact hearing aid provides direct, mechanical vibration of the lateral side of the tympanic membrane.
Regulation Medical Specialty
Ear, Nose, Throat
Review Panel
Ear, Nose, Throat
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
874.3315
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
Yes
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 874.3315 Tympanic membrane contact hearing aid

§ 874.3315 Tympanic membrane contact hearing aid.

(a) Identification. A tympanic membrane contact hearing aid is a prescription device that compensates for impaired hearing. Amplified sound is transmitted by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The patient contacting components must be demonstrated to be biocompatible.

(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, and must include:

(i) Mechanical integrity testing;

(ii) Electrical and thermal safety testing;

(iii) Software verification, validation, and hazard analysis;

(iv) Reliability testing consistent with expected device life;

(v) Electromagnetic compatibility testing; and

(vi) Validation testing of device output and mechanical force applied to the tympanic membrane in a clinically appropriate model.

(3) Clinical performance testing must characterize any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.

(4) Professional training must include the ear impression procedure, correct placement, fitting, monitoring, care, and maintenance of the device.

(5) Labeling must include the following:

(i) A detailed summary of the adverse events and effectiveness outcomes from the clinical performance testing;

(ii) Detailed instructions on how to fit the device to the patient;

(iii) Instructions for periodic cleaning of any reusable components;

(iv) Information related to electromagnetic compatibility; and

(v) Patient labeling that includes:

(A) A patient card that identifies if a patient has been fitted with any non-self- removable components of the device and provides relevant information in cases of emergency;

(B) Information on how to correctly use and maintain the device;

(C) The potential risks and benefits associated with the use of the device; and

(D) Alternative treatments.

[81 FR 3326, Jan. 21, 2015]

Tympanic Membrane Direct Contact Hearing Aid

Page Type
Product Code
Definition
A tympanic membrane contact hearing aid is intended to compensate for impaired hearing by transmitting amplified sound by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.
Physical State
A tympanic membrane contact hearing aid consists of an externally worn audio processor, and a custom fitted tympanic membrane receiver/transducer.
Technical Method
A tympanic membrane contact hearing aid is a device worn by the patient that uses an external audio processor to transmit amplified sound by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane
Target Area
A tympanic membrane contact hearing aid provides direct, mechanical vibration of the lateral side of the tympanic membrane.
Regulation Medical Specialty
Ear, Nose, Throat
Review Panel
Ear, Nose, Throat
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
874.3315
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
Yes
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 874.3315 Tympanic membrane contact hearing aid

§ 874.3315 Tympanic membrane contact hearing aid.

(a) Identification. A tympanic membrane contact hearing aid is a prescription device that compensates for impaired hearing. Amplified sound is transmitted by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The patient contacting components must be demonstrated to be biocompatible.

(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, and must include:

(i) Mechanical integrity testing;

(ii) Electrical and thermal safety testing;

(iii) Software verification, validation, and hazard analysis;

(iv) Reliability testing consistent with expected device life;

(v) Electromagnetic compatibility testing; and

(vi) Validation testing of device output and mechanical force applied to the tympanic membrane in a clinically appropriate model.

(3) Clinical performance testing must characterize any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.

(4) Professional training must include the ear impression procedure, correct placement, fitting, monitoring, care, and maintenance of the device.

(5) Labeling must include the following:

(i) A detailed summary of the adverse events and effectiveness outcomes from the clinical performance testing;

(ii) Detailed instructions on how to fit the device to the patient;

(iii) Instructions for periodic cleaning of any reusable components;

(iv) Information related to electromagnetic compatibility; and

(v) Patient labeling that includes:

(A) A patient card that identifies if a patient has been fitted with any non-self- removable components of the device and provides relevant information in cases of emergency;

(B) Information on how to correctly use and maintain the device;

(C) The potential risks and benefits associated with the use of the device; and

(D) Alternative treatments.

[81 FR 3326, Jan. 21, 2015]