INTERACOUSTICS MODEL MT 10, HANDHELD IMPEDANCE AUDIOMETER
K983832 · Idem (Int'L Dist. of Electronics For Medicine) · ETY · Jan 25, 1999 · Ear, Nose, Throat
Device Facts
Record ID
K983832
Device Name
INTERACOUSTICS MODEL MT 10, HANDHELD IMPEDANCE AUDIOMETER
Applicant
Idem (Int'L Dist. of Electronics For Medicine)
Product Code
ETY · Ear, Nose, Throat
Decision Date
Jan 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1090
Device Class
Class 2
Indications for Use
The Interacoustics MT10 Handheld Impedance Audiometer is an easy to use handheld diagnostic middle ear analyzer that is well suited for providing quick and reliable results in both diagnostic and screening settings. The tympanometry test features a user selectable extended pressure range that is capable of -600daPa. Four ipsilateral reflexes at different frequencies may be tested at each ear. Reflex testing operates either at a user defined fixed level or uses an automatic intensity search function. It assists in the diagnosis of now sible otologic disorders.
Device Story
Handheld impedance audiometer for diagnostic and screening middle ear analysis; used in clinical settings by healthcare professionals. Device inputs include acoustic signals via probe tone (226 Hz) and pressure sweeps (+300 to -600 daPa). Performs tympanometry, ipsilateral acoustic reflex testing (fixed level or automatic intensity search), and screening audiometry. Produces digital measurements of compliance and reflex thresholds; provides hardcopy printouts for objective documentation of otitis media and middle ear disorders. Supports PC connectivity via Windows-based software (laBase 95, Printview, NOAH) for monitoring and data management. Benefits include quick, reliable assessment of middle ear function to aid clinical diagnosis.
Clinical Evidence
Bench testing only. Device performance verified against IEC 1027-1995, ANSI 3.60-1989, ANSI 3.60-1985, and IEC 601-1-1988 safety standards. No clinical trial data provided.
Technological Characteristics
Handheld impedance audiometer; digital display; NiMH or AA NiCa battery power. Probe tone: 226 Hz (+/-3%). Pressure range: +300 to -600 daPa/sec. Compliance range: 0.0 to 2.5 cc (displayed). Connectivity: PC interface via Windows software (laBase 95, Printview, NOAH). Standards: IEC 1027-1995, ANSI 3.60-1989/1985, IEC 601-1-1988.
Indications for Use
Indicated for patients requiring diagnostic or screening middle ear analysis to assist in the diagnosis of otologic disorders.
Regulatory Classification
Identification
An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.
Special Controls
*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
Predicate Devices
Grason Stadler GSI- 37 Auto Tymp
Submission Summary (Full Text)
{1}------------------------------------------------
K983832
Page 1 of 2
JAN 25 1999
## 510(k) SUMMARY
| PREPARED BY: | INTERNATIONAL DISTRIBUTORS OF<br>ELECTRONICS FOR MEDICINE, INC.<br>(IDEM)<br>4814 East Second Street<br>Benicia, CA 94510 |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Donna Ward, President |
| TELEPHONE: | 707-746-6334 |
| DATE ON WHICH THE SUMMARY<br>WAS PREPARED: | October 1, 1998 |
| NAME IF DEVICE: | Interacoustics Model MT10 Handheld<br>Impedance Audiometer |
| COMMON NAME: | Impedance Audiometer |
| PREDICATE DEVICE: | Grason Stadler GSI- 37 Auto Tymp |
DESCRIPTION OF DEVICE: Our device, the Model MT10 Handheld Impedance Audiometer is an easy to use handheld diagnostic middle ear analyzer that is well suited for providing quick and reliable results in both diagnostic and screening settings. The tympanometry test features a user selectable extended pressure range that is capable of -600daPa. Four ipsilateral reflexes at different frequencies may be tested at each ear. Reflex testing operates either at a user defined fixed level or uses an automatic intensity search function. It assists in the diagnosis of possible otologic disorders.
Comparison of the Interacoustics Model MT10 Handheld Impedance Audiometer and the Grason Stadler GSI-37 Auto Tymp.
