The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic reference device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.
Device Story
Audioscan Verifit VF-1 is a hearing aid analysis system; used by hearing aid manufacturers and dispensers. Device inputs include acoustic signals from hearing aids; processed via digital signal processing (DSP) to assess electroacoustic performance (frequency, intensity, distortion). System comprises a display/control unit and a physically separate test chamber. Operates by measuring output of hearing aids in a test chamber or via probe microphone in a patient's ear canal. Output provided to clinician via color display/printer; assists in fitting and calibration of hearing aids. Benefits include optimized hearing aid performance and patient sound quality. Safety ensured by F-Type Applied Part classification for probe assembly; no direct electrical patient connection.
Clinical Evidence
Bench testing only. Device performance verified against ANSI S3.46-1997 and ANSI S3.22-1996 standards. Safety verified against IEC 60601-1, EN60601-1, UL-2601, and UL 544. Software validation performed per FDA guidance.
Technological Characteristics
System uses digital signal processing (DSP) for analysis. Features color SVGA display, dual stimulus channels, and independent speakers. Probe microphone assembly uses silicone tubing. Connectivity includes Ethernet, USB, RS232, and parallel ports. Powered by electrical mains; meets IEC 60601-1 safety standards. Test chamber includes induction coils and battery simulators per ANSI S3.22-1996.
Indications for Use
Indicated for manufacturers and dispensers of hearing aids to calibrate and assess electroacoustic frequency and sound intensity characteristics of hearing aids, either coupled to an ear simulator in a test chamber or worn by a patient.
Regulatory Classification
Identification
A hearing aid calibrator and analysis system is an electronic reference device intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid, master hearing aid, group hearing aid or group auditory trainer. The device consists of an acoustic complex of known cavity volume, a sound level meter, a microphone, oscillators, frequency counters, microphone amplifiers, a distoration analyzer, a chart recorder, and a hearing aid test box.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
Predicate Devices
Audioscan® Real Ear Measurement System RMS00 (K884046/A)
Submission Summary (Full Text)
{0}------------------------------------------------
Edi E
sion incorporated
41 Byron Avenue, Dorchester, Ontario NOL 1G0 Tel: (519) 268-3313 Fax: (519) 268-3256
SEP 2 1 2001
K012306
## SECTION E - 510(K) SUMMARY OF SAFETY & EFFECTIVENESS
## Audioscan® Verifit™ Model VF-1
| Classification Name: | Hearing-Aid Calibrator & Analysis System |
|-----------------------------|--------------------------------------------|
| Common/Usual Name: | Hearing Aid Analyzer |
| Proprietary Name: | Audioscan <sup>®</sup> Verifit™ Model VF-1 |
| Establishment Registration: | 8022229 |
| Medical Specialty: | Panel 77, Ear, Nose, & Throat |
| Product Code: | ETW |
| Regulation Number: | 21 CFR 874.3310 |
| Device Classification: | II |
| Contact Person: | William A Cole, President |
#### Indications for Use:
The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic reference device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.
## Performance Standards Conformance:
- 1. The VF-1 meets or exceeds the equipment requirements of the following standards: American National Standards Institute, Methods of Measurement of Real-Far Performance Characteristics of Hearing Aids, (ANSI S3.46-1997), American National Standards Institute, New York, 1997, and, American National Standards Institute, Specification of Hearing Ald Characteristics, (ANSI S3.22-1996), American National Standards Institute, New York, 1996.
- 2. The VF-1 meets or exceeds the applicable requirements of the following standards for safety: IEC 60601-1, EN60601-1, UL-2601 and UL 544.
{1}------------------------------------------------
Etymonic Design Incorporated Applied Microelectronics
- The VF-1 meets or exceeds the Food and Drug Administration radiation safety 3. performance standards prescribed in the Code of Federal Regulations, Title 21, Chapter I. Subchapter J.
