K972317 · Boston Medical Products, Inc. · MIX · Sep 18, 1997 · Ear, Nose, Throat
Device Facts
Record ID
K972317
Device Name
MONTGOMERY THYROPLASTY IMPLANT SYSTEM
Applicant
Boston Medical Products, Inc.
Product Code
MIX · Ear, Nose, Throat
Decision Date
Sep 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Boston Medical Products, Inc. Montgomery® Thyroplasty Implant is indicated for medialization thyroplasty in patients with unilateral vocal cord paralysis to improve voice quality.
Device Story
Montgomery® Thyroplasty Implant System is a surgical implant used for medialization thyroplasty in patients with unilateral vocal cord paralysis. The device consists of a pre-molded, one-piece silicone implant featuring a three-tiered base and an obtuse triangular top, available in five sizes for both males and females. The implant is designed with a softer top and a firmer base to better match surrounding tissue characteristics compared to traditional solid silicone blocks. The system includes accessory surgical instruments, such as measuring kits for size selection and procedure kits for positioning and marking the thyroid cartilage. The device is used by surgeons in an operating room setting. By providing a consistent, pre-molded shape, the implant eliminates the need for intraoperative carving of silicone blocks, reducing variability in implant precision and surface finish. This standardization aims to improve surgical outcomes and voice quality for patients suffering from vocal cord paralysis.
Clinical Evidence
Clinical data from a cohort of 44 patients demonstrated safety and effectiveness. Outcomes were measured via patient surveys, acoustic analysis, and stroboscopy, showing improvement in voice quality and a low complication rate.
Technological Characteristics
One-piece molded silicone implant with dual-durometer construction (softer top, firm base). Radiopaque. Provided in five sizes for males and females. Includes sterile, single-use measuring and procedure kits. Manual surgical instruments (elevators, hooks) included. Non-electronic, mechanical device.
Indications for Use
Indicated for medialization thyroplasty in patients with unilateral vocal cord paralysis to improve voice quality.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{0}------------------------------------------------
K97223/7
# 510(k) Summary Boston Medical Products, Inc. Montgomery® Thyroplasty Implant System (per 21 CFR 807.92)
10077 18
June 18, 1997 DATE OF PREPARATION: 1. Boston Medical Products, Inc. SPONSOR/APPLICANT: 2. 117 Flanders Road Westborough, MA 10581
> Stuart Montgomery, President Telephone: 508-898-9300
#### DEVICE NAME: 4.
3.
CONTACT NAME:
Montgomery® Thyroplasty Implant System Trade/Proprietary Name: Thyroplasty Implant and Accessories Common/Usual Name: Ear, nose, and throat synthetic polymer material and Classification Name: manual surgical instruments
# IDENTIFICATION OF THE PREDICATE OR LEGALLY MARKETED DEVICE(S) TO 5. WHICH EQUIVALENCE IS BEING CLAIMED:
The Montgomery® Thyroplasty Implant is substantially equivalent to legally marketed silicone block, as carved at the time of surgery, specifically the following:
| • Duralastic Silicone | Allied Biomedical | K955368, K955370, K955433 |
|-----------------------|-------------------|---------------------------|
| • Silicone Block | Bentec Medical | Not identified |
| • Silicone Block | Xomed | K970910 |
### DEVICE DESCRIPTION 6.
The Montgomery® Thyroplasty Implant System consists of the Implant and associated accessory surgical instruments which are used for selection of implant size and for the surgical thyroplasty procedure.
The Montgomery® Thyroplasty Implant is a one-piece device consisting of a threetiered base and an obtuse triangular top. It is provided in five sizes for males and five for females. The difference in size relates to the height of the triangular portion of the Implant. There is a single base size for all five male implants and a smaller base size for all five female implants.
