PFO · Active Implantable Bone Conduction Hearing System
Ear, Nose, Throat · 21 CFR 874.3340 · Class 2
Overview
| Product Code | PFO |
|---|---|
| Device Name | Active Implantable Bone Conduction Hearing System |
| Regulation | 21 CFR 874.3340 |
| Device Class | Class 2 |
| Review Panel | Ear, Nose, Throat |
| Implant | Yes |
Identification
An active implantable bone conduction hearing system is a prescription device consisting of an implanted transducer, implanted electronics components, and an audio processor. The active implantable bone conduction hearing system is intended to compensate for conductive or mixed hearing losses by conveying amplified acoustic signals to the cochlea via mechanical vibrations on the skull bone.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
In combination with the general controls of the FD&C Act, the active implantable bone conduction hearing system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must characterize any adverse events observed during implantation and clinical use, and must also demonstrate that the device performs as intended under anticipated conditions of use. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following: (i) Performance data must validate force output in a clinically relevant model. (ii) Impact testing in a clinically relevant anatomic model must be performed. (iii) Mechanical integrity testing must be performed. (iv) Reliability testing consistent with expected device life must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the patient-contacting components of the device. (5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (6) Performance data must demonstrate the wireless compatibility, electromagnetic compatibility, and electrical safety of the device. (7) Software verification, validation, and hazard analysis must be performed. (8) Labeling must include: (i) A summary of clinical testing conducted with the device that includes a summary of device-related complications and adverse events; (ii) Instructions for use; (iii) A surgical guide for implantation, which includes instructions for imaging to assess bone dimensions; (iv) A shelf life, for device components provided sterile; (v) A patient identification card; and (vi) A patient user manual.
Recent Cleared Devices (11 of 11)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K240614 | Sentio Ti Implant Kit; Sentio 1 Mini; Genie Medical BAHS | Oticon Medical AB | Jul 10, 2024 | SESE |
| K240155 | Cochlear Osia System | Cochlear Americas | Apr 18, 2024 | SESE |
| K231204 | Cochlear™ Osia® System; Cochlear™ Osia® OSI300 Implant; Cochlear™ Magnet Cassette; Cochlear™ Non-Magnetic Cassette; Cochlear™ Osia® 2(I) Sound Processor; Cochlear™ Osia® Fitting Software 2; Cochlear™ Osia® Smart App | Cochlear | Aug 18, 2023 | SESE |
| K220922 | Cochlear Osia 2 System, Cochlear Osia OSI200 Implant, Cochlear Osia 2 Sound Processor, Cochlear Osia Fitting Software 2, Cochlear MRI Kit | Cochlear Americas | Jul 27, 2022 | SESE |
| K201983 | BCI 602 Lifts (1 mm), BCI Lifts (1 mm), BCI Lifts (2 mm & 3 mm), BCI Lifts (4 mm) | Med-El Elektromedizinische Geraete GmbH | Oct 9, 2020 | SESE |
| K200504 | BONEBRIDGE System, SAMBA 2 BB, SYMFIT 8.0, SAMBA 2 GO, SAMBA 2 Remote | Med-El Elektromedizinische Geraete GmbH | Jul 29, 2020 | SESE |
| K191921 | Cochlear Osia OSI200 Implant, Cochlear Osia 2 Sound Processor, Osia Fitting Software 2.0, Osia SmartApp, Cochlear Osia Surgical Instruments | Cochlear Americas | Nov 15, 2019 | SESE |
| K191457 | BONEBRIDGE System, Bone Conduction Implant Kit (BCI 602 Implant Kit), BCI 602 Sizer Kit, BCI 602 Lifts (1mm) | Med-El Elektromedizinische Geraete GmbH | Sep 18, 2019 | SESE |
| K190589 | Cochlear Osia OSI100 Implant, Cochlear Osia Sound Processor, Osia Intraoperative Test Software, Osia Fitting Software, Cochlear Osia Surgical Instruments | Cochlear Americas | Jul 3, 2019 | SESE |
| K183373 | Bonebridge | Med-El Elektromedizinische Geraete GmbH | Mar 7, 2019 | SESE |
| DEN170009 | Bonebridge | Med-El Elektromedizinische Geraete GmbH | Jul 20, 2018 | DENG |
Top Applicants
- Med-El Elektromedizinische Geraete GmbH — 5 clearances
- Cochlear Americas — 4 clearances
- Cochlear — 1 clearance
- Oticon Medical AB — 1 clearance