#### Substantial Equivalence:
Determination of substantial equivalence of the Audioscan® Verifit™ VF-1 instrument to the currently marketed Audioscan® Real Ear Measurement System RMS00, 510(k) No. K884046/A, February 13, 1989, is based on the following safety and effectiveness information:
- The Audioscan® Verifit™ device is manufactured and delivered completely . assembled to the health care professional using materials and techniques widely used by other manufacturers of devices as with the predicate device.
- Patient safety is preserved in both the Audioscan® Verifit™ and predicate . devices. There is no direct electrical connection to the patient. The silicone probe microphone assembly that has contact with the patient's ear canal during the testing of the hearing aid, has no exposed metal parts and is classed as an F-Type Applied Part in UL2601-1. Thus, there is no possibility of electrical conductance.
- Regarding physical configurations, the system is integrated with the hearing test ● chamber physically part of the display/control unit in the predicate device, whereas the Audioscan® Verifit™ hearing aid test chamber is physically separated from the display/control unit.
- Software algorithms are essentially equivalent for both the Audioscan® Verifit™ t and predicate devices.
- To establish the safety and effectiveness of the software that controls the . instrument, the development, verification, validation, and testing procedures are in accordance with FDA guidances for devices containing software 12, Quality Assurance procedures are adhered to and test results demonstrate that both system specifications and function requirements are met.
<sup>1</sup> Guidance Principles of Software Validation, Draft Guidance Version 1.1, June 9, 1997, Center for Devices and Radiological Health, Food and Drug Administration
<sup>2</sup> ODE Guidance for the Content of Premarket Submission for Medical Devices Containing Software, May 29, 1998, Center for Devices and Radiological Health, Food and Drug Administration
{2}------------------------------------------------
Etymonic Design Incorporated Applied Microelectronics
41 Byron Avenue, Dorchester, Ontario NoL. 1G0 Tel: (519) 268-3313 Fax: (519) 268-3256
- The maximum sound level output of the Audioscan® Verifit™ speakers, if used . as intended, remains below +90 dB SPL throughout the frequency range of 200 to This demonstrates that the Audioscan® Verifit™ unit emits sounds 8000 Hz. well within OSHA permissible exposure limits of 8 hours at 90 dB, The maximum sound level output of the predicate device speakers is 85 dB SPL if used as intended.
- The signal processing and analysis differ between the Audioscan® Verifit™ and . the predicate device. Signals are processed and analyzed by digitally controlled analog amplifiers, attenuators and filters in the predicate device. In the Audioscan® Verifit™, with the exception of the microphone preamplifiers and output amplifiers, all signal processing and analysis is carried out digitally.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "edi" in a bold, sans-serif font. The letters are composed of small squares, giving them a pixelated appearance. The letters are closely spaced together.
vmon n Incorporated
41 Byron Avenue, Dorchester, Ontanio NOL 1G0
| CHARACTERISTIC | MODEL RM500 | VERIFIT™ MODEL<br>VF-1 |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| GENERAL | | |
| Overall dimensions | 15.5" x 13.2" x 6.2" | Rear main unit 14.5" x 16" x 6.5"<br>Test chamber 14" x 14.4" x 4" |
| Weight | 15 lbs | 16 lbs |
| Display type | fluorescent backlit blue on<br>white | fluorescent backlit active color |