{1}------------------------------------------------
Accessories to the System include the sterile, single use Measuring Kits for selection of the correct implant size at the time of surgery, the Procedure Kit for determining the correct position of the Implant and marking the window in the thyroid cartilage, and various elevators, hooks, etc. which are provided to facilitate both the surgical procedure and the placement of the Implant.
#### INTENDED USE 7.
The Montgomery® Thyroplasty Implant is indicated for medialization thyroplasty in patient with unilateral vocal cord paralysis to improve quality of vocalization.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 8.
The Montgomery® Thyroplasty Implant is equivalent to silicone block, as carved at the time of surgery. It is composed of an inner, softer top and firm base which secures the implant in the thyroid laminotomy. Bentec silicone block is available in four durometers: Soft (20A), Medium (35A), Firm (50A), and Extra firm (65A). Allied Biomedical silicone block is provided as Very soft (5A), Soft (10A), Medium (20-30A), and Hard (40A and above). While Bentec and Allied Biomedical product designations for durometer are similar, they are not equal. Unlike the use of solid silicone block of a single durometer, the durometers used for the Montgomery Thyroplasty Implant are more consistent with the characteristics of the surrounding tissue.
The Montgomery® Thyroplasty Implant is radiopaque, while the radiopacity of the Bentec and Allied Biomedical silicone blocks is not specified. Bentec does not specify coloration or radiopacity of their block. Allied Biomedical silicone block is generally supplied clear; however custom colors of white, opaque, and flesh tone are available as special order items.
The Montgomery® Thyroplasty Implant is supplied sterility status of Bentec and Allied Biomedical silicone block is not specified in product literature.
While the Montgomery® Thyroplasty Implant is molded, implants fashioned from Bentec and Allied Biomedical silicone block must be carved at time of surgery. Appropriate sizing of the Montgomery Thyroplasty Implant is accomplished using the sterile, single use, and disposable Measuring Kits. Sizing of silicone blockderived medialization implants is dependent on surgeon skill.
Tolerances for the Montgomery® Thyroplasty Implant provide consistency across devices whereas the precision for medialization thyroplasty implants carved from silicone block is limited and dependent on surgeon skill.
The surface of the Montgomery® Thyroplasty Implant has smooth edges and radii. The finish quality of medialization thyroplasty implants which are fashioned from silicone block at the time of surgery is variable and dependent on surgeon skill.
{2}------------------------------------------------
Based on the above discussion of the use of carved silicone block for medialization thyroplasty, Boston Medical Products, Inc. believes that the Montgomery® Thyroplasty Implant is substantially equivalent to solid silicone block as carved at the time of thyroplasty surgery and that differences do not adversely effect safety and effectiveness.
## CLINICAL STUDIES 9.
Clinical data for a cohort of 44 patients treated with the Montgomery® Thyroplasty Implant demonstrate that the device is safe, as evidenced by the low number of complications, and effective for improvement of voice quality, as demonstrated by responses to the patient survey, acoustic analysis, and stroboscopy.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features the department's name in a circular arrangement on the left side of the image. To the right of the text is a symbol that appears to be an abstract representation of a human figure or a stylized flame-like design.
DEPARTMENT OF HEALTH & HUMAN SERVICES
SEP 1 8 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Boston Medical Products, Inc. c/o Rosina Robinson, RN, MEd, RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760
Re: K972317 Montgomery® Thyroplasty Implant System Dated: June 19, 1997 Received: June 20, 1997 Regulatory Class: II 21 CFR 874.3620/Procode: 77 KHJ
Dear Ms. Robinson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
h.J.Liau Yui
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): _K973317
Device Name: Boston Medical Products, Inc. Montgomery® Thyroplasty Implant System
Indications For Use:
The Boston Medical Products, Inc. Montgomery® Thyroplasty Implant is indicated for medialization thyroplasty in patients with unilateral vocal cord paralysis to improve voice quality.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K972317
Boston Medical Products, Inc. 510(k) CONFIDENTIAL 6/18/97 Page v
Montgomery® Thyroplasty Implant System
6/18/97
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.