OTOMAG BONE CONDUCTION HEARING SYSTEM MODEL ALPHA 1 (S) AND ALPHA 1 (M)
K102199 · Sophono, Inc. · LXB · May 23, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K102199
Device Name
OTOMAG BONE CONDUCTION HEARING SYSTEM MODEL ALPHA 1 (S) AND ALPHA 1 (M)
Applicant
Sophono, Inc.
Product Code
LXB · Ear, Nose, Throat
Decision Date
May 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K102199 · May 23, 2011
OTOMAG BONE CONDUCTION HEARING SYSTEM MODEL ALPHA 1 (S) AND ALPHA 1 (M)
Sophono, Inc.
Retrospective clinical patient cohort
Clinical data from 57 patients (86 devices) were used to evaluate the safety and effectiveness of the Alpha 1(M) configuration, specifically monitoring for adverse events (infections, tissue necrosis, pressure points) and hearing performance (speech understanding, free field gain).
Retrospective clinical cohort; Follow-up/Duration: Mean 1.6 years (range 0.5 years)
Patients with conductive or mixed hearing loss, or profound sensorineural hearing loss in one ear; ages 5 to 67 years; Sample Size: 57 patients (86 devices)
Not applicable for this study
Surgical/post-surgical adverse events, tissue health, device dislodgement, speech understanding, free field gain
Indications for Use
The Otomag Alpha 1 Sound Processor is intended for use with the Otomag Headband or Otomag Softband (no age limitations), or with the Otomag Magnetic Implant (patients 5 years of age and up) for the following patients and indications: - Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz). - Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies. - Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
Device Story
Otomag Bone Conduction Hearing System consists of sound processors and accessories using bone conduction of sound vibrations. Alpha 1 (S) configuration uses headband/softband; Alpha 1 (M) uses surgically implanted magnet assembly. Sound processor attaches magnetically to head; vibration transduced through skin to bone. Audiologist uses software to program device based on patient hearing loss. Device helps patients with conductive/mixed hearing loss or single-sided deafness by amplifying sound. Clinical benefit includes improved speech understanding. Used in clinical settings; operated by audiologists and patients.
Clinical Evidence
Clinical trial of 86 Alpha 1 (M) devices in 57 patients (ages 5-67). Mean follow-up 1.6 years. Results: 0% surgical/post-surgical infections or tissue necrosis. 8% of subjects experienced minor initial redness, resolved by adjusting magnetic spacer strength. No processor dislodgement reported. Aided speech understanding improved from 4% (unaided) to 86% (aided). Mean free field gain 38 dB.
Technological Characteristics
Materials: Titanium (ASTM F136, ASTM F67). Principle: Bone conduction of sound vibrations. Configuration: Sound processor with magnetic attachment. Connectivity: Programmable via software. Sterilization: Moist heat (ISO 17665-1) for implant. Software: Rule-based programming for individual hearing loss adjustment.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (BC threshold better than 45 dB HL) or single-sided profound sensorineural hearing loss with normal hearing in the opposite ear (AC threshold better than 20 dB HL). Headband/softband use has no age limit; magnetic implant use is for patients 5 years and older. Contraindicated for sudden/rapidly progressing hearing loss, ear pain, sudden/unilateral tinnitus, asymmetrical hearing loss of unknown origin, vertigo, or patients requiring MRI.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Predicate Devices
Otomag Bone Conduction Hearing System, Unmodified Version (K100193)
{0}------------------------------------------------
K102199
MAY 2 3 2011
# 510(k) Summary: Otomag Bone Conduction Hearing System
This summary of substantial equivalence information is being submitted in accordance with the requirements set forth in 21 CFR 807.92.
Submitter: Sophono, Inc.
#### Establishment Registration Number: 3008514292
#### Contact Information:
Sophono, Inc. 3022 Sterling Circle Boulder, Colorado 80301 Ph: 303-546-1663 Fax: 720-222-2432 · Contact: Jim Kasic Email: Jim.Kasic@sohpono.com
Date Prepared: May 20, 2011
Name of Device: Otomag Bone Conduction Hearing System
Common Name: Bone Conduction Hearing System
Classification Name: Hearing Aid
#### Device Classification:
- . Classification: II
- Classification Panel: Ear, Nose and Throat .
- Regulation Number: 874.3300 .
- Product Code: LXB .