| Display size | 6.4" diagonal | 12.1" diagonal |
| Display pixels (resolution) | 320 x 200 (VGA) | 800 x 600 (SVGA) |
| Printer type | thermal line printer | SAME |
| Printer resolution | 200 dots per inch | SAME |
| Paper width | 3" (80 mm) | SAME |
| Connectors | 1 - external monitor (15HD)<br>1-RS232 (9D)<br>1 - external speaker (RCA)<br>1 - real-ear/coupler mic. (RJ11) | 1 - mouse (PS2-6 pin)<br>1 - QWERTY keyboard (PS2-6pin)<br>1 - external monitor (15HD)<br>1 - parallel printer (25D)<br>1 - RS232 serial (9D)<br>1 - ethernet (RJ45)<br>2 - USB<br>2 - external speakers (RCA)<br>2 - external speakers (1/4" mono))<br>2 - real-ear mics. (3.5 mm st)<br>1 - RECO transducer (3.5 mm st)<br>1 - monitor headset (3.5 mm st)<br>1 - test chamber ref. mic. (3.5 mm st)<br>1 - coupler microphone (3.5 mm st)<br>1 - battery substitute (3.5 mm st) |
| Headphone monitor amplifier | 1 watt into 16 ohms | SAME |
| Power amplifiers | 1 - 1.7 watts | 2 - 5 watts each |
| Stimulus channels | 1 | 2 |
| Measurement channels | 2 | SAME |
| HEARING AID TEST CHAMBER | | |
| Working space | 5.5" x 3.5" x 2" | 8" x 5" x 1.5" (approx.) |
| Speakers | 1 - 3.5" diameter | 2 - 2" x 3" Independent |
| Induction coils | 1 - Telephone Magnetic Field<br>Simulator (TMFS ANSI S3.22 -<br>1996)<br>1 - 20 cm diameter test loop | SAME |
| Battery simulator | per ANSI S3.22 - 1996 | SAME |
| Frequency range | 200 to 8000 Hz | SAME |
| Test stimuli | tone | tone, broad-band noise, recorded<br>speech |
| CHARACTERISTIC | MODEL RM500 | VERIFIT™ MODEL<br>VF-1 |
| Test stimulus levels | 40 to 90 dB SPL in 5 dB steps | SAME |
| Test stimulus levels (inductive) | 31.6 mA/m per ANSI S3.22 -1996 | SAME |
| Test stimulus distortion | <2% at 90 dB SPL<br><0.5% at 70 dB SPL | SAME |
| Test stimulus accuracy at reference<br>microphone for tones (200 - 2000 Hz) | ± 1.5 dB SPL | SAME |
| Test stimulus accuracy at reference<br>microphone for tones (2000 -8000 Hz) | ± 2.5 dB SPL | SAME |
| Equalization method | pressure method | SAME |
| Analysis frequencies per octave | 12 | SAME |
| Analysis filter bandwidth | 1/7 octave | 1/12 octave |
| Measurement accuracy at 1 kHz | ± 1 dB | SAME |
| Measurement accuracy re 1 kHz | ± 1 dB (200 - 5000 Hz)<br>± 2.5 dB (5000 - 8000 Hz) | SAME |
| Measurement range | 30 - 140 dB SPL | SAME |
| Harmonic distortion measurement | 2nd plus 3rd | 2nd and 3rd or 2nd plus 3rd |
| Harmonic distortion range | 200 to 2670 Hz | 200 to 4000 Hz |
| Harmonic distortion accuracy | ± 1% | SAME |
| Battery drain range | 0-20 mA | SAME |
| Battery drain accuracy | ± .01 mA | SAME |
| ANSI S3.22-1996 tests available | ➤ OSPL90<br>➤ Full-on Gain<br>➤ Reference Test Gain<br>➤ Frequency Response<br>➤ Frequency Range<br>➤ Maximum OSPL90<br>➤ Harmonic Distortion<br>➤ Attack & Release time<br>➤ Equivalent Input Noise<br>➤ Input/Output Curves<br>➤ Coupler SPL - Telephone<br>Simulator<br>➤ Simulated Telecoil<br>➤ Sensitivity<br>➤ Battery drain | SAME<br>With the added features:<br>➤ Coupler SPL - Vertical<br>➤ Magnetic Field<br>➤ Test Loop Sensitivity |
| Other tests available | ➤ coupler SPL vs frequency<br>➤ spectral analysis<br>➤ manual measurement of<br>➤ output, gain and distortion | SAME<br>With the added features:<br>➤ coupler gain vs frequency<br>➤ distortion vs frequency |
| REAL-EAR MEASUREMENT | | |
| Speakers | 1 - 3.5" diameter | 2 - 2" x 3" ducted ports |
| CHARACTERISTIC | MODEL RM500 | VERIFIT™ MODEL |
| | | VF-1 |
| Probe microphone tube | Silicone 1.0mm diameter x<br>75mm | SAME |
| Frequency range | 200 to 8000 Hz | SAME |
| Test stimuli | frequency-modulated tone, tone<br>burst, amplitude-modulated tone | frequency-modulated tone, tone<br>burst, broad-band noise, recorded<br>speech |