Indication for use - Identical for both units.
Similarities and differences:
| Model MT10 Handheld Impedance<br>Audiometer | Equivalent - Grason Stadler GSI-37<br>Auto Tymp |
|---------------------------------------------|-------------------------------------------------|
| Display Description: Digital | Digital |
| Lightweight, handheld | Lightweight, handheld |
{2}------------------------------------------------
| Hardcopy printout provides objective<br>documentation of otitis media and<br>other middle ear disorders. | Hardcopy printout provides objective<br>documentation of otitis media and<br>other middle ear disorders. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Intensity: 10dBHL to 50dBHL | Not applicable |
| Pressure range: +300 -600daPa/sec. | +200 -400 daPa/sec. |
| Compliance range: .0 to 2.5 cc<br>displayed; .0 to 5.0 cc numeric | .0 to 1.5cm³ displayed |
| Tests: tympanometry, acoustic reflex<br>and screening audiometry | Tympanometry |
| Compatible Windows Software:<br>laBase 95 database program;<br>Printview for on-line PC monitoring and<br>printing; NOAH hearing aid fitting<br>software | None |
| Standards: meets or exceeds<br>standards specified in IEC 1027-1995,<br>ANSI 3.60 -1989, ANSI 3.60-1985,<br>Safety IEC 601-1-1988 | Meets ANSI S3.39 - 1987 for Type 4<br>Instrument and IEC 601-1-1988 |
| Available frequencies: .5kHz, 1kHz,<br>2kHz, 3kHz & 4kHz | Not applicable |
| Probe tone: frequency: 226 Hz, +/-3% | 226 Hz, +/-3% |
| Amplitude: 85dB SPL +/-3dB | 85.5 dB SPL, +/-2.0 dB |
| Power: NiMH batteries or Standard AA<br>NiCa batteries | NiCad batteries |
| Direction of Pressure Sweep: positive<br>to negative | Positive to negative |
| Size and Weight: Probe: 4" x 10" x 5";<br>1 lb.; printer/charger: 12" x 9" x 4"; 4<br>lbs. 11.6 oz. | Probe: 2.4" x 8.3" x 4.3"; 10.5 oz;<br>printer/charger 9" x 7.6" x 4"; 3 lbs. |
| External Power Supply: (optional) 1.8<br>Ibs (recharger power supply EPS11<br>power transformer-11 volts) | External Power Supply: (optional) 1.5<br>Ibs. (recharger power supply -9 volts) |
## SAFETY AND EFFECTIVENESS:
The Interacoustics Model MT10 Handheld Impedance Audiometer is in compliance with the following performance and safety standards:
> Standards in IEC 1027-1995, ANSI 3.60-1989, ANSI 3.60-1985, Safety IEC 601-1-1988. Approved for Medical CE-marking by Danish Notified Body D.G.M. Identification No. 0543, Quality System Certificate DGM 016 (CE 0543)
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 25 1999
Donna Ward President International Distributors of Electronics for Medicine, Inc. 4814 East Second Street Benicia, CA 94510
Re: K983832
Interacoustics Model MT10 Handheld Impedance Audiometer Dated: October 28, 1998 Received: October 30, 1998 Regulatory class: II: 21 CFR 874.1090/Procode: 77 ETY
Dear Ms. Ward:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): 49838383
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Usc:
The Interacoustics MT10 Handheld Impedance Audiometer is an easy to use handheld diagnostic middle ear analyzer that is well suited for providing quick and reliable results in both diagnostic and screening settings. The tympanometry test features a user selectable extended pressure range that is capable of -600daPa. Four ipsilateral reflexes at different frequencies may be tested at each ear. Reflex testing operates either at a user defined fixed level or uses an automatic intensity search function. It assists in the diagnosis of now sible otologic disorders.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------|-----------------------------------------------------------------------|--|
| | (Division Sign-Off) | |
| | Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
| 510(k) Number | K983832 | |
| Prescription Use | OR | Over-The-Counter Use |
|----------------------|----|----------------------|
| (Per 21 CFR 801.109) | | (Optional Format) |
(Optional Formal 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.