Auris series hearing aids (AE, TL, HS, AC, MM, COE, CIC) are air-conduction devices designed to amplify and transmit sound to the ear. Devices utilize DSD K-amp or Intrigue programmable circuitry to process sound based on individual audiogram requirements. Devices are assembled from standard components common to the hearing aid industry. Operated by the patient, the devices are worn in or on the ear (various form factors including all-in-the-ear, half-shell, and canal styles). Power is provided by standard batteries (sizes 13, 312, 10A, or 5A). Healthcare providers fit the devices by adjusting frequency response to match the patient's audiogram. The output provides amplification to assist patients with hearing loss. Users are advised of potential electromagnetic interference from digital cellular telephones.
Clinical Evidence
No clinical data provided; bench testing only. Technical specifications comply with S3.2-1987 ANSI Standards.
Technological Characteristics
All-in-the-ear, half-shell, and canal-style hearing aids. Programmable circuitry (DSD K-amp or Intrigue). Powered by standard batteries (13, 312, 10A, 5A). Technical specifications comply with S3.2-1987 ANSI Standards. Frequency response is customized via individual audiogram.
Indications for Use
Indicated for patients with mild to severe conductive, mixed, or sensoryneural hearing loss. Specific models indicated for cosmetic concerns (HS, MM, CIC, AC, TL) or specific high-frequency loss patterns (COE).
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
Submission Summary (Full Text)
{0}------------------------------------------------
K971399
- TUN I 3 Nagt
NAME OF DEVICE: Auris AE
TYPE OF DEVICE: All-in-the-Ear
To amplify and transmit sound to the INTENDED USE : ear.
DSD K-amp or Intrigue programmable FEATURES : circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications comply with S3.2-1987 ANSI Standards" Frequency response dictated by FIT:
individual audiogram
Volume control similar to other devices. CONTROLS : Standard size 13 battery POWER :
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{1}------------------------------------------------
NAME OF DEVICE: Auris TL
TYPE OF DEVICE: All-in-the-Ear
To amplify and transmit sound to the INTENDED USE : ear.
DSD K-amp or Intrigue programmable FEATURES : ي تي پي په پ circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications comply with S3.2-1987 ANSI Standards" Frequency response dictated by FIT: individual audiogram Volume control similar to other devices. CONTROLS :
Standard size 312 battery POWER:
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{2}------------------------------------------------
NAME OF DEVICE: Auris HS
TYPE OF DEVICE: Half-shell-in-the-Canal
To amplify and transmit sound to the INTENDED USE : ear.
DSD K-amp or Intrigue programmable FEATURES : ﺔ ﺗﻮﻓﻴﺮ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications comply with S3.2-1987 ANSI Standards" Frequency response dictated by FIT: individual audiogram
Volume control similar to other devices. CONTROLS :
POWER: Standard size 312 battery
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{3}------------------------------------------------
NAME OF DEVICE: Auris AC
TYPE OF DEVICE: All-in-the-Canal
To amplify and transmit sound to the ear. INTENDED USE : DSD K-amp or Intrigue programmable FEATURES :
circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid ته ته ته ته ته ا manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications
comply with S3.2-1987 ANSI Standards"
Frequency response dictated by individual FIT: audiogram
Volume control similar to other devices. CONTROLS :
Standard size 312 battery POWER:
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{4}------------------------------------------------
NAME OF DEVICE: Auris MM
FIT:
TYPE OF DEVICE: Micro-Mini (All-in-the-Canal)
To amplify and transmit sound to the ear. INTENDED USE : DSD K-amp or Intrigue programmable FEATURES : circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications
comply with S3.2-1987 ANSI Standards"
Frequency response dictated by
individual audiogram
Volume control similar to other devices. CONTROLS :
Standard size 10A battery POWER:
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{5}------------------------------------------------
NAME OF DEVICE: Auris COE
TYPE OF DEVICE: Canal-Open-Ear (All-in-the-Canal)
To amplify and transmit sound to the INTENDED USE : ear.
DSD K-amp or Intrigue programmable FEATURES : نيو سي پيتي جي تي circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications comply with S3.2-1987 ANSI Standards" Frequency response dictated by FIT: individual audiogram
Volume control similar to other devices. CONTROLS : Standard size 10A battery POWER:
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{6}------------------------------------------------
NAME OF DEVICE: Auris CIC
TYPE OF DEVICE: Completely-in-the-Canal
To amplify and transmit sound to the ear. INTENDED USE :
DSD K-amp or Intrigue programmable FEATURES : ﻨﻲ circuitry
Assembled from standard components that ASSEMBLY : are widely used by other hearing aid manufacturers.