#### Predicate Device:
Otomag Bone Conduction Hearing System, Unmodified Version - K100193
Xomed Audiant Bone Conductor - K855059 / K861971
## Device Description:
The Otomag Bone Conduction Hearing System is a family of sound processors and accessories that operate on the principle of bone conduction of sound vibrations. The Otomag System consists of two distinct configurations; Alpha 1 (S) and Alpha 1 (M). This 510(k) is being submitted to add the Alpha 1 (M) configuration to the existing product family.
In the Otomag Alpha 1 (S), the Otomag Sound Processor is attached magnetically to a headband or softband. The headband or softband holds the sound processor against the head and vibration is transduced through direct contact with the patient's skin and the bone below.
In the Otomag Alpha 1 (M), the Otomag Sound Processor is attached magnetically to an implanted magnet assembly. The magnetic field holds the sound processor against the head and vibration is transduced through direct contact with the patient's skin and the bone below. A variety of magnetic
{1}------------------------------------------------
spacers, each with a different magnetic strength, are provided with the Alpha 1(M) configuration to allow adjustment of the magnetic field strength that holds the sound processor against the head.
The magnetic implant of the system is provided clean and non-pyrogenic, but is not sterile. The implant is intended to be sterilization at the healthcare facility immediately prior to implementation.
The Otomag System is designed for use for those patients with conductive hearing loss, those patients who have sensorineural hearing loss up to 45 dB in combination with their conductive loss, and single sided deafness as defined in the indications for use. The prescriptive formula and adjustments available to the audiologist in the software allow for programming the Otomag System for individual patient hearing loss.
#### Technological Characteristics:
The technological characteristics of the version of the Otomag Bone Conduction Hearing System are equivalent to those of the predicate device. Items that are identical between the modified version and the predicate are:
- Overall Device Performance .
- . Fundamental Technology (bone conduction)
- . Software
Technological characteristics that are different between the modified device and the predicate are as follows:
- Patient contact materials: Addition of a titanium cased magnetic implant for attachment of . the sound processor assembly to the patient's skin.
- . Method of Sound Processor Attachment: Addition of a titanium cased magnetic implant for attachment.
. .
Non-clinical and clinical performance testing and evaluation were used to support the safety and effectiveness of the modified device.
#### Device Indications For Use:
生 《 The Otomag Albha 1 Sound Processor is intended for use with the Otomag Headband or Otomag Softband (no age limitations), or with the Otomag Magnetic Implant (patients 5 years of age and up) for the following patients and indications:
- Patients with conductive or mixed hearing losses, who can still benefit from amplification of . sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz).
- Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed . hearing loss. The difference between the left and right sides' BC thresholds should be less than 10dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies.
- . Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
{2}------------------------------------------------
## Contraindications
Any factor that would cause a clinician to refer the patient for medical assessment will temporarily, or in some cases permanently, halt the process of hearing aid fitting. These factors include:
- A hearing loss of sudden onset; .
- . A rapidly progressing hearing loss;
- Pain in either ear; .
- Tinnitus of sudden recent onset, or unilateral tinnitus; .
- Unilateral or markedly asymmetrical hearing loss of unknown origin; .
- . Vertigo (e.g. dizziness)
Patients using the Alpha 1(M) configuration of the device, which utilizes a magnetic implant, should not undergo magnetic resonance imaging (MRI) without having the implant removed prior to scanning.
## Performance Standards:
The design of the Otomag Bone Conduction Hearing System conforms to the following voluntary standards:
- . ISO 14971:2007: Medical devices - application of risk management to medical devices;
- . FDA Memorandum #G95-1: Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing
- ISO 17665-1:2006: Sterilization of Healthcare Products Moist Heat Part 1 Requirements . for the Development, Validation, and Routine Control of Sterilization Process for Medical Devices
- ISO 17664-1:2004: Sterilization of medical devices Information to be provided by the . manufacturer for the processing of resterilizable medical devices
- . ISO 15223-1:2007: Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied
- . AAMI TIR 12:2004 Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A quide for medical device manufacturers
- AAMI/ANSI ST8:2008, Hospital steam sterilizers .
- AAMI/ANSI ST67:2003/(R) 2008,Sterilization of health care products Requirements for . products labeled "STERILE"
- AAMI / ANSI ST77:2006, Containment devices for reusable medical device sterilization .