| Frequency modulation | triangular ±5% at 36 Hz | SAME |
| Test stimulus levels at reference<br>microphone for tones at 1 m | 40 to 85 dB SPL in 5 dB steps | 40 to 90 dB SPL in 5 dB steps |
| Test stimulus accuracy at reference<br>microphone for tones (200 - 2000 Hz) | ± 1.5 dB SPL | SAME |
| Test stimulus accuracy at reference<br>microphone for tones (2000 -8000 Hz) | ± 2.5 dB SPL | SAME |
| Equalization method | pressure method | SAME |
| Analysis frequencies per octave | 12 | SAME |
| Analysis filter bandwidth | 1/7 octave | 1/12 octave |
| Measurement accuracy at 1 kHz | ± 1 dB | SAME |
| Measurement accuracy re 1 kHz | ± 1 dB (200 - 5000 Hz)<br>± 2.5 dB (5000 - 8000 Hz) | SAME |
| Measurement range | 20-135 dB SPL (200 - 2500<br>Hz)<br>30 - 140 dB SPL (2500 - 8000<br>Hz) | SAME |
| ANSI S3,46-1997 tests available | > Real-Ear Unaided<br>Response<br>> Real-Ear Aided Response<br>> Real-Ear Occluded<br>Response<br>> Real-Ear Insertion Gain | SAME |
| Other tests available | > Real-ear harmonic<br>distortion<br>> Real-ear spectral analysis<br>> manual measurement of<br>output, gain and distortion | SAME |
| Prescriptive fitting methods available | > National Acoustics Labs<br>Revised<br>> Pogo II<br>> Berger<br>> Libby<br>> Desired Sensation Level | SAME |
# Specification Comparison Table
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Etymonic Design Incorporated and Applied Microelectronics. The text is in black and is centered on the page. The words "Etymonic Design Incorporated" are on the top line, and the words "Applied Microelectronics" are on the bottom line. The logo is simple and professional.
.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Etymonic Design Incorporated, Applied Microelectronics. The logo consists of the letters "edi" in a stylized font on the left, followed by the company name in a more standard font on the right. The words "Applied Microelectronics" are written in a smaller font below the company name.
41 Byron Avenue, Dorchester, Ontario NOL 1G0 Tel: (519) 288-3313 Fax: (519) 268-3256
NOTES:
The couplers used in the RM500 hearing aid test chamber are also used in the Verifit™ VF-1. 1)
2) The probe tubes used with the RM500 real-ear microphone assembly are also used with the VF-1.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 1 2001
William A. Cole, President Etymonic Design Incorporated 41 Byron Avenue Dorchester, Ontario N0L 1G0
Re: K012306
Trade Name: Audioscan Verifit™ Regulation Number: 21 CFR 874.3310 Regulation Name: Hearing Aid Calibrator and Analysis System Regulatory Class: Class II Product Code: ETW Dated: July 20, 2001 Received: July 23, 2001
Dear Mr. Cole
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
### Page 2 - William A. Cole, President
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "edi" in a pixelated font. The letters are formed by small squares, giving it a retro, digital appearance. The font is simple and blocky, making the letters easily readable despite the pixelation.
Etymonic Design Incorporated Applied Microelectronics
41 Byron Avenue, Dorchester, Ontario N0L 1G0 Fax: (519) 268-3256 Tel: (519) 268-3313
# SECTION D - INDICATIONS FOR USE STATEMENT
er/ 3 - 4/24/96
Applicant: Etymonic Design Incorporated
510(k) Number (if known):
Device Name: Audioscan® Verifit™
Indications For Use
The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic referenc device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.
Ama H. Kane Ph.D. 9/17/01
Division Sign-Off
Division Sign-Off) Division of Ophthalmic Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.