MEROGEL INJECTABLE BIORESORBABLE NASAL DRESSING AND SINUS STENT AND OTOLOGIC GEL
Applicant
Medtronic Xomed
Product Code
NHB · Ear, Nose, Throat
Decision Date
Oct 31, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process. MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, tympanoplasty, and stapes and mastoid surgery. The device is indicated following nasal / sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.
Device Story
MeroGel is a sterile, transparent, viscoelastic, bioresorbable gel composed of cross-linked hyaluronic acid polymers. Supplied in a syringe, it is applied by a physician to ENT cavities following surgery or trauma. The gel acts as a space-occupying stent to separate mucosal surfaces, prevent adhesions, and provide a tamponade effect to control minimal bleeding. It aids the natural healing process and is either naturally eliminated or aspirated by the physician. Used in clinical settings (OR/clinic) by ENT specialists.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including formulation analysis (viscosity, HA concentration), filling process validation, and terminal sterilization process validation.
Technological Characteristics
Material: Cross-linked hyaluronic acid polymers. Form: Viscoelastic gel in a syringe. Bioresorbable. Sterile. Class II device (21 CFR 874.3620).
Indications for Use
Indicated for patients undergoing middle ear or external ear canal surgery (canalplasty, myringoplasty, tympanoplasty, stapes, mastoid) or nasal/sinus surgery/trauma. Used to prevent mucosal adhesions and middle turbinate lateralization, and to control minimal post-operative bleeding.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{0}------------------------------------------------
K1a2434
OCT 31 2012
# 510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS FOR
# MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel
| 510(k) Owner | Medtronic Xomed, Inc<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216-0980 USA<br>904-296-9600<br>904-296-2386 (FAX) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Name | Antoine Kouchakjy<br>Principal Regulatory Affairs Specialist<br>Medtronic Xomed, Inc |
| Date Summary Prepared | August 9, 2012 |
| Proprietary Name | MeroGel Injectable Bioresorbable Nasal Dressing and Sinus<br>Stent and Otologic Gel |
| Common Name | Polymer, Ear, Nose and Throat, Synthetic, Absorbable |
| Classification Name | Ear, nose and throat synthetic polymer material<br>(21 CFR 874.3620, Product Code NHB, Class II) |
# Marketed device claiming equivalence to
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel (MeroGel) is equivalent to Medtronic Xomed's MeroGel Injectable Bioresorbable Stent, K070886.
# Device Description
MeroGel is a sterile, transparent, viscoelactic, bioresorbable gel composed of cross-linked polymers of hyaluronic acid. The MeroGel device fills ENT cavities following surgery or trauma to keep tissues or structures separate during the healing process. During this time, the tamponade effect helps control minimal bleeding normally associated with routine ENT surgery. MeroGel leaves the site of placement by natural elimination, or it may be aspirated from the cavity earlier at the discretion of the physician.
# Intended Use
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.
# Indications for Use
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, tympanoplasty, and stapes and mastoid surgery. The device is indicated following nasal / sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.
{1}------------------------------------------------
# Summary of Changes
- Manufacturing site change. .
- . Secondary sterilization method change
- Sterilization pouch material change. .
# Summary of Testing
The following tests were conducted in order to demonstrate that the MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is substantially equivalent to the MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel of K070886.
- Formulation for viscosity and HA concentration .
- . Filling process
- Terminal Sterilization processes (primary and secondary) .
- Assembly .
- Biocompatibility .
- Labeling and Packaging .
{2}------------------------------------------------
| Device Name | MeroGel Injectable Bioresorbable Nasal<br>Dressing and Sinus Stent and Otologic Gel<br>Medtronic Xomed<br>510(k) PROPOSED | MeroGel Injectable Bioresorbable Stent<br>Medtronic Xomed<br>K070886 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | ENT synthetic polymer material | ENT synthetic polymer material |
| Intended Use / Indication | NHB<br>Same as K070886:<br>MeroGel Injectable Bioresorbable Nasal<br>Dressing and Sinus Stent and Otologic Gel<br>is a space occupying gel stent intended to<br>separate and prevent adhesions between<br>mucosal surfaces, help control minimal<br>bleeding following surgery or nasal<br>trauma, and act as an adjunct to aid in the<br>natural healing process.<br>MeroGel Injectable Bioresorbable Nasal<br>Dressing and Sinus Stent and Otologic Gel<br>is indicated for use in the middle ear and<br>external ear canal following canalplasty,<br>myringoplasty, tympanoplasty, and stapes<br>and mastoid surgery. The device is<br>indicated following nasal / sinus surgery or<br>trauma to prevent lateralization of the<br>middle turbinate and nasal adhesions<br>during the post operative period. | NHB<br>MeroGel Injectable Bioresorbable Stent is<br>a space occupying gel stent intended to<br>separate and prevent adhesions between<br>mucosal surfaces, help control minimal<br>bleeding following surgery or nasal<br>trauma, and act as an adjunct to aid in the<br>natural healing process.<br>MeroGel Injectable Bioresorbable Stent is<br>indicated for use in the middle ear and<br>external ear canal following canalplasty,<br>myringoplasty, tympanoplasty, and stapes<br>and mastoid surgery. The device is<br>indicated following nasal / sinus surgery or<br>trauma to prevent lateralization of the<br>middle turbinate and nasal adhesions<br>during the post operative period. |
| Material | Derivative hyaluronic acid | Derivative hyaluronic acid |
| Bioresorbable | YES | YES |
| Sterile | YES | YES |
| Product Matrix | Gel in a syringe | Gel in a syringe |
.
.
Section 08 - 510(k) Summary : MeroGel Injectable
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
OCT 31 2012
Medtronic Xomed, Inc. % Mr. Antoine Kouchakjy Principal Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville, FL 32216-0980
Re: K122434
Trade/Device Name: MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel
Regulation Number: 21 CFR 874.3620
Regulation Name: Ear, nose, and throat synthetic polymer material
Regulatory Class: Class II
Product Code: NHB
Dated: September 28, 2012
Received: October 1, 2012
Dear Mr. Kouchakjy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Makin D. Ficklen, M.D.
Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
KI22434
Device Name:
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel
Indications For Use:
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.
MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, and stapes and mastoid surgery. The device is indicated following nasal / sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Page 1 of
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel C. Clupper
Division Sign-Off) · >>vision of Ophthalmic, Neurological and Ear, ivose and Throat Devices
K122434
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.