- AAMI/ANSI ST79:2006,A1:2008,A2:2009 Comprehensive guide to steam sterilization and . sterility assurance in health care facilities
- . AAMI/ANSI ST81:2004, Information to be provided by the manufacturer for the processing of resterilizable medical devices
- ASTM F136 -- 08e1: Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI . (Extra Low Interstitial) Alloy for Surgical Implant Applications
- . ASTM F67-06: Standard Specification for Unalloyed Titanium, for Surgical Implant Applications
{3}------------------------------------------------
- IEC 60601-1-2:2007: Medical Electrical Equipment Part 1-2: General Requirements for . Safety - Collateral Standard: Electromagnetic Compatibility - Immunity (ESD, Radiated RF Electromagnetic Field Amplitude Modulated, Power Frequency Magnetic Field)
#### Non-Clinical Performance Testing:
The Otomag Bone Conduction Hearing System has been subjected to extensive safety and performance testing. Testing of the Otomag Bone Conduction Hearing System included verification and validation of the following;
- Simulated Use: Device Retention .
- Simulated Use: Vibration Transmission Comparison Alpha 1(M) vs. Alpha 1(S) .
- Software Validation .
- . Sound Processor Performance: Frequency Response
- . Immunity (IEC 60601-1-2)
- . Implant Performance after 3x Sterilization
- � Autoclave Cycle Performance (Sterilization)
- . Accelerated Aging: Magnet Retention
- Dislodgement Force: Normal and Tangential ◆
- . Accelerated Aging: Screw Retention
The testing information presented in this submission demonstrates the device's performance is substantially equivalent to the predicate devices.
## Clinical Performance Testing:
Clinical data on eighty-six (86) Alpha 1(M) devices implanted into 57 patients ranging in age from 5 vears to 67 years was submitted in support of the safety and effectiveness of the Alpha 1(M) configuration of the device. The mean follow up time for the study was 1.6 years from the time of implant, with a range of 0.5 years. The study showed the following important results:
- No surgical or post-surgical infections or adverse events associated with implant were . reported as a part of the trial.
- No tissue necrosis or other significant tissue problems were reported in any of the 57 . subjects at any time in the trial.
- . A small percentage of subjects (8%) showed some initial redness under the magnetic spacer after installing the Alpha 1 (M) sound processor assembly. These issues were resolved by changing the strength of the magnetic spacers. Data showed none of these patients had recurring problems with pressure points or reddening.
- . None of the subjects complained of issues, either mild or severe, of external processor dislodgement.
- . Unaided and aided hearing performance is shown in the following table
{4}------------------------------------------------
.
| Measure | Unaided<br>Air<br>Conduction<br>Thresholds | Unaided<br>Bone<br>Conduction<br>Thresholds | Unaided<br>Speech<br>Understanding | Aided Speech<br>Understanding | Free<br>Field<br>Gain | Aided Free<br>Field<br>Thresholds |
|----------------------|--------------------------------------------|---------------------------------------------|------------------------------------|-------------------------------|-----------------------|------------------------------------------------------------------------|
| Mean<br>and<br>Stdev | 54+/- 12dB | 16+/-10dB | 4+/- 10% | 86+/- 17% | 38+/-<br>8db | 16dB<br>(calculated from<br>unaided thresholds<br>and free field gain) |
## Conclusion:
Based on the clinical and non-clinical performance testing, the modified version of the Otomag Bone Conduction Hearing System is safe and effective and considered to be substantially equivalent to the predicate Otomag device cleared in K100193 and the Audiant cleared K855059 and K861971.
·
ార్లు గ్రామం
మూలాలు
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird with three lines representing its wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 2 3 2011
Sophono Inc. c/o Clay Anselmo Reglera LLC 555 Zang Street Lakewood, CO 80228
Re: K102199
Trade/Device Name: Sophono Otomag Bone Conduction Hearing System . Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: May 10, 2011 Received: May 13, 2011
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Ruhm
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number: K102199
Device Name: Otomag Bone Conduction Hearing System
Indications for Use:
The Otomag Alpha I Sound Processor is intended for use with the Otomag Headband or Olomag Softband (no age limitations), or with the Otomag Osseointegrated Magnetic Implant (patients 5 years of age and up) for the following patients and indications:
- · Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz).
- · Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies.
- Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
**Prescription Use** *X*
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ZartBosin
Page 1 of 1
(Division Sign-Off) Division of Ophthalmic, Neurological and Nose and Throat Devices
510(k) Number K102199
Page 20 of 192
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.