K163094 · Jiangsu Betterlife Medical Co., Ltd. · KLW · May 17, 2017 · Ear, Nose, Throat
Device Facts
Record ID
K163094
Device Name
TinniLogic Mobile Tinnitus Management Device
Applicant
Jiangsu Betterlife Medical Co., Ltd.
Product Code
KLW · Ear, Nose, Throat
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TinniLogic Mobile Tinnitus Management Device is indicated for use in the temporary relief of tinnitus symptoms. The device is a player to play customized sounds and display professional counseling or education to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is adults (18 years or older). This is a medical device and should only be used with the advice of a physician, audiologist or other hearing healthcare professional.
Device Story
Device is a sound player system comprising a tablet, specialized headset, and charger; used by patients for tinnitus management. Healthcare professionals customize sound therapies and counseling/education content for the patient. Device delivers sound stimulation during sleeping or waking hours. Inputs include customized sound profiles; outputs include sinusoidal amplitude modulated (SAM) tinnitus pitch-matched sounds, narrow-band noise (centered at tinnitus frequency), broad-band noise, pure tones, white noise, natural sounds, and relax sounds. Device features individual volume control per ear and data logging of patient usage. Intended for use in clinical or home settings under professional supervision. Benefits include temporary relief of tinnitus symptoms through sound therapy and patient education.
Clinical Evidence
No clinical data. Bench testing only, including IEC 60601-1 (safety) and IEC 60601-1-2 (EMC) compliance, plus output performance testing (maximum output, harmonic distortion, frequency response).
Technological Characteristics
Sound player system; tablet-based with specialized headset. Energy source: rechargeable Li-Ion battery. Connectivity: USB and wireless. Output: 50Hz-15KHz, max 90 dB SPL. Standards: IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014. Software-based sound generation and data logging.
Indications for Use
Indicated for temporary relief of tinnitus symptoms in adults (18+). Must be used under the guidance of a physician, audiologist, or hearing healthcare professional.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
SoundCure™ Serenade™ Tinnitus Management System (K111293)
Submission Summary (Full Text)
{0}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract design featuring three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2017
Jiangsu Betterlife Medical Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN
Re: K163094
Trade/Device Name: TinniLogic Mobile Tinnitus Management Device Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: April 14, 2017 Received: April 17, 2017
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S 2017.05.17 13:21:17 -04'00'
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### Device Name
Mobile Tinnitus Management Device
## Models:BTG-A5/ BTG-A6/ BTG-B5/ BTG-B6/ BTG-C5/ BTG-C6
Indications for Use (Describe)
The TimiLogic Mobile Timitus Management Device is indicated for use in the temporary relice of timitus symptoms. The device is a plaver to play customized sounds and display professional counseling or education to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is adults (18 years or older). This is a medical device and should only be used with the advice of a physician, audiologist or other hearing healthcare professional.
Type of Use (Select one or both, as applicable)
| <div></div> | <span>Form Designed to Meet ATBCB/ADA Guidelines</span> |
|-------------|---------------------------------------------------------|
| <div></div> | <span>Form Designed to Meet WCAG 2.0 Guidelines</span> |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K163094
- 1. Date of Preparation: 4/14/2017
- 2. Sponsor Identification
# Jiangsu BetterLife Medical Co., Ltd.
No. 11 Si Hai Road, Changshu Economic and Technological Development Park, Jiangsu Province, 215513, China.
Establishment Registration Number: 3010197022
Contact Person: Mr. Hongwei Xu Position: Management Representative Tel: +86-512-52269378 Fax: +86-512-52269366 Email: jerry.xu(@goblm.com.cn
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Subject Device
Trade Name: TinniLogic Mobile Tinnitus Management Device Common Name: Tinnitus Masker Models: BTG-A5, BTG-A6, BTG-B5, BTG-B6, BTG-C5 and BTG-C6
### Regulatory Information
Classification Name: Tinnitus Masker; Classification: II; Product Code: KLW; Regulation Number: 21CFR 874.3400 Review Panel: Ear Nose& Throat;
### Indications for Use
The TinniLogic Mobile Tinnitus Management Device is indicated for use in the temporary relief of tinnitus symptoms. The device is a player to play customized sounds and display professional counseling or education to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is adults (18 years or older). This is a medical device and should only be used with the advice of a physician, audiologist or other hearing healthcare professional.
### Device Description
The TinniLogic Mobile Tinnitus Management Device is a sound player which is designed to assist the qualified healthcare professional in delivering customized sound therapies and professional counseling or education to the patient for treatment. The TinniLogic Mobile Tinnitus Management Device provides idiopathic timitus sound treatment based upon sound stimulation during sleeping or waking hours. The TinniLogic Mobile Tinnitus Management Device uses amplitude modulated (SAM) timitus pitch matched sounds, narrow-band noise centered at the tinnitus frequency. broad-band noise, pure tone, white noise, natural sounds, and combination of these sounds.
The subject device consists of a tablet, a specialized headset, a charger and a charger cable.
#### Identification of Predicate Devices ર .
Predicate Device 1 510(k) Number: K140845 Product Name: LEVO Tinnitus Masking Software Device
Predicate Device 2 510(k) Number: K111293 Product Name: SoundCure™ Serenade™ Tinnitus Management System
{5}------------------------------------------------
Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- > IEC 60601-1-2: 2014 Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- A IEC 60601-1:2005+A1:2012 Medical electrical equipment- Part 1: General requirements for basic safety, and essential performance, including the US National Differences
- A Output performance testing including the testing of Maximum output, harmonic distortion and Output frequency; additionally, the output performance testing is conducted to demonstrate that the device calibration and the maximum output reamin valid in various settings of three possible volume controls.
