The Baha implant system is intended for treatment of patients who have conductive or mixed hearing loss as a result of certain medical conditions such as bilateral atresia and chronic supportive otitis media, and for those who have Single-Sided Deafness (SSD) caused by a congenital condition, surgery, trauma, or disease. The intended use of the system is to provide an implanted magnet, and a transcutaneous magnetic connection of an external sound processor, in this instance, the Cordelle II Sound Processor. Patients should have sufficient bone quality and quantity to support successful implant placement.
Device Story
Cochlear Baha Attract system uses transcutaneous magnetic coupling to transmit sound to inner ear via bone conduction; bypasses external auditory canal/middle ear. System components: titanium implant (osseointegrated into skull), titanium-encased magnet pair, external magnet pair with soft pad, and Cordelle II sound processor. Cordelle II processor uses analog signal processing with K-Amp circuitry to provide high gain/output. Used in clinical settings; operated by patients. Sound processor attaches to external magnet via snap coupling; vibrations travel through skull to stimulate inner ear nerve fibers. Benefits patients with conductive/mixed hearing loss or SSD by restoring auditory perception through bone transmission.
Clinical Evidence
Bench testing only. Technical feedback and force transfer measurements compared the Baha Attract System with Cordelle II Sound Processor against the Baha Softband with Cordelle II Sound Processor. Results support substantial equivalence in performance.
Technological Characteristics
System components: titanium implant, titanium-encased magnet pair, external magnet pair, replaceable soft pad, and Cordelle II sound processor. Uses transcutaneous magnetic coupling for bone conduction. Cordelle II processor utilizes analog signal processing with K-Amp circuitry. No new materials or components introduced.
Indications for Use
Indicated for patients aged 5+ with conductive/mixed hearing loss (bone-conduction threshold ≤ 65 dB HL for Cordelle II) or unilateral sensorineural deafness (SSD) with normal hearing in the contralateral ear (air-conduction threshold ≤ 20 dB HL). Bilateral fitting indicated for symmetric moderate-to-severe conductive/mixed loss. Also indicated for patients who cannot/will not use air-conduction CROS hearing aids.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Predicate Devices
Cochlear™ Baha® Cordelle II Sound Processor with Baha Softband (K080363)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 22, 2015
Cochlear Americas Ms. Laura Blair AuD Regulatory Manager 13059 E Peakview Avenue Centennial, Colorado 80111
Re: K150751
Trade/Device Name: Cochlear™ Baha® Cordelle II Sound Processor Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: March 20, 2015 Received: March 23, 2015
Dear Ms. Blair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
FDA Form 3514, Section G
510(k) Number (if known): K150751
Device Name: Cordelle II Sound Processor
#### Indications for Use:
Conductive, mixed and single sided deafness. Patients should have sufficient bone quality and quantity to support successful implant placement.
The Cochlear Baha® Attract is intended for the following patients and indications for use:
- Patients aged 5 and older ●
- . Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3kHz) should be better than or equal to 45 dB HL for use with the BP100, Baha 4 and Baha 5 sound processors. 55 dB HL for use with the BP110 sound processor. and better than or equal to 65 dB HL for use with the Cordelle II sound processor.
- Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conductive thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1.2. and 3 kHz). or less than a 15dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single-sided deafness: SSDTM). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
- Baha for SSD is also indicated for any patient who is indicated for an air conduction ● contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Prescription Use Subpart D) (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{3}------------------------------------------------
# 4. 510(K) Summary
# A. Submitter Information
| Submitted by: | Cochlear Americas<br>13059 East Peakview Ave.<br>Centennial, CO 80111 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| On behalf of the manufacturer: | Cochlear Bone Anchored Solutions AB<br>Konstruktionsvagen 14<br>SE-435 33 Molnlycke<br>Sweden<br>(Establishment Number 9616024) |
| Contact: | Laura Blair<br>Regulatory Manager<br>Cochlear Americas<br>lblair@cochlear.com<br>(303) 524-6852 (o)<br>(303) 524-6825 (f) |
| B. Date Prepared | 3/20/2015 |
| C. Device Class | II |
| D. Device Name | |
| Trade/Proprietary Name: | CochlearTM Baha® Cordelle II Sound Processor |
| Common/Usual Name: | Auditory Osseointegrated Implant |
| Classification Name: | Hearing Aid (Bone Conduction)<br>21 CFR 874.3300, Class II |
| Classification Panel | Ear, Nose, and Throat |
| Product Code: | LXB |
| E. Predicate Devices | |
| Trade/Proprietary Name: | CochlearTM Baha® Cordelle II Sound Processor<br>with Baha Softband |
| Common/Usual Name: | Auditory Osseointegrated Implant Sound Processor<br>with Softband |
| Classification Name: | Hearing Aid (Bone Conduction)<br>21 CFR 874.3300, Class II |
| Classification Panel | Ear, Nose, and Throat |
| Product Code: | LXB |
| 510(k): | K080363 |
| Trade/Proprietary Name: | CochlearTM Baha® Attract |
| Common/Usual Name: | Auditory Osseointegrated Implant Sound Processor |
| Classification Name: | Hearing Aid (Bone Conduction) |
| Classification Panel | 21 CFR 874.3300, Class II |
| Product Code: | Ear, Nose, and Throat |
| 510(k): | LXB |
| | K131240 |
{4}------------------------------------------------
#### F. Purpose of Submission
The purpose of this Traditional 510(k) is to expand the fitting indications of the Baha Attract System to include the use of the Cordelle II Sound Processor and the fitting range that is currently indicated for the Cordelle II Sound Processor with the titanium abutment and Baha softband. The current indications for the Cordelle II Sound Processor were cleared 4/10/2008 under 510(k) K080363.