TECHNICAL CHARACTERISTICS: "Technical specifications comply with S3.2-1987 ANSI Standards" Frequency response dictated by individual FIT: audiogram
CONTROLS : Volume control similar to other devices.
POWER: Standard size 10a or 5a battery
A USERS' MANUAL AND OTHER INFORMATION
IS SUPPLIED WITH EACH HEARING AID.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### 1661 8 1 NUTT
Roger P. Juneau President General Hearing Instruments, Inc. -P.O. Box 61010 New Orleans, LA 70161-1010
学
Re: K971399
Programmable Option for Auris AE, Auris TL, Auris HS, Auris AC, Auris MM, Auris COE and CIC Air Conduction Hearing Aids Dated: April 25, 1997 Received: April 28, 1997 Regulatory Class: I 21 CFR 874.3300/Procode: 77 ESD
Dear Mr. Juneau:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the exactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent . determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device in a classification for your device and thus, permits your device to proceed to the market.
While your device has been deemed substantially equivalent to other legally marketed hearing ands, please be advised that electromagnetic interference from digital cellular telephones, as well as from other sources, is increasingly becoming a concern. Typically, this interference takes the form of a buzzing sound that can range from annoying to very loud and may render a hearing aid temporarily ineffective for the wearer. Because electromagnetic interference may affect your device, you may be asked to test for electromagnetic compatibility in the future. In this interim period, we encourage you to modify your device labeling to inform practitioners and users of the potential for electromagnetic interference. Please be aware that a 510(k) submission is required for any claims that infer that your device is compatible with potential sources of electromagnetic interference, such as "compatible with digital cellular telephones", and that data supporting such claims is necessary.
{8}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address ----------------------------------------------------"http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
.
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
..............................................................................................................................................................................
{9}------------------------------------------------
510(k) Number (if known): _K971399
Device Name: Auris AE-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to severe conductive, mixed or sensoryneural hearing loss
发:大
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Drive
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Harriet M. Bergman
ivision Sign-Off) · vision of Reproductive, Abdominal, ENT, d Radiological Devices 0(k) Number _
{10}------------------------------------------------
Page 2_of_7
510(k) Number (if known): K971399
Device Name: Auris HS-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to severe conductive, mixed or sensoryneural hearing loss when cosmetics are of some concern
乐 > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > >
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Seine
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Schmidt Ch. Sejrem
(Division Sign-Off)
( - Choice Of Reproductive, Abdominal, ENT,
and Radiological Devices and Radiological Device
510(k) Number K971344
{11}------------------------------------------------
Page 3 of 7
510(k) Number (if known): _K971399
Device Name: Auris MM-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to moderate conductive, mixed or sensoryneural hearing loss when cosmetics are of major concern, but the ear can not accommodate a complete in the canal style
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Service
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Clifford W. Kellison
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices
510(k) Number K9711394
{12}------------------------------------------------
Page 4 of 7
510(k) Number (if known): K971399
Device Name: Auris CIC-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to moderate conductive, mixed or sensoryneural hearing loss when cosmetics are of major concern.
乐心--
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Service
Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96)
Varick A. Wynen
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number _
{13}------------------------------------------------
Page 5_of _
510(k) Number (if known):_K971399
Device Name: Auris TL-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to severe conductive, mixed or sensoryneural hearing loss when a more cosmetic AE is desired.
ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Sample. \$\checkmark\$
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Daniel C. Segerson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number
{\rtf1\ansi\deff0
{\fonttbl{\f0 Arial;}}
{\colortbl ;\red0\green0\blue0;}
\viewkind4\uc1\pard\cf1\f0\fs24 }
{14}------------------------------------------------
510(k) Number (if known): _K971399
Device Name: Auris AC- (DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Mild to severe conductive, mixed or sensoryneural hearing loss when cosmetics are a major concern but the ear prevents use of a CIC or MM.
ﺮ ﺍﻟﻤﻮﺍﻗﻊ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Device
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
David A. Siyonn
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devi 510(k) Number
{15}------------------------------------------------
510(k) Number (if known):_K971399
Device Name: Auris COE-(DSD K-amp) or (Intrigue-Pro) Option
Indications For Use:
Normal low and middle frequency hearing with a sloping mild to moderate high frequency sensory neural or mixed loss.
4
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restricted Device ✓
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Daniel A. Lippman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devi 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.