- 6. Clinical Test Conclusion
No clinical study is included in this submission.
{6}------------------------------------------------
#### 7. Substantially Equivalent (SE) Comparison
| Item | Subject Device | Predicate Device 1<br>K140845 | Predicate Device 2<br>K111293 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | KLW | KLW | KLW |
| Regulation Number | 21 CFR 874.3400 | 21CFR 874.3400 | 21 CFR 874.3400 |
| Indications for Use | The TinniLogic Mobile Tinnitus<br>Management Device is indicated<br>for use in the temporary relief of<br>tinnitus symptoms. The device is a<br>player to play customized sounds<br>and display professional<br>counseling or education to relieve<br>patients suffering from tinnitus and<br>can be used in a tinnitus<br>management program. The target<br>population is adults (18 years or<br>older). This is a medical device<br>and should only be used with the<br>advice of a physician, audiologist<br>or other hearing healthcare<br>professional. | The Levo System is<br>indicated for use in the<br>temporary relief of<br>tinnitus symptoms. The<br>device is a tool to<br>generate customized<br>sounds to relieve patients<br>suffering from tinnitus<br>and can be used in a<br>tinnitus management<br>program. The target<br>population is adults (18<br>years or older).<br>This is a medical device<br>and should only be used<br>with the advice of a<br>physician, audiologist or<br>other hearing healthcare<br>professional. | The SoundCure Serenade<br>Tinnitus Treatment System is<br>indicated for use in the<br>temporary relief of tinnitus<br>symptoms. The device is a tool<br>to generate customized sounds<br>to relieve patients suffering<br>from tinnitus and can be used<br>in a tinnitus management<br>program. The target<br>population is adults (18 years<br>or older). This is a medical<br>device and should only be<br>used with the advice of a<br>physician, audiologist or other<br>hearing healthcare<br>professional. |
| Components | TinniLogic Mobile Tinnitus<br>Management Device<br>with software pre-installed<br>Earphones<br>Accessories (adaptor, power cable,<br>USB cable) | Levo Manager software<br>supplied pre-installed on<br>iPad or iPad Air<br>Levo Patient software<br>supplied pre-installed on<br>iPod touch<br>Ear buds<br>Accessories (standard<br>Apple charger with<br>Apple device) | Serenade Treatment Software<br>(to be used with user's<br>commercially available<br>computer)<br>Serenade patient device with<br>software pre-installed<br>Earphones<br>Accessories (power supply,<br>power cord, USB cable) |
| Sounds | Sounds customized to the patient<br>by qualified healthcare<br>professional.<br>The subject device just only<br>displays the following sounds and | Sounds customized to the<br>patient by qualified<br>health care professional.<br>From 1 to ten (10)<br>combinations of the<br>following sounds: | Sounds customized to the<br>patient by qualified health care<br>professional. Individual<br>patient selectable from the<br>following sounds:<br>Sinusoidal amplitude |
| | | | |
| | the combination of them:<br>Sinusoidal amplitude modulated<br>(SAM) tinnitus pitch matched<br>sounds<br>Narrow-band noise centered at the<br>tinnitus frequency<br>Broad-band noises<br>Pure tone<br>White noises<br>Natural sounds<br>Relax sounds | Pure Tone<br>White Noises<br>Narrow Band Noises<br>(bandwidth selectable by<br>HCP)<br>Includes the ability to<br>create sinusoidal<br>amplitude modulated<br>(SAM) tones. | modulated (SAM) tones (S-<br>Tones) (choice of 2 predefined<br>by HCP)<br>White Noises<br>Band Noises |
| Data Logging | Data logging of patient use. | Data logging of patient<br>use. | Data logging of patient use. |
| Volume Control | Individual volume control per ear<br>of the application software;<br>Physical volume controls | Individual volume<br>control per ear of the<br>application software;<br>Physical volume controls | Individual volume control per<br>ear of the application software;<br>Physical volume controls |
| Maximum Sound<br>Loudness Output | 90 dB SPL | 85 dB SPL | 92 dB SPL |
| Sound<br>Output<br>Frequency | 50Hz~15KHz | 100Hz- 16KHz | 1 KHz to 14 KHz |
| Power | Rechargeable lithium-Ion (Li-Ion)<br>Battery, external power supply<br>with power cord for recharging | Rechargeable lithium-Ion<br>(Li-Ion) Battery, external<br>power supply with power<br>cord for recharging.<br>(Battery and charger<br>provided by Apple for<br>Apple devices) | Rechargeable lithium-Ion<br>(Li-Ion) Battery, external<br>power supply with power cord<br>for recharging. |
| Communication | USB cable and/or Wireless | Wireless | USB cable |
| Patient<br>contact<br>material | Earphones | Earphones | Earphones |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
# Table 1 Comparison of Technology Characteristics
{7}------------------------------------------------
The subject device is as safe and effective as the two predicate devices. The subject device has the similar indications for use, technological characteristics as its predicate devices. The subject device and predicate devices have the same target population and audio player functions.
of the subject device are The Maximum sound loudness output and Output sound frequency of similar as those of the predicate devices. This minor difference on output performance will not raise new problem on the safety and effectiveness of the subject device. The output performance he subject device have been conducted to demonstrate that the device calibration and the um output reamin valid in various settings of three possible volume controls.
{8}------------------------------------------------
#### 8. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.