## G. Device Description
The technology base for the Baha implant was originally derived from the 1952 discovery by Dr. Per-Ingvar Brånemark that titanium was biocompatible with bone, leading to the term "osseointegration". The Brånemark System, as it would later be known, formed the basis for the rapid development and widespread implementation of root form endosseous dental implants, now safely and effectively used by many hundreds of thousands of people worldwide. Root form dental implants share many characteristics with the Baha implant component including material, conformation, and surgical implantation procedures.
The Cochlear Baha Attract auditory implant system consists of three parts: a titanium implant, a titanium-encased magnet pair, and an external magnet pair with a replaceable soft pad. The existing range of Baha sound processors attach to this external magnet via a snap coupling. The system works by utilizing natural bone transmission as a pathway for sound to travel to the inner ear, bypassing the external auditory canal and middle ear. After surgical placement, the titanium implant naturally integrates with the skull bone over time through a process known as osseointegration. The external sound processor transmits sound vibrations through the transcutaneous magnetic coupling to the titanium implant. The vibrating implant creates vibrations within the skull that stimulate the nerve fibers of the inner ear, allowing hearing.
The change proposed in this Traditional 510(k) is to expand the fitting indications of the Baha Attract to include the use of the Cordelle II Sound Processor and the fitting range that is currently indicated for the Cordelle II Sound Processor with a titanium abutment or softband. The fundamental operating principle of the device (hearing rehabilitation through bone conduction) remains unchanged.
## H. Intended Use
The Baha implant system is intended for treatment of patients who have conductive or mixed hearing loss as a result of certain medical conditions such as bilateral atresia and chronic supportive otitis media, and for those who have Single-Sided Deafness (SSD) caused by a congenital condition, surgery, trauma, or disease. The intended use of the system is to provide an implanted magnet, and a transcutaneous magnetic connection of an external sound processor, in this instance, the Cordelle II Sound Processor. Patients should have sufficient bone quality and quantity to support successful implant placement.
{5}------------------------------------------------
The Cochlear Baha® Attract is intended for the following patients and indications for use:
- Patients aged 5 and older ●
- . Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5. 1. 2. and 3kHz) should be better than or equal to 45 dB HL for use with the BP100. Baha 4 and Baha 5 sound processors, 55 dB HL for use with the BP110 sound processor, and better than or equal to 65 dB HL for use with the Cordelle II Sound Processor.
- . Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conductive thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1,2, and 3 kHz), or less than a 15dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (Le. Single-sided deafness: SSDTM). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
- Baha for SSD is also indicated for any patient who is indicated for an air conduction ● contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
## I. Technological Characteristics
The Cochlear Baha Attract consists of five basic components, none of which are new to this submission:
- An implanted magnet
- . An implanted fixture to act as an anchoring point for the implanted magnet (The BI300, cleared in K100360)
- An external magnet for attachment to the Sound Processor (referred to as the SP Magnet) in a variety of magnetic strengths
- A patient-replaceable SP Magnet soft pad to equalize peak skin pressures across the interface between the SP Magnet and the patient's skull
- An existing, cleared Baha sound processor (currently consisting of either the BP100, ● Baha 4, Baha 5 or BP110, cleared in K090720, K132278, K142907 and K110996 respectively)
The Baha Cordelle II Sound Processor is an external sound processor that utilizes analog signal processing with K-Amp circuitry (Killion, 1993) for use with the Baha auditory osseointegrated implant. The Cordelle II Sound Processor is often chosen for a patient who has greater gain needs since it offers the highest output levels and K-Amp circuitry that helps prevent saturation.
## J. Materials
The Baha Cordelle II Sound Processor has not changed in design, materials, or manufacturing since the 510(k) clearance of the 65dB indication with the titanium implant and softband (K080363). This 510(k) premarket notification is solely for an indication change, and therefore, this section does not apply to this submission.
{6}------------------------------------------------
#### K. Performance Data
Technical feedback and force transfer measurements with the Cordelle II Sound Processor used with the Baha Attract System were completed in support of this 510k submission. The focus of the test report was on comparing the Baha Attract System with Cordelle II Sound Processor against the Baha Softband with Cordelle II Sound Processor. The results of the bench testing support the conclusion that the performance of the Cordelle II Sound Processor with Baha Attract is substantially equivalent to the Cordelle II Sound Processor used with the Baha Softband.
#### L. Conclusion
Based on the indications for use, technological characteristics, and substantial equivalence comparison to the predicate devices, the Cordelle II Sound Processor has been shown to be safe and effective for its proposed intended use with the Baha Attract System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.